(29 days)
No
The summary describes standard medical image viewing, manipulation, quantification, analysis, and comparison software. There is no mention of AI, ML, or related terms, nor any description of training or test sets typically associated with AI/ML development.
No
The device is a diagnostic application for image viewing, manipulation, and analysis to aid in disease management and treatment planning, not for therapy itself.
Yes
Explanation: The "Intended Use / Indications for Use" section explicitly states, "MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images..." and "MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function." This clearly identifies its purpose as aiding in diagnosis.
Yes
The device description explicitly states "MI View&GO is a software-only medical device".
Based on the provided text, MI View&GO is NOT an IVD (In Vitro Diagnostic) device.
Here's why:
- IVD devices are used to examine specimens derived from the human body (like blood, urine, tissue) in vitro (outside the body). The purpose is to provide information about a physiological state, health, disease, or congenital abnormality.
- MI View&GO is a software application that processes and analyzes medical images obtained from imaging modalities like PET, NM, CT, and MR. These images are generated in vivo (within the body) and represent the internal structures and functions of the patient.
- The intended use and device description clearly state that MI View&GO is for viewing, manipulating, quantifying, analyzing, and comparing medical images. It aids healthcare professionals in managing diseases based on these images.
Therefore, MI View&GO falls under the category of medical imaging software or a picture archiving and communication system (PACS) component, not an IVD.
N/A
Intended Use / Indications for Use
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, neurology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
Product codes (comma separated list FDA assigned to the subject device)
LLZ
Device Description
MI View&GO is a software-only medical device which will be delivered with the new generation of Siemens SPECT/CT and PET/CT scanners.
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
positron emission tomography (PET), nuclear medicine (NM), computed tomography (CT), magnetic resonance (MR)
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
appropriately trained health care professionals
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
Siemens Medical Solutions USA, Inc. % Clayton Ginn Regulatory Technical Specialist 2501 N Barrington Road HOFFMAN ESTATES IL 60192
Re: K201202
Trade/Device Name: MI View&GO VA10A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 1, 2020 Received: May 4, 2020
Dear Clayton Ginn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
June 2, 2020
1
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
K201202
Device Name MIView&GO VA10A
Indications for Use (Describe)
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, neurology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
Form Approved: 0MB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
3
Section 05
510(k) Summary
as required by 21 CFR Part 807.87(h)
ldentification of the Submitter
Primary Contact: | Alternate Contact: | |
---|---|---|
Submitter: | Clayton Ginn | |
Regulatory Technical | ||
Specialist | ||
Siemens Medical Solutions | ||
USA, Inc. | ||
810 Innovation Drive | ||
Knoxville, TN 37932 | Veronica Padharia | |
Regulatory Technical | ||
Specialist | ||
Siemens Medical Solutions | ||
USA, Inc. | ||
2501 N. Barrington Road | ||
Hoffman Estates, IL 60192 | ||
Telephone Number: | (865) 898-2692 | (630) 877-5761 |
Fax Number: | (865) 218-3227 | (847) 304-6023 |
Name / Address of | ||
Manufacturer | Siemens Medical Solutions USA, Inc | |
Molecular Imaging | ||
2501 N. Barrington Road | ||
Hoffman Estates, IL 60192 | ||
USA | ||
Date of Submission: | May 1st, 2020 | |
Identification of the product | ||
Device Proprietary Name: | MI View&GO VA10A | |
Common Name: | Image Processing Software | |
Classification Name: | Picture Archiving and Communication System per 21 CFR | |
892.2050 | ||
Product Code: | LLZ | |
Classification Panel: | Radiology | |
Device Class: | Class II | |
Device Proprietary Name: | syngo.via MI Workflows VB40A | |
Common Name: | Image Processing Software | |
Classification Name: | Picture Archiving and Communication System per 21 CFR | |
892.2050 | ||
Product Code: | LLZ | |
Classification Panel: | Radiology | |
Device Class: | Class II | |
Manufacturer: | Siemens Medical Solutions | |
USA, inc. | ||
510(k) Number: | K191309 (July 2019) | |
Reference Predicate Device | ||
Device Proprietary Name: | syngo.CT View&GO VB20A | |
Common Name: | Image Processing Software | |
Classification Name: | Picture Archiving and Communication System per 21 CFR | |
892.2050 | ||
Product Code: | LLZ | |
Classification Panel: | Radiology | |
Device Class: | Class II | |
Manufacturer: | Siemens Medical Solutions | |
USA, Inc. | ||
510(k) Number: | K170952 (April 2017) | |
Reference Predicate Device | ||
Device Proprietary Name: | syngo.CT Extended Functionality VB40A | |
Common Name: | Image Processing Software | |
Classification Name: | System, X-ray, Tomography, Computed per 21 CFR | |
Product Code: | JAK | |
Classification Panel: | Radiology | |
Device Class: | Class II | |
Manufacturer: | Siemens Medical Solutions | |
USA, Inc. | ||
510(k) Number: | K192402 (September 2019) |
4
Primary Predicate Device
Classification Name:
892.1750
5
syngo.via MI Workflows is deemed the primary predicate device due to it being the most similar to the device under review of this submission with respect to indications for use and technical characteristics. syngo.CT View&GO and syngo.CT Extended Functionality are considered reference predicate devices because MI View&GO has integrated technical characteristics initially cleared within these devices.
