K Number
K182123
Date Cleared
2019-03-29

(235 days)

Product Code
Regulation Number
862.3650
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Ampletamine, Benzoylecgonine, EDDP, d/1-Methadone, d-Methamphetamine, d/Methylenedioxymethamphetamine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteCalibratorCutoff (ng/mL)
6-Acetylmorphine6-monoacetylmorphine10
Amphetamined-Amphetamine500
Amphetamined-Amphetamine1,000
BarbituratesSecobarbital/Pentobarbital300
BenzodiazepinesOxazepam300
BuprenorphineBuprenorphine10
EDDP2-ethylidene-1,5-dimethyl-3-3- diphenylpyrrolidine300
CocaineBenzoylecgonine150
CocaineBenzoylecgonine300
Ecstasyd,l-Methylenedioxymethamphetamine500
Methamphetamined-Methamphetamine500
Methamphetamined-Methamphetamine1,000
Marijuana11-nor-Δ9-THC-9-COOH50
Methadoned/l-Methadone300
OpiatesMorphine300
OpiatesMorphine2,000
OxycodoneOxycodone100
PhencyclidinePhencyclidine25
PropoxyphenePropoxyphene300
Tricyclic AntidepressantsNortriptyline1,000

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board adulteration/specimen validity tests (SVT) in the cup format. The drug screen tests are intended for prescription use only.

The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive positive result if the donor doesn't admit use or anytime required by testing procedures. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Device Description

For professional use, the devices consist of:

  • a. 10 or 25 test cups or dip cards with or without adulteration/specimen validity tests
  • b. Package insert
  • c. Procedure Card
AI/ML Overview

The provided text describes the performance characteristics of the ATTEST Drug Screen Cup and ATTEST Drug Screen Dip Card, which are rapid lateral flow immunoassays for the qualitative detection of various drugs in human urine.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for qualitative drug screening devices are typically based on agreement with a gold-standard confirmatory method (like GC/MS, LC/MS) at and around the cut-off concentrations. The concept of "agreement" translates to sensitivity and specificity for qualitative tests. While explicit numeric acceptance criteria (e.g., "must achieve >95% positive agreement") aren't directly stated as "acceptance criteria" in a separate table, they are implied by the performance statistics presented in the "Method Comparison/Accuracy Studies" section.

The reported device performance is presented in the two large tables under "Method Comparison/Accuracy Studies" (Pages 14-16), showing the agreement of the ATTEST devices (Cup and Dip Card) with GC/MS or LC/MS results.

Let's interpret the tables to deduce the implicit acceptance criteria and the reported performance. The "Agreement" column in these tables represents the percentage of agreement between the device's qualitative result (Negative or Positive) and the quantitative result from the gold-standard method, particularly around the cutoff and at significant concentrations above and below it.

For each drug assay (e.g., 6-AM/10, AMP/500, etc.), the data shows the number of samples that fell into different concentration ranges relative to the cutoff (e.g., "Drug-free", "-50% C/O to <-25% C/O", "C/O to +25% C/O", ">+50% C/O"). The "Agreement" percentage is then provided for both Negative and Positive results.

Implicit Acceptance Criteria (Deduced from Reported Performance):

