(134 days)
The LYHER® Urine Multi-Drug Test Kit (Cup), LYHER® Urine Multi- Drug Test Kit (Cassette), and LYHER® Urine Multi-Drug Test Kit (Dipcard) are rapid lateral flow immunoassays for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 1000 |
| Cocaine (COC) | Benzoylecgonine | 300 |
| Marijuana (THC) | 11-nor-Δ9-THC-9-COOH | 50 |
| Methamphetamine (MET) | d-Methamphetamine | 1000 |
| Opiates(OPI) | Morphine | 2000 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The LYHER® Urine Multi-Drug Test Kit (Cup) device consists of 25 or 40 cup devices and a package insert. The LYHER® Urine Multi-Drug Test Kit (Dipcard) device consists of 10/15/20/25 Dip Card devices, a package insert. The LYHER® Urine Multi-Drug Test Kit (Cassette) device consists of 10/15/20/25 cassette devices, 10/15/20/25 droppers, a package insert.
Here's an analysis of the acceptance criteria and study details from the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the device are implicitly defined by the performance observed in the precision and method comparison studies. The document states that for precision, all test results for samples at and below -25% of the cut-off were negative, and all results for samples at and above +25% of the cut-off were positive. This indicates the device accurately identifies samples substantially below and above the cut-off.
For the method comparison, the device's performance is presented through the concordance with GC/MS results, with discordant results explicitly listed. The overall performance is demonstrated by the number of samples correctly identified as negative or positive across various concentration ranges relative to the cut-off.
Since the document does not explicitly state quantitative acceptance criteria (e.g., minimum sensitivity, specificity, or overall agreement percentage), the performance reported in the tables below represents how the device met the implied expectation of accurate qualitative detection around the cut-off.
| Test (Analyte) | Cut-off (ng/mL) | Acceptance Criteria (Implicit) | Reported Device Performance Range (Cassettes, Dipcards, Cups) |
|---|---|---|---|
| Precision Study | |||
| AMP, COC, THC, MET, OPI, PCP | (Various, see doc) | - Samples at/below -25% cut-off: All Negative - Samples at/above +25% cut-off: All Positive | - Samples at/below -25% cut-off: All Negative - Samples at/above +25% cut-off: All Positive |
| Method Comparison (Qualitative Performance around Cut-off) | |||
| AMP | 1000 | Concordance with GC/MS for negative, positive, and near-cut-off samples. | Cassettes: - Negative Urine, <-50% cut-off: 43 Neg, 0 Pos - -50% cut-off ~ cut-off: 12 Neg, 0-2 Pos - cut-off ~ +50% cut-off: 22-23 Pos, 0-1 Neg - >+50% cut-off: 17 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.) |
| COC | 300 | Concordance with GC/MS for negative, positive, and near-cut-off samples. | Cassettes: - Negative Urine, <-50% cut-off: 42 Neg, 0 Pos - -50% cut-off ~ cut-off: 13-14 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 37-38 Pos, 0-1 Neg - >+50% cut-off: 5 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.) |
| MET | 1000 | Concordance with GC/MS for negative, positive, and near-cut-off samples. | Cassettes: - Negative Urine, <-50% cut-off: 43 Neg, 0 Pos - -50% cut-off ~ cut-off: 10-11 Neg, 1-2 Pos - cut-off ~ +50% cut-off: 23-24 Pos, 0-1 Neg - >+50% cut-off: 16 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.) |
| OPI | 2000 | Concordance with GC/MS for negative, positive, and near-cut-off samples. | Cassettes: - Negative Urine, <-50% cut-off: 45 Neg, 0 Pos - -50% cut-off ~ cut-off: 9-10 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 32-33 Pos, 0-1 Neg - >+50% cut-off: 8 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.) |
| PCP | 25 | Concordance with GC/MS for negative, positive, and near-cut-off samples. | Cassettes: - Negative Urine, <-50% cut-off: 45 Neg, 2 Pos - -50% cut-off ~ cut-off: 11-12 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 33-34 Pos, 1-2 Neg - >+50% cut-off: 8 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.) |
| THC | 50 | Concordance with GC/MS for negative, positive, and near-cut-off samples. | Cassettes: - Negative Urine, <-50% cut-off: 44 Neg, 4 Pos - -50% cut-off ~ cut-off: 16-17 Neg, 0-1 Pos - cut-off ~ +50% cut-off: 23-25 Pos, 1-2 Neg - >+50% cut-off: 15 Pos, 0 Neg (Similar results for Dipcards & Cups with minor variations, see original tables for full detail.) |
2. Sample Size Used for the Test Set and Data Provenance
The studies used unaltered clinical samples for the method comparison. The samples were collected and tested as follows for each analyte and device type (Cassette, Dipcard, Cup):
- AMP: 43 Negative urine samples, 0 samples <-50% cut-off, 12 samples -50% cut-off ~ cut-off, 23 samples cut-off ~ +50% cut-off, 17 samples >+50% cut-off. (Total = 95 samples per operator/device)
- COC: 42 Negative urine samples, 0 samples <-50% cut-off, 14 samples -50% cut-off ~ cut-off, 38 samples cut-off ~ +50% cut-off, 5 samples >+50% cut-off. (Total = 99 samples per operator/device)
- MET: 43 Negative urine samples, 0 samples <-50% cut-off, 12 samples -50% cut-off ~ cut-off, 24 samples cut-off ~ +50% cut-off, 16 samples >+50% cut-off. (Total = 95 samples per operator/device)
- OPI: 45 Negative urine samples, 0 samples <-50% cut-off, 10 samples -50% cut-off ~ cut-off, 33 samples cut-off ~ +50% cut-off, 8 samples >+50% cut-off. (Total = 96 samples per operator/device)
- PCP: 45 Negative urine samples, 2 samples <-50% cut-off, 12 samples -50% cut-off ~ cut-off, 35 samples cut-off ~ +50% cut-off, 8 samples >+50% cut-off. (Total = 102 samples per operator/device)
- THC: 44 Negative urine samples, 4 samples <-50% cut-off, 17 samples -50% cut-off ~ cut-off, 25 samples cut-off ~ +50% cut-off, 15 samples >+50% cut-off. (Total = 105 samples per operator/device)
The total number of unique samples is not directly reported, as these numbers represent the distribution of samples across different concentration ranges. Since "unaltered clinical samples" were used, this suggests they were retrospective in nature. The provenance (country of origin) of the data is not explicitly stated, but the applicant company is based in Hangzhou, China, suggesting the studies were likely performed there.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The ground truth for the test set was established using Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS). These are "preferred confirmatory methods" as stated in the Indications for Use. The document does not specify the number of experts or their qualifications for operating these instruments or interpreting their results; it implies the use of standard laboratory practices for these confirmatory methods.
