AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentaryl Citrate, and Select Other Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

These gloves were tested for use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large, Extra Extra Large. Gloves meet the specification of ASTM D6319- 19 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate, and select other drugs as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.

AI/ML Overview

The provided documentation describes the non-clinical performance testing of a medical device, specifically Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs (K233598). The study aims to demonstrate that the device meets acceptance criteria and is substantially equivalent to a predicate device (K223713).

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily defined by various ASTM and ISO standards, and the reported performance indicates that the device passed these criteria. The key performance metrics are related to physical dimensions, physical properties (tensile strength, elongation), freedom from holes, residual powder, permeation by chemotherapy drugs and fentanyl citrate, and biocompatibility.

Methodology / Test PerformedAcceptance CriteriaReported Device Performance (Result)
ASTM D6319-19: Physical Dimensions (Length)Minimum 220mm for size XS and S; 230mm for size M, L, XLPass
ASTM D6319-19: Physical Dimensions (Palm Width)XS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass
ASTM D6319-19: Physical Dimensions (Thickness)Finger: 0.05mm (min); Palm: 0.05mm (min)Pass
ASTM D6319-19, ASTM D412-16 (2021): Physical PropertiesBefore aging: 14MPa, min (Tensile Strength); 500%, min (Ultimate Elongation) After aging: 14MPa, min (Tensile Strength); 400%, min (Ultimate Elongation)Pass
ASTM D6319-19, ASTM D5151-19: Water Leak TestG-I, AQL 2.5 (ISO 2859-1)Pass
ASTM D6319-19, ASTM D6124-06 (2017): Powder ResidueMax 2mg/glovePass
ASTM D6978-05 (2019): Permeation by Chemotherapy DrugsMost Drugs: >240 minutes (e.g., Bendamustine HCl, Bleomycin sulfate, Carboplatin, etc.) Carmustine (3.3 mg/ml): 1.7 minutes ThioTepa (10 mg/ml): 0 minutesMeets stated values (see table)
ASTM D6978-05 (2019): Permeation by Fentanyl CitrateFentanyl Citrate Injection (100 mcg/2ml): >240 minutes>240 min.
ASTM D6978-05 (2019): Permeation by Select Other DrugsChloroquine (50mg/ml), Cyclosporin A (100mg/ml), Retrovir (10 mg/ml): >240 minutes>240 min.
ISO 10993-10:2021: Skin SensitizationNo Skin SensitizationThe sample extract elicits no sensitization reactions under the test conditions.
ISO 10993-23:2021: IrritationNo Skin IrritationUnder the conditions of this study, the irritation response category of the test article is classified as Negligible for polar extract and Negligible for non-polar extract.
ISO 10993-11:2017: Acute Systemic Toxicity StudySubject showed no adverse biological reactionNo death or acute systemic toxicity was observed in the mice which received injection of the extract.

Important Note on Chemotherapy Drugs: The document explicitly states: "Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 1.7 minutes and Thiotepa: 0 minute. Warning: Do not use with Carmustine and Thiotepa." This means that while the glove was tested against these drugs, its performance for them is below the typical acceptable threshold for chemotherapy gloves (which is usually >240 minutes for other drugs), and this is a critical aspect of its labeling and intended use.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the exact sample sizes (number of gloves or individual tests) used for each performance test. It only refers to the standards (e.g., ASTM D6319-19, ASTM D6978-05), which would specify the required sample sizes for testing. For instance, ASTM D6319-19 defines AQL (Acceptable Quality Level) for certain tests (e.g., G-I, AQL 2.5 for water leak test), which implies specific sampling plans.
  • Data Provenance: The tests were non-clinical performance studies. The data provenance is linked to the manufacturer, Anhui Intco Medical Products Co., Ltd., China. The tests are described as having been "conducted to verify that the proposed device met all design specifications," implying a prospective testing approach within a laboratory setting, rather than retrospective analysis of existing data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of device and study. The "ground truth" for glove performance (e.g., tensile strength, breakthrough time, freedom from holes) is established through standardized laboratory testing methods (ASTM, ISO), not through expert consensus or clinical judgment of images or patient data. The results are objective, measurable physical properties and chemical resistance.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies, particularly in image interpretation, where human readers might disagree, and a consensus or majority opinion is needed to establish ground truth. For the physical and chemical performance tests of gloves, the results are quantitative and objective, determined by laboratory equipment and standardized procedures.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers interpret medical images or data. The device in question is a medical glove, a physical barrier, not a diagnostic tool requiring human interpretation with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, this is not applicable. The device is a physical product (a glove), not an algorithm or software. Therefore, there is no "algorithm only" performance to evaluate. The tests performed are on the physical properties and chemical resistance of the glove itself.