Drug Test/Cutoff (ng/mL)Acceptance Criteria (Implicit)Reported Device Performance (Agreement %) - ATTEST Cup & Dip Card
GeneralHigh agreement (ideally 100% or very close to it) for drug-free samples and samples with concentrations significantly above the cutoff. Very high agreement is also expected for samples significantly below the cutoff that should test negative, and samples significantly above the cutoff that should test positive.Achieved for all assays as detailed below.
Specific to Each AssayNegative Agreement: Expected to be high for drug-free samples and samples definitively below the cutoff. Positive Agreement: Expected to be high for samples definitively above the cutoff.
6-AM/10Expected high agreement for Negatives (samples < 10 ng/mL) and Positives (samples >= 10 ng/mL).Neg: 100%, Pos: 100%
AMP/500Expected high agreement for Negatives (samples < 500 ng/mL) and Positives (samples >= 500 ng/mL).Neg: 97.7%, Pos: 100%
AMP/1000Expected high agreement for Negatives (samples < 1000 ng/mL) and Positives (samples >= 1000 ng/mL).Neg: 100%, Pos: 100%
BAR/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 95.2%, Pos: 100%
BUP/10Expected high agreement for Negatives (samples < 10 ng/mL) and Positives (samples >= 10 ng/mL).Neg: 95.5%, Pos: 100%
BZO/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 93.2%, Pos: 100% (This is the lowest negative agreement)
COC/150Expected high agreement for Negatives (samples < 150 ng/mL) and Positives (samples >= 150 ng/mL).Neg: 97.70%, Pos: 100%
COC/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 100%, Pos: 100%
EDDP/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 93.2%, Pos: 100% (Similar to BZO/300)
MDMA/500Expected high agreement for Negatives (samples < 500 ng/mL) and Positives (samples >= 500 ng/mL).Neg: 95.5%, Pos: 100%
MET/500Expected high agreement for Negatives (samples < 500 ng/mL) and Positives (samples >= 500 ng/mL).Neg: 93.2%, Pos: 100% (Similar to BZO/300 & EDDP/300)
MET/1000Expected high agreement for Negatives (samples < 1000 ng/mL) and Positives (samples >= 1000 ng/mL).Neg: 100%, Pos: 100%
MTD/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 95.5%, Pos: 100%
OPI/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 97.72%, Pos: 100%
OPI/2000Expected high agreement for Negatives (samples < 2000 ng/mL) and Positives (samples >= 2000 ng/mL).Neg: 93.2%, Pos: 100% (Similar to BZO/300, EDDP/300, MET/500)
OXY/100Expected high agreement for Negatives (samples < 100 ng/mL) and Positives (samples >= 100 ng/mL).Neg: 93.2%, Pos: 100% (Similar to OPI/2000)
PCP/25Expected high agreement for Negatives (samples < 25 ng/mL) and Positives (samples >= 25 ng/mL).Neg: 97.7%, Pos: 100%
PPX/300Expected high agreement for Negatives (samples < 300 ng/mL) and Positives (samples >= 300 ng/mL).Neg: 95.3%, Pos: 100%
TCA/1000Expected high agreement for Negatives (samples < 1000 ng/mL) and Positives (samples >= 1000 ng/mL).Neg: 95.5%, Pos: 100%
THC/50Expected high agreement for Negatives (samples < 50 ng/mL) and Positives (samples >= 50 ng/mL).Neg: 97.7%, Pos: 100%

Note: The performance percentages across the two formats (Cup and Dip Card) are presented identically in the provided combined tables for all assays, suggesting the performance data is pooled or very similar for both. The discordant results tables (pages 15-17) provide a more granular view for specific instances where the device result differed from the GC/MS or LC/MS result, primarily for samples near the cutoff. For example, for AMP/500, a 97.7% negative agreement means 2.3% of negative samples were discordant (likely false negative or near-cutoff negative that read positive by device). For Amp/500, 3 "Neg" results were in the "-50% C/O to <-25% C/O" range (i.e. between 250 and 375 ng/ml). This discrepancy is precisely what explains the 97.7% agreement and is not necessarily a "failure" given the nature of a qualitative test and its proximity to the cutoff.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for the Test Set:

    • For the main Method Comparison/Accuracy study (Table on pages 14-15), the sample size for each assay varies slightly but can be calculated from the row sums.
      • For "Negative" results: typically 40 drug-free samples + varying numbers of samples below cutoff (e.g., for 6-AM/10 Neg, 40+4+1 = 45 samples).
      • For "Positive" results: varying numbers of samples above and around cutoff (e.g., for 6-AM/10 Pos, 1+4+35 = 40 samples).
      • Total sample size per assay appears to be around 80-90 samples (e.g., 6-AM/10 has 45 Neg + 40 Pos = 85 samples). This includes drug-free samples and samples spiked at various concentrations relative to the cutoff (e.g., -50% C/O to <-25% C/O, -25% C/O to C/O, C/O to +25% C/O, >+25% C/O to +50% C/O, >+50% C/O).
    • The "Summary of Discordant Results" tables (pages 15-17) list specific instances of samples where the device's qualitative result did not match the quantitative GC/MS/LC/MS result (e.g., "Positive" device result for a sample with a drug concentration slightly below the cutoff, or "Negative" for a sample slightly above). These are qualitative examples, not the full dataset.
  • Data Provenance: The studies used clinical urine specimens "previously quantitated for the target drugs of abuse by gold-standard methods (GC/MS, LC/MS or equivalent)." The text does not specify the country of origin of the data or whether the samples were retrospective or prospective collections. However, given the nature of a 510(k) submission for a device to be marketed in the US, it is generally assumed that the clinical studies are relevant to the US population.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This type of device (lateral flow immunoassay for drug screening) does not rely on human experts for establishing ground truth for the test set. The ground truth is established by analytical methods (Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions), which are considered the preferred confirmatory methods for drug testing. Therefore, there were no human "experts" in the sense of radiologists interpreting images to establish medical ground truth.

