K Number
K240940
Date Cleared
2024-05-03

(28 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vial2Bag Advanced® 20mm Admixture Device is indicated to serve as a connection between a 50, 100 or 250mL IV bag, vial with 20mm closure, and an external IV administration set. The integrated Vial Adapter makes it possible to reconstitute and/or admix drugs prior to administration to the patient. Indicated for adolescent and adult patients only.

Device Description

The Vial2Bag Advanced® 20mm Admixture Device is a single-use, sterile, needle-less, nonpyrogenic, fluid transfer device that allows for reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution, through the IV bag administration port. The device consists of the body, Protector, IV Port, and an integrated vial adapter. The device is intended to be used with standard drug vials with a 20mm closure and an elastomeric stopper. The Vial2Bag Advanced® 20mm device is designed to work with a standard 50, 100, or 250mL IV bag and an external IV infusion set.

AI/ML Overview

The provided text describes a medical device, the Vial2Bag Advanced® 20mm Admixture Device, and its comparison to a predicate device for substantial equivalence. It is not an AI/ML device, therefore, many of the requested fields are not applicable.

Here's an analysis based on the information provided:

1. A table of acceptance criteria and the reported device performance:

The document outlines performance testing, but the explicit acceptance criteria are not tabulated with specific performance results. Instead, it states that testing was conducted to ensure the device "met the applicable design and performance requirements." For some tests, it mentions the basis for acceptance criteria or sample size, but not the actual criteria set.

TestAcceptance Criteria (Explicitly Stated)Reported Device Performance (Explicitly Stated)
Vial Adaptor Tensile Detachment ForceNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Vial Adaptor Torque TestNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Detachment Force of Vial Adapter from VialNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Attachment Force of Vial Adapter to the VialNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Leakage TestingNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Internal Diameter of the Upper Skirt for Vial AdapterNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
1m Drop DurabilityNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
FragmentationBased on EN ISO 8536-2, section 6.2.2Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Mass TransferNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Residual VolumeNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
Dose ConcentrationNot explicitly stated (In-house test method)Met applicable design and performance requirements; subject device is as safe and effective as the predicate.
BiocompatibilityIn accordance with ISO 10993-1, classified as externally communicating device with prolonged contact duration (>24 hours to 30 days)Test reports from the reference device (K201415), which uses an identical colorant, were leveraged and deemed applicable to demonstrate biological safety.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Fragmentation Test: "sample size based on EN ISO 7864, Annex B, Section B.4." (Specific number not provided, but the standard is cited.)
  • Sample Size for other tests: Not explicitly stated.
  • Data Provenance: The tests are in-house, suggesting they were conducted by the manufacturer, West Pharma. Services IL, Ltd., in Ra'anana, Israel. There is no mention of retrospective or prospective data in the context of performance testing on the device itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is not applicable as the device is a physical medical device (fluid transfer device) and not an AI/ML diagnostic or predictive tool that requires expert-established ground truth on a test set. The performance testing is based on engineering and material standards.

4. Adjudication method for the test set:

Not applicable for a physical device's performance testing. The outcomes of the tests are objective measurements against established standards or in-house criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI/ML device, and no MRMC study was conducted or is relevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. This is not an AI/ML device.

7. The type of ground truth used:

The performance criteria are established through:

  • In-house test methods (presumably based on engineering principles and intended function).
  • International standards (e.g., EN ISO 8536-2 for Fragmentation, ISO 10993-1 for Biocompatibility).
  • Comparison to a predicate device to demonstrate "substantial equivalence."