7. The Type of Ground Truth Used

The ground truth used for these tests is based on:

  • Standardized Measurement: Directly measurable physical properties (e.g., length, width, thickness, tensile strength, elongation, residual powder).
  • Controlled Chemical Permeation Tests: Breakthrough detection times for specific hazardous chemicals (chemotherapy drugs, fentanyl citrate).
  • Biological Testing: In-vivo (animal) and in-vitro tests for biocompatibility (skin irritation, sensitization, acute systemic toxicity) as defined by ISO standards.

Essentially, the "ground truth" is defined by the established and validated protocols of the cited ASTM and ISO standards for testing medical gloves.

8. The Sample Size for the Training Set

This information is not applicable. This submission is for a physical medical device (a glove) that undergoes manufacturing processes and performance testing, not an artificial intelligence (AI) or machine learning (ML) algorithm. Therefore, there is no "training set" in the context of AI/ML.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

January 26, 2024

Anhui Intco Medical Products Co., Ltd. Max Li Quality Specialist No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park Suixi County Huaibei City, Anhui 235000 China

Re: K233598

Trade/Device Name: Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, QDO, OPJ Dated: November 9, 2023 Received: November 9, 2023

Dear Max Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director

{2}------------------------------------------------

DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233598

Device Name

Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentaryl Citrate, and Select Other Drugs

Indications for Use (Describe)

Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentaryl Citrate, and Select Other Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

These gloves were tested for use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Test Chemotherapy drug & ConcentrationMinimum Breakthrough Detection Times (Minutes)
Bendamustine HCI; 5 mg/ml>240
Bleomycin sulfate; 15 mg/ml>240
Busulfan; 6 mg/ml>240
Carboplatin; 10 mg/ml>240
Carfilzomib; 2 mg/ml>240
Carmustine; 3.3 mg/ml1.7
Cetuximab; 2 mg/ml>240
Cisplatin; 1 mg/ml>240
Cladribine; 1 mg/ml>240
Cyclophosphamide; 20 mg/ml>240
Cytarabine (Cytosine); 100 mg/ml>240
Cytovene (Ganciclovir); 10 mg/ml>240
Dacarbazine; 10 mg/ml>240
Daunorubicin HCL; 5 mg/ml>240
Decitabine; 5 mg/ml>240
Docetaxel; 20 mg/ml>240
Doxorubicin HCL; 2 mg/ml>240
Epirubicin HCL; 2 mg/ml>240
Etoposide; 20 mg/ml>240
Fludarabine Phosphate; 25 mg/ml>240
Fluorouracil; 50 mg/ml>240
Fulvestrant; 50 mg/ml>240
Gemcitabine HCI; 38 mg/ml>240
Idarubicin HCI; 1 mg/ml>240
Ifosfamide; 50 mg/ml>240
Irinotecan HCI; 20 mg/ml>240
Mechlorethamine HCL; 1 mg/ml>240
Melphalan HCI; 5 mg/ml>240
Methotrexate; 25 mg/ml>240
Mitomycin C; 0.5 mg/ml>240
Mitoxantrone HCI; 2 mg/ml>240
Oxaliplatin; 5 mg/ml>240
Paclitaxel; 6 mg/ml>240