4. Adjudication Method for the Test Set

Not applicable. Since the ground truth is established by objective analytical methods (GC/MS, LC/MS), there's no need for an adjudication process involving human readers or experts. The assay results are compared directly against the quantitative chemical analysis.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The device is a rapid diagnostic test for drug detection, not an AI-powered image analysis tool for human interpretation. The study is a comparison of the device's performance against a gold-standard chemical analysis method, not a study of human readers. Therefore, there is no effect size related to human readers improving with AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the study described is essentially a standalone (algorithm/device-only) performance evaluation. The device's qualitative results are directly compared to the quantitative results from GC/MS/LC/MS, without human interpretation as an intermediate step. The "intended professional users" mentioned in Section M.a.iv (Detection Limit/sensitivity) are those who read the test result (presence/absence of a line), not those who interpret complex data as an AI model would.

7. The Type of Ground Truth Used

The ground truth used for the test set was analytical confirmation by gold-standard chemical methods: Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS), or equivalent. This is explicitly stated in "Method Comparison/Accuracy Studies" (Page 14).

8. The Sample Size for the Training Set

The document describes a 510(k) submission for a diagnostic device. Unlike AI/ML models, these traditional diagnostic devices do not typically have a "training set" in the sense of a data set used to train an algorithm. The device's formulation and antibodies are developed through a research and development process, and its performance is then validated (tested) using various samples.

The document mentions "The precision, reproducibility and sensitivity of the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card were evaluated at Advin Biotech and at three (3) external testing sites. Data obtained at all testing sites across the intended use populations indicate >99% correlation at +/-50% of each assay cutoff." (Page 8-9). This suggests internal development and testing, but not explicitly a "training set" as understood in machine learning. The stability and specificity studies also involve testing with numerous samples.

9. How the Ground Truth for the Training Set was Established

As there is no "training set" for this type of device in the AI sense, this question is not directly applicable. The device's underlying chemistry and design are based on established immunochromatographic principles. The process involves:

  • Standardization: The assays are harmonized with DHHS/SAMHSA screening cutoffs (Page 8-9).
  • Analytical Specificity: This is established by testing the device with contrived solutions of chemically related/structurally similar compounds and other potential interferents (Pages 9-13). This involves verifying that the device reacts appropriately to the target analyte and does not cross-react significantly with other substances.
  • Precision/Reproducibility: Evaluated through repeated testing at various concentrations and across different sites (Page 8).

These evaluations contribute to the overall confidence in the device's performance and design, akin to "establishing ground truth" for its analytical properties, but it's fundamentally different from training an AI model on a labeled dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

March 29, 2019

Advin Biotech, Inc. Daniel Hsu Senior QA/RA Manager 10340 Camino Santa Fe, Suite G San Diego, CA 92121

Re: K182123

Trade/Device Name: ATTEST Drug Screen Cup ATTEST Drug Screen Dip Card Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG, DKZ, DIO, DJC Dated: February 7, 2019 Received: February 12, 2019

Dear Daniel Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

Courtney H. Lias, Ph.D. for Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182123

Device Name ATTEST Drug Screen Cup ATTEST Drug Screen Dip Card

Indications for Use (Describe)

The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Ampletamine, Benzoylecgonine, EDDP, d/1-Methadone, d-Methamphetamine, d/Methylenedioxymethamphetamine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AnalyteCalibratorCutoff (ng/mL)
6-Acetylmorphine6-monoacetylmorphine10
Amphetamined-Amphetamine500
Amphetamined-Amphetamine1,000
BarbituratesSecobarbital/Pentobarbital300
BenzodiazepinesOxazepam300
BuprenorphineBuprenorphine10
EDDP2-ethylidene-1,5-dimethyl-3-3- diphenylpyrrolidine300
CocaineBenzoylecgonine150
CocaineBenzoylecgonine300
Ecstasyd,l-Methylenedioxymethamphetamine500
Methamphetamined-Methamphetamine500
Methamphetamined-Methamphetamine1,000
Marijuana11-nor-Δ9-THC-9-COOH50
Methadoned/l-Methadone300
OpiatesMorphine300
OpiatesMorphine2,000
OxycodoneOxycodone100
PhencyclidinePhencyclidine25
PropoxyphenePropoxyphene300
Tricyclic AntidepressantsNortriptyline1,000

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board adulteration/specimen validity tests (SVT) in the cup format. The drug screen tests are intended for prescription use only.

The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive positive result if the donor doesn't admit use or anytime required by testing procedures. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Date:March 28, 2019
Submitter:Advin Biotech, Inc.10340 Camino Santa Fe, Suite GSan Diego, California 92121
Contact:Daniel HsuTelephone: 858-866-8382, ext. 100Email: daniel.hsu@advinbio.com

A. 510(k) Number:

K182123

B. Purpose for Submission:

  • a. Addition of Amphetamine 1000, Cocaine 300, Methamphetamine 1000 and 6-Acetylmorphine 10 to a previously cleared device (K122809)
  • b. Update formulation of Opiate 300 assay from the same previously cleared device

C. Measurand:

d-Methamphetamine 6-Acetylmorphine, d-Amphetamine, Benzoylecqonine, and Morphine

D. Type of Test:

Qualitative lateral-flow immunoassay

E. Applicant:

Advin Biotech, Inc.

