K Number
K241733
Date Cleared
2025-04-16

(303 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HEBE is intended to provide topical heating to elevate tissue temperature for the treatment of selected medical conditions, including pain relief, muscle spasm reduction, and increased local circulation.

Device Description

HEBE is a device indicated for radiofrequency (RF) treatments at 448kHz. This frequency is associated with pain reduction and improved functionality in patients with musculoskeletal disorders. This wavelength promotes ion mobilization between the intracellular and extracellular matrix and restores cell membrane permeability. The energy generated at 448 kHz improves cell membrane permeability, enhancing intracellular and extracellular exchange and tissue regeneration. Furthermore, high-intensity radiofrequency treatment at 448 kHz can induce significant hyperthermia, which can lead to a considerable increase in cell metabolism and accelerated recovery. These hyperthermal adaptations to radiofrequency treatment are based on vasodilatation, causing a targeted local increase of blood perfusion and decreases in muscle tension and spasm, increased oxygen and nutrient supply, and healing acceleration.
RF treatment can be applied through either capacitive or resistive electrodes that pass radiofrequency to the human body.
HEBE has 3 different technologies:

  • Monopolar Capacitive Radiofrequency
  • Monopolar Resistive Radiofrequency
  • Bipolar Resistive Radiofrequency
AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the HEBE device primarily focus on substantial equivalence to predicate devices through comparisons of features and performance specifications. It does not contain a detailed study demonstrating that the device meets specific acceptance criteria based on clinical outcomes or a rigorous statistical evaluation of its effectiveness.

The document highlights bench testing and compliance with safety and electrical standards (IEC 60601 series, IEC 62304) and biocompatibility standards (ISO 10993 series). This indicates the device's safety and fundamental operational aspects were verified, but not its clinical efficacy against pre-defined performance metrics in a clinical setting.

Therefore, many of the specific questions about acceptance criteria and clinical study details cannot be answered from the provided text, as the application relies on substantial equivalence and non-clinical performance data, rather than a direct clinical performance study with acceptance criteria for clinical outcomes.

Here's an attempt to answer the questions based only on the provided information, noting where information is absent:


Acceptance Criteria and Device Performance (Based on Provided Data)

The concept of "acceptance criteria" for clinical performance is not explicitly defined or measured in this document. Instead, the "performance specifications" are a comparison to predicate devices, indicating similarity rather than a proven threshold for efficacy.

1. A table of acceptance criteria and the reported device performance

Based on the provided information, there are no clinically-defined acceptance criteria with corresponding device performance data for efficacy. The "performance specifications" listed are comparative to the predicate and reference devices regarding technical parameters.

Feature (as "Performance Specification")Acceptance Criteria (Implicit: "Same as predicate/reference")Reported Device Performance (HEBE)
Treatment Temperature RangeImplicit: Capable of achieving temperatures similar to predicate/reference devices for therapeutic effect. The footnote indicates an ability to reach 40-45°C.+43 °C – +45 °C
FrequencyImplicit: Operate within a similar radiofrequency range for therapeutic effect.448 KHz
Output RF PowerImplicit: Capable of delivering comparable power output for therapeutic effect.Max 200 W
VoltageImplicit: Operates within standard electrical voltage range.100-240 V

Note on "Acceptance Criteria" for Performance Features: For a 510(k) submission based on substantial equivalence, the "acceptance criteria" for these technical performance specifications is typically that they are "the Same" or "Equivalent" to the predicate device(s) without raising new questions of safety or effectiveness. The document asserts that the HEBE device meets this implicitly by stating "Same" or providing values within a comparable range.


Study Details (Based on Provided Data)

The document does not describe a clinical study that proves the device meets specific acceptance criteria for clinical outcomes (e.g., pain relief, muscle spasm reduction). The performance data cited are related to bench testing for safety, electrical, and biocompatibility standards, and a comparison of technical specifications to predicate devices.

Therefore, for most of the following questions, the answer will be that the information is not provided in the clearance letter.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not provided. The document describes "Verification and validation were performed on the device using the established methods by using the predicate device" and lists standards for safety and electrical performance. This refers to bench testing, not a clinical human test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No clinical test set with human data requiring expert ground truth establishment is described. The "experts" would be engineers and testers performing bench tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No clinical test set requiring adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is an electrosurgical cutting and coagulation device for topical heating, not an AI-assisted diagnostic imaging device that would typically undergo an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical medical device (radiofrequency treatment), not an algorithm with standalone performance. Its "performance" relates to its physical output and safety.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the bench testing: The "ground truth" would be the specifications and requirements defined by the IEC and ISO standards (e.g., electrical safety limits, biocompatibility thresholds, specified temperature output).
  • For clinical efficacy: This information is not provided. The 510(k) relies on substantial equivalence and the known mechanism of action of similar devices for its intended use, rather than presenting new clinical outcomes data.

