K Number
K241851

Validate with FDA (Live)

Date Cleared
2024-11-25

(151 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Versus™ Catheter is indicated for:

The controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary artery vasculature in adult patients.

The assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring.

Device Description

The Versus™ Pulmonary Artery Catheter is a dual-tip infusion catheter. Two lumens access and deliver physician-specified fluids, including thrombolytics, into the pulmonary arteries of each lung via a single access site. The Secondary Catheter tip telescopes independently to facilitate infusion into the contralateral lung. The device may allow for assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Pressure is measured through an interface between the catheter's fluid lumen and an externally located, FDA-cleared, pressure transducer. The device may include an additional lumen that is in communication with a distal flow-directed balloon.

AI/ML Overview

The provided text does not contain information about the acceptance criteria or a study proving that a device meets acceptance criteria related to AI/algorithm performance.

Instead, the document is a 510(k) summary for a medical catheter (Versus™ Catheter (VS110-8B)), comparing it to predicate devices to establish substantial equivalence.

The performance testing section (Section VII) outlines several non-clinical tests performed, including:

  • Sterilization and Shelf Life: Validated per ISO 11135:2014, with a sterility assurance level of 10^-6 and a shelf life of 3 months established through accelerated aging studies.
  • Biocompatibility: Evaluated against ISO 10993 and FDA Draft Guidance, including tests for acute systemic toxicity, complement activation, cytotoxicity, hemolysis, heparinized blood platelet and leukocyte count, intracutaneous irritation, material-mediated pyrogenicity, maximization sensitization, partial thromboplastin time (PTT), and particulate infrared spectroscopy.
  • Human Factors Testing (Usability): Conducted per IEC 62366-1:2015 and FDA Guidance, demonstrating that intended users could perform critical tasks in a simulated environment. The results met predefined acceptance criteria, but the specific criteria and performance metrics are not detailed in this document.
  • Performance Testing - Bench: Included tests for kink radius, trackability, advancement force, retraction force, infusion uniformity, pressure lumen measurement, dimensional and visual verification, luer hub compatibility, balloon inflation and integrity, corrosion resistance, and radiopacity.
  • Performance Testing – Animal: A GLP porcine study comparing the Versus™ Catheter to the primary predicate device. The study met all success criteria established.

Crucially, there is no mention of:

  • AI/algorithm performance: The device is a physical catheter, not an AI software.
  • Test set sample size and data provenance for AI: Not applicable as it's not an AI device.
  • Number of experts and their qualifications for ground truth: Not applicable for device testing of this nature.
  • Adjudication method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not applicable.
  • Standalone (algorithm only) performance: Not applicable.
  • Type of ground truth (e.g., expert consensus, pathology, outcomes data): Not applicable in the context of AI. The ground truth for biocompatibility and bench testing is established through standardized testing methodologies and material properties. For the animal study, the ground truth would be the physiological and pathological outcomes observed in the animals.
  • Training set sample size and how ground truth was established for the training set: Not applicable as it's not an AI/machine learning device.

Therefore, I cannot populate the table or answer the questions related to AI/algorithm performance as the provided text pertains to a physical medical device (catheter) and its substantial equivalence testing, not an AI or software as a medical device.

The only "acceptance criteria" mentioned are that the human factors testing and animal study "met the predefined acceptance/success criteria," but the specific numerical or qualitative details of those criteria are not provided in this summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 25, 2024

Liquet Medical Inc. Carrie Kuehn Regulatory Affairs Consultant Evergreen Strategic Consulting 108 N Rolling RD Catonsville, Maryland 21228

Re: K241851

Trade/Device Name: Versus™ Catheter (VS110-8B) Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: OEY, KRA, DOO, DYG Dated: June 20, 2024 Received: June 27, 2024

Dear Carrie Kuehn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ariel G. Ash- Digitally signed by
shakoor -S Ariel G. Ash-shakoor -S
Date: 2024.11.25
09:26:01 -05'00'

For

Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241851

Device Name Versus™ Catheter (VS110-8B)

Indications for Use (Describe) The Versus™ Catheter is indicated for:

The controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary artery vasculature in adult patients.

The assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY - K241581

VERSUS™ CATHETER (VS110-8B)

l. SUBMITTER'S NAME

Liquet Medical Inc. 5619 Country Hills Ln Glen Allen VA 23059, United States

Contact Name: Carrie M. Kuehn, Head of Regulatory Email: c.kuehn@liquetmedical.com Phone: 301-337-8159 Date prepared: November 21, 2024

II. Device Information

Proprietary Name: Versus™ Catheter (VS110-8B) Common Name: Embolectomy Catheter Classification Name: Mechanical Thrombolysis Catheter Regulation Number: 870.5150 Classification Codes: QEY Associated Product Codes: KRA, DQO, DYG

III. Predicate Devices

Primary Predicate Device: Bashir N-X Endovascular Catheter (K183290) Secondary Predicate Device: Swan Ganz Catheters (K160084)

IV. Device Description

The Versus™ Pulmonary Artery Catheter is a dual-tip infusion catheter. Two lumens access and deliver physician-specified fluids, including thrombolytics, into the pulmonary arteries of each lung via a single access site. The Secondary Catheter tip telescopes independently to facilitate infusion into the contralateral lung. The device may allow for assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Pressure is measured through an interface between the catheter's fluid lumen and an externally located, FDA-cleared, pressure transducer. The device may include an additional lumen that is in communication with a distal flow-directed balloon.

V. Indications for Use

The Versus™ Catheter is indicated for:

The controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary artery vasculature in adult patients.

{5}------------------------------------------------

The assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring.