{4}------------------------------------------------

Pemetrexed ; 25 mg/ml>240
Rituximab ; 10 mg/ml>240
ThioTepa ; 10 mg/ml0
Topotecan HCL ; 1 mg/ml>240
Trisonex ; 1 mg/ml>240
Velcade (Bortezomib) ; 1 mg/ml>240
Vidaza (Azacytidine) ; 25 mg/ml>240
Vincrinstine Sulfate ; 1 mg/ml>240
Vinorelbine Tartrate, 10 mg/ml>240
Zoledronic Acid ; 0.8 mg/ml>240

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU): 1.7 minutes and Thiotepa:0 minute

Warning: Do not use with Carmustine and Thiotepa

Fentanyl Citrate & Other DrugsMinimum Breakthrough Detection Times
Fentanyl Citrate Injection(100 mcg/2ml)>240
Chloroquine ; 50mg/ml>240
Cyclosporin A; 100mg/ml>240
Retrovir ; 10 mg/ml>240
Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui Province, China

K233598

510(K)SUMMARY

(As required by 21 CFR 807.92) K233598

Date Prepared: Jan. 25. 2024

1. Owner's Identification:

Applicant Name: ANHUI INTCO MEDICAL PRODUCTS CO., LTD.

Location: No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui Province ,China

Contact Person: Max Li Tel: +86-18918364816

2. Name of the Device:

Trade / Product Name: Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs

Common Name: Exam Gloves Classification Name: Non-Powdered Patient Examination Glove Specialty Regulation: 21 CFR 880.6250 Product Code: LZA, LZC, QDO, OPJ Classification Panel: General Hospital Device Class: Class I

3. Predicate Device Information:

Better Care Plastic Technology Co., Ltd. Powder Free Nitrile Examination Glove (Grey) Tested for Use with Chemotherapy Drugs and Fentany) Citrate (K223713) Product code LZA, LZC, QDO, OPJ.

4. Device Description:

Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large, Extra Extra Large. Gloves meet the specification of ASTM D6319- 19 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate, and select other drugs as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.

5. Indications for Use:

{6}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui Province, China

K233598

Powder Free Nitrile Examination Glove 2.8 Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

These gloves were tested for use with Chemotherapy Drugs, Fentanyl Citrate, and Select Other Drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugsMinimum Breakthrough Detection Time(minutes)
Bendamustine HCl; 5 mg/ml>240 min.
Bleomycin sulfate; 15 mg/ml>240 min.
Busulfan; 6 mg/ml>240 min.
Carboplatin; 10 mg/ml>240 min.
Carfilzomib; 2 mg/ml>240 min.
Carmustine; 3.3 mg/ml1.7
Cetuximab; 2 mg/ml>240 min.
Cisplatin; 1 mg/ml>240 min.
Cladribine; 1 mg/ml>240 min.
Cyclophosphamide; 20 mg/ml>240 min.
Cytarabine (Cytosine); 100 mg/ml>240 min.
Cytovene (Ganciclovir); 10 mg/ml>240 min.
Dacarbazine; 10 mg/ml>240 min.
Daunorubicin HCL; 5 mg/ml>240 min.
Decitabine; 5 mg/ml>240 min.
Docetaxel; 20 mg/ml>240 min.
Doxorubicin HCl; 2 mg/ml>240 min.
Epirubicin HCl; 2 mg/ml>240 min.
Etoposide; 20 mg/ml>240 min.
Fludarabine Phosphate; 25 mg/ml>240 min.
Fluorouracil; 50 mg/ml>240 min.
Fulvestrant; 50 mg/ml>240 min.
Gemcitabine HCl; 38 mg/ml>240 min.
Idarubicin HCl; 1 mg/ml>240 min.
Ifosfamide; 50 mg/ml>240 min.
Irinotecan HCl; 20 mg/ml>240 min.
Mechlorethamine HCL; 1 mg/ml>240 min.
Melphalan HCl; 5 mg/ml>240 min.
Methotrexate; 25 mg/ml>240 min.
Mitomycin C; 0.5 mg/ml>240 min.
Mitoxantrone HCl; 2 mg/ml>240 min.
Oxaliplatin;5 mg/ml>240 min.
Paclitaxel; 6 mg/ml>240 min.
Pemetrexed; 25 mg/ml>240 min.
Rituximab; 10 mg/ml>240 min.
Thio Tepa; 10 mg/ml0
Topotecan HCl; 1 mg/ml>240 min.
Trisenox (Arsenic Trioxide); 1 mg/ml>240 min.
Velcade (Bortezomib); 1 mg/ml>240 min.
Vidaza (Azacytidine); 25mg/ml>240 min.
Vincristine Sulfate; 1 mg/ml>240 min.
Vinorelbine Tartrate; 10 mg/ml>240 min.
Zoledronic Acid; 0.8 mg/ml>240 min.