8. The sample size for the training set

  • Not applicable. This is not a machine learning or AI device that would have a "training set" in the conventional sense. The "training" for such devices is their design and manufacturing to meet the specified performance and safety standards.

9. How the ground truth for the training set was established

  • Not applicable. (See point 8.)

Summary of what the document does provide:

The 510(k) summary focuses heavily on:

  • Substantial Equivalence: Comparing the HEBE device's features, indications for use, principle of action, and technical specifications (frequency, power) to a primary predicate device (NuEra Tight RF Family) and a reference device (Deep Care). The conclusion is that it is "safe and effective and are substantially equivalent."
  • Bench Testing & Standard Compliance: The device underwent verification and validation tests in accordance with a range of established international standards for medical electrical equipment (IEC 60601 series, IEC 62304) and biocompatibility (ISO 10993 series). These tests verify the device's electrical safety, usability, electromagnetic compatibility, and material safety.

This type of submission is common for Class II devices where a new device is similar enough to existing legally marketed devices that extensive new clinical trials demonstrating efficacy are not required, provided new questions of safety or effectiveness are not raised.

FDA 510(k) Clearance Letter - K241733

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

April 16, 2025

Novasonix Technology S.L
℅ Shilpa Gampa
Senior Manager
Freyr Inc
150 College Rd W #102
Princeton, New Jersey 08540

Re: K241733
Trade/Device Name: Hebe (np0000763); Hera (np0000706)
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical cutting and coagulation device and accessories
Regulatory Class: Class II
Product Code: PBX, GEX
Dated: March 13, 2025
Received: March 13, 2025

Dear Shilpa Gampa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K241733 - Shilpa Gampa Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensiveregulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K241733 - Shilpa Gampa Page 3

Sincerely,

James H. Jang -S Digitally signed by James H. Jang -S Date: 2025.04.16 07:26:47 -04'00'

For
Long Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K241733

Device Name: HEBE

Indications for Use (Describe):
HEBE is intended to provide topical heating to elevate tissue temperature for the treatment of selected medical conditions, including pain relief, muscle spasm reduction, and increased local circulation

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Contact Details

510(k) NumberK241733
510(k) TypeTraditional
Applicant InformationNOVASONIX TECHNOLOGY S.Lc/ Pla de Ramasar 52, Pol. Ind, Carrer Pla del RamassarGranollers, Barcelona, 08402, Spain
ContactMr. Carlos Navarro+34 931 356 541cnavarro@novasonix.es
Data Prepared13-March-2025
Device Name(s)HEBE
Common NameMassager, Vacuum, Radio Frequency Induced Heat
Regulatory ClassClass II (21CFR 878.4400)
Product CodePBX
Regulation NamesElectrosurgical cutting and coagulation device and accessories

Predicate and Reference Devices

510(k) RefPro Code/Reg NoTrade NameApplicant
Primary Predicate Device
K210867PBX - Massager, Vacuum, Radio Frequency Induced HeatNuEra Tight RF FamilyBios s.r.l.
Reference Device
K161458PBX -Massager, Vacuum, Radio Frequency Induced HeatDeep CareIndiba USA Inc.

Device Description

HEBE is a device indicated for radiofrequency (RF) treatments at 448kHz. This frequency is associated with pain reduction and improved functionality in patients with musculoskeletal disorders. This wavelength promotes ion mobilization between the intracellular and extracellular matrix and restores cell membrane permeability. The energy generated at 448 kHz improves cell membrane permeability, enhancing intracellular and extracellular exchange and tissue regeneration. Furthermore, high-intensity radiofrequency treatment at 448 kHz can induce significant hyperthermia, which can lead to a considerable increase in cell metabolism and accelerated recovery. These hyperthermal adaptations to radiofrequency treatment are based on vasodilatation, causing a targeted local increase of blood perfusion and decreases in muscle tension and spasm, increased oxygen and nutrient supply, and healing acceleration.