4. Adjudication Method for the Test Set
The document does not describe a formal adjudication method for the test set results. The comparison is directly between the device's qualitative result (positive/negative) and the quantitative GC/MS or LC/MS result. Discordant results are noted but no further adjudication (e.g., by multiple experts reviewing the same case) is mentioned.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study evaluated the standalone performance of the device against a gold standard (GC/MS/LC/MS), not its effectiveness in assisting human readers.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
Yes, the studies conducted were standalone performance evaluations. The device (LYHER® Urine Multi-Drug Test Kit) is a rapid lateral flow immunoassay, which produces a visual result (line present or absent). The "operators" in the study simply interpret this visual result. The results presented are the device's performance as interpreted by an operator, not an AI algorithm. So, while it involves human interpretation of the device's output, it is a standalone performance of the device itself without additional AI assistance beyond the inherent design of the test kit.
7. The Type of Ground Truth Used
The ground truth used was quantitative analytical results from Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS). These are considered highly accurate confirmatory methods for drug detection and concentration.
8. The Sample Size for the Training Set
The document describes "Precision studies" involving spiking drugs into negative samples to achieve concentrations at various percentages around the cut-off. However, this is described as part of the analytical performance characterization and not as a "training set" for an algorithm. This device is a lateral flow immunoassay, not a machine learning or AI-based device, and therefore does not have a "training set" in the conventional sense of an AI model.
9. How the Ground Truth for the Training Set was Established
As this is not an AI/ML device, there isn't a "training set" that requires ground truth establishment. For the precision studies mentioned (which are part of analytical performance validation), the "ground truth" was established by spiking known concentrations of the target drug into negative urine samples. These spiked concentrations were then "confirmed by LC/MS" to ensure accuracy of the prepared samples.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hangzhou Laihe Biotech Co., Ltd. % Ethan Liu RA specialist Shanghai Thinkwell Consulting Co., Ltd Xinling Rd., 211/6F Shanghai, Shanghai 201100 China
Re: K232597
Trade/Device Name: LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG, DIO, DKZ, LDJ, DJC, LCM Dated: November 27, 2023 Received: December 1, 2023
Dear Ethan Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely, Digitally signed by Jos Joseph A. A Kotarek -S Date: 2024.01.09 14:04:36 Kotarek -S Joseph Kotarek, PhD Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K232597
Device Name
L YHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard)
Indications for Use (Describe)
The LYHER® Urine Multi-Drug Test Kit (Cup), LYHER® Urine Multi- Drug Test Kit (Cassette), and LYHER® Urine Multi-Drug Test Kit (Dipcard) are rapid lateral flow immunoassays for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 1000 |
| Cocaine (COC) | Benzoylecgonine | 300 |
| Marijuana (THC) | 11-nor-Δ9-THC-9-COOH | 50 |
| Methamphetamine (MET) | d-Methamphetamine | 1000 |
| Opiates(OPI) | Morphine | 2000 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the logo for "LYHER" in green with the "Y" in yellow. A registered trademark symbol is next to the "R". Below the logo, the text "Hangzhou Laihe Biotech Co., Ltd." is written in black.
510(k) Summary 510(k) Number: K232597
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
5.1 Submitter
| Submitted by: | Hangzhou Laihe Biotech Co., Ltd. | |
|---|---|---|
| Address: Room 401-406, F1-3, Building 1, No.425Miaohouwang Road, Xixing Street, Binjiang District,Hangzhou, 310051 Zhejiang P.R. China | ||
| Tel.: +86 571 8765 3090 | ||
| Fax: +86 571 8665 8000 | ||
| www.lyher.com | ||
| Contact Person: | Ethan Liu | |
| Shanghai Thinkwell Consulting Co., Ltd | ||
| Address:Xinling Rd., 211/6F Shanghai, 201100,P.R. China | ||
| Phone:0086-15216699240 | ||
| Email:xtdeepwater@126.com | ||
| Date Prepared: | Dec 29, 2023 |
5.2 Device
LYHER® Urine Multi-Drug Test Kit(Cup) LYHER® Urine Multi-Drug Test Kit(Cassette) LYHER® Urine Multi-Drug Test Kit(Dipcard)
Classification:
| Test | Product Code | CFR # | Panel |
|---|---|---|---|
| Opiates(OPI) | DJG | 862.3650/Opiate test system. | Toxicology |
| Cocaine (COC) | DIO | 862.3250/Cocaine andcocaine metabolite testsystem. | Toxicology |
| Amphetamine (AMP) | DKZ | 862.3100/Amphetamine TestSystem | Toxicology |
| Marijuana (THC) | LDJ | 862.3870/Cannabinoid testsystem. | Toxicology |
| Methamphetamine(MET) | DJC | 862.3610/Methamphetaminetest system. | Toxicology |
| Phencyclidine (PCP) | LCM | Unclassified | Toxicology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for LYHER, a company named Hangzhou Laihe Biotech Co., Ltd. The LYHER logo is in green and yellow. The company name is in black text and is located below the logo.
