(30 days)
The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.
The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).
The Millipede 088 Access Catheter is a single lumen, reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.
This document is a 510(k) Premarket Notification from the FDA regarding a medical device called the "Millipede 088 Access Catheter." It is a submission for a new version of an existing device (K231802). The document focuses on demonstrating that the new subject device (K233648) is "substantially equivalent" to this predicate device. This means the primary goal is to show the new device is as safe and effective as the previously cleared one, not to prove its efficacy from scratch through a clinical trial.
Therefore, the requested information about device performance, particularly related to clinical acceptance criteria, expert ground truth, MRMC studies, standalone algorithm performance, and training/test sets, does not directly apply to this type of device submission. This document describes a physical medical device (catheter), not a software algorithm or an AI-powered diagnostic tool. The "performance testing" described refers to mechanical, material, and functional bench tests, not clinical performance metrics in the way one would evaluate an AI diagnostic tool.
The "acceptance criteria" for this device are met by demonstrating substantial equivalence to the predicate device through:
- Identical Indications for Use.
- Similar technological characteristics.
- Successful completion of non-clinical (bench) performance testing and biocompatibility testing.
Here's a breakdown of the closest applicable information from the provided text, while acknowledging that many of the original prompt's requests are not relevant to a physical catheter's 510(k) submission:
1. Table of Acceptance Criteria and Reported Device Performance
The "acceptance criteria" here are design specifications and regulatory standards for the physical and biological properties of the catheter. The "reported device performance" indicates whether the device met these specifications and standards.
| Test (Acceptance Criteria Category) | Test Method | Reported Device Performance (Conclusion) |
|---|---|---|
| Dimensional Inspection | The device dimensions were measured to confirm conformance to the specifications. | The device met established specifications. |
| Visual Inspection | Device surface characteristics were assessed to confirm freedom from defects that could affect clinical use. | The device surface characteristics are suitable for its intended use. |
| Simulated Use Testing | Deliverability and compatibility with accessory devices were evaluated in a neurovascular model. | The device performs as intended under simulated use conditions. |
| Hydrophilic Coating Integrity | The integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles. | The hydrophilic coating integrity is suitable for its intended use. |
| Particulate Recovery | The purpose of this test was to quantify the particulate sizes and counts generated during simulated use of the test article. | The particulate generation was similar to control devices. |
| Kink Resistance | Test specimen segments were formed into a defined bend diameter to evaluate kink resistance. | The device met established specifications. |
| Tip Stiffness | The bending stiffness of the catheter tip was measured to confirm conformance to the specification derived from comparator devices. | The device met established specifications. |
| Air Leakage | Tested per ISO 10555-1:2013 Annex D. | The device integrity is suitable for its intended use. |
| Liquid Leakage | Tested as per ISO 10555-1:2013 Annex C. | The device integrity is suitable for its intended use. |
| Static Burst | Tested as per ISO 10555-1 Annex F. | The device integrity is suitable for its intended use. |
| Catheter Joint Tensile Testing | The tensile strength was evaluated for the bonds between sections of the catheter. | The device met established specifications. |
| Torque Strength | The test specimens were rotated with the distal end constrained from movement to evaluate integrity after rotation. | The device met established specifications. |
| Flow Rate Characterization | The flow rate of saline and a contrast-saline solution was characterized when injected through the catheter. | The flow rate was characterized. (Implies it met acceptable ranges/expectations) |
| Biocompatibility | Cytotoxicity (ISO MEM Elution), Sensitization (ISO Guinea Pig Maximization), Irritation (ISO Intracutaneous Reactivity), Acute Systemic Toxicity (ISO Acute Systemic Injection), Material-Mediated Pyrogenicity, Hemocompatibility (Complement Activation, Partial Thromboplastin Time, ASTM Hemolysis, Thromboresistance) based on ISO 10993-1:2018. | All tests passed; suitable for intended use, non-cytotoxic, non-sensitizing, non-irritating, non-pyrogenic, non-hemolytic, similar thromboresistance to control. |
| Shelf Life | Functional and performance testing after 12-month accelerated aging according to ASTM F1980. | Device performance verified after 12-month accelerated aging. |
| Sterilization | Confirmatory EO residual testing. | EO residuals were within limits specified in ISO 10993-7. |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each bench test (e.g., how many catheters were kink-tested or burst-tested). It refers to standard "test specimens" and "device dimensions."
