(28 days)
No
The summary describes standard image viewing, manipulation, quantification, analysis, and comparison functionalities. There is no mention of AI, ML, deep learning, or any related terms, nor is there any description of training or test sets typically associated with AI/ML model development. The performance studies focus on software verification and validation, not AI/ML model performance metrics.
No.
The device is described as a medical diagnostic application for viewing, manipulating, quantifying, analyzing, and comparing medical images, and aids in the management of diseases and radiotherapy treatment planning. It does not directly treat or provide therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states, "MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images..." and "MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function." This indicates its use in aiding diagnosis and patient management.
Yes
The device description explicitly states "MI View&GO is a software-only medical device".
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The description clearly states that MI View&GO is a software application for viewing, manipulating, quantifying, analyzing, and comparing medical images (PET, NM, CT, MR, SPECT). It works with data acquired directly from imaging scanners.
- No Sample Analysis: There is no mention of the device analyzing biological samples or performing tests on bodily fluids or tissues. Its function is solely related to processing and interpreting medical images.
Therefore, while it is a medical diagnostic device, it falls under the category of medical imaging software rather than an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
MI View&GO is a medical diagnostic application for viewing, manipulation, guantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
Product codes (comma separated list FDA assigned to the subject device)
QIH
Device Description
MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.
The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:
- Review acquired and reconstructed images at the scanner console
- Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
- Prepare images for reading
- Perform a basic read
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
positron emission tomography (PET), nuclear medicine (NM), computed tomography (CT), magnetic resonance (MR), SPECT
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
appropriately trained health care professionals, physician, technologist
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The device under application, MI View&GO, did not conduct any additional performance testing. Verification and validation testing substantiates all requirement and functional specifications, including specifications related to device hazards, and supports the claim of substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 30, 2024
Siemens Medical Solutions USA, Inc. Brian Wui Official Correspondent 2501 North Barrington Road Hoffman Estates, Illinois 60192
Re: K242300
Trade/Device Name: MI View&GO Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: August 2, 2024 Received: August 2, 2024
Dear Brian Wui:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
1
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FDA
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Indications for Use
Submission Number (if known)
Device Name
MI View&GO
Indications for Use (Describe)
MI View&GO is a medical diagnostic application for viewing, manipulation, guantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
510(k) Summary
1. Identification of the Submitter
| Submitter / Primary Contact
Person | Brian Wui
Regulatory Affairs
hansong.wui@siemens-healthineers.com
+1 (865) 367-4337 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Secondary Contact Person | Clayton Ginn
Regulatory Affairs
clayton.ginn@siemens-healthineers.com
+1 (865) 898-2692 |
| Applicant Name and Address | Siemens Medical Solutions, Inc. USA
2501 North Barrington Road
Hoffman Estates IL, 60192, USA
Establishment Registration Number: 1423253 |
2. Device Name and Classification
Product Trace Name: | MI View&GO VA30 |
---|---|
Common Name: | Medical image management and processing system |
Classification Name: | Automated Radiological Image Processing Software |
Classification Panel: | Radiology |
CFR Section: | 21 CFR §892.2050 |
Device Class: | Class II |
Product Code: | QIH |
3. Predicate Devices
Primary Predicate Device:
Product Trace Name: | MI View&GO VA20 |
---|---|
510(k) Number: | K201202 |
Clearance Date: | 07/21/2022 |
Common Name: | Medical image management and processing system |
Classification Name: | Automated Radiological Image Processing Software |
Classification Panel: | Radiology |
CFR Section: | 21 CFR §892.2050 |
Device Class: | Class II |
Product Code: | QIH |
4
Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
Reference Device
Trade Name: | syngo.via MI Workflows VB80 |
---|---|
510(k) Number: | K232000 |
Clearance Date: | 11/28/2023 |
Common Name: | Medical image management and processing system |
Classification Name: | Automated Radiological Image Processing Software |
Classification Panel: | Radiology |
CFR Section: | 21 CFR §892.5050 |
Device Class: | Class II |
Product Code: | QIH |
4. Device Description
MI View&GO is a software-only medical device which will be delivered in conjunction with Siemens SPECT/CT and PET/CT scanners. MI View&GO software provides additional specific capabilities for handling of PET and SPECT as well as CT and MR data directly at the acquisition console.
