K Number
K240806

Validate with FDA (Live)

Date Cleared
2024-05-21

(57 days)

Product Code
Regulation Number
878.4850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The safety lancet is intended for capillary blood sampling.

Device Description

The safety lancet is single use medical device, which is designed to collect capillary blood sample.
The intended users include Healthcare personnel, patients and lay persons.
According to surface difference or structure, the Safety lancets have 7 models: XXVIII, XXIX, XXX, XXXI, XXXII, XXXIII, XXXIV.
XXIX, XXX model have sub model according to surface difference.
For Model XXVIII and XXX, the safety lancets consist of needle core, button, housing and spring. For Model XXIX, the safety lancets consist of needle core, button, housing, spring, and Depth adjuster ring. For Model XXXI, the safety lancets consist of needle core, top cover, bottom and spring. For Model XXXII, the safety lancets consist of needle core, top cover, bottom, small lid and spring. For Model XXXIII, the safety lancets consist of needle core, button, housing, inner sleeve and spring. And for Model XXXIV, the safety lancets consist of needle core, button, housing, bottom, Protective cap and spring.
The sterile part of the safety lancet is the needle tip.The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf-life of the product is 5 years.

AI/ML Overview

This document is a 510(k) summary for a medical device called "Safety Lancet," and it does not contain information about an AI/ML powered device. The information provided is for a physical medical device (safety lancets) used for capillary blood sampling. Therefore, many of the requested fields related to AI/ML device studies (such as MRMC studies, standalone algorithm performance, training set details, and expert ground truth establishment for AI) are not applicable.

However, I can extract the relevant information regarding the acceptance criteria and the studies performed for this physical device.

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Launch performance should be good, launch button press smoothly, no jamMeet the requirement
The needle tip of the needle should have good puncture ability.Meet the requirement
Visual, should not be visible droplets. (Lubricant)Meet the requirement
Safety lancet should be single use, no second launch after used. (Disposable)Meet the requirement
The force to activate the safety feature: 4 - 15NMeet the requirement
Test access to the sharp: the needle shall not touch the sphere.Meet the requirement
Not produce cytotoxicity (Biocompatibility - In Vitro Cytotoxicity)No Cytotoxicity
Not produce skin sensitization (Biocompatibility - Skin Sensitization)No Skin sensitization
Not produce irritation (Biocompatibility - Intracutaneous reactivity)No irritation
Not cause acute systemic toxicity (Biocompatibility - Acute Systemic Toxicity)No Acute Systemic Toxicity
Not cause pyrogenic reaction (Biocompatibility - Pyrogenicity)No thermogenic reaction
Safety mechanism evaluation (Simulated Clinical Use)Met the pre-established criteria

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Simulated Clinical Use Study: 300 device samples each for the Safety Lancet (various models).
  • Data Provenance: Not specified in the provided document (e.g., country of origin, retrospective/prospective). This was a bench test and simulated use study, not a human data study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The "ground truth" for this physical device was based on objective physical measurements and established biological safety standards from ISO 10993 and FDA guidance, not expert interpretation of data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This was a non-clinical bench and simulated use study, not a study requiring adjudication of expert opinions.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML powered device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML powered device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For bench testing: Objective physical measurements and adherence to functional specifications.
For biocompatibility testing: Laboratory results following ISO 10993 standards.
For simulated clinical use: Evaluation of the safety mechanism against pre-established criteria defined by FDA Guidance (Medical Device with Sharps Injury Prevention Feature) and ISO 23908.

8. The sample size for the training set

Not applicable. This is not an AI/ML powered device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML powered device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".

May 21, 2024

Tianjin Huahong Technology Co., Ltd. Wang Ningning, Registered Engineer A01, Plant B, No. 278, Hangkong Road, Tianjin Pilot Free Trade Zone (Air Port Industrial Park) Tianiin, 300308 China

Re: K240806

Trade/Device Name: Safety lancet (XXVIII, XXIX, XXX, XXXI, XXXII, XXXIII, XXXIV.) Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: March 18, 2024 Received: March 25, 2024

Dear Wang Ningning:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long H. Chen =S Digitally signed by Long H.Chen-S
Date: 2024.05.21 12:24:35 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K240806

Device Name

Safety lancet (XXVIII, XXIX, XXX, XXXI, XXXII, XXXIII, XXXIV. )

Indications for Use (Describe)

The safety lancet is intended for capillary blood sampling

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) summary

l Submitter

Tianjin Huahong Technology Co., Ltd. A01, Plant B, No.278, Hangkong Road, Tianjin Pilot Free Trade Zone (Air Port Industrial Park), 300308 Tianjin, China

Establishment Registration Number: 3009498536

Contact person: Ms. Ningning Wang Registered Engineer Tel.: +86-13021381776 E-mail: ningning.wang@hh-technology.com

Preparation date: Mar 18, 2024

II Proposed Device

Trade Name of Device:Safety Lancet (XXVIII, XXIX, XXX, XXXI, XXXII, XXXIII, XXXIV)
Common name:Single Use Only Blood Lancet With An Integral Sharps InjuryPrevention Feature
Regulation Number:21 CFR 878.4850
Regulatory Class:Class II
Product code:FMK
Review Panel:General & Plastic Surgery

III Predicate Devices

510(k) Number:K220370
Trade name:Safety Lancet

Classification: Class II FMK Product Code: Manufacturer Tianjin Huahong Technology Co., Ltd.