Data Provenance: The testing was conducted by Perfuze Ltd., which is based in Ireland, suggesting the provenance of the bench test data is from their internal testing or accredited labs they contracted. The tests are "bench" tests, meaning they are performed in a lab setting, not on patient data.
Retrospective or Prospective: These are laboratory bench tests, so the concept of retrospective or prospective data collection doesn't apply as it would for clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This concept is irrelevant for a physical device's bench testing. Ground truth for these tests is established by objective physical measurements against engineering specifications (e.g., using calipers for dimensional inspection, pressure sensors for burst tests, standardized biological assays for biocompatibility).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Bench tests follow predefined protocols and objective measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is for a physical medical device, not an AI diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for a physical medical device, not an AI diagnostic tool.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the bench tests is based on engineering specifications, industry standards (e.g., ISO, ASTM), and validated test methods. For biocompatibility, it's based on biological responses in standardized assays.
8. The sample size for the training set
Not applicable. This is a physical device, not an AI model that undergoes "training."
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
December 14, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Perfuze Ltd. Anne-Marie Gannon Director of Regulatory Affairs Unit 6, Galway Business Park, Dangan Galway, H91 W7CP, Ireland
Re: K233648
Trade/Device Name: Millipede 088 Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: November 14, 2023 Received: November 14, 2023
Dear Anne-Marie Gannon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K233648
Device Name Millipede 088 Access Catheter
Indications for Use (Describe)
The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K233648
Submitter Information
Submitter's Name: Address:
Contact Person: Telephone: Date Prepared:
Subject Device
Proprietary Name: Common/Usual Name: Classification Name: Regulatory Class: Requlation: Product Code:
Predicate Device
- Proprietary Name: Common/Usual Name: Classification Name: Regulatory Class: Requlation: Product Code: Manufacturer: 510(k) Number:
Perfuze Ltd. Unit 6. Galway Business Park, Dangan. Galway, H91 W7CP, Ireland Anne-Marie Gannon +353 91 428083 December12th, 2023
Millipede 088 Access Catheter Guide Catheter Catheter, Percutaneous, Neurovasculature ll 21 CFR 870.1250 QJP
Millipede 088 Access Catheter Guide Catheter Catheter, Percutaneous, Neurovasculature = 21 CFR 870.1250 QJP Perfuze Ltd. K231802
Device Description
The Millipede 088 Access Catheter consists of the catheter, a rotating hemostasis valve (RHV) and a valve crossing tool. The catheter, RHV and valve crossing tool are provided sterile. They are sterilized by ethylene oxide (EO).
The Millipede 088 Access Catheter is a single lumen, reinforced, variable stiffness catheter. The distal segment has a hydrophilic coating for navigation through the vasculature. The catheter has a radiopaque marker located at its distal end for visualization under fluoroscopy. The valve crossing tool is used to open the valve of the access sheath and to facilitate insertion of the Millipede 088 Access Catheter through the access sheath without damage. The RHV is assembled onto the hub of the Millipede 088 Access Catheter and is used to maintain hemostasis during infusion of saline and contrast agent and insertion of other devices through the Millipede 088 Access Catheter.
Indications for Use
The Millipede 088 Access Catheter is indicated for use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the neurovasculature.
The Indications for Use statement for the subject Millipede 088 Access Catheter is identical to the predicate device.