VI. Comparison of Technological Characteristics with the Predicate Devices

Table 1. Comparison of Subject Device to Predicate Devices
------------------------------------------------------------------
CategorySubject Device –Predicate Device –Predicate Device –Comparison
Versus™ Catheter (VS110-8B)Bashir Endovascular Catheter ModelSwan-Ganz Catheters
K2418517201, Bashir N-X EndovascularK160084
Catheter, Mode 7200
K183290
ClassIIIIIISame
Product CodeQEY - Mechanical ThrombolysisQEY - Mechanical ThrombolysisDYG - Catheter, Flow DirectedSimilar
CatheterCatheterDQE - Catheter, Oximeter, FiberopticA combination of
KRA - Catheter, Continuous FlushKRA - Catheter, Continuous FlushDQO - Catheter, Intravascular,the predicates'
DQO - Catheter, IntravascularDiagnosticproduct codes
DiagnosticKRA - Catheter, Continuous Flush
DYG - Catheter, Flow Directed
DeviceThe Versus™ Pulmonary ArteryThe Bashir N-X Endovascular CatheterThe Swan-Ganz thermodilutionSimilar
DescriptionCatheter is a dual-tip infusion catheter(Ref. No. 7200) is a device intended forcatheters provide diagnostic
Two lumens access and deliverthe localized infusion of physician-information to rapidly determine
physician-specified fluids, includingspecified fluids, into the peripheralhemodynamic pressures and cardiac
thrombolytics, into the pulmonaryvasculature, including the pulmonaryoutput when used with a compatible
arteries of each lung via a single accessarteries. The distal infusion segment ofcardiac output computer.
site. The Secondary Catheter tipthe device is 12.50 cm (4.94 in) longIn addition to the standard distal
telescopes independently to facilitateand consists of a segment with mini-(pulmonary artery) and injectate
infusion into the contralateral lung.infusion catheters, each with multiplelumens, the Swan-Ganz VIP
The device may allow for assessmentinfusion holes. It is used for thethermodilution catheter (Models
of a patient's hemodynamic conditiondelivery of the physician-specified831F75, 831F75P and 831VF75P) has
through direct intracardiac andfluids at multiple cross-sectional pointsan additional lumen that provides
pulmonary artery pressure monitoring.of the target vessel location. Thedirect access to the right atrium. The
Pressure is measured through aninfusion line connector is located onSwan-Ganz VIP+ tri-lumen infusion
interface between the catheter's fluidthe handle.thermodilution catheter (Model
lumen and an externally located, FDA-The catheter is advanced over a834F75) is equipped with a right atrium
cleared, pressure transducer. Theguidewire using standard endovascularlumen and an additional lumen. The
device may include an additionalinterventional techniques and iscatheter's right ventricular (RV) lumen
lumen that is in communication with acompatible with standard infusionterminates 19 cm from the catheter tip
distal flow-directed balloon.connectors, accessories andand the right atrial (RA) lumen
equipment.terminates at 31 cm. The VIP lumen
allows for continuous infusion, even
during cardiac output determinations.
Intended UseThe Versus™ Catheter is intended forThe Bashir N-X Endovascular CatheterSwan-Ganz thermodilution cathetersSame
the controlled and selective infusion ofis intended for the controlled andare indicated for the assessment of aThe subject device
physician-specified fluids, includingselective infusion of physician-specifiedpatient's hemodynamic conditionincludes one
thrombolytics, into the pulmonaryfluids into the peripheral andthrough direct intracardiac andintended use from
artery vasculature in adult patients.pulmonary artery vasculature.pulmonary artery pressure monitoring,the Primary
The Versus™ Catheter is also intendedcardiac output determination, and forPredicate and one
for the assessment of a patient'sinfusing solutions.intended use from
hemodynamic condition through directThe distal (pulmonary artery) port alsothe Secondary
intracardiac and pulmonary arteryallows sampling of mixed venous bloodPredicate.
pressure monitoring.for the assessment of oxygen transport
balance and the calculation of derived
parameters such as oxygen
consumption, oxygen utilization
coefficient, and intrapulmonary shunt
fraction.
Indications forThe Versus™ Catheter is indicated for:The Bashir N-X Endovascular CatheterSwan-Ganz thermodilution cathetersSimilar
Useis intended for the controlled andare indicated for the assessment of a
The controlled and selectiveinfusion of physician-specified fluids, includingthrombolytics, into thepulmonary arteryvasculature in adultpatients. The assessment of apatient's hemodynamiccondition through directintracardiac and pulmonaryartery pressure monitoring.selective infusion of physician-specified fluids into the peripheral andpulmonary artery vasculature.patient's hemodynamic conditionthrough direct intracardiac andpulmonary artery pressure monitoring,cardiac output determination, and forinfusing solutions.The distal (pulmonary artery) port alsoallows sampling of mixed venous bloodfor the assessment of oxygen transportbalance and the calculation of derivedparameters such as oxygenconsumption, oxygen utilizationcoefficient, and intrapulmonary shuntfraction.The indications foruse contains anindication fromeach predicatedevice.
Anatomical site(s)Insertion through venous access topulmonary arteriesInsertion through venous access topulmonary arteries or peripheralvasculatureInsertion through venous access topulmonary arteriesSame