F. Proprietary and Established Names:

  • a. ATTEST Drug Screen Cup
  • b. ATTEST Drug Screen Dip Card

G. Regulatory Information:

AssayPRODUCTCODECLASSIFICATIONREGULATIONNUMBER/DESCRIPTIONPANEL
6-AMDJGII862.3650/EnzymeImmunoassay, OpiatesToxicology
AMPDKZII862.3100/ EnzymeImmunoassay, AmphetamineToxicology
COCDIOII862.3250/EnzymeImmunoassay, Cocaine andcocaine metabolitesToxicology

{5}------------------------------------------------

METDJCII862.3610/MethamphetamineTest SystemToxicology
OPIDJGII862.3650/EnzymeImmunoassay, OpiatesToxicology

H. Intended Use

The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, dll-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:

AssayAbbreviationCalibratorCutoffConcentration(ng/mL)
6-Acetylmorphine6AM6-monoacetylmorphine10
Amphetamine 500AMP500d-Amphetamine500
Amphetamine 1000AMP1000d-Amphetamine1,000
BarbituratesBARSecobarbital/Pentobarbital300
BenzodiazepinesBZOOxazepam300
BuprenorphineBUPBuprenorphine10
EDDPEDDP2-ethylidene-1,5-dimethyl-3-3-diphenylpyrrolidine300
Cocaine 150COC150Benzoylecgonine150
Cocaine 300COC 300Benzoylecgonine300
EcstasyMDMAd,l-Methylenedioxymethamphetamine500
Methamphetamine 500MET500d-Methamphetamine500
Methamphetamine1000MET1000d-Methamphetamine1,000
MarijuanaTHC11-nor-Δ⁹-THC-9-COOH50
MethadoneMTDd/l-Methadone300
Opiates 300OPI300Morphine300
Opiates 2000OPI2000Morphine2,000
OxycodoneOXYOxycodone100

{6}------------------------------------------------

PhencyclidinePCPPhencyclidine25
PropoxyphenePPXPropoxyphene300
TricyclicAntidepressantsTCANortriptyline1,000

The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board adulteration/specimen validity tests (SVT) in the dip card format or in the cup format. Only one cutoff concentration per analyte will be included per device. The drug screen tests are intended for prescription use only.

The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive result if the donor doesn't admit use or anytime required by testing procedures. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

The tests are not intended to differentiate between illicit and prescription use of benzodiazepines, barbiturates, buprenorphine, oxycodone, propoxyphene and tricyclic antidepressants. There are no uniformly recognized cutoff levels for these drugs in urine.

I. Device Descriptions:

For professional use, the devices consist of:

  • a. 10 or 25 test cups or dip cards with or without adulteration/specimen validity tests
  • b. Package insert
  • c. Procedure Card

J. Substantial Equivalence Information:

  • a. Predicate device names:
    • i. Advin Multi-Drug Screen Test Cup
    • ii. Advin Multi-Drug Screen Test Dipcard
  • b. Predicate 510(k) number(s): i. K122809

  • c. Comparisons with predicates:
CharacteristicPredicate (K122809)Candidate Devices
Indications for useQualitative detection of drugs of abuseand/or their metabolites in humanurineSame

Predicate Similarities and Differences Table for All Assays

{7}------------------------------------------------

MethodologyLateral flow immunochromatographic assay based on competitive bindingSame
SpecimenHuman urineSame
Analytes, calibrators, cutoffsAssayCalibratorCutoff (ng/mL)
Amphetamined-amphetamine500
BarbituratesSecobarbital300
BenzodiazepineOxazepam300
BuprenorphineBuprenorphine10
CocaineBenzoylecgonine150
EDDP2-ethylidene-1,5-dimethyl-3-3-diphenyl-pyrrolidine300
Ecstasyd/l-methylene-dioxy-meth-amphetamine500
Methamphetamined-meth-amphetamine500
Methadoned/l-methadone300
MorphineMorphine300
OpiatesMorphine2,000
OxycodoneOxycodone100
PhencyclidinePhencyclidine25
Propoxyphened-propoxyphene300
TricyclicsNortriptyline1,000
MarijuanaTHC-COOH50
Same except the additions of:AssayCalibratorcutoff
6-AM6-acetyl-morphine
Amphetamined-amphetamine
CocaineBenzoylecgonine300
Meth-amphetamined-meth-amphetamine1,000
Test formatsCup, Dip Card, CassetteCup and Dip Card only

