K Number
K241823
Date Cleared
2024-08-30

(67 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ToxiSeal™ Vial Adaptor mechanically prohibits environments from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The device also prevents the introduction of microbial contaminations into the drug or fluid path for up to 168 hours (or 7 days) when used as intended.

Device Description

The ToxiSeal™ Vial Adaptor devices are single-use, sterile, non-pyrogenic CSTD drug vial adaptors that are fitted to the drug vials and are sealed against the closures of the vials. They are used as sterile interfaces between the drug vials and the ProSeal™ Injector or the ProSeal™ Injector Plus (both are syringe adaptors) for the injection of diluents into the drug vials and/or withdrawal of liquids from the vials. Changes proposed in this Special 510(k) Submission are as follow: Removed external balloon and added activated carbon filter.

AI/ML Overview

The provided document is a 510(k) summary for the ToxiSeal™ Vial Adaptor (K241823). It describes the device, its intended use, a comparison to a predicate device, and performance data supporting its substantial equivalence. However, the document does not contain information about an AI/software device or a comparative effectiveness study involving human readers with and without AI assistance.

Therefore, I cannot provide information for points 5, 8, and 9 of your request as they are not present in the document.

Based on the available information regarding the medical device itself (ToxiSeal™ Vial Adaptor), here's the information related to acceptance criteria and the studies conducted:

The ToxiSeal™ Vial Adaptor is a physical medical device, not an AI/software product. Therefore, the "acceptance criteria" and "study that proves the device meets acceptance criteria" are related to its functional performance, biocompatibility, and sterility, rather than AI performance metrics.

Acceptance Criteria and Device Performance (for a physical device)

The document outlines the various tests performed to demonstrate the device's conformance to recognized standards. The "acceptance criteria" are implied by the conformance to these standards and the findings that the device "met the acceptance criteria therein" or "did not raise any new or different questions of safety or effectiveness."

Here's a table summarizing the acceptance criteria (standards/tests) and the reported device performance.

CategoryAcceptance Criteria (Standards/Tests)Reported Device Performance
Functional Performance- ISO 8536-2:2010 (Infusion equipment for medical use – Part 2: Closures for infusion bottles) - Evaluated to be in conformance.
- ISO 8536-4:2019 (Infusion equipment for medical use Part 4: Infusion sets for single use, gravity feed) - Evaluated to be in conformance.
- ISO 8871-5:2016 (Elastomeric parts for parenterals and for devices for pharmaceutical use Part 5: Functional requirements and testing) - Evaluated to be in conformance.
- ISO 22413:2021 (Transfer sets for pharmaceutical preparations – Requirements and test methods) - Evaluated to be in conformance.
- Leak integrity testing (air- and liquid-tightness) per ISO 8536-4:2019, paragraph 7.2 and Annex A.3 - Performed successfully on both Subject and predicate devices; differences in design (activated carbon filter, no external balloon) "did not raise any new or different questions of safety or effectiveness."
- Fragmentation study to ISO 22413 & ISO 8871-5 - Performed successfully on predicate device (K241476), implying substantial equivalence for the subject device.
- Vial Adaptor penetration force testing to ISO 22413 - Performed successfully on predicate device (K241476), implying substantial equivalence for the subject device.
- Tests for leakages to ISO 8536-4:2019, Annex A.3.3 - Performed successfully on the Subject device; differences in design "did not raise any new or different questions of safety or effectiveness."
- Testing to (draft) NIOSH CSTD Test Protocol and NIOSH 2016, Performance Test Protocol for Closed System Transfer Devices Used During Pharmacy Compounding and Administration of Hazardous Drugs - Performed successfully on the Subject device and a previously cleared device (K222929); differences in design "did not raise any new or different questions of safety or effectiveness."
- Microbial ingress/ microbiological integrity testing - Conducted successfully on a previously cleared device (K222929), indicating prevention of microbial contamination for up to 168 hours (7 days).
Biocompatibility- ISO 10993-1: 2018 (Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process) - Classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). - Referencing previous successful testing on existing devices of the ProSeal™ CSTD system (K192075, K192075/S001, K222929) for: Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Intracutaneous Reactivity (ISO 10993-10), Acute Systemic Toxicity (ISO 10993-11), 14-day Subacute/ Subchronic Acute Systemic Toxicity (ISO 10993-11), In-vitro Hemolysis Assessment (ISO 10993-4), Material Mediated Pyrogenicity (ISO 10993-11), Chemical Characterization and Toxicological Risk Management (ISO 10993-18 and ISO 10993-17). The subject device is deemed biocompatible based on these references.
- USP <788> Particulate Matter in Injections - Testing conducted on devices under K192075 and K222929, and "found to have met the acceptance criteria therein." This performance is leveraged for the subject device.
Sterility & Shelf-Life- ISO 11135:2014 (Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices) - "Comply with sterilization requirements."
- Package Integrity Tests per ASTM F1980-21 (Standard guide for accelerated aging of sterile barrier systems for medical devices), ASTM F88/F88M-21 (Seal strength), ASTM F1929-23 (Dye Penetration), EN 868-5:2009 (Heat and self-sealable pouches and reels) - Performed on the proposed device (referencing K222929) and ensure package integrity.
- Pyrogen Tests per ANSI/AAMI ST72/2019, USP 40<151>, USP-NF <161>, USP-NF <85> - Conducted under K222929, with testing to be conducted on every lot.
- Validated shelf-life using ASTM 1980-21 - A shelf-life of 3 years (36 months) has been validated.