8. The sample size for the training set:

Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established:

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 3, 2024

West Pharmaceutical Services, Inc. % Fred Cowdery Director Regulatory Affairs 530 Herman O. West Drive Exton, Pennsylvania 19431

Re: K240940

Trade/Device Name: Vial2Bag Advanced® 20mm Admixture Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: March 19, 2024 Received: April 5, 2024

Dear Fred Cowdery:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a

change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Pasane Bennett

Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240940

Device Name

Vial2bag Advanced® 20mm Admixture device

Indications for Use (Describe)

The Vial2Bag Advanced® 20mm Admixture Device is indicated to serve as a connection between a 50, 100 or 250mL IV bag, vial with 20mm closure, and an external IV administration set. The integrated Vial Adapter makes it possible to reconstitute and/or admix drugs prior to administration to the patient. Indicated for adolescent and adult patients only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K240940 - 510(K) SUMMARY

SUBMITTER

Applicant:

West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297

Manufacturer:

West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297

Contact Person:

Fred Cowdery Director, Regulatory Affairs Phone: 267-205-1273 Fax: 610-717-0668 E-mail: fred.cowdery@westpharma.com

Date Prepared: 05-Apr-2024

DEVICE

Trade Name:Vial2Bag Advanced® 20mm Admixture Device
Common/Usual Name:I.V. Fluid Transfer Set
Regulation Name:Intravascular administration set
Product Code:LHI
Regulation No.:880.5440
Class:II
Panel Identification:General Hospital Panel

{4}------------------------------------------------

PREDICATE DEVICE

Vial2Bag Advanced® 13mm Admixture Device (K230988).

A reference device, Vial2Bag Advanced® 20mm Admixture Device; K201415 is also included in this submission.

DEVICE DESCRIPTION

Device Design and Operation

The Vial2Bag Advanced® 20mm Admixture Device is a single-use, sterile, needle-less, nonpyrogenic, fluid transfer device that allows for reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution, through the IV bag administration port. The device consists of the body, Protector, IV Port, and an integrated vial adapter. The device is intended to be used with standard drug vials with a 20mm closure and an elastomeric stopper. The Vial2Bag Advanced® 20mm device is designed to work with a standard 50, 100, or 250mL IV bag and an external IV infusion set.

Users should not attach a Vial2Bag Advanced® device to another Vial2Bag Advanced® Device.

The device does not contain any medicinal substances and no additional accessories are required nor provided with the Vial2Bag Advanced® 20mm Device to meet its intended purpose.

Principle of Operation

The Vial2Bag Advanced® 20mm Admixture Device is operated by manual process. The Vial Adapter is first attached to the drug vial, and after removing the Protector, the IV spike is then connected to the administration port of the IV bag. Fluid is transferred from the IV bag to the drug vial to reconstitute/dilute the drug prior to being transferred back to the IV bag. The IV administration set is then connected to the device's IV Port followed by administration to the patient.

INDICATION FOR USE

The Vial2Bag Advanced® 20mm Admixture Device is indicated to serve as a connection between a 50, 100 or 250mL IV bag, vial with 20mm closure, and an external IV administration set. The integrated Vial Adapter makes it possible to reconstitute and/or admix drugs prior to administration to the patient. Indicated for adolescent and adult patients only.

Both the subject and predicate devices have the same intended use except for the specified vial closure diameter. While the predicate device is interface with vials having a 13mm closure, the subject device is intended to interface with vials having a 20mm closure.

{5}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALAENCE

The Vial2Bag Advanced® 20mm Admixture Device, subject of this Special 510(k), is substantially equivalent in its intended use, design, technology, principles of operation, materials (with exception of Vial Adapter colorant additive), and performance to the predicate device, Vial2Bag Advanced® 13mm Admixture Device, cleared under K230988.

A summary comparing the subject device and the predicate device are provided in the table below.