5.3 Predicate Device:
ATTEST Drug Screen Cup, ATTEST Drug Screen Dip Card, K182123
5.4 Device Description
The LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The LYHER® Urine Multi-Drug Test Kit (Cup) device consists of 25 or 40 cup devices and a package insert. The LYHER® Urine Multi-Drug Test Kit (Dipcard) device consists of 10/15/20/25 Dip Card devices, a package insert. The LYHER® Urine Multi-Drug Test Kit (Cassette) device consists of 10/15/20/25 cassette devices, 10/15/20/25 droppers, a package insert.
5.5 Indication for Use:
The LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are rapid lateral flow immunoassays for the qualitative detection of d-Amphetamine, d-Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 1000 |
| Cocaine (COC) | Benzoylecgonine | 300 |
| Marijuana (THC) | 11-nor-\u03949-THC-9-COOH | 50 |
| Methamphetamine (MET) | d-Methamphetamine | 1000 |
| Opiates(OPI) | Morphine | 2000 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 6 analytes. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive positive result. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/Mass Spectrometry (LC/MS) and their tandem massspectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
5.8 Substantial Equivalence
A summary comparison of features of the LYHER® Urine Multi-Drug Test Kit(Cup/Cassette/Dipcard) and the predicate devices is provided in following tables.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "LYHER" in green and yellow letters. The "L", "H", "E", and "R" are green, while the "Y" is yellow. There is a registered trademark symbol next to the "R".
Section 5 510(k) Summary
Hangzhou Laihe Biotech Co., Ltd.
| Item | Proposed Device | Predicate Device-K182123 |
|---|---|---|
| Indicationsfor Use | For the qualitative determination of drugs ofabuse and/or their metabolites in human urine. | Same |
| Methodology | Lateral flow immunochromatographic assaybased on competitive binding | Same |
| Type of Test | Qualitative | Same |
| SpecimenType | Human Urine | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for LYHER. The "LY" is in yellow, and the "HER" is in green. There is a registered trademark symbol next to the "R".
Section 5 510(k) Summary
Hangzhou Laihe Biotech Co., Ltd.
| Cut-off Values | |||
|---|---|---|---|
| Test | Calibrator | Cut-off ng/mL | |
| Opiates(OPI) | Morphine | 2000 | |
| Cocaine (COC) | Benzoylecgonine | 300 | |
| Amphetamine (AMP) | d-Amphetamine | 1000 | |
| Marijuana (THC) | 11-nor-Δ9-THC-9-COOH | 50 | |
| Methamphetamine (MET) | d-Methamphetamine | 1000 | |
| Phencyclidine (PCP) | Phencyclidine | 25 | |
| Test | Calibrator | Cutoff ng/ml | |
| 6-Acetylmorphine | 6-monoacetylmorphine | 10 | |
| Amphetamine 500 | d-amphetamine | 500 | |
| Amphetamine 1000 | d-amphetamine | 1000 | |
| Barbiturates | Secobarbital | 300 | |
| Benzodiazepine | Oxazepam | 300 | |
| Buprenorphine | Buprenorphine | 10 | |
| Cocaine 150 | Benzoylecgonine | 150 | |
| Cocaine 300 | Benzoylecgonine | 300 | |
| EDDP | 2-ethylidene-1,5-dimethyl-3-3-diphenyl-pyrrolidine | 300 | |
| Ecstasy | d/l-methylenedioxy-methamphetamine | 500 | |
| Methamphetamine 500 | d-methamphetamine | 500 | |
| Methamphetamine 1000 | d-methamphetamine | 1000 | |
| Methadone | d/l-methadone | 300 | |
| Morphine | Morphine | 300 | |
| Opiates | Morphine | 2,000 | |
| Oxycodone | Oxycodone | 100 | |
| Phencyclidine | Phencyclidine | 25 | |
| Propoxyphene | d-propoxyphene | 300 | |
| Tricyclics | Nortriptyline | 1,000 | |
| Marijuana | THC-COOH | 5 | |
| Intended Use | For prescription uses. | Same | |
| Configurations | Cup,Dip Card, Cassette | Cup, Dip Card |
5.9 Test Principle
Each device employs lateral flow immunochromatographic technology and is based on the principle of competitive binding. The product contains membrane strips coated with drug-protein conjugates on the T zone, goat anti-rabbit polycolonal antibody at the C zone, and a conjugate pad which 5-4
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" is in yellow, while the "HER" is in green. There is a registered trademark symbol next to the "R".
contains colloidal gold particles coated with mouse monoclonal antibodies specific against to the corresponding drugs and colloidal gold particles coated with rabbit IgG.
If the level of drug in the urine specimen is below the cutoff concentration, the T line appears as a visible burgundy line. If the level of drug in the urine specimen is above the cutoff, no T line develops.The control line (C line) serves as an internal quality control line should always appear as a burgundy-colored band regardless of the drug, if enough sample volume has been added to the test and if the sample has correctly migrated up the test strip. Testing is based on the principle of a competitive immunochemical reaction between a chemically labeled drug (drug-protein conjugate) and the drug or drug metabolites which may be present in the urine sample competing for the limited antibody binding sites.