Now addressing the specific points of your request based on the provided document:

  1. A table of acceptance criteria and the reported device performance

    • See the table above. Note that for a physical medical device, "acceptance criteria" are typically defined by conformance to established performance standards and successful completion of specified tests.
  2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Sizes: The document does not specify the exact sample sizes used for each performance test (e.g., number of devices tested for leak integrity, biocompatibility, etc.). It mentions that tests were "performed on the Subject devices," "on devices under K241476," or "on device under K222929," implying samples were used from these device types.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given it's a 510(k) submission to the FDA, the data would typically be generated by the manufacturer or contract labs following international and US standards. The studies, being part of device verification and validation prior to market clearance, are generally prospective in nature for the tests conducted on the subject device.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable to this type of medical device submission. Ground truth established by experts (like radiologists for imaging AI) is relevant for diagnostic accuracy studies of software/AI. For a physical device like a vial adaptor, the "ground truth" is based on objective measurements against established engineering and biological standards. There are no human "experts" establishing a diagnostic ground truth here.
  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • This question is not applicable as there is no diagnostic test set or human interpretation being adjudicated. The tests are quantitative and involve laboratory measurements.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This document describes a physical medical device, not an AI software. No MRMC study was conducted or is relevant for this device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm/AI device. The device's performance is standalone in the sense that its mechanical and biological properties are tested independently.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this medical device is the objective measurement of physical properties, chemical properties, and biological safety against predefined engineering, material, and biocompatibility standards. Examples include:
      • Meeting specific leakage rates.
      • Absence of fragmentation.
      • Demonstrating specified penetration force.
      • Absence of microbial ingress.
      • Absence of cytotoxicity, sensitization, systemic toxicity, pyrogenicity, and meeting particulate matter limits.
      • Maintaining sterility over shelf-life.
    • It's based on quantitative laboratory testing results compared to acceptance criteria defined by recognized standards (ISO, ASTM, USP, NIOSH).
  8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device; there is no "training set."
  9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI model, this question does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

August 30, 2024

Epic Medical Pte. Ltd. Freddie Lee CEO/md 105 Cecil Street #20-04, The Octagon Singapore, 069534 Singapore

Re: K241823

Trade/Device Name: ToxiSeal™ Vial Adaptor Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: June 24, 2024 Received: July 31, 2024

Dear Freddie Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Porsche Bennett
Porsche Bennett
for David Wolloscheck, Ph.D
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K241823

Device Name ToxiSeal™ Vial Adaptor

Indications for Use (Describe)

The ToxiSeal™ Vial Adaptor mechanically prohibits environments from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The device also prevents the introduction of microbial contaminations into the drug or fluid path for up to 168 hours (or 7 days) when used as intended.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

cp10

K241823 – 510(k) Summary

I. Submitter

Epic Medical Pte. Ltd. 105 Cecil Street #20-04, The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292

Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director Date Prepared: August 30, 2024 Content and Format: Prepared in accordance with 21 CFR 807.92 Type of Submission: Special

II. Subject Device

510(k) Number:K241823
Trade/ Device Name:ToxiSeal™ Vial Adaptor
Common/Usual Name:Closed Antineoplastic and Hazardous Drug Reconstitution andTransfer Device
Regulation Number:21 CFR 880.5440
Regulation Name:Intravascular administration set
Regulatory Class:Class: II
Classification Product Code:ONB