Areas forComparisonSubject DeviceVial2Bag Advanced®20mm Admixture DevicePredicate Device(K230988)Vial2Bag Advanced®13mm Admixture DeviceComparison
Indications for UseTo serve as a connectionbetween a 50, 100, or250ml IV bag, vial with20mm closure, and anexternal IV administrationset. The integrated VialAdapter makes it possibleto reconstitute and/oradmix drugs. Indicated foradolescent and adultpatients only.To serve as a connectionbetween a 50, 100, or250ml IV bag, vial with13mm closure, and anexternal IV administrationset. The integrated VialAdapter makes it possibleto reconstitute and/oradmix drugs. Indicated foradolescent and adultpatients only.Difference #1: Although thesubject device Vial Adaptersize differs from the predicatedevice, the reference deviceperformance test resultsconfirm the vial adapter sizedoes not raise new or differentquestions of safety andeffectiveness.
ContraindicationsNone knownNone knownIdentical
Intended UserPopulationIntended for use byHealthcare Professionals(HCPs)Intended for use byHealthcare Professionals(HCPs)Identical
Intended UseEnvironmentIntended for use inhospitals, outpatientnursing units and othersuitable clinicalenvironmentsIntended for use inhospitals, outpatientnursing units and othersuitable clinicalenvironmentsIdentical
Device ProductCode, Class &Classification NameLHI, Class II, I.V. FluidTransfer SetLHI, Class II, I.V. FluidTransfer SetIdentical
Regulation Number/ Name21CFR 880.5440IntravascularAdministration Set21CFR 880.5440IntravascularAdministration SetIdentical
Prescription UseYesYesIdentical
Single UseYesYesIdentical
Operation PrincipleManualManualIdentical
DesignMade of plastic material,featuring a Vial Adapterto access the drug contentMade of plastic material,featuring a Vial Adapter toaccess the drug content in aDifference #1:Although the subject deviceVial Adapter size differs from
Areas forComparisonSubject DeviceVial2Bag Advanced®20mm Admixture DevicePredicate Device(K230988)Vial2Bag Advanced®13mm Admixture DeviceComparison
in a 20mm vial, an IVSpike integrated forconnection to an IV bag,and a twist off which,after removal, opens theIV port for connection tothe IV administration set.13mm vial, an IV Spikeintegrated for connectionto an IV bag, and a twistoff which, after removal,opens the IV port forconnection to the IVadministration set.the predicate device, thereference device performancetest results confirm the vialadapter size difference does notraise new or different questionsof safety and effectiveness.
Materials ofConstruction20mm Vial Adapter:Polycarbonate (PC) +Blue colorant masterbatch13mm Vial Adapter:Polycarbonate (PC) +Orange colorantmasterbatchDifference #2: Although thesubject device Vial Adaptercolorant differs from thepredicate device, the referencedevice Biocompatibility testreports confirm subject devicecolorant does not impact deviceintended use, clinicaleffectiveness, or safety profile.
IV Port Twist Off: PC +Polyvinyl Chloride (PVC)3224 (non-DEHP)IV Port Twist Off: PC +Polyvinyl Chloride (PVC)3224 (non-DEHP)Identical
Spike Protector: LowDensity Polyethylene(LDPE)Spike Protector: LowDensity Polyethylene(LDPE)Identical
Compatible VialSize20mm13mmDifference #1: Although thesubject device Vial Adaptersize differs from the predicatedevice, the reference deviceperformance test resultsconfirm the vial adapter sizedoes not raise new or differentquestions of safety andeffectiveness.
Bag Size50, 100, 250mL50, 100, 250mLIdentical
Dose Concentration200%200%Identical
Single/inlineconfigurationSingle configuration onlyDo not attach one deviceto another device.Single configuration onlyDo not attach one device toanother device.Identical
Vial Adapter FitVial first, snap fit to vialVial first, snap fit to vialIdentical
Performance onfluid transferTransfer of vial contentsto the IV bag and to theadministration set wasquantified to establishproduct requirement.Transfer of vial contents tothe IV bag and to theadministration set wasquantified to establishproduct requirement.Identical
BiocompatibilityISO 10993-1:2018External Communicating,Prolonged Indirect BloodContact (>24hr to 30 days)ISO 10993-1:2018External Communicating,Prolonged Indirect BloodContact (>24hr to 30 days)Identical
Areas forComparisonSubject DeviceVial2Bag Advanced®20mm Admixture DevicePredicate Device(K230988)Vial2Bag Advanced®13mm Admixture DeviceComparison
Contact (>24hr to 30days)
Non-pyrogenicYesYesIdentical
SterilitySterileSterileIdentical
Sterilization MethodEthylene OxideEthylene OxideIdentical
Sterility AssuranceLevelSAL of 10-6SAL of 10-6Identical
PackagingSterile Barrier packagematerials: PETG blisterwith Tyvek® sealSterile Barrier packagematerials: PETG blisterwith Tyvek® sealIdentical

