K Number
K161458
Manufacturer
Date Cleared
2016-10-03

(130 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Indiba Diathermia Radiofrequency Devices are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation.

The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.

Device Description

The Indiba Diathermia Radiofrequency Device is a therapeutic device for deep, non-invasive diathermy. The device consists of a console which generates a radiofrequency current which is delivered to the patient, in monopolar form, through two different types of electrodes: stainless steel conductive resistive electrodes, and thin-layer insulated capacitive electrodes. The electrodes are inserted into a handle/handpiece, one handle for each kind of electrode, and the handle is connected to the console by means of a 2-metre cable.

In resistive mode the system delivers a high-frequency current of 448 kHz directly to the patient's skin surface. In capacitive mode, the electrode coating creates a layer between the electrode and the human tissue, forming a capacitor that allows a high-frequency current to pass. The high frequency current ranges between 400 kHz and 449 kHz and is automatically tuned by the equipment according to the patient's impedance. Current returns through the neutral return electrode. The Indiba Diathermia Radiofrequency Device is provided with an electroconductive media which is applied to the patients' skin prior to each treatment session.

The RF energy generates a heating profile that produces a moderate temperature rise in the subcutaneous tissue.

The temperature on the skin is measured using a separate IR (Infra-red) thermometer and there is an integrated massage device that can be used to massage the skin during cellulite treatment.

AI/ML Overview

This document is a 510(k) summary for the Indiba Diathermia Radiofrequency Devices. It describes the device, its intended use, and provides evidence of substantial equivalence to legally marketed predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" against "reported device performance" in the typical format of a clinical study. However, it details the performance characteristics tested and confirms they were met. The primary performance characteristic demonstrated by the in-house study is the ability of the device to elevate and maintain skin surface temperature.

Characteristic TestedAcceptance Criteria (Implied)Reported Device Performance
Temperature ElevationRaise and maintain skin surface temperature to 40-45 °CIndividual volunteers' maximum temperatures on capacitive treatment were 41-44 °C and on combined capacitive and resistive were 41-43 °C. This was achieved and maintained after a minimum of 5 minutes.
Electrical Output (Output frequency, voltage, power)Within stated performance characteristics specification"The results for each tested output characteristic for all the Indiba Diathermia Radiofrequency Devices passed the acceptance criteria." (Specific numerical values for acceptance are not provided.)
BiocompatibilityMaterials must be biocompatible for intended usePatient contacting materials confirmed to be biocompatible for their intended use as per ISO10993-1:2009.
Device Safety & SoftwareCompliance with relevant IEC standards (60601-1, 60601-1-2, 60601-2-2) and FDA software guidanceDevices are in compliance with IEC/EN 60601-1:2005+ Corr. 2006, IEC/EN60601-1-2:2007+AC:2010, and applicable sections of IEC 60601-2-2: 2009. Software verification and validation conducted to IEC 62304: 2006.

2. Sample size used for the test set and data provenance

  • Sample Size (for temperature elevation study): 9 healthy volunteers.
  • Data Provenance: The study was an "in-house study" on healthy volunteers. The country of origin is not explicitly stated, but the applicant and C/O applicant addresses are in the USA and Spain, respectively. It was a prospective study, as it involved actively treating volunteers and recording data.

3. Number of experts used to establish the ground truth for the test set and their qualifications

The document does not mention the use of experts to establish a "ground truth" for the temperature elevation study. The ground truth for this study was directly measured skin surface temperature using an Infrared Thermometer.

4. Adjudication method for the test set

Not applicable. The study involved direct measurement of physical parameters (temperature) rather than subjective interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device for therapeutic heating, not an AI-assisted diagnostic device. The document does not describe an MRMC study or AI assistance for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not an AI algorithm. The device is a radiofrequency energy delivery system. The temperature elevation study assessed the standalone performance of the device in achieving its intended therapeutic effect (raising tissue temperature).

7. The type of ground truth used

For the temperature elevation study, the "ground truth" was measured skin surface temperature using an Infrared Thermometer. This is an objective, directly measured physiological parameter.