The MI View&GO software integrates molecular imaging more efficiently in the clinical environment by providing an interface for its users to review, post-process and read medical images immediately after acquisition. The purpose of the MI View&GO is to allow the technologist and reading physician to:
- Review acquired and reconstructed images at the scanner console
- Determine that the acquired data is of sufficient quality for reading, so the patient can be released.
- Prepare images for reading
- Perform a basic read
5. Indications for Use
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
6. Indications for Use Comparison to the Predicate Device
The indications for use are the same between the subject device and the primary predicate device.
5
Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
7. Comparison of Technological Characteristics with the Pro
MI View&GO with software version VA30 software provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the legally marketed predicate device since all devices are software only devices.
The software features have been modified in comparison to the predicate device to support enhanced device functionality.
The intended use, indications for use, and fundamental scientific technology for the subject device remains unchanged from the predicate device. No features present from the predicate device have been de-scoped.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Data Supported (PET, SPECT, CT, MR)
- . Server/Client architecture
- . Workflow Activities (preprocessing, evaluation and reading, reporting and storage)
- . Feature Licensing Structure
- SUV values calculated
The following technological differences exist between the subject device and predicate devices.
MI View&GO VA30:
- MI Features: ●
- O Brain AC-PC(Anterior Commissure – Posterior Commissure)
- O PERCIST
- VOI Isocontour O
- Shared Software Common Tools and CT Features: .
- Layout Editor O
- O Average
- Stroke Layout O
- . Layout Improvements:
- MIP (Maximum Intensity Projection) Rotation O
- o Volume Stripe Image Order
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation are completed. Test results show that the subject devices are comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices.
6
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a cluster of orange dots arranged in a circular pattern.
8. Non-Clinical and/or Clinical Test Summary & Conclusion
The following performance data were provided in support of the substantial equivalence determination.
Software Verification and Validation
'Enhanced' software documentation per FDA's guidance document "Content of Premarket Submissions for Device Software Functions" issued on June, 2023 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The testing supports that all software specifications have met the predetermined acceptance criteria. Verification and validation testing substantiates all requirement and functional specifications, including specifications related to device hazards, and supports the claim of substantial equivalence.
Performance Testing
The device under application, MI View&GO, did not conduct any additional performance testing.
Risk Analysis
The risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the predetermined acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Standards
Siemens hereby certifies that MI View&GO meets the following FDA Recognized Consensus standards listed below:
| Designation
Number and
Edition/Date | Title | Standards
Development
Organization | Recognition
Number |
|------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------|-----------------------|
| PS 3.1 - 3.20
2023e | Digital Imaging and Communications in Medicine
(DICOM) Set | NEMA | 12-349 |
| 62304
Edition 1.1
2015-06
CONSOLIDATED
VERSION | Medical Device Software – Software Life Cycle
Processes | AAMI, ANSI,
IEC | 13-79 |
| 14971 Third
Edition
2019-12 | Medical devices – Application of risk management to
medical devices | ISO | 5-125 |
7
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words are a series of orange dots arranged in a circular pattern.
| 62366-1
Edition 1.1
2020-06
CONSOLIDA
TED
VERSION | Medical devices - Part 1: Application of usability
engineering to medical devices | AAMI, ANSI,
IEC | 5-129 |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-------|
| 15223-1
Fourth
edition
2021-07 | Medical devices - Symbols to be used with medical
device labels, labelling, and information to be
supplied - Part 1: General requirements | ISO | 5-134 |
| 20417 First
edition
2021-04
Corrected
version
2021-12 | Medical devices - Information to be supplied by the
manufacturer | ISO | 5-135 |
Conclusion
There are no differences in the Indications for Use, Intended Use, or Fundamental Technological Characteristics of the MI View&GO VA30 software as compared to the currently commercially available MI View&GO VA20 software (K222172).
Both the current and predicate devices are used for viewing, manipulation, quantification, analysis, and comparison of medical images from single or multiple imaging modalities with one or more time-points.
Additionally, the new features implemented within this release do not raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens' opinion that the MI View&GO software with the modifications outlined in this application—is substantially equivalent to the predicate device.