{4}------------------------------------------------
Comparison to the Predicate Device
The intended use of the subject device is identical to the predicate device. The subject and predicate devices have similar technological characteristics as shown in the following table.
| Attribute | Predicate DeviceMillipede 088 Access Catheter (K231802) | Subject DeviceMillipede 088 Access Catheter(K233648) |
|---|---|---|
| RegulationNumber | 21 CFR 870.1250 | Same |
| Regulation Name | Percutaneous catheter | Same |
| Classification | Class II | Same |
| Product Code | QJP | Same |
| Indications forUse | The Millipede 088 Access Catheter isindicated for use in facilitating the insertionand guidance of microcatheters into aselected blood vessel in theneurovasculature. | Same |
| Prescription/Over-the-Counter Use | Prescription | Same |
| DeviceDescription | Single-use, variable stiffness, wire-reinforced catheter with a single lumen.The catheter is comprised of a hollowcylindrical tube bonded at the proximal endto a standard luer fitting. The wall of thetube is constructed using metals andpolymers. A radiopaque marker providesvisual confirmation of the distal tip locationunder fluoroscopy. | Same |
| Principle ofOperation | May be used with support catheters toassist in accessing the targetneurovasculature. | Same |
| Techniques forUse | Standard percutaneous, interventionaltechniques, including access sitepreparation, introduction of the catheterinto the access vessel, advancing thecatheter under fluoroscopy, withdrawingthe catheter, and closing the access site. | Same |
| Materials | Polymers and metals commonly used inthe manufacture of medical devices. | Proximal end of catheter: Samepatient-contacting materials.Distal end of catheter: Samemetal reinforcement withadditional polymer materials indistal tip.Accessories and packaging:Same |
| Distal Tip | Soft, flexible. | Same |
| Catheter WallConstruction | Coil and braid reinforced, with ribbedsurface at distal section. | Same |
| Coating | Hydrophilic coating | Same |
| Catheter Profile | 8 Fr | Same |
| Inner Diameter | Distal: 0.088"Proximal: 0.088" | Same |
| Outer Diameter | Distal: 0.104"Proximal: 0.108" | Same |
| Working Length | 119 cm | 120 cm |
| Attribute | Predicate DeviceMillipede 088 Access Catheter (K231802) | Subject DeviceMillipede 088 Access Catheter(K233648) |
| Packaged Accessories | RHV and Valve Crossing Tool | Same |
| Condition Supplied | Sterile and single use | Same |
| Sterilization Method | Ethylene Oxide (EO), Sterility Assurance Level 10-6 | Same |
| Packaging Configuration | The catheters are placed in a protective polyethylene tube, mounted with accessory RHV and valve crossing tool onto a cardboard packaging card, placed into a pouch, sealed, and labeled. The sealed pouch and Instructions for Use are placed in a labeled shelf carton box. | Same |
| Labelled Shelf Life | 8 months | 12 months |
{5}------------------------------------------------
Performance Testing (Bench)
The successful completion of the performance testing listed in the table below demonstrates that the subject Millipede 088 Access Catheter meets the defined design specifications and is suitable for its intended use.