5.10 Performance Characteristics
5.10.1 Analytical Performance
- a. Precision
Precision studies were carried out for samples with concentrations of -100% cut off, -75%cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off, These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. Each concentration has three lots, and 6 operators have joined the performance. Each specimen was analyzed in 50 replicates by each of the testing personnel at each Lab site by using three lots of the product separately.
The test results of the specimens at the concentrations at and below -25% of the cut off obtained by all the three personnel were all negative while the test results of the specimens at the concentration at and above +25% of the cut off value were all positive.
- b. Linearity
Not applicable
Stability ﻥ
The devices are stable at 2-30 °C for 24 months based on the accelerated stability study.
d. Cut-off
Dilute the solution by negative urine to the concentration of +50% cut off, +25%cut off, -25%cut off, -50%cut off respectively. To test all these above specimens by three different batches of the candidate product and each concentration will be tested for 25 replicates.
The following cut-off values for the candidate devices have been verified.
| Test | Calibrator | Cut-off(ng/mL) |
|---|---|---|
| Opiates(OPI) | Morphine | 2000 |
| Cocaine (COC) | Benzoylecgonine | 300 |
| Amphetamine(AMP) | d-Amphetamine | 1000 |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "Y" is orange, while the rest of the letters are green. There is a small registered trademark symbol to the right of the "R".
| Marijuana (THC) | 11-nor-Δ9-THC-9-COOH | 50 |
|---|---|---|
| Methamphetamine(MET) | d-Methamphetamine | 1000 |
| Phencyclidine(PCP) | Phencyclidine | 25 |
e. Interference
Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and target drugs urine with concentrations at 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different devices
Amphetamine:
| Substance | Substance | Substance |
|---|---|---|
| Acetaminophen | Erythromycin | Penicillin-G |
| Acetone | Ethanol | Pheniramine |
| Acetylsalicylic acid | Furosemide | Phenothiazine |
| Albumin | Gabapentin | Pregablin |
| Ampicillin | Glucose | Procaine |
| Ascorbic acid | Guaiacol glyceryl ether | Propoxyphene |
| Aspartame | Hemoglobin | Quinidine |
| Atropine | Ibuprofen | Ranitidine |
| Benzocaine | Isoproterenol | Riboflavin |
| Bilirubin | Isoproterenol | Sertraline |
| Caffeine | Ketamine | Sodium chloride |
| Chloroquine | Lidocaine | Sulindac |
| Chlorpheniramine | l-Phenylephrine | Theophylline |
| Creatine | Methadone | Tyramine |
| Dexbrompheniramine | Methylephedrine | β-Phenylethylamine |
| Dextromethorphan | Naproxen | Phencyclidine(PCP) |
| Dimenhydrinate | Niacinamide | Cocaine |
| Dimethylaminoantipyrine | Nicotine | Morphine |
| Diphenhydramine | Norephedrine | Δ9-tetrahydrocannabinol |
| Dopamine | Oxalic acid | |
| EDDP | Oxycodone |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for LYHER. The letters are in a sans-serif font, with the "LY" in yellow and the "HER" in green. There is a registered trademark symbol next to the "R".
| Ephedrine | Pantoprazole |
|---|---|
| ----------- | -------------- |
Cocaine(COC)
| Substance | Substance | Substance |
|---|---|---|
| Acetaminophen | Erythromycin | Penicillin-G |
| Acetone | Ethanol | Pheniramine |
| Acetylsalicylic acid | Furosemide | Phenothiazine |
| Albumin | Gabapentin | Pregablin |
| Ampicillin | Glucose | Procaine |
| Ascorbic acid | Guaiacol glyceryl ether | Propoxyphene |
| Aspartame | Hemoglobin | Quinidine |
| Atropine | Ibuprofen | Ranitidine |
| Benzocaine | Isoproterenol | Riboflavin |
| Bilirubin | Isoproterenol | Sertraline |
| Caffeine | Ketamine | Sodium chloride |
| Chloroquine | Lidocaine | Sulindac |
| Chlorpheniramine | l-Phenylephrine | Theophylline |
| Creatine | Methadone | Tyramine |
| Dexbrompheniramine | Methylephedrine | β-Phenylethylamine |
| Dextromethorphan | Naproxen | Phencyclidine(PCP) |
| Dimenhydrinate | Niacinamide | Morphine |
| Dimethylaminoantipyrine | Nicotine | Methylamphetamine |
| Diphenhydramine | Norephedrine | MDMA |
| Dopamine | Oxalic acid | Δ9-tetrahydrocannabinol |
| EDDP | Oxycodone | Amphetamine |
| Ephedrine | Pantoprazole |
MET
| Substance | Substance | Substance |
|---|---|---|
| Acetaminophen | Erythromycin | Penicillin-G |
| Acetone | Ethanol | Pheniramine |
| Acetylsalicylic acid | Furosemide | Phenothiazine |
| Albumin | Gabapentin | Pregablin |
| Ampicillin | Glucose | Procaine |
| Ascorbic acid | Guaiacol glyceryl ether | Propoxyphene |
| Aspartame | Hemoglobin | Quinidine |
| Atropine | Ibuprofen | Ranitidine |
| Benzocaine | Isoproterenol | Riboflavin |
| Bilirubin | Isoproterenol | Sertraline |
| Caffeine | Ketamine | Sodium chloride |
| Chloroquine | Lidocaine | Sulindac |
| Chlorpheniramine | l-Phenylephrine | Theophylline |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "L" and "HER" are in green, while the "Y" is in orange. There is a registered trademark symbol next to the "R".