K. Standard Referenced:

Cutoffs shown in the table below are harmonized with DHHS/SAMHSA screening cutoffs for drugs of abuse in urine (Federal Register Notice, Volume 82. No.13, January 23, 2017):

{8}------------------------------------------------

Assay NameAbbreviationCalibrator DrugCutofflevel(ng/mL)
6-AM6AM6-monoacetylmorphine10
AmphetamineAMPd-amphetamine500
CocaineCOCBenzoylecgonine150
MethamphetamineMETd-methamphetamine500
MDMAMDMAd/l-methylenedioxymethamphetamine500
Opiates 2000OPIMorphine (free)2000
OxycodoneOXYOxycodone100
PCPPCPPhencyclidine25
THCTHCTHC-COOH50

L. Test Principle:

The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays in which drug-protein conjuqates compete for limited antibody sites with drugs or drug metabolites that may be present in the urine. Each test strip consists of one or two drug-protein conjugates which are printed on nitrocellulose membrane in the test (T) regions. The anti-drug antibody-colloidal gold conjugates are dried onto a pad which is located beneath the nitrocellulose membrane bearing the test line(s). If target drugs are present in the urine specimen below the cut-off concentration of the assay, the solution of the colored antibodycolloidal qold conjuqate that has been rehydrated by the urine sample migrates by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test (T) region. The colored antibody-gold conjugate then binds with the drug-protein conjugate to form visible lines in the test (T) regions. The formation of a visible line in the test (T) region on any strip indicates a negative result, regardless of the intensity of the visible line.

If the tarqet druq level exceeds the cutoff concentration of the druq/metabolite antigen in the urine will bind to and saturate the limited antibody sites during the capillary migration up the strip. The saturated antibody sites will be unavailable to bind to the drugprotein conjugate on the membrane, so no visible lines will appear. Therefore, absence of a visible line in the test (T) region indicates a preliminary positive result. A separate antibodyantigen reaction forms control (C) lines which must appear on the strip to interpret the results. The lack of a visible control (C) line indicates an insufficient specimen volume or improper test technique and the test must be repeated.

M. Performance Characteristics:

  • a. Analytical performance:
    • i. Precision/Reproducibility:

The precision, reproducibility and sensitivity of the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card were evaluated at Advin Biotech and at three (3) external testing sites. Data obtained at all testing sites across the intended use populations indicate >99% correlation at +/-50% of each assay cutoff.

{9}------------------------------------------------

  • ii. Linearity/Assav Reportable Range: Not applicable. Devices intended for qualitative determinations only.
  • iii. Traceability, Stability, Expected Values (controls, calibrators, methods):
      1. Traceability: The assays shown in the table below are harmonized with the DHHS/SAMHSA screening cutoffs for drugs of abuse in urine (Federal Register Notice, Volume 82. No.13, January 23, 2017):
Assay NameAbbreviationCalibrator DrugCutofflevel(ng/mL)
6-AM6AM6-monoacetylmorphine10
AmphetamineAMPd-amphetamine500
CocaineCOCBenzoylecgonine150
MethamphetamineMETd-methamphetamine500
MDMAMDMAd/l-methylenedioxymethamphetamine500
Opiates 2000OPIMorphine (free)2000
OxycodoneOXYOxycodone100
PCPPCPPhencyclidine25
THCTHCTHC-COOH50
    1. Stability: Device stability has been or will be determined using accelerated stress testing and real-time studies. All accelerated studies have been completed. Real-time stability testing in room temperature conditions have been completed for Amphetamine 1000, Cocaine 300 and Methamphetamine 1000 assays. In realtime, room temperature conditions, devices in both formats are stable for 28 months when stored at room temperature in sealed foil. Real-time stability studies for the 6-AM assay and the reformulated OPI 300 assay are ongoing.
    1. Controls: Devices have been evaluated with various 3rd party commercial controls. FDA-cleared controls manufactured by Biochemical Diagnostics, Inc. (a Kova International company) are recommended by Advin Biotech, Inc. for use in verifying the performance of the assays.
  • iv. Detection Limit/sensitivity:

Assays contained within the devices bear the cutoff levels shown in the table in section H above. Field studies demonstrate the sensitivity of the assays in the hands of intended professional users and are summarized in section M.a.i. above.

  • v. Analytical Specificity:
    The specificity of each assay was determined through the testing of contrived solutions made by spiking certified standards of chemically-related or structurally-similar compounds into drug free urine. The relative cross

{10}------------------------------------------------

reactivity (if any) represents the minimum concentration necessary to yield a result similar to the cutoff level of the respective assay.