Substantial Equivalence Comparison Table

{6}------------------------------------------------

{7}------------------------------------------------

  • Difference #1- Although the subject device Vial Adapter size differs from the predicate . device, the reference device, performance test results confirm the vial adapter size does not raise new or different questions of safety and effectiveness.
  • . Difference #2 - Although the subject device Vial Adapter colorant differs from the predicate device, the reference device Biocompatibility and Chemical test reports confirm subject device colorant does not impact device intended use, clinical effectiveness, or safety profile.

Although the subject and reference devices are similar in design and technology, the reference device materials contain a DEHP plasticizer additive, which is not present in the subject device components. Functionally, the subject device, Vial2Bag Advanced® 20mm Admixture Device, is otherwise the same as the reference device cleared under K201415.

{8}------------------------------------------------

PERFORMANCE DATA

Performance Testing

Performance testing was conducted to ensure that the Vial2Bag Advanced® 20mm Admixture Device met the applicable design and performance requirements throughout its shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device.

Risks associated with the proposed changes to the subject device do not alter or introduce new device risks. The Risk Management Files prepared for both the predicate and reference devices have been maintained and remain applicable to the subject device.

This change does not raise new or different questions of safety and effectiveness. Therefore, the subject device is as safe and effective as the predicate device.

TestTest Method/ Standard
Vial Adaptor Tensile Detachment ForceIn-house test method
Vial Adaptor Torque TestIn-house test method
Detachment Force of Vial Adapter from VialIn-house test method
Attachment Force of Vial Adapter to the VialIn-house test method
Leakage TestingIn-house test method
Internal Diameter of the Upper Skirt for Vial AdapterIn-house test method
1m Drop DurabilityIn-house test method
FragmentationAcceptance criteria based on EN ISO 8536-2, section 6.2.2 and sample size based on EN ISO 7864, Annex B, Section B.4.
Mass TransferIn-house test method
Residual VolumeIn-house test method
Dose ConcentrationIn-house test method

Summary of Performance Testing

Biocompatibility Testing

In accordance with ISO 10993-1. the subject device, Vial2Bag Advanced® 20mm Admixture Device, is classified as an externally communicating device with prolonged contact duration (>24 hours to 30 days) and blood path indirect contact.

The subject device materials are identical to materials included in the predicate device submission (K230988), except for the vial adapter component colorant additive. While the predicate device contains an orange colorant, the subject device contains a blue colorant identical to the reference device colorant.

Since the subject device vial adapter colorant is identical to the colorant used in the reference device vial adapter component, the reference device (K201415) biocompatibility test reports are deemed applicable and are being leveraged for the subject device to demonstrate the subject device biological safety.

{9}------------------------------------------------

CLINICAL DATA

No clinical trial was performed for the subject device, Vial2Bag Advanced® 20mm Admixture Device, nor required to support the proposed changes.

CONCLUSION

In summary, the Vial2Bag Advanced® 20mm Admixture Device, the subject of this Premarket Notification, is as safe, as effective, and is substantially equivalent in its intended use, technology/principle of operation, and performance to the predicate device, Vial2Bag Advanced® 13mm Admixture Device (K230988).

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.