Page 6

510(k) Number K241733

RF treatment can be applied through either capacitive or resistive electrodes that pass radiofrequency to the human body.

HEBE has 3 different technologies:

  • Monopolar Capacitive Radiofrequency
  • Monopolar Resistive Radiofrequency
  • Bipolar Resistive Radiofrequency

Indications of Use

HEBE is intended to provide topical heating to elevate tissue temperature for the treatment of selected medical conditions, including pain relief, muscle spasm reduction, and increased local circulation.

Page 7

510(k) Number K241733

Predicate Device Comparison

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
1Device NameHEBENuEra Tight RF FamilyDeep CareN/A
2Device ManufacturerNovasonix Technology S.L.Bios s.r.l.Indiba USA Inc.N/A
3510(K) NumberK210867K161458N/A
4Product CodePBX - Massager, Vacuum, Radio Frequency Induced HeatPBX - Massager, Vacuum, Radio Frequency Induced HeatPBX -Massager, Vacuum, Radio Frequency Induced HeatSame
5Regulation21 CFR 878.440021 CFR 878.440021 CFR 878.4400Same

Page 8

510(k) Number K241733

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
6Indications for UseHEBE medical device is intended to provide topical heating to elevate tissue temperature for the treatment of selected medical conditions, including pain relief, muscle spasm reduction, and increased local circulation.The NuEra Tight RF Family is intended:To provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as temporary relief of pain, muscle spasms, and increase local circulation; To provide, with a massage device, a temporary reduction in the appearance of cellulite.The Indiba Diathermia Radiofrequency Devices are intended to provide topical heating for the purpose of elevating tissues temperature for treatment of selected medical conditions such as relief of pain, muscle spasms, increase in local Circulation.The Massage device provided is intended to provide a temporary reduction in the appearance of cellulite.Same

Page 9

510(k) Number K241733

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
7Principle of actionRadio frequency waves penetrate in exposed tissue and produce heat, increasing the temperature on a certain part of the body for therapeutic purposes.Electromagnetic waves penetrate in exposed tissue and produce heat, increasing the temperature on a certain part of the body for therapeutic purposes.The RF energy generates a heating profile that produces a moderate temperature rise in the subcutaneous tissue.The temperature on the skin is measured using a separate IR (Infra-red) thermometer and there is an integrated massage device that can be used to massage the skin during cellulite treatment.Same
8Clinical usePrescription UsePrescription UsePrescription UseSame
9Electrical ProtectionClass I Type BFClass I type BFClass I type BFSame
10User InterfaceTouch ScreenTouch ScreenTouch ScreenSame
11Firmware ControlledYesYesYesSame
12Type of energyRadiofrequency wavesRadiofrequency wavesRadiofrequency wavesSame
13Temperature ControlYes with external IR thermometerYesYes with external IR thermometerSame
14Applicator TypesReusable Electrodes:Monopolar Capacitive (Coating material: Polyamide)Monopolar Resistive (Material: Stainless steel)Reusable Electrodes:Monopolar CapacitiveBipolar ResistiveSmall handpiece tip provided sterile and disposableReusable Electrodes:Monopolar Capacitive (Coating material: Polyamide)Monopolar Resistive (Material: Stainless steel)Same with K161458. With the exception that the predicate device has no

Page 10

510(k) Number K241733

FeatureSubject DevicePredicate DeviceReference DeviceSimilarity
Bipolar Resistive (Material: Stainless steel)The electrodes are inserted into a handle/handpieceElectrodes dimensions and Power density:Capacitive: Size in diameter (Power density)There are four sizes 20 mm (63.69 W/cm²), 40mm (15.89 W/cm²), 60mm (28.26 W/cm²) and 80-mm (4.37 W/cm²), and 5.5 mm thickness.Resistive: Size in diameter (Power density)There are three sizes.40mm (9.93 W/cm²), 60mm (4.42 W/cm²)80mm (2.48 W/cm²) and 5.5 mm thickness.Bipolar: Size in diameter (Power density)There are two sizes.30mm (48.82 W/cm²), 70mm (6.9 W/cm²), and 5.5 mm thickness).The electrodes are inserted into a handle/handpiece.Electrodes Dimensions for Monopolar Capacitive:ø20 mm, ø30 mm, ø40 mm,ø60 mm, ø70 mm, ø80 mm,ø100 mm, ø40 mmThe electrodes are inserted into a handle/handpiece.Electrode dimensionsCapacitive:Flat circular shape and sized at 30-, 40-, 55- and 65-mm diameter.Resistive:Flat circular shape and sized at 35-, 50- and 65-mm diameter.bipolar applicatorSame with K210867 with the exception of the small area handpiece tip provided EtO sterile for single use and does not use Monopolar resistive type of applicator.Applicator area for Bipolar electrode not reported. Power density not reported for K210867 & K161458