III. Predicate

510(k) Number:K241476
Trade/ Device Name:ToxiSeal™ Vial Adaptor with External Flip Balloon
Common/Usual Name:Closed Antineoplastic and Hazardous Drug Reconstitution andTransfer Device
Regulation Number:21 CFR 880.5440
Regulation Name:Intravascular administration set
Regulatory Class:Class: II
Classification Product Code:ONB

{4}------------------------------------------------

IV. Device Description

The purpose of this Special 510(k) Submission is to seek FDA clearance prior to the introduction in the US, of a modified design of a ToxiSeal™ Vial Adaptor - the Subject ToxiSeal™ Vial Adaptor. This device is an addition to the ProSeal™ CSTD system of devices (cleared K240433). The nominated Predicate device is the existing ToxiSeal™ Vial Adaptor with External Flip Balloon (cleared K241476).

The ToxiSeal™ Vial Adaptor devices are single-use, sterile, non-pyrogenic CSTD drug vial adaptors that are fitted to the drug vials and are sealed against the closures of the vials. They are used as sterile interfaces between the drug vials and the ProSeal™ Injector or the ProSeal™ Injector Plus (both are syringe adaptors) for the injection of diluents into the drug vials and/or withdrawal of liquids from the vials.

The ToxiSeal™ Vial Adaptor devices, designed to be used with the cleared ProSeal™ Injector or Injector Plus within the ProSeal™ CSTD system (K240433), are variants of CSTD vial adaptor component device offerings to the ProSeal™ CSTD system. The additions enhance the completeness of the portfolio of ProSeal™ CSTD devices system.

Changes proposed in this Special 510(k) Submission are as follow:

  • Removed external balloon and added activated carbon filter .

V. Indications for Use Statement

The ToxiSeal™ Vial Adaptor mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The device also prevents the introduction of microbial contaminations into the drug or fluid path for up to 168 hours (or 7 days) when used as intended.

VI. Comparison of Technological Characteristics

The Subject device and the Predicate device share the following key characteristics:

  • System components of the ProSeal™ CSTD .
  • Closed System Technology
  • Interface between system components ●
  • Membrane material ●
  • Hydrophobic filter ●
  • Intended Injectors ●
  • Sterile barrier packaging ●
  • Sterilization process ●
  • Shelf-life .
  • . Single-use

An overview table summarizing the comparison between the key characteristics between the Subject and the Predicate device is provided hereunder.