8. The sample size for the training set

Not applicable. This is a medical device that delivers radiofrequency energy, not a machine learning model requiring a "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with three lines forming the shape of a staff with a snake winding around it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2016

Indiba USA Inc. % Mr. Howard Beaumont Indiba S.A. 13 - Pol. Ind. Casablancas Sant Quirze Del Valles, Barcelona, Spain 08192

Re: K161458

Trade/Device Name: Indiba Diathermia Radiofrequency Device - Activ; Indiba Diathermia Radiofrequency Device - Deep Care Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX Dated: August 20, 2016 Received: August 26, 2016

Dear Mr. Beaumont:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161458

Device Name

Indiba Diathermia Radiofrequency Devices

Indications for Use (Describe)

The Indiba Diathermia Radiofrequency Devices are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation.

The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Food and Drug AdministrationOffice of Chief Information OfficerPRAStaff(@fda.hhs.govDepartment of Health and Human ServicesPaperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection ofinformation unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "INDIBA" in a bold, sans-serif font. The letters are black against a white background. There is a degree symbol after the letter A.

SECTION 06 - 510(k) SUMMARY

Contact Details

510(k) NumberK161458
Applicant:Indiba USA Inc.
1717 Embarcadero Road
Palo Alto
CA 94303
USA
Applicant Contact:Carles Janer
Regulatory Affairs Manager
C/O Applicant:Indiba S.A.
C / Moianés
13 - Pol. Ind. Casablancas
08192 Sant Quirze del Vallés
Barcelona
Spain
Application Correspondent:Howard Beaumont
Regulatory Affairs Consultant
Date Prepared18th May 2016
Device Details
Common Name:Radiofrequency Device
Device Trade NameIndiba Diathermia Radiofrequency Device
Classification Name:Electrosurgical, cutting and coagulation accessories
Regulation Number:878.4400
Product CodePBX
Device Class:2
Panel:General & Plastic Surgery

{4}------------------------------------------------

K161458

Indiba Diathermia Radiofrequency Devices.

Legally Marketed Predicate Device(s)

510(k) NumberProduct CodeTrade NameApplicant
K133739PBX, 878.4400TruSculptCutera Inc.
K132949PBX, 878.4400PelleFirm SystemEllman International(Cynosure Inc.)

Device Description

The Indiba Diathermia Radiofrequency Device is a therapeutic device for deep, non-invasive diathermy. The device consists of a console which generates a radiofrequency current which is delivered to the patient, in monopolar form, through two different types of electrodes: stainless steel conductive resistive electrodes, and thin-layer insulated capacitive electrodes. The electrodes are inserted into a handle/handpiece, one handle for each kind of electrode, and the handle is connected to the console by means of a 2-metre cable.

In resistive mode the system delivers a high-frequency current of 448 kHz directly to the patient's skin surface. In capacitive mode, the electrode coating creates a layer between the electrode and the human tissue, forming a capacitor that allows a high-frequency current to pass. The high frequency current ranges between 400 kHz and 449 kHz and is automatically tuned by the equipment according to the patient's impedance. Current returns through the neutral return electrode. The Indiba Diathermia Radiofrequency Device is provided with an electroconductive media which is applied to the patients' skin prior to each treatment session.

The RF energy generates a heating profile that produces a moderate temperature rise in the subcutaneous tissue.

The temperature on the skin is measured using a separate IR (Infra-red) thermometer and there is an integrated massage device that can be used to massage the skin during cellulite treatment.

Indications for Use

The Indiba Diathermia Radiofrequency Devices are intended to provide topical heating for the purpose of elevating tissue temperature for treatment of selected medical conditions such as: relief of pain, muscle spasms, increase in local circulation.

Indiba 510(k): K161458

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "INDIBA" in all caps. The letter "A" is stylized with a caret symbol above it. There is a small circle to the right of the caret symbol.

The massage device provided is intended to provide a temporary reduction in the appearance of cellulite.

Technical Characteristics

CharacteristicValue / Description
Input voltage supply :(100 – 130) V~ 50/60 Hz
Temperature range for operation :+10°C to +40°C
Temperature range for storage andtransport :-20°C to +50°C
Timer range :0 – 99 minutes
Display :5.7 inch TFT color 320 x 240 pixels
Other features :- Output protected against accidentalshort-circuit between neutral plate andactive electrode.- Output over-voltage limitation- Output over-power limitation- Contact detection with patient skin- Internal over-temperature protection

Substantial Equivalence

The indications for use, functionality, type of application, type of electrodes, skin temperature sensing and compliance to safety standards of the Indiba Diathermia Radiofrequency Device are the same or very similar to the legally marketed, claimed predicate devices for the purpose of this 510(k) submission. Technological differences between the Indiba Diathermia Radiofrequency Device and the predicate devices are limited to small variations in the maximum power (200 W, while predicate devices have a range from 120 W to 300 W) and the RF frequency (400 kHz - 449 kHz, while the predicate device has a range of 300 kHz - 50 MHz). Also, the patient contacting materials of the Indiba Diathermia Radiofrequency Device are different to those of the predicate devices, but are biocompatible for their intended purpose.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "INDIBA" in all caps. The letters are in a sans-serif font and are black. There is a degree symbol to the right of the letter A.