6
Device Description
MI View&GO is a software-only medical device which will be delivered with the new generation of Siemens SPECT/CT and PET/CT scanners.
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function.
| Feature | New
Functionality | Modification Description |
|------------------------------------------------------------------------------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Auto Viewing | Yes | Stream data into the application immediately
following acquisition. |
| Layout Gallery with
Dynamic Summing,
Automatic Layouts, and
Remove Images | Yes | Generate Automatic Layouts at runtime,
Automatically sum frames for review in the
layout gallery, and remove images from a
display |
| Configurable User Presets | Yes | Configuration UI for preferred presets and
layouts |
| Tool Gallery | No | Functionality Cleared in K170952 |
| Organ Processing | No | Functionality Cleared in K191309 |
| MI Cardiology Reading and
Third Party Software
(Cedars 2017, Corridor
4DM 2018) | Yes | Restart applications and filter the cardiology
data passed to an application. |
| Series Panel | No | Functionality Cleared in K170952, K191309 |
| Basic Operation in the
Viewer | Yes | Correct slice positions, panning, rotation on
synchronized segments |
| Automatic Reconstruction,
Auto Ranges, and Cardiac
Auto Ranges | Yes | Generate parallel ranges with Auto Ranges and
generate a series of images in cardiac planes
Cardiac Auto Ranges |
| Display Types and Image
Fusion | No | Functionality Cleared in K170952, K191309 |
| 3D Visualization and
Orientation | No | Functionality Cleared in K170952, K191309 |
| Restricting Volumes and
Cardiac Masking | Yes | Mask the heart and exclude background regions
that can have an impact on statistical
calculations in an examination with Cardiac
Masking. |
| Image Optimization and
Preparation | No | Functionality Cleared in K170952, K191309 |
| Toggle Image Overlay | Yes | Configure customized image text |
| Spine and Rib Labeling | No | Functionality Cleared in K191309 |
MI View&GO includes the following features:
7
Neuro DSA | No | Functionality Cleared in K192402 |
---|---|---|
Osteo Extension | No | Functionality Cleared in K192402 |
Oncology Extension | No | Functionality Cleared in K192402 |
Vessel Extension | No | Functionality Cleared in K192402 |
ROI HU Threshold | No | Functionality Cleared in K192402 |
Dual Energy ROI | No | Functionality Cleared in K192402 |
Endoscopic View | No | Functionality Cleared in K192402 |
Printing of Images | No | Functionality Cleared in K170952 |
Technological Characteristics
The MI View&GO software modifications are based on the commercially available syngo.via MI Workflows software (K191309), syngo.CT View&GO software (K170952), and syngo.CT Extended Functionality (K192402). The features into this software device had no impact on the technological characteristics already present in the commercially available predicate system.
Intended Use
An individual software program, or group of programs, routines, or algorithms that add specific image processing and/or analysis capabilities to a positron emission tomography (PET) and Single Photon Emission Computed Tomography (SPECT) imaging system configuration. A basic set of application programs and routines is included with such computer controlled imaging systems and they can be upgraded to correct programming errors or to add new system capabilities. Some application software routines or groups of routines (packages) must be combined with specific hardware or firmware accessories or configurations in order to function as intended. Application program packages are typically identified by a proprietary name and "version" or "upgrade" number.
The intended use for MI View&GO compared to the primary and reference device has not changed.
Indications for Use
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
8
Performance Testing / Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management has been ensured via risk analyses in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards for development including EN ISO 13485 and IEC 62304.
Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the software and external devices.
Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
The device is designed and manufactured in accordance with Quality System Regulations as outlined in 21 CFR 820. The FDA recognized standards are listed as follows:
- . Recognition Number 13-79: IEC 62304 Edition 1.1 2015-06
- . Recognition Number 12-300: NEMA PS 3.1 – 3.20 (2016)
- Recognition Number 5-40: ISO 14971:2007 Second Edition ●
- . Recognition Number 5-114: IEC 62366-1 Edition 1.0 2015
- . Recognition Number 5-117: ISO 15223-1 Third Edition 2016
Statement Regarding Substantial Equivalence:
There are no differences in the Intended Use or Fundamental Technological Characteristics of the MI View&GO software as compared to the currently commercially available syngo.via MI Workflows software (K191309). The Indications for Use for MI View&GO is a condensed version of the Indications for Use of syngo.via MI Workflows and does not expand the intended use from that of the predicate device. Both the current and predicate devices are used for viewing, manipulation, 3D- visualization and comparison of medical images from multiple imaging modalities and/or multiple time-points
Additionally, there have been no changes that raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens opinion that the MI View&GO software-with the modifications outlined in this application-is substantially equivalent to the predicate device.