{5}------------------------------------------------

Side-by-side comparison of technology & design characteristics
CharacteristicscomparedPredicate Device (K241476)ToxiSeal™ Vial Adaptorwith External Flip BalloonSubject Device (K241823)ToxiSeal™ Vial AdaptorDetailed discussion andcomparative evaluation
Intended useandIndications for Usestatement"The ToxiSeal™ Vial Adaptor withExternal Flip Balloon mechanicallyprohibits environmental contaminantsfrom entering the system and the escapeof drug or vapor concentrations from thesystem, thereby minimizing individualand environmental exposure to drugvapor, aerosols, and spills. The devicealso prevents the introduction ofmicrobial contaminations into the drug orfluid path for up to 168 hours (or 7 days)when used as intended.""The ToxiSeal™ Vial Adaptormechanically prohibits environmentalcontaminants from entering the systemand the escape of drug or vaporconcentrations from the system, therebyminimizing individual and environmentalexposure to drug vapor, aerosols, andspills. The device also prevents theintroduction of microbial contaminationsinto the drug or fluid path for up to 168hours (or 7 days) when used as intended."Same
Intended userpopulationAdequately trainedhealth careprofessionals or pharmacistsAdequately trainedhealth careprofessionals or pharmacistsSame
Intended useenvironmentClinical settingClinical settingSame
Intended drug typeParenteral drugsParenteral drugsSame
Single use or reusableSingle use onlySingle use onlySame
Prescription use orover-the-counter useR onlyR onlySame
LabelingspecificationsMeets the requirementsspecified in 21 CFR 801Meets the requirementsspecified in 21 CFR 801Same
CharacteristicscomparedPredicate Device (K241476)ToxiSeal™ Vial Adaptorwith External Flip BalloonSubject Device (K241823)ToxiSeal™ Vial AdaptorDetailed discussion andcomparative evaluation
System componentdevices1. ProSeal™ Injector (model 421010)(K241476)2. ProSeal™ Injector Plus (model 421050)(K241476)3. ProSeal™ Connector (422010)(K241476)4. ProSeal™ Injection Site Extended MaleLuer Lock (K241476)5. ProSeal™ Vial Adaptor (3 models4200X0, for options of vial neck sizesand pressure equalizations) K241476)6. ToxiSeal™ Vial Adaptor with ExternalBalloon (K241476)7. ToxiSeal™ Vial Adaptor with ExternalFlip Balloon (K241476)8. eZSURE™ Empty Fluid Container withProSeal™ Injection Site (K241476)9. ProSeal™ Closed System AdministrationSet (K230343)10. ProSeal™ Assembly Fixture (Fixture forProSeal™ Vial Adaptor assembly,model 424010) (K241476)1. ProSeal™ Injector (model 421010)(K241476)2. ProSeal™ Injector Plus (model 421050)(K241476)3. ProSeal™ Connector (422010)(K241476)4. ProSeal™ Injection Site Extended MaleLuer Lock (K241476)5. ProSeal™ Vial Adaptor (3 models4200X0, for options of vial neck sizesand pressure equalizations) (K241476)6. ToxiSeal™ Vial Adaptor with ExternalBalloon (K241476)7. ToxiSeal™ Vial Adaptor with ExternalFlip Balloon (K241476)8. ToxiSeal™ Vial Adaptor (the Subjectdevice)9. eZSURE™ Empty Fluid Container withProSeal™ Injection Site (K241476)10. ProSeal™ Closed SystemAdministration Set (K230343)11. ProSeal™ Assembly Fixture (Fixture forProSeal™ Vial Adaptor assembly,model 424010) (K241476)Different,the Subject systemcomponent device (#8 inthe column SubjectDevice ), is an addition tothe existing ten (10)cleared componentdevices of ProSeal™CSTD (K241476) and theProSeal™ Closed SystemAdministration Set(K230343), indicated inbold face texts .Differences are addressedin Comments #1 and #2under this table.
Closed SystemTechnologyPhysical barrier to preventall drug massesfrom crossing the system boundaryPhysical barrier to preventall drug massesfrom crossing the system boundarySame
Interfacebetween systemcomponentsElastomericresealing double-membraneElastomericresealing double-membraneSame
Intended Injector foruse with the SubjectdeviceProSeal™ Injector (model 421010) orProSeal™ Injector Plus (model 421050)(cleared K241071 & K240433),both Male Luer Lock tipProSeal™ Injector (model 421010) orProSeal™ Injector Plus (model 421050)(cleared K241071 & K240433),both Male Luer Lock tipSame
CharacteristicscomparedPredicate Device (K241476)ToxiSeal™ Vial Adaptorwith External Flip BalloonSubject Device (K241823)ToxiSeal™ Vial AdaptorDetailed discussion andcomparative evaluation
Membrane materialIsoprene rubber (IR)Isoprene rubber (IR)Same
Hydrophobic filter(non-fluid path)Polytetrafluoroethylene (PTFE)Polytetrafluoroethylene (PTFE)Same
Activated carbonfilterNone (carbon filter was removed inToxiSeal™ Vial Adaptor with ExternalFlip Balloon (K241476))Activated carbonDifferent.See Comment #1
External balloon(non-fluid path)Polyethylene terephthalate -Polyethylene(PET/PE) laminateand Polyethylene (PE)NoneDifferent.See Comment #2
Primarypackage top webMedical grade paper andmedical plastic film, heat sealedMedical grade paper andmedical plastic film, heat sealedSame
Sterilization methodEthylene Oxide, EO, SAL 10-6Ethylene Oxide, EO, SAL 10-6Same
Validatedshelf life3 years (36 months)3 years (36 months)Same
Reuseor single-useSingle-use onlySingle-use onlySame

Side-by-side comparison of technology & design characteristics

{6}------------------------------------------------

Side-by-side comparison of technology & design characteristics

{7}------------------------------------------------

Side-by-side comparison of technology & design characteristics

Discussion of differences in technological characteristics Comment #1

An activated carbon filter, which was not predicate (K241476) is now included for the Subject device. The filter adsorbs drug vapors within the vial adaptor, ensuring safety for healthcare workers and minimizing contamination risk. This differed and addressed in the verification testing of the Subject device over the shell life, to the following standards: 1.3 and NIOSH 2016, Performance Test Protocol for Closed System Transfer Devices Used During Pharmacy Compounding and Administration of Hazardous Drugs, and found it did not raise any new or different questions of safety or effectiveness. Details are provided in section VII.