| Creatine | Methadone | Tyramine |
|---|---|---|
| Dexbrompheniramine | Methylephedrine | β-Phenylethylamine |
| Dextromethorphan | Naproxen | Phencyclidine(PCP) |
| Dimenhydrinate | Niacinamide | Cocaine |
| Dimethylaminoantipyrine | Nicotine | Morphine |
| Diphenhydramine | Norephedrine | Δ9-tetrahydrocannabinol |
| Dopamine | Oxalic acid | |
| EDDP | Oxycodone | |
| Ephedrine | Pantoprazole |
OPI
| OPT | ||
|---|---|---|
| Substance | Substance | Substance |
| Acetaminophen | Erythromycin | Penicillin-G |
| Acetone | Ethanol | Pheniramine |
| Acetylsalicylic acid | Furosemide | Phenothiazine |
| Albumin | Gabapentin | Pregablin |
| Ampicillin | Glucose | Procaine |
| Ascorbic acid | Guaiacol glyceryl ether | Propoxyphene |
| Aspartame | Hemoglobin | Quinidine |
| Atropine | Ibuprofen | Ranitidine |
| Benzocaine | Isoproterenol | Riboflavin |
| Bilirubin | Isoproterenol | Sertraline |
| Caffeine | Ketamine | Sodium chloride |
| Chloroquine | Lidocaine | Sulindac |
| Chlorpheniramine | I-Phenylephrine | Theophylline |
| Creatine | Methadone | Tyramine |
| Dexbrompheniramine | Methylephedrine | β-Phenylethylamine |
| Dextromethorphan | Naproxen | Phencyclidine(PCP) |
| Dimenhydrinate | Niacinamide | Cocaine |
| Dimethylaminoantipyrine | Nicotine | Methylamphetamine |
| Diphenhydramine | Norephedrine | MDMA |
| Dopamine | Oxalic acid | Amphetamine |
| EDDP | Oxycodone | Δ9-tetrahydrocannabinol |
| Ephedrine | Pantoprazole |
РСР
| Substance | Substance | Substance |
|---|---|---|
| Acetaminophen | Erythromycin | Penicillin-G |
| Acetone | Ethanol | Pheniramine |
| Acetylsalicylic acid | Furosemide | Phenothiazine |
| Albumin | Gabapentin | Pregablin |
| Ampicillin | Glucose | Procaine |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in orange, while the "HER" portion is in green. The letters are bold and have a rounded appearance. There is a registered trademark symbol next to the "R".
| Ascorbic acid | Guaiacol glyceryl ether | Propoxyphene |
|---|---|---|
| Aspartame | Hemoglobin | Quinidine |
| Atropine | Ibuprofen | Ranitidine |
| Benzocaine | Isoproterenol | Riboflavin |
| Bilirubin | Isoproterenol | Sertraline |
| Caffeine | Ketamine | Sodium chloride |
| Chloroquine | Lidocaine | Sulindac |
| Chlorpheniramine | I-Phenylephrine | Theophylline |
| Creatine | Methadone | Tyramine |
| Dexbrompheniramine | Methylephedrine | β-Phenylethylamine |
| Dextromethorphan | Naproxen | Cocaine |
| Dimenhydrinate | Niacinamide | Morphine |
| Dimethylaminoantipyrine | Nicotine | Methylamphetamine |
| Diphenhydramine | Norephedrine | MDMA |
| Dopamine | Oxalic acid | Amphetamine |
| EDDP | Oxycodone | Δ9-tetrahydrocannabinol |
| Ephedrine | Pantoprazole |
THC
| Substance | Substance | Substance |
|---|---|---|
| Acetaminophen | Erythromycin | Penicillin-G |
| Acetone | Ethanol | Pheniramine |
| Acetylsalicylic acid | Furosemide | Phenothiazine |
| Albumin | Gabapentin | Pregablin |
| Ampicillin | Glucose | Procaine |
| Ascorbic acid | Guaiacol glyceryl ether | Propoxyphene |
| Aspartame | Hemoglobin | Quinidine |
| Atropine | Ibuprofen | Ranitidine |
| Benzocaine | Isoproterenol | Riboflavin |
| Bilirubin | Isoproterenol | Sertraline |
| Caffeine | Ketamine | Sodium chloride |
| Chloroquine | Lidocaine | Sulindac |
| Chlorpheniramine | l-Phenylephrine | Theophylline |
| Creatine | Methadone | Tyramine |
| Dexbrompheniramine | Methylephedrine | β-Phenylethylamine |
| Dextromethorphan | Naproxen | Phencyclidine(PCP) |
| Dimenhydrinate | Niacinamide | Cocaine |
| Dimethylaminoantipyrine | Nicotine | Morphine |
| Diphenhydramine | Norephedrine | Methylamphetamine |
| Dopamine | Oxalic acid | MDMA |
| EDDP | Oxycodone | Amphetamine |
| Ephedrine | Pantoprazole |
{12}------------------------------------------------
f. Specificity
To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below. There were no differences observed for different devices.