| Test | Test Method | Conclusions |
|---|---|---|
| DimensionalInspection | The device dimensions were measured toconfirm conformance to the specifications. | The device met establishedspecifications. |
| Visual Inspection | Device surface characteristics wereassessed to confirm freedom from defectsthat could affect clinical use. | The device surfacecharacteristics are suitable forits intended use. |
| Simulated UseTesting | Deliverability and compatibility withaccessory devices were evaluated in aneurovascular model. | The device performs asintended under simulated useconditions. |
| HydrophilicCoating Integrity | The integrity of the hydrophilic coating wasevaluated after multiple insertion andwithdrawal cycles. | The hydrophilic coating integrityis suitable for its intended use. |
| ParticulateRecovery | The purpose of this test was to quantify theparticulate sizes and counts generatedduring simulated use of the test article. | The particulate generationwas similar to control devices. |
| Kink Resistance | Test specimen segments were formed into adefined bend diameter to evaluate kinkresistance. | The device met establishedspecifications. |
| Tip Stiffness | The bending stiffness of the catheter tipwas measured to confirm conformance tothe specification derived from comparatordevices. | The device met establishedspecifications. |
| Air Leakage | Tested per ISO 10555-1:2013 Annex D. | The device integrity is suitablefor its intended use. |
| Liquid Leakage | Tested as per ISO 10555-1:2013 Annex C. | The device integrity is suitablefor its intended use. |
| Static Burst | Tested as per ISO 10555-1 Annex F. | The device integrity is suitablefor its intended use. |
| Catheter JointTensile Testing | The tensile strength was evaluated for thebonds between sections of the catheter. | The device met establishedspecifications. |
| Torque Strength | The test specimens were rotated with thedistal end constrained from movement toevaluate integrity after rotation. | The device met establishedspecifications. |
{6}------------------------------------------------
| Test | Test Method | Conclusions |
|---|---|---|
| Flow RateCharacterization | The flow rate of saline and a contrast-saline solution was characterized wheninjected through the catheter. | The flow rate was characterized |
Biocompatibility
The subject device includes some material differences compared to the predicate device. To ensure that all new materials were assessed, relevant biocompatibility endpoints were addressed through biocompatibility testing of the subject Millipede 088 Access Catheter. The testing was completed in accordance with ISO 10993-1:2018. The table below summarizes the biocompatibility testing completed.
| Test | Results |
|---|---|
| Cytotoxicity - ISO MEM Elution | The test article is non-cytotoxic. |
| Sensitization – ISO Guinea PigMaximization Sensitization Test | The test article did not elicit a sensitization response. |
| Irritation - ISO IntracutaneousReactivity | Requirements of the ISO intracutaneous reactivity testwere met for the test article. |
| Acute Systemic Toxicity – ISO AcuteSystemic Injection | Requirements of the ISO acute systemic injection testwere met for the test article. |
| Material-Mediated Pyrogenicity | The test article is non-pyrogenic. |
| Hemocompatibility – ComplementActivation (SC5b-9) | The test article is not considered to be a potentialactivator of the complement system. |
| Hemocompatibility – PartialThromboplastin Time | The test article is not considered to be an activator ofthe intrinsic coagulation pathway. |
| Hemocompatibility – ASTM Hemolysis | The test article is considered non-hemolytic. |
| Hemocompatibility – Thromboresistance | The test article had similar thromboresistancecharacteristics as the control device. |
Shelf Life and Packaging
The subject device has a labeled shelf life of 12 months, compared to 8 months for the predicate device. Device performance was verified by functional and performance testing after 12-month accelerated aging according to ASTM F1980.
Packaging validation for the predicate device included testing on units subjected to accelerated aging to simulate a 12-month shelf life as per ASTM F1980. This was used to support the labelled 8-month shelf life for the predicate device. The packaging is identical for the subject and predicate devices. Therefore, the original packaging testing for the predicate device remains valid and supports the subject device's 12-month shelf life. No further packaging testing is required for the subject device.
Sterilization
The sterilization method is identical for the subject and predicate devices. Confirmatory EO residual testing was repeated on the subject Millipede 088 Access Catheter to confirm that the design differences did not impact residual EO levels of the testing confirmed that EO residuals were within the limits specified in ISO 10993-7.
Animal Study
No animal study was deemed necessary to demonstrate substantial equivalence between the subject and predicate devices.
Clinical Data
The non-clinical performance data presented were determined to be sufficient to support the substantial equivalence of the subject and predicate devices. Therefore, no clinical study was conducted.
{7}------------------------------------------------
Conclusion
The indications for use of the Millipede 088 Access Catheter are identical to the predicate device. The subject Millipede 088 Access Catheter and the predicate device use the same operating principles and have a similar design. The differences identified in this submission do not raise different or new questions of safety or effectiveness. The successful completion of performance and biocompatibility testing demonstrates that the subject Millipede 088 Access Catheter is substantially equivalent to the predicate device.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).