Non-Clinical Testing

The Indiba Diathermia Radiofrequency Devices have been tested and are in compliance with IEC/EN 60601-1:2005+ Corr. 2006, Medical electrical equipment. Part 1: General requirements for basic safety and essential performance, IEC/EN60601-1-2:2007+AC:2010 Medical electrical equipment. Part 1: General requirements for safety - Collateral standard: Electromagnetic Compatibility -Requirements and tests, and applicable sections of IEC 60601-2-2: 2009 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. In addition to the electrical safety testing performed, software verification and validation was conducted to IEC 62304: 2006 - Medical device software - Software Life-Cycle Processes, and FDA guidance on software validation. The results of this testing conclude the software has met these requirements.

Patient contacting materials have been evaluated according to the requirements of ISO10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing, and confirmed to be biocompatible for their intended use.

Design verification and validation was also performed on the Indiba Diathermia Radiofrequency Devices in compliance with internal design control procedures.

The electrical output characteristics of the Indiba Diathermia Radiofrequency Devices have been tested to demonstrate the devices meet the stated performance characteristics. The Output Characteristic testing measured output frequency, output voltage and output power at 100% (3 resistance loads). The results for each tested output characteristic for all the Indiba Diathermia Radiofrequency Devices passed the acceptance criteria. The testing concluded that each output characteristic for each device in the range was within the stated performance characteristics specification.

Clinical Evaluation

Clinical investigations were not conducted to support substantial equivalence. An independent search, review and analysis of published and un-published

Indiba 510(k): K161458

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "INDIBA" in a bold, sans-serif font. The letters are black against a white background. The "A" in "INDIBA" is stylized with a line through the middle, resembling the Greek letter lambda.

literature to support the claim that the Indiba Diathermia Radiofrequency Devices are safe and effective was conducted. For the purpose of the evaluation of the literature, publications pertaining to RF applications under the indications for: relief of pain, relief of muscle spasms, increase in local circulation and temporary reduction in the appearance of cellulite, were sought. The independent evaluation concluded that there was supportive evidence that the Indiba family of devices, delivering RF energy in resistive and capacitive modes, can achieve a significantly beneficial effect on ache/pain, local circulation, cellulite and the relief of muscle spasm.

Indiba Diathermia Radiofrequency Devices have been validated in an in-house study on healthy volunteers to demonstrate that the Indiba Diathermia Radiofrequency Devices increase skin surface temperature to the range 40-45℃.

The study was conducted on nine healthy volunteers, in accordance with Indiba Diathermia Radiofrequency Device standard treatment protocols, with treatment applied to the middle third of the anterior right thigh. Each session consisted of a combination of application of the capacitive electrode followed by the resistive electrode. Prior to each session electroconductive media was applied to the area of skin to be treated. Treatment times were approximately 5-6mins capacitive followed by 7-11mins resistive. Skin surface temperature was recorded using an Infrared Thermometer.

Tolerance of the treatment was good, there were no reports of discomfort and there were no requirements to adjust the treatment powers. Mild transitory erythema (skin reddening) was present in all volunteers in temperatures around 42-43 °C.

The results confirmed that individual volunteers' maximum temperatures on capacitive treatment were 41-44 °C and on capacitive and resistive were 41-43 °C.

The validation confirmed that after a minimum of 5 minutes the Indiba Diathermia Radiofrequency Device can raise, and maintain, the skin temperature to the range of 40-45 °C under capacitive and resistive applications.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "INDIBA" in all caps. The font is sans-serif and the letters are bold. There is a degree symbol after the letter A.

Conclusion

The results of external and in-house testing, and the conclusions of the independent review of literature, summarised above, demonstrate that the Indiba Diathermia Radiofrequency Devices are safe and effective and are substantially equivalent to the predicate devices in terms of performance.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.