Table 1

{7}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui Province, China

K233598

*Please note that the following drugs have extremely low permeation times:

Carmustine: 1.7 minutes, Thio Tepa: 0 minutes

Warning: Do not use with Carmustine and Thio Tepa.

Fentanyl Citrate & Other DrugsMinimum Breakthrough Detection Time(minutes)
Fentanyl Citrate Injection(100 mcg/2ml)>240 min.
Chloroquine; 50mg/ml>240 min.
Cyclosporin A; 100mg/ml>240 min.
Retrovir; 10 mg/ml>240 min.

6. Comparison of Subject Device and Predicate Device:

The following tables are summaries of the technological characteristics, biocompatibility and

performance testing of the proposed and predicate devices. General Comparison Table:

DeviceProposed DevicePredicate DeviceComparison
510K #K233598K223713-
Product NamePowder Free Nitrile ExaminationGlove 2.8 Blue Color, Non-Sterile,and Tested for Use withChemotherapy Drugs, FentanylCitrate, and Select Other DrugsPowder Free Nitrile ExaminationGlove (Grey) Tested for Use withChemotherapy Drugs and FentanylCitrate-
Product CodeLZA, LZC, QDO, OPJLZA, LZC, QDO, OPJSame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
Indications forusePowder Free Nitrile ExaminationGlove 2.8 Blue Color, Non-Sterile,and Tested for Use withChemotherapy Drugs, FentanylCitrate, and Select Other Drugs is adisposable device intended formedical purposes that is worn on theexaminer's hand or finger to preventcontamination between patient andexaminer.These gloves were tested for usewith Chemotherapy Drugs,Fentanyl Citrate, and Select OtherDrugs as per ASTM D6978-05(Reapproved 2019) StandardPractice for Assessment ofThe glove is a disposable deviceintended for medical purposes that isworn on the examiner's hand toprevent contamination betweenpatient and examiner. Gloves havebeen tested for use with chemotherapydrugs and Fentanyl Citrate.Same

{8}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui Province, China

K233598

Medical Gloves to Permeation byChemotherapy Drugs.
Powder freeYesYesSame
Design featureAmbidextrousAmbidextrousSame
MaterialNitrileNitrileSame
ColorBlueGreyDifferent
SizeXS, S, M, L, XL, XXLXS, S, M, L, XLDifferent
SterileNon-SterileNon-SterileSame
UseSingle useSingle useSame
ChemotherapyDrugs,FentanylCitrate, andSelect OtherDrugs ClaimSee below comparison tableSee below comparison tableSimilar

Analysis: The proposed device has size XS, S, M, L, XL, XXL and Blue color, while the predicate device has size XS, S, M, L, XL and Grey color. But safety and performance testing has been done to the proposed device and the results showed that the device meets the requirements of standard ASTM D6319-19. Therefore, this difference does not raise any new safety or performance questions.

Technological Characteristic and Biocompatibility Comparison Table:

ItemStandardProposed devicePredicate device K223713Result
DimensionASTM D6319-19Minimum 220mm for sizeXS and S,230mm for size M, L, XLWidth(mm):XS: 70±10S: 80±10M: 95±10L: 110±10XL: 120±10XXL: 130±10Thickness(mm):Palm: Minimum 0.05Finger: Minimum 0.05Length: Min 230mmWidth(mm):XS: 70±10S: 80±10M: 95±10L: 110±10XL: 120±10XXL: /Thickness:(mm)Palm: Minimum 0.05Finger: Minimum 0.05Similar
Physical propertiesBefore agingTensile strengthUltimate elongationAfter agingTensile strengthUltimate elongationASTM D6319-19ASTM D412-1614MPa, min500%, min14MPa, min400%, min14MPa, min500%, min14MPa, min400%, minSame

