K Number
K240529
Date Cleared
2024-10-02

(222 days)

Product Code
Regulation Number
870.1250
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 8F Modified Sheath System is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Device Description

The 8F Modified Sheath System is comprised of a Sheath, a Navigating Catheter, and an RHV (rotating hemostasis valve). The Sheath is a single-lumen, variable stiffness catheter with a radiopaque marker on the distal end. The inner lumen of the catheter is compatible with 8F or smaller catheters. The Dilator may be placed within the Sheath to facilitate percutaneous introduction of the Sheath into a femoral artery. The Dilator has a radiopaque marker at the distal tip. The Navigating Catheter is a singlelumen, variable stiffness catheter with a radiopaque marker at the distal tip. The Navigating Catheter is compatible with the Sheath and has a shaped distal end to facilitate placement. All of the catheters are coated with hydrophilic coating.

AI/ML Overview

The provided text describes the 8F Modified Sheath System and its substantial equivalence to a predicate device (Base Camp® Sheath System, K191717) for FDA clearance. The document focuses on non-clinical testing to demonstrate performance.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document lists various performance tests and their outcomes, uniformly indicating "PASS" and that "The pre-determined acceptance criteria were met." However, it does not explicitly state the specific quantitative acceptance criteria for each test. It only reports that the device met them.

TestTest MethodReported Device Performance
Dimensional VerificationDevice dimensions were measured to confirm conformance to the specifications.PASS
Luer IntegrityTested per ISO 80369-7:2016.PASS
Tensile StrengthThe tensile strength of the catheter sections and bonds was tested.PASS
Kink ResistanceTest specimen segments were formed into a defined bend diameter to evaluate kink resistance.PASS
Torsion ResistanceThe test specimens were rotated to evaluate integrity after rotation.PASS
Tip FlexibilityTest specimens were tested for tip flexibility.PASS
Air LeakageTested per ISO 10555-1:2013 Annex D.PASS
Liquid LeakageTested per ISO 10555-1:2013 Annex C.PASS
Hydrophilic Coating IntegrityThe integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles in a vascular model.PASS
Simulated Use TestingDeliverability and compatibility with accessory devices were evaluated in a vascular model.PASS
Static Burst Pressure TestingTested per ISO 10555-1:2013 Annex F.PASS
Particulate TestingThe size and quantity of particulate matter generated during simulated use testing were determined.PASS
Delivery/Retrieval Force TestingThe forces to deliver and retrieve the subject device in an anatomical model were measured.PASS

Biocompatibility Testing Acceptance

For biocompatibility, the acceptance criteria are generally qualitative for these types of tests and are implicitly met when the "Conclusion" states no adverse reactions or that requirements were met.

TestConclusion
SensitizationTest article did not elicit a sensitization response.
IrritationRequirements of the ISO intracutaneous reactivity test have been met for the test article.
Acute Systemic ToxicityRequirements of the ISO acute systemic injection test have been met for the test article.
Material Mediated PyrogenicityThe test article met the requirements for the absence of pyrogens.
CytotoxicityThe test article is non-cytotoxic.
Hemocompatibility – ASTM HemolysisThe test article is considered non-hemolytic.
Hemocompatibility - Thrombogenicity: ASTM Partial Thromboplastin Time AssayThe test article met the requirements of the test.
Hemocompatibility - Complement Activation AssayThe test article is a non-activator of the complement system.
Hemocompatibility - Thrombogenicity: ASTM Platelet/Leukocyte CountThe test article met the requirements of the test.
Hemocompatibility - Thrombogenicity: Comparative Surface AssessmentThe test article does not have any features that would increase thrombogenicity risk.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample sizes used for each of the performance or biocompatibility tests. It refers to "test specimens" or animals (e.g., "guinea pig," "rabbits," "mice") for biocompatibility, but without specific numbers.

  • Data Provenance: The data provenance is from non-clinical bench testing and laboratory animal testing. There is no information provided about the country of origin, but given it's an FDA submission, the tests were conducted to international standards (e.g., ISO, ASTM) which are globally recognized. All testing appears to be prospective (conducted specifically for this submission).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This is a submission for a medical device (catheter system) where "ground truth" in the clinical sense (e.g., diagnosis, outcome) is not relevant for the non-clinical and biocompatibility studies presented. The "truth" or acceptance is determined by predefined engineering specifications and regulatory standards (ISO, ASTM). The document does not mention the use of human experts to establish ground truth for this type of testing.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human interpretation of data (e.g., medical imaging) to resolve discrepancies. The studies described are objective performance and biocompatibility tests where a result either meets or does not meet a predefined standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device (catheter system), not an AI/software device. No MRMC study was conducted or is relevant.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical medical device, not an algorithm or AI system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

As mentioned in point 3, the concept of "ground truth" in the clinical diagnostic sense is not applicable here. For the performance and biocompatibility studies, the "ground truth" is defined by:

  • Pre-determined engineering specifications: For dimensional verification, tensile strength, kink resistance, etc.
  • International standards and regulatory guidelines: Such as ISO 80369-7:2016 for Luer Integrity, ISO 10555-1:2013 for leakage and burst pressure, and ISO 10993-1 for biocompatibility.

8. The Sample Size for the Training Set

Not applicable. There is no "training set" in the context of this traditional medical device submission. Training sets are relevant for machine learning or AI models.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).