AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Device Description

The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system.

The subject device is categorized into two types of models according to its usage.

Reusable Blood Pressure Cuff models:

1201B1,12L2B1,1202B1,12L3B1,1204B1,1205B1,1206B1,1207B1-1#,1207B1-2#,1207B1-

3#,1207B1-4#,1207B15#,1201W2,12L2W2,1202W2,12L3W2,1203W2,

1204W2,1205W2,1206W2 and 1207W2

Disposable Blood Pressure Cuff models:

1201D1,12L2D1,1202D1,12L3D1,1204D1,1205D1,1204D1,1207D1-1#,1207D1-2#,1207D1-

3#,1207D1-4#,1207D1-5#

Connector models: PLINMA-03-127-003, PLINMA-03-T, PLINMA-03-Z3

AI/ML Overview

The provided text is a 510(k) summary for a Blood Pressure Cuff. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study with detailed performance metrics. The information about a study proving device performance is limited.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily refers to meeting ISO standards and does not present a table of specific acceptance criteria (e.g., specific accuracy ranges for blood pressure measurements) versus reported device performance in a clinical setting. It states that the device was "tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type". However, it does not provide the results of this testing in a quantifiable form or against specific, pre-defined acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance:

The document states, "The subject and predicate devices are exactly the same. Since the blood pressure cuffs are identical, no further clinical testing is necessary." This implies that no separate test set with human subjects was used for this 510(k) submission to evaluate the performance of the device in terms of blood pressure measurement accuracy. The data provenance is assumed to be from the predicate device's testing, but details are not provided here.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

Since no separate clinical performance testing was conducted for this submission, no experts were used to establish ground truth from newly collected data for this specific device.

4. Adjudication Method for the Test Set:

Not applicable, as no dedicated clinical performance test set was used for this submission.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable, as this is a blood pressure cuff and not an AI-powered diagnostic imaging device. The document does not describe any human reader involvement or AI assistance.

6. Standalone Performance Study:

A standalone performance study focused on technical compliance with ISO 81060-1 was mentioned ("The subject device has been tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type"). However, the results of this testing, which would quantify its performance against the ISO standard's requirements, are not provided in this summary. The summary only attests that the device "met the requirements."

7. Type of Ground Truth Used:

For the "performance data" mentioned, the ground truth would be the reference measurements and methods defined within the ISO 81060-1 standard for non-invasive sphygmomanometers. For the biocompatibility testing, the ground truth would be the established scientific methods and criteria for evaluating cytotoxicity, sensitization, and irritation.

8. Sample Size for the Training Set:

Not applicable, as this is a medical device (blood pressure cuff) and not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable.

Summary of Acceptance Criteria and Study (based on available information):

Acceptance Criteria (Implied)Reported Device Performance (Summary)
Compliance with ISO 81060-1 for non-automated measurement type"The subject device has been tested according to ISO 81060-1... and met the requirements."
Biocompatibility (Cytotoxicity, Sensitization, Irritation)"The test results show that the proposed device has no issue of cytotoxicity, sensitization, or skin irritation."
Functional equivalence to predicate devicesStated that "The subject and predicate devices are exactly the same."
Limb Circumference Range (as per ISO 81060-1 and labeling)"the size of limb circumference of subject device met the requirements of ISO 81060-1, and the limb circumference was marked in the labeling."
Tube Configuration (meeting ISO 81060-1)"The subject device met the requirements of ISO 81060-1."

Study Details:

  • Test Set Sample Size: Not applicable for clinical performance testing, as it was deemed unnecessary due to the device being "exactly the same" as predicates. For technical testing to ISO 81060-1, the sample size is implicitly defined by the standard's requirements for testing a physical device.
  • Data Provenance: Not specified for the ISO 81060-1 technical testing, but typically conducted in a laboratory setting.
  • Experts for Ground Truth: Not applicable for clinical performance as no new human data was collected. Ground truth for technical testing is established by standard operating procedures and reference equipment.
  • Adjudication Method: Not applicable for clinical performance.
  • MRMC Study: No.
  • Standalone Study: Yes, a technical standalone study was performed to ISO 81060-1 and biocompatibility standards, with results indicating compliance, although specific performance metrics are not listed in this summary.
  • Type of Ground Truth: ISO standard requirements and established biocompatibility testing protocols.
  • Training Set Sample Size: Not applicable (not an AI/ML device).
  • Training Set Ground Truth Establishment: Not applicable.