| Compound | Concentration(ng/mL) | %Cross Reactivity |
|---|---|---|
| Amphetamine (AMP) | ||
| D-Amphetamine | 1000 | 100% |
| D,L-Amphetamine sulfate | 2500 | 40% |
| β-Phenylethylamine | 60000 | 1.67% |
| (±)3,4-Methylenedioxyamphetamine (MDA) | 1500 | 66.7% |
| Phentermine | 2500 | 40% |
| Tryptamine | 60000 | 1.67% |
| Tyramine | 5000 | 20% |
| L-amphetamine | 100000 | 1% |
| COCAINE (COC) | ng/ml | |
| Benzoylecgonine | 300 | 100% |
| Cocaine | 750 | 40% |
| Cocaethylene | 10000 | 3% |
| Ecgonine | 20000 | 1.5% |
| Ecgonine methylester | 50000 | 0.6% |
| MARIJUANA (THC) | ||
| Cannabinol | 20000 | 0.25% |
| 11-nor-Δ8-THC-9 COOH | 30 | 167% |
| 11-nor-Δ9 -THC-9 COOH | 50 | 100% |
| Δ8 -THC | 15000 | 0.33% |
| Δ9 -THC | 15000 | 0.33% |
| (±)-11-Hydroxy-Δ9-THC | 5000 | 1% |
| Cannabidiol | 20000 | 0.25% |
| METHAMPHETAMINE (MET) | ||
| p-Hydroxymethamphetamine | 30000 | 3.33% |
| D-Methamphetamine | 1000 | 100% |
| L-Methamphetamine | 8000 | 12.5% |
| (±)-3,4-Methylenedioxymethamphetamine | 2000 | 50% |
| Mephentermine | 50000 | 2% |
| D-Amphetamine | 100000 | 1% |
| Compound | Concentration(ng/mL) | %Cross Reactivity |
| L-Amphetamine | 100000 | 1% |
| (1R, 2S)-(-)-Ephedrine | 100000 | 1% |
| (+/-) 3,4-Methylenedioxy-nethylamphetamine (MDEA) | 30000 | 3.33% |
| OPIATES (OPI) | ||
| Morphine | 2000 | 100% |
| Codeine | 1500 | 133% |
| Ethylmorphine | 30000 | 6.67% |
| Hydrocodone | 80000 | 2.5% |
| Hydromorphone | 10000 | 20% |
| Levophanol | 5000 | 40% |
| 6-Monoacetylmorphine | 2000 | 100% |
| Morphine 3-β-D-glucuronide | 5000 | 40% |
| Norcodeine | 30000 | 6.67% |
| Normorphone | 100000 | 2% |
| Oxycodone | 100000 | 2% |
| Oxymorphone | 100000 | 2% |
| Procaine | 60000 | 3.33% |
| Thebaine | 30000 | 6.67% |
| 6-Acetylmorphine (6-AM) | 2000 | 100% |
| Diacetylmorphine (heroin) | 5000 | 40% |
| Phencyclidine (PCP) | ||
| Phencyclidine | 25 | 100% |
| 4-hydroxy-PCP | 12500 | 0.2% |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" is in orange, while the "HER" is in green. There is a registered trademark symbol to the right of the "R".
g. Effect of Urine Specific Gravity and Urine pH
To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.003 to 1.032 specific gravity or urine samples with pH 3 to 9 were spiked with target
drugs at 25% below and 25% above Cut-off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different devices.
5.10.2 Method comparison
Method comparison studies for the LYHER® Urine Multi-Drug Test were performed in-house with three laboratory assistants for each device. Operators ran samples unaltered clinical samples. The samples were blind labeled and compared to GC/MS results are presented in the tables below:
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows the logo for LYHER. The logo is green, except for the Y, which is orange. There is a registered trademark symbol to the right of the word.
| AMPCassettes | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 43 | 0 | 12 | 1 | 0 |
| Positive | 0 | 0 | 0 | 22 | 17 | |
| Operator 2 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 23 | 17 | |
| Operator 3 | Negative | 43 | 0 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 23 | 17 |
Discordant Results of AMP Cassettes
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 59 | 202300501 | 994 | Positive |
| Operator 1 | 75 | 202300517 | 1008 | Negative |
| AMPDipcards | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 43 | 0 | 12 | 1 | 0 |
| Positive | 0 | 0 | 0 | 22 | 17 | |
| Operator 2 | Negative | 43 | 0 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 23 | 17 | |
| Operator 3 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 23 | 17 |
Discordant Results of AMP Dipcards
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 3 | 59 | 202300501 | 994 | Positive |
| Operator 1 | 75 | 202300517 | 1008 | Negative |
| AMP Cups | Test result | Negative urine | <-50% cut off | -50% cut off~cut off | Cut off~+50%cut off | >+50%cut off |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 43 | 0 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 23 | 17 | |
| Operator 2 | Negative | 43 | 0 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 23 | 17 | |
| Operator 3 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 23 | 17 |
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in orange, while the "HER" portion is in green. There is a registered trademark symbol to the right of the word.
Discordant Results of AMP Cups
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 3 | 59 | 202300501 | 994 | Positive |
coc
| COCCassettes | Testresult | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cut off~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 42 | 0 | 13 | 1 | 0 |
| Positive | 0 | 0 | 1 | 37 | 5 | |
| Operator 2 | Negative | 42 | 0 | 14 | 0 | 0 |
| Positive | 0 | 0 | 0 | 38 | 5 | |
| Operator 3 | Negative | 42 | 0 | 13 | 0 | 0 |
| Positive | 0 | 0 | 1 | 38 | 5 |
Discordant Results of COC Cassettes
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 17 | 202300680 | 298 | Positive |
| Operator 3 | 17 | 202300680 | 298 | Positive |
| Operator 1 | 21 | 202300684 | 303 | Negative |
| COCDipcards | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 42 | 0 | 14 | 1 | 0 |
| Positive | 0 | 0 | 0 | 37 | 5 | |
| Operator 2 | Negative | 42 | 0 | 13 | 0 | 0 |
| Positive | 0 | 0 | 1 | 38 | 5 | |
| Operator 3 | Negative | 42 | 0 | 13 | 0 | 0 |
| Positive | 0 | 0 | 1 | 38 | 5 |
Discordant Results of COC Dipcards
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 17 | 202300680 | 298 | Positive |
| Operator 3 | 17 | 202300680 | 298 | Positive |
| Operator 1 | 21 | 202300684 | 303 | Negative |
| COC Cups | Test result | Negative urine | <-50% cut off | -50% cut off~cut off | Cut off~+50%cut off | >+50%cut off |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 42 | 0 | 14 | 0 | 0 |
| Operator 1 | Positive | 0 | 0 | 0 | 38 | 5 |
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "L" and "Y" are in yellow, while the "HER" is in green. There is a small registered trademark symbol to the right of the "R".