Comment #2

The Subject device does not include an external balloon - a pressure equalization mechanism. This difference of not having an external balloon together with the inclusion of an activated and addressed in the verfication tests described in Comment #1. Bench performance verification testing of the Subject device over the shelf life was conducted to the following standards: 180 8556-42019, subclause A.3.3 and NOSH 2016, Performance Test Protocol for Closed System Transfer Devices Used During Pharmacy Compounding and Administration of Hazardous Drugs. Testing found it did not raise any new or effectiveness. Further details are provided in section VII hereunder.

{8}------------------------------------------------

Performance Data Supporting Substantial Equivalence VII.

A. Functional Performance

The sterile, single-use, non-pyrogenic ToxiSeal™ Vial Adaptor devices described in this Summary were evaluated to be in conformance with the following ISO and FDA recognized standards:

  • . ISO 8536-2:2010. Infusion equipment for medical use – Part 2: Closures for infusion bottles
  • ISO 8536-4:2019, Infusion equipment for medical use Part 4: Infusion sets for single use, ● gravity feed
  • ISO 8871-5:2016, Elastomeric parts for parenterals and for devices for pharmaceutical use Part 5: Functional requirements and testing
  • ISO 22413:2021, Transfer sets for pharmaceutical preparations – Requirements and test methods

Bench performance verifications and validations referred to and performed on the Subject device as detailed hereunder:

  • Leak integrity testing (air- and liquid-tightness) per ISO 8536-4:2019, paragraph 7.2 and ● Annex A.3 performed on the Subject devices, and on devices under K241476
  • Fragmentation study to ISO 22413 & ISO 8871-5, performed on device under K241476
  • . Vial Adaptor penetration force testing to ISO 22413, on device under K241476
  • Tests for leakages to ISO 8536-4:2019, Annex A.3.3, performed on the Subject device ●
  • . Testing to (draft) NIOSH CSTD Test Protocol, performed on the Subject device, and on device under K222929
  • . Microbial ingress/ microbiological integrity testing conducted on device under K222929

B. Biocompatibility

In accordance with ISO 10993-1: 2018, the Subject devices, are classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The following testing are referred to, from testing on existing devices of the ProSeal™ CSTD system:

  • Cytotoxicity to ISO 10993-5 under K192075
  • Sensitization to ISO 10993-10 under K192075 .
  • Intracutaneous Reactivity to ISO 10993-10 under K192075 .
  • Acute Systemic Toxicity to ISO 10993-11 under K192075
  • . 14-day Subacute/ Subchronic Acute Systemic Toxicity to ISO 10993-11 under K192075 and under K222929
  • . In-vitro Hemolysis Assessment to ISO 10993-4 under K192075 and K192075/S001
  • Material Mediated Pyrogenicity to ISO 10993-11 under K192075 .
  • Chemical Characterization and Toxicological Risk Management to ISO 10993-18 and ISO 10993-17 under K192075/S001

Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections, and found to have met the acceptance criteria therein, on devices under K192075 and K222929.

{9}------------------------------------------------

C. Sterility, Shipping, and Shelf-Life

The Subject devices, comply with sterilization requirements of ISO 11135:2014, Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices and the following testing/ evaluations conducted under K222929:

  • . Package Integrity Tests per ASTM F1980-21, Standard guide for accelerated aging of sterile barrier systems for medical devices and Sterile Barrier Packaging Testing performed on the proposed device: Seal strength - ASTM F88/F88M-21, Standard test method for seal strength of flexible barrier materials; Dye Penetration – ASTM F1929-23, Standard test method for detecting seal leaks in porous medical device packaging by dye penetration; EN 868-5:2009, Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods
  • Pyrogen Tests per ANSI/AAMI ST72/2019, Bacterial endotoxins – Test methods, routing monitoring, and alternatives to batch testing, USP 40<151>, Pyrogen test (USP rabbit test), USP-NF <161>, Medical Devices-Bacterial Endotoxin and Pyrogen Tests, USP-NF <85>, Bacterial Endotoxins Test and testing will be conducted on every lot
  • Shelf-life of 3 years has been validated using the FDA recognized standard, ASTM 1980-21, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

VIII. Clinical Tests

Not applicable.

IX. Conclusion

The differences between the Predicate and the Subject device do not raise any new or different questions of safety or effectiveness. The Subject ToxiSeal™ Vial Adaptor is substantially equivalent to the Predicate, the ToxiSeal™ Vial Adaptor with External Flip Balloon, with respect to the indications for use, principles of operation and technological characteristics.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.