The document primarily relies on demonstrating substantial equivalence to predicate devices and adherence to relevant ISO standards, rather than presenting a detailed clinical performance study specific to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 29, 2023

Shenzhen Plinma Technology Co. Ltd. Zhang Mongdi QA Manager 401, Xinheng Building No.22, Row 3, Shiling Road Jinsha Community, Kengzi Street, Pingshan District Shenzhen, Guangdong 518122 China

Re: K232772

Trade/Device Name: Blood Pressure Cuff (1201D1, 12L2D1, 1202D1, 12L3D1, 1203D1, 1204D1, 1205D1, 1206D1, 1207D1-1#, 1207D1-2#, 1207D1-3#, 1207D1-4#, 1207D1-5#, 1201B1,12L2B1,1202B1,12L3B1,1203B1,1204B1,1205B1,1206B1, 1207B1-1#, 1207B1-2#, 1207B1-3#, 1207B1-4#, 1207B1-5#, 1201W2, 12L2W2, 1202W2, 12L3W2, 1203W2, 1204W2, 1205W2, 1206W2 and 1207W2) Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: September 9, 2023 Received: September 11, 2023

Dear Zhang Mongdi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known) K232772

Device Name

Blood Pressure Cuff

(1201D1,12L2D1,1202D1,12L3D1,1203D1,1204D1,1205D1,1206D1,1207D1-1#,1207D1-2#,1207D 1-3#,1207D1-4#,1207D1-5#,1201B1,12L2B1,1202B1,1203B1,1204B1,1205B1,1206B1,120 7B1-1#,1207B1-2#,,1207B1-3#,1207B1-4#,1207B1-5#,1201W2,12L2W2,1203W2,1203W2 1204W2,1205W2,1206W2 and 1207W2)

Indications for Use (Describe)

The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared DateNovember 1, 2023
SponsorName: Shenzhen Plinma Technology Co.,LtdAddress: 401, Xinheng Building, No.22, Row 3, Shiling Road, Jinsha Community, Kengzi Street, Pingshan District, Shenzhen, Guangdong,China
Contact personSandra ZhaiE-mail: Master@plinma.comTel: +86-755- 28097345
Device Trade NameBlood pressure cuff
Device Common NameBlood pressure cuff
Classification NamesBlood pressure cuff
Regulation Number21 CFR 870.1120
Product CodeDXQ
Regulatory ClassClass II
Predicate Device 1UNIMED MEDICAL SUPPLIES INC. Unimed Blood Pressure Cuffunder K112544
Predicate Device 2UNIMED MEDICAL SUPPLIES INC. Unimed Disposable BloodPressure Cuff under K120364

Device Description

The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system.

The subject device is categorized into two types of models according to its usage.

Reusable Blood Pressure Cuff models:

1201B1,12L2B1,1202B1,12L3B1,1204B1,1205B1,1206B1,1207B1-1#,1207B1-2#,1207B1-

3#,1207B1-4#,1207B15#,1201W2,12L2W2,1202W2,12L3W2,1203W2,

1204W2,1205W2,1206W2 and 1207W2

Disposable Blood Pressure Cuff models:

1201D1,12L2D1,1202D1,12L3D1,1204D1,1205D1,1204D1,1207D1-1#,1207D1-2#,1207D1-

3#,1207D1-4#,1207D1-5#

Connector models: PLINMA-03-127-003, PLINMA-03-T, PLINMA-03-Z3

Intended use & Indication for use

The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

{5}------------------------------------------------

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

FeaturesSubject device-DisposablePredicate device(K120364)Comment
Intended Use&Indications foruseThe disposable blood pressurecuff is an accessory used inconjunction with noninvasiveblood pressure measurementsystems. The cuff is non-sterileand for single-patient use. It isavailable in neonatal, pediatricand adult sizes. The cuff is notdesigned, sold, or intended foruse except as indicated.The Unimed disposable bloodpressure cuff is an accessory used in conjunction withnoninvasive blood pressuremeasurement systems. The cuffis non-sterile and for single-patient use. It is available inneonatal, pediatric and adultsizes. The cuff is not designed,sold, or intended for use exceptas indicated.Same
Prescription/over-the-counter- usePrescriptionPrescriptionSame
TargetpopulationAdults/Pediatrics/neonatalAdults/Pediatrics/neonatalSame
ApplicationsiteArm or legArm or legSame
LimbCircumference(cm)Neonate 1 (3-6)Neonate 2 (4-8)Neonate 3 (6-11)Neonate 4 (7-13)Neonate 5 (8-15)Infant (12.3-16.8)Child (15.8-21.3)Small Adult (20.5-28.5)Adult (27.5-36.5)Adult Long (27.5-36.5)Large Adult (35.5-46)Neonate 1 (3-6)Neonate 2 (4-8)Neonate 3 (6-11)Neonate 4 (7-13)Neonate 5 (8-15)Infant (9-14.8)Child (13.8-21.5)Small Adult (20.5-28.5)Adult (27.5-36.5)Adult Long(27.5-36.5)Large Adult (35.5-46)Difference 1
FeaturesSubject device-DisposablePredicate device(K120364)Comment
Thigh (44-56)Thigh (45-56)
MaterialSleeve of disposable cuff: PVCSleeve of disposable cuff: PVCSame
composite nonwovencomposite nonwoven
Tubing: PVCTubing: PVC
Hook: Molded NylonHook: Molded Nylon
Loop: NylonLoop: Nylon
TubeConfigurationOneOne or two tubeDifference 2
Pressure limits0-300mmHg0-300mmHgSame
SterileNon-sterileNon-sterileSame
Standards metISO 81060-1ISO 81060-1Same
ISO 10993-1ISO 10993-1
ISO 10993-5ISO 10993-5
ISO 10993-10ISO 10993-10