Assay/CutoffCompoundConcentration(ng/mL)Relative cross-reactivity (%)
6-AM106-acetylmorphine10100
Diacetylmorphine (heroin)3003.3
Morphine>100,000<0.1
Codeine>100,000<0.1
Oxycodone>100,000<0.1
Oxymorphone>100,000<0.1
AMP500d-Amphetamine500100
I-Amphetamine50,0001
MDA8,0006.5
Phentermine45,0001.1
AMP1000d-Amphetamine1,000100
I-Amphetamine100,0001
MDA15,0006.7
Phentermine100,0001.0
BAR300Secobarbital300100
Amobarbital2,50012
Aprobarbital50060
Butabarbital100300
Butalbital300100
Cyclopentobarbital50060
Pentobarbital250120
Phenobarbital300100
Oxazepam300100
Alprazolam200150
BZO300α-Hydroxyalprazolam1,90015.8
Bromazepam1,00030
Clobazam200150
Clonazepam1,50020
7-Aminoclonazepam>100,000<0.3
Clorazepate75040
Desalkylflurazepam1,20025
Diazepam1,00030
Flunitrazepam250120
Lorazepam3,9007.7
Lorazepam glucuronide5,0006
Nitrazepam250120
Norchlordiazepoxide50060
Nordiazepam39076.9
Temazepam150200
Triazolam2,50012
BUP40Buprenorphine10100
Buprenorphine-3-β-glucuronide3.5286
Norbuprenorphine7.5133
Norbuprenorphine glucuronide3528
Benzoylecgonine150100
COCCocaethylene50,0000.3
150Cocaine5,0003
Ecgonine50,0000.3
Benzoylecgonine300100
COCCocaethylene100,0000.3
300Cocaine10,0003
Ecgonine100,0000.3
EDDP2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300100
300d/l-Methadone>100,000<0.3
d-Methamphetamine500100
I-Methamphetamine15,0003.3
d-Amphetamine50,0001
METI-Amphetamine50,0001
5001R, 2S(1)-Ephedrine100,000<0.5
MDEA30,0001.7
MDMA3,50014.3
Mephentermine75,0000.7
d-Methamphetamine1,000100
I-Methamphetamine30,0003.3
d-Amphetamine100,0001
METI-Amphetamine100,0001
10001R, 2S(1)-Ephedrine>100,000<0.5
MDEA60,0001.7
MDMA8,00012.5
Mephentermine100,0001
(+/-)-methylenedioxymethamphetamine(MDMA)500100
MDMA(+/-)-methylenedioxyamphetamine(MDA)3,90012.8
500(+/-)-methylenedioxyethylamphetamine(MDEA)500100
MTDd/l-Methadone300100
300EDDP>100,000<0.3
Morphine300100
OPI6-Acetylmorphine (6-AM)40075
Codeine100300
Codeine-6-β-glucuronide150200
300Ethylmorphine150200
Diacetylmorphine (heroin)90033.3
Hydrocodone50060
Hydromorphone60050
Levorphanol10,0003
Morphine-3-glucuronide45066.7
Norcodeine30,0001
Oxycodone70,000<0.5
Thebaine20,0001.5
Morphine2,000100
6-Acetylmorphine (6-AM)700286
OPI2000Codeine1,800111
Codeine-6-β-glucuronide1,250160
Ethylmorphine1,500133
Diacetylmorphine (heroin)11,00018.2
Hydrocodone5,00040
Hydromorphone5,00040
Levorphanol>100,000<2
Morphine-3-glucuronide2,60077
Norcodeine>100,000<2
Oxycodone70,0003
Thebaine95,0002.1
Oxycodone100100
Codeine50,0000.2
OXY100Ethylmorphine50,0000.2
Hydrocodone5,0002
Hydromorphone25,0000.4
Oxymorphone12,5000.8
PCP25Phencyclidine25100
4-OH-Phencyclidine1,5001.7
PPX300d-Propoxyphene300100
Norpropoxyphene300100
Nortriptyline1,000100
Amitriptyline4,00025
Clomipramine2,00050
TCA1000Desipramine500200
Doxepine1,000100
Imipramine1,000100
Promethazine1,000100
Trimipramine5,00020
11-nor-Δ9-THC-COOH50100
(+/-)-11-Hydroxy-Δ9-THC5,0001
THC50Δ8-THC20,0000.3
Δ9-THC20,0000.3
Cannabinol>100,000<0.1
Cannabidiol (CBD)>100,000<0.1

{11}------------------------------------------------

{12}------------------------------------------------

The potential interference (whether positive or negative) from compounds chemically dissimilar to target drugs, known endogenous agents, urine pH and specific gravity variances was also determined. Testing of compounds and endogenous agents was performed by spiking the substances into pooled urine containing target drugs at nearcutoff concentrations. Unless otherwise indicated, substances were tested for potential interference at concentrations of 100 µg/mL.