{9}------------------------------------------------

No. 6, Haitang South Road, Suixi Wedical Products Co.,Ltd.
No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui
Province, Province,

vince, China

Freedom from holesASTM D6319-19ASTM D5151-19G-I, AQL2.5G-I, AQL2.5Same
Residual PowderASTM D6319-19ASTM D6124-06≤2 mg per glove≤2 mg per gloveSame
BiocompatibilityIrritationISO 10993-23Under the conditions of thisstudy, the irritation responsecategory of the test article isclassified as Negligible forpolar extract and Negligiblefor non-polar extract.Based on the criteria, the testarticle extracts wereconsidered negligible irritants.Similar(1)
Sensitization ISO10993-10The sample extract elicits nosensitization reactions underthe test conditionsBased on the criteria andconditions of the study, thetest article is classified as anon-sensitizerSimilar
Acute SystemicToxicity Test ISO10993 -11No death or systemic toxicitywas observed in the micewhich received injection ofthe extractBased on the criteria, the testarticle meets therequirementsSimilar
  1. The new version of standard for Skin Irritation we refer is ISO 10993-23:2021, the test method and process is the old version ISO 10993-10

Chemotherapy Permeation, Fentanyl Citrate, and Select Other Drugs Comparison Claim:

Chemotherapy Drugs, Fentanyl Citrate,and Select Other DrugsMinimum BDT (Minutes)
Proposed devicePredicate deviceK223713Remark
Bendamustine HCI ; 5 mg/ml>240 min.>240 min.Same
Bleomycin sulfate ; 15 mg/ml>240 min.>240 min.Same
Busulfan ; 6 mg/ml>240 min.>240 min.Same
Carboplatin; 10 mg/ml>240 min.>240 min.Same
Carfilzomib ; 2 mg/ml>240 min.>240 min.Same
Carmustine ; 3.3 mg/ml1.713.9Different*
Cetuximab ; 2 mg/ml>240 min.>240 min.Same
Chloroquine ; 50mg/ml>240 min.--Different*
Cisplatin ; 1 mg/ml>240 min.>240 min.Same
Cladribine ; 1 mg/ml>240 min.--Different*
Cyclophosphamide; 20 mg/ml>240 min.>240 min.Same
Cyclosporin A; 100mg/ml>240 min.--Different*
Cytarabine (Cytosine) ; 100 mg/ml>240 min.>240 min.Same
Cytovene (Ganciclovir); 10 mg/ml>240 min.>240 min.Same
Dacarbazine ; 10 mg/ml>240 min.>240 min.Same
Daunorubicin HCL; 5 mg/ml>240 min.>240 min.Same
Decitabine ; 5 mg/ml>240 min.>240 min.Same
Docetaxel ; 20 mg/ml>240 min.>240 min.Same
Province, China
K233598
Doxorubicin HCL ; 2 mg/ml>240 min.>240 min.Same
Epirubicin HCL ; 2 mg/ml>240 min.>240 min.Same
Etoposide ; 20 mg/ml>240 min.>240 min.Same
Eribulin Mesylate; 0.5 mg/ml-->240 min.Different*
Fludarabine Phosphate; 25 mg/ml>240 min.>240 min.Same
Fluorouracil ; 50 mg/ml>240 min.>240 min.Same
Fulvestrant ; 50 mg/ml>240 min.>240 min.Same
Gemcitabine HCI ; 38 mg/ml>240 min.>240 min.Same
Idarubicin HCI ; 1 mg/ml>240 min.>240 min.Same
Ifosfamide ; 50 mg/ml>240 min.>240 min.Same
Irinotecan HCI ; 20 mg/ml>240 min.>240 min.Same
Mechlorethamine HCL; 1 mg/ml>240 min.>240 min.Same
Melphalan HCI; 5 mg/ml>240 min.>240 min.Same
Methotrexate ; 25 mg/ml>240 min.>240 min.Same
Mitomycin C ; 0.5 mg/ml>240 min.>240 min.Same
Mitoxantrone HCI ; 2 mg/ml>240 min.>240 min.Same
Oxaliplatin ;5 mg/ml>240 min.>240 min.Same
Pacilitaxel ; 6 mg/ml>240 min.>240 min.Same
Pemetrexed ; 25 mg/ml>240 min.>240 min.Same
Pertuzumab; 30 mg/ml-->240 min.Different*
Raltitrexed monohydrate-->240 min.Different*
Retrovir ; 10 mg/ml>240 min.>240 min.Same
Rituximab ; 10 mg/ml>240 min.>240 min.Same
Temsirolimus ; 25mg/ml-->240 min.Different*
ThioTEPA ; 10 mg/ml044.4Different*
Topotecan HCL ; 1 mg/ml>240 min.>240 min.Same
Trastuzumab ; 21mg/ml-->240 min.Different*
Trisenox (Arsenic Trioxide); 1 mg/ml>240 min.>240 min.Same
Velcade (Bortezomib); 1 mg/ml>240 min.>240 min.Same
Vidaza (Azacytidine); 25mg/ml>240 min.>240 min.Same
Vinblastine ; 1mg/ml-->240 min.Different*
Vincristine Sulfate ; 1 mg/ml>240 min.>240 min.Same
Vinorelbine Tartrate; 10 mg/ml>240 min.>240 min.Same
Zoledronic Acid ; 0.8 mg/ml>240 min.>240 min.Same
Fentanyl; 100mcg/2ml>240 min.>240 min.Same
Simulated Gastric Acid------
Fluid/Fentanyl Citrate Injection Mix50/50 Solution-->240 min.Different*

