(236 days)
Not Found
No
The summary describes a physical medical device (angiographic catheter) used for delivering contrast media and potentially for anatomical measurements. There is no mention of software, algorithms, or data processing that would suggest the use of AI or ML. The performance studies focus on the physical properties and functionality of the catheter itself.
No
The device is used for delivering radiopaque media for diagnostic procedures and anatomical measurements, not for treating a disease or condition.
Yes.
The device's Intended Use states it is for "delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures" and that "Angiographic catheters with marker bands may also be used for anatomical measurements", indicating its role in diagnosis. The Device Description also explicitly calls them "intravascular diagnostic catheters".
No
The device description clearly describes a physical catheter, which is a hardware component, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to deliver radiopaque media to selected sites in the vascular system for diagnostic procedures (angiography). This is an in vivo procedure, meaning it's performed within the living body.
- Device Description: The device is an intravascular catheter designed for administering contrast for fluoroscopic studies. This further confirms its use within the body.
- Lack of IVD Characteristics: IVD devices are designed to examine specimens (like blood, urine, tissue) outside the body to provide information about a person's health. This device does not perform any analysis on specimens in vitro.
Therefore, the Impress Angiographic Catheter is a medical device used for in vivo diagnostic imaging procedures, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The Impress Angiographic Catheter is designed to be used for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures. Angiographic catheters with marker bands may also be used for anatomical measurements.
Product codes (comma separated list FDA assigned to the subject device)
DQO
Device Description
The Merit Impress Angiographic Catheters with/without hydrophilic coating are intravascular diagnostic catheters that are intended for the administration of contrast for conducting fluoroscopic studies. The catheters are available in a variety of 4F and 5F wire-braided and non-braided configurations ranging in length from 40cm to 125cm. The device may include a marker band to assist anatomical measurements. The devices are available in a variety of tip shapes to cater to variations in physician preference and patient anatomy.
The catheters consist of a shaft with a molded hub assembly. It is the shaft that may be offered with or without a wire-braided reinforcement. The distal tip of the device is tapered for efficient volume flow and dispersion of the contrast media. The distal tip is flexible so as to minimize the potential for vessel trauma.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
vascular system
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance testing of the subject Impress Angiographic Catheter was conducted based on the risk analysis and based on the requirements of several international standards (ISO 10555-1:2013-1, ISO 80369-7:2021, ISO 10993-1:2018, ISO 10993-4:2017, ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-11:2017, ASTM F756-13, ISO 14971:2019). The tests were performed on both representative Impress Catheter samples, including sterile devices (following ethylene oxide processing) and aged devices (3-year simulated). All samples were manufactured in accordance with existing and validated processes. All tested samples met pre-established performance criteria and were deemed acceptable.
Safety & Performance Tests (Design Verification Studies):
- Shaft O.D. meets Specifications
- Confirmation of Working Length Catheter
- Pressure Assessment for Maximum Flow Rate
- Static Pressure Rating
- Force at Break - Tensile; Hub-to-Shaft
- Peak Tensile; Hub-to-Shaft
- I.D. of Assembly / Compatibility with Mandrel
- Hub Defects Assessment
- Kinking Assessment of Strain Relief
- Printing on Hub
- Stiffness/Torque-ability after 37°C water bath
- Catheter Flex Assessment
- Shaft O.D. after Hydration of Hydrophilic Coating in water bath
- Coating Degradation
- Tip Free of Defects
- Tip Curves adherence to templates
- Friction Assessment of Coated Shafts
- Overall Assessment for defects
- Leak Test
- Air Leakage to Hub Assembly
Design Validation Studies (Clinician feedback following assessment of design changes were collected):
- Optical Density
- Catheter Push-ability
- Catheter Track-ability
- Catheter Torque-ability
- Kink Resistance
- Catheter Stiffness
- Tip Flexibility
- Catheter Compatibility
- Withdrawal from Model
- Simulated Use in accord with IFU
- Useability Assessment
Biocompatibility Studies:
- Cytotoxicity
- Intracutaneous Irritation (Extract)
- Sensitization ISO Guinea Pig Maximization
- Acute Systemic Toxicity
- Pyrogenicity Material Mediated: Rabbit Study
- Hemolysis
- Complement Activation (SCb5-9)
- In Vitro Blood Loop Assasy w/ Comparison Article (3xAnimal)
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
No reference devices are identified for this submission.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1200 Diagnostic intravascular catheter.