Page 11

510(k) Number K241733

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
15Use of Return PlateYes, except in treatments with bipolar electrode.Characteristics:FlexibleSize: 20x14 cm; Area covered 280 cm² (small)Size 28x20 cm; Area covered 560 cm² (large).Material: Stainless SteelYes, except in treatments with bipolar electrode.Characteristics:Disposable return platesYesCharacteristics:FlexibleSize: 20x26 cmArea covered 520 cm²Material: Stainless SteelSame except that area covered information is not available for K210867
16Use of conductive creamYes, to provide thermal transfer assistance.Yes, to provide thermal transfer assistance.Yes, to provide thermal transfer assistance.Same
17Other AccessoriesHandles, IR thermometer, power cable, trolleyHandles, IR thermometer, power cable, trolleyHandles, IR thermometer, power cable, trolleySame
18BiocompatibilityISO 10993ISO 10993ISO 10993Same
19Type of body contactSuperficial contact on healthy skinSuperficial contact on healthy skinSuperficial contact on healthy skinSame
20SterilizationNoOnly the small area handpiece tip provided EtO sterile for single use.NoSame

Page 12

510(k) Number K241733

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
21Electrode reprocessingYes, after electrode cleaningYes, except for the small area handpiece tip provided EtO sterile for single use.Yes, after electrode cleaningSame
22Shelf life50000 Hours in continuous operationInformation not availableInformation not availableN/A

Performance specifications

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
23Treatment temperature range+43 °C – +45 °C+40 °C - +42 °C+10 °C – +40 °CPlease refer to the foot notes at the end of the table
24Frequency448 KHz470 kHz; 1 MHz; 2 MHz; 4 MHz; 6 MHz400 kHz - 449 kHzSame
25Output RF PowerMax 200 WMax 250 WMax 200 WSame
26Voltage100-240 V100 - 240 V100 - 240 VSame

Safety Features

Page 13

510(k) Number K241733

SrNoFeatureSubject DevicePredicate DeviceReference DeviceSimilarity
27Temperature and PowerThe device contains current sensor in the electronics that generates the RF signal. By controlling the current, the treatment temperature is controlled. If the measured current is out of the acceptance range the device stops and issues a warning through the GUI.-Internal over-temperature protection by means of a fan that is activated and regulates its power according to the measurement of an internal temperature sensor in the device.Controlling the current intensity generated by the electronics generating the RF also controls the output power.Information not available-Output protected against accidental short circuit between neutral plate and active electrode.-Output over-voltage limitation-Output over-power limitation-Contact detection with patient skin-Internal over-temperature protectionSame with Indiba internal overtemperature protection.

a The 510(k) Summary of the reference device K161458 stated in its performance testing "The validation confirmed that after a minimum of 5 minutes the Indiba Diathermia Radiofrequency Device can raise, and maintain, the skin temperature to the range of 40-45 °C under capacitive and resistive applications."

Performance Date and Bench Test

The Verification and validation were performed on the device using the established methods by using the predicate device.

The test was conducted in accordance with the following standards.

Page 14

510(k) Number K241733

  • IEC 60601-1; Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
  • IEC 60601-1-2; Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests;
  • IEC 60601-1-6; Medical Electrical Equipment – Part 1-6: General Requirements For Basic Safety And Essential Performance – Collateral Standard: Usability;
  • IEC 60601-2-2: 2009 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
  • IEC 62304; Medical Device – Software Life Cycle Processes

The Biocompatibility of the device was conducted in accordance with the following standards.

  • ISO 10993-1-2018: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • ISO 10993-5-2009: Biological evaluation of medical devices - Part 5 Test for invitro cytotoxicity
  • ISO 10993-10-2010: Biological evaluation of medical devices - Part 10 Test for irritation and Skin Sensitization

Conclusion

The HEBE has the same intended use and similar indications for use, technological characteristics and principles of operation as its predicate device & Reference device. The performance testing the non-clinical testing is performed and its similar to the Predicate device.

HEBE devices are safe and effective and are substantially equivalent to the predicate devices in terms of performance.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.