| Operator 2 | Negative | 42 | 0 | 13 | 1 | 0 |
|---|---|---|---|---|---|---|
| Positive | 0 | 0 | 1 | 37 | 5 | |
| Operator 3 | Negative | 42 | 0 | 13 | 0 | 0 |
| Positive | 0 | 0 | 1 | 38 | 5 |
Discordant Results of COC Cups
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 17 | 202300680 | 298 | Positive |
| Operator 3 | 17 | 202300680 | 298 | Positive |
| Operator 2 | 21 | 202300684 | 303 | Negative |
MET
| METCassettes | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 43 | 0 | 10 | 0 | 0 |
| Positive | 0 | 0 | 2 | 24 | 16 | |
| Operator 2 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 24 | 16 | |
| Operator 3 | Negative | 43 | 0 | 11 | 1 | 0 |
| Positive | 0 | 0 | 1 | 23 | 16 |
Discordant Results of MET Cassettes
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 15 | 202300331 | 932 | Positive |
| Operator 1 | 58 | 202300374 | 561 | Positive |
| Operator 2 | 58 | 202300374 | 561 | Positive |
| Operator 3 | 58 | 202300374 | 561 | Positive |
| Operator 3 | 77 | 202300393 | 1007 | Negative |
| METDipcards | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 24 | 16 | |
| Operator 2 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 24 | 16 | |
| Operator 3 | Negative | 43 | 0 | 11 | 1 | 0 |
| Positive | 0 | 0 | 1 | 23 | 16 |
Discordant Results of MET Dipcards
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the word "LYHER" in green and yellow. The "LY" is green, but the middle of the "Y" is yellow. The rest of the word is green. There is a registered trademark symbol in the upper right corner of the word.
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 58 | 202300374 | 561 | Positive |
| Operator 2 | 58 | 202300374 | 561 | Positive |
| Operator 3 | 58 | 202300374 | 561 | Positive |
| Operator 3 | 77 | 202300393 | 1007 | Negative |
| MET Cups | Test result | Negative | <-50% cut | -50%cut | Cut | >+50%cut |
|---|---|---|---|---|---|---|
| urine | Off | off~cut off | off~+50%cut | ેન્દ્રિક રીકેની વિદ્વારા નિર્માન નામાં આવેલું એક ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત | ||
| ેન્દ્રિક રીકેની વિદ્વારા નિર્માન નામાં આવેલું એક ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત | ||||||
| Operator 1 | Negative | 43 | O | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 24 | 16 | |
| Operator 2 | Negative | 43 | 0 | 11 | 0 | 0 |
| Positive | 0 | 0 | 1 | 24 | 16 | |
| Operator 3 | Negative | 43 | O | 11 | 1 | 0 |
| Positive | O | 0 | 1 | 23 | 16 |
Discordant Results of MET Cups
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 58 | 202300374 | 561 | Positive |
| Operator 2 | 58 | 202300374 | 561 | Positive |
| Operator 3 | 58 | 202300374 | 561 | Positive |
| Operator 3 | 77 | 202300393 | 1007 | Negative |
OPI
| OPICassettes | Test result | Negativeurine | <-50% cut off | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 45 | 0 | 10 | 0 | 0 |
| Positive | 0 | 0 | 0 | 33 | 8 | |
| Operator 2 | Negative | 45 | 0 | 9 | 1 | 0 |
| Positive | 0 | 0 | 1 | 32 | 8 | |
| Operator 3 | Negative | 45 | 0 | 10 | 0 | 0 |
| Positive | 0 | 0 | 0 | 33 | 8 |
Discordant Results of OPI Cassettes
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 59 | 202300180 | 2006 | Negative |
| Operator 2 | 80 | 202300201 | 1978 | Positive |
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for LYHER. The logo is green, except for the "Y" which is yellow. There is a small registered trademark symbol to the right of the "R" in LYHER.
| OPIDipcards | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 45 | 0 | 10 | 0 | 0 |
| Positive | 0 | 0 | 0 | 33 | 8 | |
| Operator 2 | Negative | 45 | 0 | 9 | 0 | 0 |
| Positive | 0 | 0 | 1 | 33 | 8 | |
| Operator 3 | Negative | 45 | 0 | 10 | 1 | 0 |
| Positive | 0 | 0 | 0 | 32 | 8 |
Discordant Results of OPI Dipcards
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 3 | 59 | 202300180 | 2006 | Negative |
| Operator 2 | 80 | 202300201 | 1978 | Positive |
| OPI Cups | Test result | Negative urine | <-50% cut off | -50% cut off~cut off | Cut off~+50%cut off | >+50%cut off |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 45 | 0 | 10 | 1 | 0 |
| Operator 1 | Positive | 0 | 0 | 0 | 32 | 8 |
| Operator 2 | Negative | 45 | 0 | 9 | 0 | 0 |
| Operator 2 | Positive | 0 | 0 | 1 | 33 | 8 |
| Operator 3 | Negative | 45 | 0 | 9 | 0 | 0 |
| Operator 3 | Positive | 0 | 0 | 1 | 33 | 8 |
Discordant Results of OPI Cups
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 59 | 202300180 | 2006 | Negative |
| Operator 2 | 80 | 202300201 | 1978 | Positive |
| Operator 3 | 80 | 202300201 | 1978 | Positive |
РСР
| PCPCassettes | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 45 | 2 | 11 | 1 | 0 |
| Positive | 0 | 0 | 1 | 34 | 8 | |
| Operator 2 | Negative | 45 | 2 | 12 | 1 | 0 |
| Positive | 0 | 0 | 0 | 34 | 8 | |
| Operator 3 | Negative | 45 | 2 | 12 | 2 | 0 |
| Positive | 0 | 0 | 0 | 33 | 8 |
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in orange, while the "HER" portion is in green. There is a registered trademark symbol next to the "R".