(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
May 22, 2024
Merit Medical Systems, Inc. Gamaliel Figueroa Manager, Regulatory Affairs 1600 West Merit Parkway South Jordan, Utah 84095
Re: K233268
Trade/Device Name: Impress Angiographic Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: DQO Dated: September 28, 2023 Received: September 29, 2023
Dear Gamaliel Figueroa:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
1
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Gregor Gregory W. Date: 2024.05.22 09:46:10 O'connell -S -04'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Indications for Use
510(k) Number (if known) K233268
Device Name Impress Angiographic Catheter
Indications for Use (Describe)
The Impress Angiographic Catheter is designed to be used for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic catheters with marker bands may also be used for anatomical measurements.
Type of Use (Select one or both, as applicable) | |
---|---|
☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
K233268 510(k) Summary
| General
Provisions | ||
---|---|---|
Submitter Name: | ||
Address: | ||
Telephone Number: | ||
Contact Person: | ||
Date Prepared: | ||
Registration Number: | Merit Medical Systems, Inc. | |
1600 West Merit Parkway | ||
South Jordan, UT 84095 | ||
(801) 208-4583 | ||
Garry A. Courtney | ||
09/28/2023 | ||
1721504 | ||
Subject Device | ||
Trade Name: | ||
Common/Usual Name: | ||
Classification Name: | ||
Regulatory Class: | ||
Product Code: | ||
21 CFR §: | ||
Review Panel: | Impress Angiographic Catheter | |
Angiographic Catheter | ||
Diagnostic Intravascular Catheter | ||
2 | ||
DQO | ||
870.1200 | ||
Cardiovascular | ||
Predicate Devices | Predicate 1 | |
Trade Name: | ||
Classification Name: | ||
Premarket Notification: | ||
Manufacturer: | Merit Impress Angiographic Catheter | |
Diagnostic Intravascular Catheter | ||
K191608 | ||
Merit Medical Systems, Inc. | ||
Predicate 2 | ||
Trade Name: | ||
Classification Name: | ||
Premarket Notification: | ||
Manufacturer: | Merit Impress Angiographic Catheter | |
Diagnostic Intravascular Catheter | ||
K053171 | ||
Merit Medical Systems, Inc. | ||
The predicate devices have not been subject to design-related recall | ||
and/or field actions. | ||
Reference Device | No reference devices are identified for this submission. |
4
| Device
Description | The Merit Impress Angiographic Catheters with/without hydrophilic
coating are intravascular diagnostic catheters that are intended for the
administration of contrast for conducting fluoroscopic studies. The
catheters are available in a variety of 4F and 5F wire-braided and non-
braided configurations ranging in length from 40cm to 125cm. The device
may include a marker band to assist anatomical measurements. The
devices are available in a variety of tip shapes to cater to variations in
physician preference and patient anatomy.
The catheters consist of a shaft with a molded hub assembly. It is the
shaft that may be offered with or without a wire-braided reinforcement.
The distal tip of the device is tapered for efficient volume flow and
dispersion of the contrast media. The distal tip is flexible so as to
minimize the potential for vessel trauma. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for
Use | There is no change in the Indications for Use Statement from the
predicate devices to the subject device.
The Impress Angiographic Catheter is designed to be used for
delivering radiopaque media to selected sites in the vascular system in
conjunction with routine diagnostic procedures. Angiographic catheters
with marker bands may also be used for anatomical measurements. |
5
The design and technological characteristics of the subject modified Impress Angiographic Catheters are equivalent to the predicate Impress Angiographic Catheters. The only differences between the subject catheters and the predicate catheters are that different materials are used in the subject (modified) devices:
- The material used in the construct of the catheter tip is changing . from one Pebax 5533 formula (SP01) to a different material. Pebax 5533 (SA01 MED). The use of the new Pebax material is fully supported by verification and validation studies, and an extensive biocompatibility assessment.
- The material used in the construct of the catheter shaft is changing o from one Pebax 7233 formula (SP01) to a different Pebax 7233 formula (SA01 MED). The use of the new Pebax material is fully supported by verification and validation studies, and an extensive biocompatibility assessment.
Comparison to Predicate Device
- The material used in the construct of the HUB is changing from a polycarbonate/Xylex resin to polycarbonate resin without Xylex. The use of the polycarbonate resin is fully supported by verification and validation studies, and an extensive biocompatibility assessment.