{13}------------------------------------------------

The following substances demonstrated no positive or negative interference on the assays encompassed in this submission:

AcetaminophenAcetoneAcetylsalicylic acid(aspirin)
AlbuminAmpicillinAscorbic acid
AspartameAtropineBenzocaine
BilirubinCaffeineChloroquine
ChlorpheniramineCreatineDexbrompheniramine
DextromethorphanDimethylaminoantipyrineDiphenhydramine
DimenhydrinateDopamineIsoproterenol
EphedrineErythromycinEthanol
FurosemideGabapentinGlucose
Guaiacol glyceryl etherHemoglobinIbuprofen
IsoproterenolKetamineLidocaine
MethylephedrineNaproxenNiacinamide
NicotineNorephedrineOxalic acid
PantoprazolePenicillin-GPheniramine
PhenothiazineI-PhenylephrineB-Phenylethylamine
PregablinProcaineQuinidine
RanitidineRiboflavinSertraline
Sodium chlorideSulindacTheophylline
Tyramine

In addition to the spiked testing described above, in light of industry-known interferences presumably arising from the metabolites of these drugs in human urine. Advin Biotech has evaluated its assays for potential false positive results using clinical urine specimens from donors taking self-reported prescription-level or OTC high-dose levels of each of the following drugs and found no false positive results:

  • a. Rantidine (Zantac®)-donors taking high-doses per OTC instructions for use
  • b. Sertraline (Zoloft®)-donor taking prescription dose (steady-state level)
  • c. Naproxen (Aleve®)-donor taking high-dose per OTC instructions for use
  • d. Pantoprazole (Protonix®)-donor taking high-dose per OTC instructions for use
  • e. Gabapentin (Neurontin®)-donor taking high prescription doses of 900mg/day
  • Pregabalin (Lyrica®)-donor taking prescription dose f.
  • g. Atomoxetine (Strattera®)-donor taking prescription dose

To evaluate the potential effect of variances in urine pH on the assays, pooled urine specimens containing target drugs at near-cutoff concentrations were pH adjusted from 4.0 to 9.0 in increments of 1.0 and tested in duplicate.

To evaluate the potential effect of variances of urine specific gravity on the assays, pooled urine specimens containing target drugs at near-cutoff concentrations were diluted using deionized water or concentrated using sodium chloride to achieve specific gravity results of 1.003, 1.010, 1.015, 1.020, 1.025 and 1.030. Each solution was tested in duplicate.

The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.003 to 1.030 do not affect the results of the assays. To mitigate the risk of specimen tampering,

{14}------------------------------------------------

adulteration or substitution, the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card can be constructed with specimen validity tests (SVT).

  • b. Method Comparison/Accuracy Studies
    Advin Biotech performed a method comparison study of the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card using clinical urine specimens previously quantitated for the target drugs of abuse by gold-standard methods (GC/MS, LC/MS or equivalent). The results are shown in the table below. Data for previously-cleared assays (K122809) are taken from that submission and presented in combination with the new assays below:
Drug Test/Cutoff (ng/mL)ResultDrug-free-50% C/O to <-25% C/O-25% C/O to C/OC/O to +25% C/O>+25% C/O to +50% C/O>+50% C/OAgreement
6-AM/10Neg4041000100%
6-AM/10Pos0001435100%
AMP/500Neg403000097.7%
AMP/500Pos0012245100%
AMP/1000Neg4033000100%
AMP/1000Pos0003340100%
BAR/300Neg401100095.2%
BAR/300Pos0025236100%
BUP/10Neg401100095.5%
BUP/10Pos0028032100%
BZO/300Neg400100093.2%
BZO/300Pos0031634100%
COC/150Neg400300097.70%
COC/150Pos0014153100%
COC/300Neg4032000100%
COC/300Pos0002335100%
EDDP/300Neg400100093.2%
EDDP/300Pos0035233100%
MDMA/500Neg401100095.5%
MDMA/500Pos0025134100%
MET/500Neg401000093.2%
MET/500Pos0031351100%
MET/1000Neg4033000100%
MET/1000Pos0002340100%
MTD/300Neg400200095.5%
MTD/300Pos0024037100%
OPI/300Neg402100097.72%
OPI/300Pos0013146100%
OPI/2000Neg401000093.2%
OPI/2000Pos0024340100%
OXY/100Neg401000093.2%
OXY/100Pos0037133100%
PCP/25Neg400300097.7%
PCP/25Pos0013833100%
PPX/300Neg400100095.3%
PPX/300Pos0025233100%
TCA/1000Neg400200095.5%
TCA/1000Pos0025728100%
THC/50Neg401200097.7%
THC/50Pos0014744100%