1

{4}------------------------------------------------

IV Device description

The safety lancet is single use medical device, which is designed to collect capillary blood sample.

The intended users include Healthcare personnel, patients and lay persons.

According to surface difference or structure, the Safety lancets have 7 models:

XXVIII, XXIX, XXX, XXXI, XXXII, XXXIII, XXXIV.

XXIX, XXX model have sub model according to surface difference.

For Model XXVIII and XXX, the safety lancets consist of needle core, button, housing and spring. For Model XXIX, the safety lancets consist of needle core, button, housing, spring, and Depth adjuster ring. For Model XXXI, the safety lancets consist of needle core, top cover, bottom and spring. For Model XXXII, the safety lancets consist of needle core, top cover, bottom, small lid and spring. For Model XXXIII, the safety lancets consist of needle core, button, housing, inner sleeve and spring. And for Model XXXIV, the safety lancets consist of needle core, button, housing, bottom, Protective cap and spring.

The sterile part of the safety lancet is the needle tip.The sterile barrier is the needle sleeve and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use only. The shelf-life of the product is 5 years.

V Indication for use

The safety lancet is intended forcapillary blood sampling.

VI Comparison of technological characteristics with the predicate devices

The comparison and discussion between the subject device and the predicate devices are listed in below table 1:

ItemProposed devicePredicate device(K220370)Discussion
Product nameSafety LancetSafety LancetSame
Product CodeFMKFMKSame
Regulation No.21 CFR § 878.485021 CFR § 878.4850Same
ClassIIIISame

Table 1 General Comparison of Safety Lancet

{5}------------------------------------------------

Prescription/over-the-counter useOver-The-Counter UseOver-The-Counter UseSame
Indication for useThe safety lancet isintended for capillaryblood sampling.The safety lancet is intendedfor capillary blood sampling.Same
Safety protectionfeaturesYesYesSame
Reuse durabilitySingle useSingle useSame
Sterilizationmethod and SALSterilized by RadiationSAL=10-6Sterilized by RadiationSAL=10-6Same
Self-life5 years5 yearsSame
Materials of partsin contact withhuman bodyNeedle core (containneedle) : PE,PP,Calcium powder, stainlesssteel (needle), silicone oil(needle);Housing, Button, Bottom,Top cover, Small lid,Depth adjuster ring,Protective cap, Innersleeve : ABS, PS, PP,Calcium powderNeedle core (containneedle) : PE,PP, Calciumpowder, stainless steel(needle), silicone oil(needle);Housing, Button, Bottom,Small lid, Depth adjuster ring,Protective cap: ABS,PSSimilar¹
BiocompatibilityConforms to therequirements of ISO10993 series standards.Conforms to therequirements of ISO 10993series standards.Same
Label/LabelingComplied with 21 CFRpart 801Complied with 21 CFR part801Same

1 The raw materials of proposed devices may be different from the predicate devices. However, these parts only contact with intact skin. All the materials are known biocompatible materials that have been used in lancets or other similar medical devices

VII Non-Clinical Testing

The bench testing performed verifies that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device in terms of critical performance characteristics. These tests are as follow.

{6}------------------------------------------------

Launch performanceLaunch performance should be good, launch button press smoothly, no jamMeet the requirement
Puncture forceThe needle tip of the needle should have good puncture ability.Meet the requirement
LubricantVisual, should not be visible droplets.Meet the requirement
DisposableSafety lancet should be single use, no second launch after used.Meet the requirement
Safety FeatureThe force to activate the safety feature : 4 - 15NTest access to the sharp: the needle shall not touch the sphere.Meet the requirement

Biocompatibility Testing:

The biocompatibility evaluations were conducted in accordance with the 2020 FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" . The tests include the following tests:

The formulation, processing, sterilization, geometry in the previously approved safety lancet products (K220370) are the same, and the nature and duration of contact with the patient are also the same.

ItemTest methodTest results
In Vitro CytotoxicityISO 10993-5: 2009No Cytotoxicity
Skin SensitizationISO 10993-10: 2010No Skin sensitization
Intracutaneous reactivityISO 10993-10: 2010No irritation
Acute Systemic ToxicityISO 10993-11: 2017No Acute Systemic Toxicity
PyrogenicityISO 10993-11: 2017no thermogenic reaction

Simulated Clinical Use

A simulated clinical use study was performed on 300 device samples each for the Safety Lancet according to FDA Guidance, Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.

{7}------------------------------------------------

VIII Clinical Testing

No clinical study is included in this submission.

IX Conclusion

The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.