Comparison to the Predicate Device Disposable

{6}------------------------------------------------

Reusable

FeaturesSubject device-ReusablePredicate device(K112544)Comment
Intended Use&Indications foruseThe reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.The Unimed Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.Same
Prescription/over-the-counter- usePrescriptionPrescriptionSame
TargetpopulationAdults/Pediatrics/neonatalAdults/Pediatrics/neonatalSame
ApplicationsiteArm or legArm or legSame
LimbCircumference(cm)For BladderNeonate (6-11)Infant (10-19)Child (15.8-21.3) cmBladderNeonatal (6-11cm)Infant (10-19cm)Pediatric (18-26cm)Difference 3
FeaturesSubject device-ReusablePredicate device(K112544)Comment
Small Adult (20-28)Small Adult (20-28cm)
Adult (25-35)Adult (25-35cm)
Adult Long (25-35)Adult Long (25-35cm)
Large Adult (33-47)Large Adult (33-47cm)
Large Adult Long (33-47)Large Adult Long (33-47cm)
Thigh (44-56)Adult Thigh (46-66cm)
For bladderless
Neonate 1 (3-6)
Neonate 2 (4-8)
Neonate 3 (6-11)
Neonate 4 (7-13)
Neonate 5 (8-15)
Infant (12.3-16.8)
Child (15.8-21.3)
Small Adult (20.5-28.5)
Adult (27.5-36.5)
Adult Long (27.5-36.5)
Large Adult (35.5-46)
Large Adult Long (35.5-46)
Thigh (44-56)
Repeatedinflation10,000 inflations3,000 hook and loop closures10,000 inflations3,000 hook and loop closureSame
MaterialSleeve of reusable cuff : PUSynthetic LeatherBladder: TransparentPolyurethane(TPU Film)Tubing: PVCHook: Molded NylonLoop: NylonCuff: PU Synthetic LeatherBladder: TransparentPolyurethane (TPU Film)Tubing: PVCHook: Molded NylonLoop: NylonSame
TubeConfigurationOneOne or two tubeDifference 4
Pressure limits0-300mmHg0-300mmHgSame
SterileNon-sterileNon-sterileSame
Standards metISO 81060-1ISO 81060-1Same
ISO 10993-1ISO 10993-1
ISO 10993-5ISO 10993-5
FeaturesSubject device-ReusablePredicate device(K112544)Comment
ISO 10993-10ISO 10993-10

{7}------------------------------------------------

{8}------------------------------------------------

Different 1&3: Both devices have slightly different limb circumference range, the size of limb circumference of subject device met the requirements of ISO 81060-1, and the limb circumference was marked in the labeling. Thus, this difference does not raise any safety or effectiveness issue.

Different 2&4: The subject device with one tube differs from that of predicate device. The subject device met the requirements of ISO 81060-1. Thus, this difference does not raise any safety or effectiveness issue.

Performance Data

The following performance data were provided in support of the substantial equivalence determination.

The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-"Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:

  • . Cytotoxicity
  • . Sensitization
  • Irritation ●

The test results show that the proposed device has no issue of cytotoxicity, sensitization, or skin irritation.

The subject device has been tested according to ISO 81060-1 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type

Clinical data

The subject and predicate devices are exactly the same. Since the blood pressure cuffs are identical, no further clinical testing is necessary.

Conclusion

The sponsor has demonstrated through performance testing and features comparison, that the subject device is substantially equivalent to the predicate device.

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).