Principles ofOperationThe Versus™ Catheter employs duallumens to facilitate the delivery oflocalized infusion of physician-specifiedfluids into the pulmonary arteries. Thedistal infusion segment consists of asegment with multiple infusion holes. Itis used for the delivery of thephysician-specified fluids at multiplecross-sectional points of the targetvessel location. The infusion lineconnector is located on the proximalhub.The catheter is advanced over aguidewire using standard endovascularinterventional techniques and iscompatible with standard infusionconnectors, accessories andequipment.Alternatively, the device may includean inflatable balloon at the tip, whichfacilitates its placement into thepulmonary artery through the flow ofblood.The Versus™ Catheter has a proximalluer that can be connected to astandard pressure transducer in linewith an IBP-compatible patientmonitor for direct intracardiac andpulmonary artery pressure monitoring.Localized infusion of physician-specified fluids, into the peripheralvasculature, including the pulmonaryarteries. The distal infusion segmentconsists multiple infusion holes. It isused for the delivery of the physician-specified fluids at multiple cross-sectional points of the target vessellocation. The infusion line connector islocated on the handle.The catheter is advanced over aguidewire using standard endovascularinterventional techniques and iscompatible with standard infusionconnectors, accessories andequipment.The Swan-Ganz catheters are wellknown pulmonary artery cathetersintended for use on critical carepatients. A Swan-Ganz Catheterincludes an inflatable balloon at the tip,which facilitates its placement into thepulmonary artery through the flow ofblood.The Swan-Ganz catheters can be usedwith compatible cardiac output patientmonitors and/or with oximetrymodules (depending on the modelnumber) to transmit signals forhemodynamic pressure monitoring,cardiac output measurements, and/oroximetry measurements. The Swan-Ganz catheters are to be used with theEdwards and/or Edwards' compatiblepatient monitors, such as the VigilanceII.SameGuidewireplacement is thesame as PrimaryPredicate.Balloon placementand pressuremonitoring is thesame as SecondaryPredicate.
Pressuremeasurementlocation(s)Image: [heart diagram with balloon]Balloon(optional)RALAPALVRVPRESSUREProximal lumen port has the capabilityfor direct blood pressure measurementat ~20cm from distal tipImage: [heart diagram]RALAPALVRVPRESSUREDistal lumen port has the capability fordirect blood pressure measurement atthe distal tipImage: [heart diagram]BalloonRALAPALVRVPRESSUREProximal lumen port has the capabilityfor direct blood pressure measurementat ~30cm from the catheter tip, andalso at the distal tipSimilar
PressuremmHg intracardiac or pulmonarymmHg distal pulmonary arterymmHg intracardiac proximal lumenSame as Primary
measurementartery pressure when connected to anpressure when connected to an in-lineport, distal pulmonary artery pressure,Predicate. Similar
outputsin-line pressure transducer external topressure transducer external to theCardiac Output (CO), Stroke Volumeto Secondary
the patientpatient(SV)Predicate
(Subject device
does not include a
thermistor to
output CO, SV
parameters)
Energy UsedThe subject device does not use orThis predicate does not use or deliverThermal energy emitted by the thermalSame as Primary
and/or Delivereddeliver any energy.any energy.filament located on the catheter.Predicate.
BiocompatibilityBoth ISO 10993 and FDA DraftBoth ISO 10993 and FDA DraftNot reported.Same as Primary
Guidance "Use of InternationalGuidance "Use of InternationalPredicate.
Standard ISO 10993, BiologicalStandard ISO 10993, Biological
Evaluation of Medical Devices Part 1:Evaluation of Medical Devices Part 1:
Evaluation and Testing within a RiskEvaluation and Testing within a Risk
Management Process"Management Process"
Working Length100 – 107 cm92.50 cm75 – 110 cmSimilar
Outer Diameter8 French7 French7 – 7.5 FrenchSimilar
RadiopaqueRadiopaque marker bands at distal andThe distal 12.50 cm (4.94 in) infusionNot reported.Similar
Markerproximal ends of infusion segments,segment is radiopaque and visible
and at bifurcation.under fluoroscopy along its full length.
GuidewireA 0.035" guidewire for placement ofA 0.035" guidewire, of the requiredA 0.025" guidewire for placementSame as Primary
Compatibilitythe Primary Catheter. Guidewire oflength to fit the sheath. Guidewire ofPredicate
0.014" for placement of the Secondary0.018" for placement.
Catheter.
TreatmentInfusion segment length of 12cm forInfusion segment length of 12.50 cm.Distal tip infusion, some catalogSimilar to Primary
zone/infusionboth Primary and Secondarynumbers include 3 or 4 infusionPredicate.
portCatheters.lumens.
Infusion flow rateSet the infusion pump to the desiredSet the infusion pump to the desiredNot relevantSame as Primary
infusion rate and activate per standardinfusion rate and activate per standardPredicate.
practice.practice.
Diameter of13 mm ±2 mmNot applicable. Device does not have a13 mm.Same as Secondary
inflated balloonballoon.Predicate.
SterilityThe devices are sterilized usingThe devices are sterilized usingSterile (EtO)Same as Primary
standard methods and the sterilizationstandard methods and the sterilizationand Secondary
cycles have been validated followingcycles have been validated followingPredicates.
international standards. Theinternational standards. The
sterilization validation complied withsterilization validation complied with
ISO 11135:2014. The sterilization cycleISO 11135:2014. The sterilization cycle
was validated to a sterility assurancewas validated to a sterility assurance
Single Uselevel of 10-6Yeslevel of 10-6.YesYesSame
Environment foruseHealthcare facility/hospitalHealthcare facility/hospitalHealthcare facility/hospitalSame as Primaryand Secondary
Predicates.
Electrical SafetyContains no electrical componentsContains no electrical componentsWarning: Compliance to IEC 60601-1 isSame as Primary
only maintained when the catheter orPredicate.
probe (CF applied part, defibrillation
proof) is connected to a patient
monitor or equipment that has a Type
CF defibrillation proof rated input
connector. If attempting to use a third
party monitor or equipment, check
with the monitor or equipment
manufacturer to ensure IEC 60601-1
compliance and compatibility with the
catheter or probe. Failure to ensure
monitor or equipment compliance to
IEC 60601-1 and
catheter or probe compatibility mayincrease the risk of electrical shock tothe patient/operator.