{10}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui
Province, China

  • Chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this difference does not raise questions of safety and effectiveness.

7. Summary of Non-Clinical Performance Data

{11}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Huaibei City, Anhui
Province, China

K233598

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

MethodologyTest PerformedAcceptance CriteriaResults
ASTM D6319-19Physical DimensionsLengthMinimum 220mm for size XSand S, 230mm for size M, L, XLPass
ASTM D6319-19Physical Dimensions PalmWidthXS: 70±10mmS: 80±10mmM: 95±10mmL:110±10mmXL: 120±10mmXXL: 130±10mmPass
ASTM D6319-19Physical DimensionsThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)Pass
ASTM D6319-19ASTM D412-16 (2021)Physical PropertiesBefore aging: 14MPa, min500%, minAfter aging: 14MPa, min400%, minPass
ASTM D6319-19ASTM D5151-19Water leak testG-I, AQL 2.5 (ISO 2859-1)Pass
ASTM D6319-19ASTM D6124-06 (2017)Powder ResidueMax 2mg/glove
ASTM D6978-05 (2019)Permeation byChemotherapy Drugsand Fentanyl CitrateRefer the above table 1
ISO 10993-10:2021Skin SensitizationNo Skin sensitizationThe sample extractelicits no sensitizationreactions under thetest conditions
ISO 10993-23:2021IrritationNo Skin irritationUnder the conditionsof this study, theirritation responsecategory of the test
article is classified asNegligible for polarextract and Negligiblefor non-polar extract.
ISO 10993-11:2017Acute systemic toxicitystudySubject showed no adversebiological reactionNo death or acute systemictoxicity was observedin the mice whichreceived injection ofthe extract

{12}------------------------------------------------

No. 6, Haitang South Road, Suixi Wuhu Modern Industrial Park, Suixi County, Hualbei City, Anhui Province, China

K233598

ASTM D6319- 19, Standard Specification for Nitrile Examination Gloves for Medical Application. .

  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D412-16 (2021) Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— Tension

ASTM D6978-05 (Reapproved 2019), Assessment of Reissuance of Medical Gloves to Permeation by Chemotherapy Drugs.

  • ISO 10993-10: 2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization.
  • ISO 10993-23:2021 Biological evaluation of medical devices Part 23 : Tests for irritation.
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity

8. Summary of Clinical Testing:

Not provided for the subject device.

9. Conclusion:

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K223713, Powder Free Nitrile Examination Glove (Grey) Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.