Discordant Results of PCP Cassettes
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 25 | 202300810 | 23.0 | Positive |
| Operator 2 | 37 | 202300822 | 25.2 | Negative |
| Operator 3 | 37 | 202300822 | 25.2 | Negative |
| Operator 1 | 100 | 202301750 | 25.2 | Negative |
| Operator 3 | 100 | 202301750 | 25.2 | Negative |
| PCPDipcards | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 45 | 2 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 35 | 8 | |
| Operator 2 | Negative | 45 | 2 | 11 | 2 | 0 |
| Positive | 0 | 0 | 1 | 33 | 8 | |
| Operator 3 | Negative | 45 | 2 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 35 | 8 |
Discordant Results of PCP Dipcards
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 25 | 202300810 | 23.0 | Positive |
| Operator 2 | 37 | 202300822 | 25.2 | Negative |
| Operator 2 | 100 | 202301750 | 25.2 | Negative |
| PCP Cups | Test result | Negative urine | <-50% cut off | -50% cut off~cut off | Cut off~+50%cut off | >+50%cut off |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 45 | 2 | 12 | 1 | 0 |
| Positive | 0 | 0 | 0 | 34 | 8 | |
| Operator 2 | Negative | 45 | 2 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 35 | 8 | |
| Operator 3 | Negative | 45 | 2 | 12 | 0 | 0 |
| Positive | 0 | 0 | 0 | 35 | 8 |
Discordant Results of PCP Cups
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 100 | 202301750 | 25.2 | Negative |
THC
| THCCassettes | Test result | Negativeurine | <-50% cutoff | -50% cutoff~cut off | Cut off~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| ------------------ | ------------- | ------------------- | ------------------ | ------------------------- | ------------------------ | ----------------- |
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the logo for LYHER. The "LY" is in orange, and the "HER" is in green. There is a registered trademark symbol to the right of the "R".
| Operator 1 | Negative | 44 | 4 | 17 | 1 | 0 |
|---|---|---|---|---|---|---|
| Positive | 0 | 0 | 0 | 24 | 15 | |
| Operator 2 | Negative | 44 | 4 | 16 | 2 | 0 |
| Positive | 0 | 0 | 1 | 23 | 15 | |
| Operator 3 | Negative | 44 | 4 | 17 | 0 | 0 |
| Positive | 0 | 0 | 0 | 25 | 15 |
Discordant Results of THC Cassettes
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 1 | 12 | 202301036 | 50.5 | Negative |
| Operator 2 | 36 | 202301060 | 52.3 | Negative |
| Operator 2 | 45 | 202301069 | 49.6 | Positive |
| Operator 2 | 92 | 202301651 | 51.1 | Negative |
| THCDipcards | Test result | Negativeurine | <-50% cut off | -50% cutoff~cut off | Cutoff~+50%cutoff | >+50%cutoff |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 44 | 4 | 16 | 0 | 0 |
| Positive | 0 | 0 | 1 | 25 | 15 | |
| Operator 2 | Negative | 44 | 4 | 17 | 3 | 0 |
| Positive | 0 | 0 | 0 | 22 | 15 | |
| Operator 3 | Negative | 44 | 4 | 17 | 1 | 0 |
| Positive | 0 | 0 | 0 | 24 | 15 |
Discordant Results of THC Dipcards
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 12 | 202301036 | 50.5 | Negative |
| Operator 2 | 36 | 202301060 | 52.3 | Negative |
| Operator 1 | 45 | 202301069 | 49.6 | Positive |
| Operator 3 | 75 | 202301099 | 52.3 | Negative |
| Operator 2 | 92 | 202301651 | 51.1 | Negative |
| THC Cups | Test result | Negative urine | <-50% cut off | -50% cut off~cut off | Cut off~+50%cut off | >+50%cut off |
|---|---|---|---|---|---|---|
| Operator 1 | Negative | 44 | 4 | 16 | 0 | 0 |
| Positive | 0 | 0 | 1 | 25 | 15 | |
| Operator 2 | Negative | 44 | 4 | 17 | 1 | 0 |
| Positive | 0 | 0 | 0 | 24 | 15 |
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the word "LYHER" in a stylized font. The "LY" portion of the word is in yellow, while the "HER" portion is in green. There is a registered trademark symbol to the right of the word.
| Operator 3 Negative | ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ | 1 - | |||
|---|---|---|---|---|---|
| Positive | CJ | 1 Pﻟ |
Discordant Results of THC Cups
| Operator | SN. | Specimen No. | LC/MS results | Results of Lyher kit |
|---|---|---|---|---|
| Operator 2 | 12 | 202301036 | 50.5 | Negative |
| Operator 3 | 36 | 202301060 | 52.3 | Negative |
| Operator 1 | 45 | 202301069 | 49.6 | Positive |
| Operator 3 | 92 | 202301651 | 51.1 | Negative |
5.11 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou Laihe Biotech Co., Ltd has demonstrated that proposed device LYHER® Urine Multi-Drug Test Kit(Cup), LYHER® Urine Multi-Drug Test Kit(Cassette), LYHER® Urine Multi-Drug Test Kit(Dipcard) are substantially equivalent to ATTEST Drug Screen Cup, ATTEST Drug Screen Dip Card, K182123.
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).