{6}------------------------------------------------

510(k) Summary – 245181

Versus™ Catheter (VS110-8B)

{7}------------------------------------------------

510(k) Summary – 245181

Versus™ Catheter (VS110-8B)

{8}------------------------------------------------

VII. Summary of Performance Testing (Non-Clinical Testing)

This 510(k) notification provides performance data to establish the substantial equivalence of the Liquet Medical Versus™ Catheter to the Primary and Secondary Predicate devices. The following is a summary of the performance data.

Sterilization and Shelf Life: The device is sterilized using standard methods and the sterilization cycles have been validated following international standards. The shelf life of the devices has been established through stability studies. The sterilization validation complied with ISO 11135:2014. The sterilization cycle was validated to a sterility assurance level of 106. The shelf life of the Versus™ Catheter was established through accelerated aging studies and support a shelf life of 3 months.

Biocompatibility: Biocompatibility evaluation has been performed to show the finished, sterilized device is biocompatible and suitable for its intended use. Both ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" were taken into account to evaluate the biocompatibility of the subject device. The testing included an evaluation of:

  • Acute Systemic Toxicity
  • Complement Activation Assay
  • Cytotoxicity
  • Hemolysis
  • Heparinized Blood Platelet and Leukocyte Count Assay
  • Intracutaneous Irritation
  • Material Mediated Pyrogenicity
  • . Maximization Sensitization Test
  • Partial Thromboplastin Time (PTT) Assay
  • Particulate Infrared Spectroscopy

Human Factors Testing: Usability testing was conducted per IEC 62366-1:2015 and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices". The study performed, using the to-be marketed device with intended users, demonstrated that the users were able to perform novel critical tasks of the Versus™ Catheter in an anatomical model in a simulated environment. The results of this testing met the predefined acceptance criteria.

Performance Testing- Bench: Performance testing was performed to characterize the Versus™ Catheter. The testing included an evaluation of:

  • . Kink radius

{9}------------------------------------------------

Versus™ Catheter (VS110-8B)

  • Trackability
  • Advancement force
  • Retraction force
  • Infusion uniformity
  • Pressure lumen measurement
  • Dimensional and visual verification
  • Luer hub compatibility
  • Balloon inflation and integrity
  • Corrosion resistance
  • Radiopacity
  • Joint tensile strengths

The performance data demonstrate that the new devices are substantially equivalent to the Primary and Secondary Predicate devices.

Performance Testing – Animal: The Versus™ Catheter was evaluated in a GLP porcine study with comparison to the primary predicate device. All success criteria established for the study were met. Use of the catheter had no adverse effects systemically or pathologically supporting substantially equivalent safety and performance outcomes of the Versus™ Catheter when used in the target vasculature.

Determination of Substantial Equivalence to Predicate Devices VIII.

The information presented in the 510(k) Submission demonstrates that the Liquet Medical Versus™ Catheter (VS110-8B) is substantially equivalent to the Primary and Secondary Predicate devices.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).