- The adhesive used in the construct of the HUB is changing from . cyanoacrylate adhesive to Loctite 4306 adhesive. The use of the new Loctite adhesive is fully supported by verification and validation studies, and an extensive biocompatibility assessment.
The comparison between the subject and the predicate devices is based on the following:
- Same intended use ●
- Same indications for use
- Same sterilization methods ●
- Same packaging scheme
- Same labeling ●
- Same design ●
- Same fundamental technology/principles of operation ●
- Similar material types (all meet ISO 10993 biocompatibility ● requirements)
6
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. Performance testing of the subject Impress Angiographic Catheter was conducted based on the risk analysis - and based on the requirements of the following international standard:
-
- ISO 10555-1:2013-1, Intravascular Catheters Sterile and single-use catheters – Part 1: General requirements.
-
- ISO 80369-7:2021, Small Bore Connectors for Liquids and Gases in Healthcare Applications – Part 7: Connectors for Intravascular or Hypodermic Applications.
-
- ISO 10993-1:2018, Biological Evaluation of Medical Devices Part I : Evaluation and Testing within a risk management process, and FDA guidance Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices.
-
- ISO 10993-4:2017, Biological Evaluation of Medical Devices Part 4: Selection of Tests for Interaction with Blood.
-
- ISO 10993-5:2009, Biological Evaluation of Medical Devices Part 5: Tests for In-Vitro Cytotoxicity
-
- ISO 10993-10:2021, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization
-
- ISO 10993-11:2017, Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicitv
-
- ASTM F756-13, Tests for Interaction with Blood: Assessment of Hemolytic Properties of Materials
-
- ISO 14971:2019, Medical Devices Application of Risk Management to Medical Devices.
7
| | The tests listed below were performed to demonstrate that the modified
device meets product specification criteria, and to demonstrate there
were no unacceptable/unmitigated risks associated with the changes
made to the device. |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The tests were performed on both representative Impress Catheter
samples. Further, the testing was performed on sterile devices (following
ethylene oxide processing) and aged devices (3-year simulated). All
samples were manufactured in accordance with existing and validated
processes and are representative of product that Merit Medical intends
to commercialize. |
| | All tested samples met pre-established performance criteria and were
deemed acceptable. |
| Safety &
Performance
Tests | Design Verification Studies
• Shaft O.D. meets Specifications
• Confirmation of Working Length Catheter
• Pressure Assessment for Maximum Flow Rate
• Static Pressure Rating
• Force at Break - Tensile; Hub-to-Shaft
• Peak Tensile; Hub-to-Shaft
• I.D. of Assembly / Compatibility with Mandrel
• Hub Defects Assessment
• Kinking Assessment of Strain Relief
• Printing on Hub
• Stiffness/Torque-ability after 37°C water bath
• Catheter Flex Assessment
• Shaft O.D. after Hydration of Hydrophilic Coating in water bath
• Coating Degradation
• Tip Free of Defects
• Tip Curves adherence to templates
• Friction Assessment of Coated Shafts
• Overall Assessment for defects
• Leak Test
• Air Leakage to Hub Assembly
|
8
Design Validation Studies
Clinician feedback following assessment of design changes were collected for the following:
- o Optical Density
- Catheter Push-ability .
- Catheter Track-ability ●
- Catheter Torque-ability ●
- Kink Resistance ●
- Catheter Stiffness ●
- Tip Flexibility ●
- Catheter Compatibility .
- Withdrawal from Model .
- Simulated Use in accord with IFU .
- Useability Assessment .
Biocompatibility Studies
- Cytotoxicity ●
- Intracutaneous Irritation (Extract) .
- Sensitization ISO Guinea Piq Maximization .
- Acute Systemic Toxicity .
- Pyrogenicity Material Mediated: Rabbit Study ●
- Hemolysis ●
- Complement Activation (SCb5-9) ●
- In Vitro Blood Loop Assay w/ Comparison Article (3xAnimal) .
Summary of Substantial Equivalence
Based on the intended use, product indications, design, materials, safety and performance testing (verification and validation), and materials, the subject Impress Angiographic Catheters are deemed to be substantially equivalent to the predicate Impress Angiographic Catheters (K191608 and K053171). Both the subject devices and the predicate devices are legally manufactured by Merit Medical Systems, Inc.