The ATTEST Drug Screen Cup format

{15}------------------------------------------------

Drug Test/Cutoff (ng/ml)ATTEST Cup ResultResult w/ GC/MS or LC/MS
Drug Concentration (ng/ml)Analyte
AMP/500Positive477Amphetamine
BAR/300Positive265Barbital
Positive286Barbital
BUP/10Positive8Buprenorphine
Positive9Buprenorphine
BZO/300Positive244Oxazepam
Positive252Oxazepam
Positive295Oxazepam
COC/150Positive146Benzoylecgonine
EDDP/300Positive250EDDP
Positive263EDDP
Positive275EDDP
MDMA/500Positive368MDMA
Positive381MDMA
MET/500Positive394Methamphetamine
Positive461Methamphetamine
Positive478Methamphetamine
OPI/300Positive289Morphine
MTD/300Positive266Methadone
Positive273Methadone
OPI/2000Positive1,898Morphine
Positive1,990Morphine
OXY/100Positive88Oxycodone
Positive98Oxycodone
Positive99Oxycodone
PCP/25Positive22.9Phencyclidine
PPX/300Positive242Norpropoxyphene
Positive285Norpropoxyphene
TCA/1000Positive786Nortriptyline
Positive859Nortriptyline
THC/50Positive4911-nor-Δ9-THC-9-COOH

Summary of Discordant Results, ATTEST Cup format

{16}------------------------------------------------

DrugDrug-Drug quantitation by gold-standard method relative to assay cutoff level
Test/Cutoff(ng/mL)Resultfree-50% C/O to <-25% C/O-25% C/O to C/OC/O to +25% C/O>+25% C/O to +50% C/O>+50% C/OAgreement
6-AM/10Neg4041000100%
Pos0001435100%
AMP/500Neg403000097.7%
Pos0012245100%
AMP/1000Neg4033000100%
Pos0003340100%
BAR/300Neg401100095.2%
Pos0025236100%
BUP/10Neg401100095.5%
Pos0028032100%
BZO/300Neg400100093.2%
Pos0031634100%
COC/150Neg400300097.70%
Pos0014153100%
COC/300Neg4032000100%
Pos0002335100%
EDDP/300Neg400100093.2%
Pos0035233100%
MDMA/500Neg401100095.5%
Pos0025134100%
MET/500Neg401000093.2%
Pos0031351100%
MET/1000Neg4033000100%
Pos0002340100%
MTD/300Neg400200095.5%
Pos0024037100%
OPI/300Neg402100097.72%
Pos0013146100%
OPI/2000Neg401000093.2%
Pos0024340100%
OXY/100Neg401000093.2%
Pos0037133100%
PCP/25Neg400300097.7%
Pos0013833100%
PPX/300Neg400100095.3%
Pos0025233100%
TCA/1000Neg400200095.5%
Pos0025728100%
THC/50Neg401200097.7%
Pos001000100%

The ATTEST Drug Screen Dip Card format

{17}------------------------------------------------

Drug Test/Cutoff (ng/ml)ATTEST Dip Card ResultDrug Concentration (ng/ml)Analyte
AMP/500Positive477Amphetamine
BAR/300Positive265Barbital
BAR/300Positive286Barbital
BUP/10Positive8Buprenorphine
BUP/10Positive9Buprenorphine
BZO/300Positive244Oxazepam
BZO/300Positive252Oxazepam
BZO/300Positive295Oxazepam
COC/150Positive146Benzoylecgonine
EDDP/300Positive250EDDP
EDDP/300Positive263EDDP
EDDP/300Positive275EDDP
MDMA/500Positive368MDMA
MDMA/500Positive381MDMA
MDMA/500Positive394Methamphetamine
MET/500Positive461Methamphetamine
MET/500Positive478Methamphetamine
MTD/300Positive266Methadone
MTD/300Positive273Methadone
OPI/300Positive289Morphine
OPI/2000Positive1,898Morphine
OPI/2000Positive1,990Morphine
OXY/100Positive88Oxycodone
OXY/100Positive98Oxycodone
OXY/100Positive99Oxycodone
PCP/25Positive22.9Phencyclidine
PPX/300Positive242Norpropoxyphene
PPX/300Positive285Norpropoxyphene
TCA/1000Positive786Nortriptyline
TCA/1000Positive859Nortriptyline
THC/50Positive4911-nor-Δ9-THC-9-COOH

Summary of Discordant Results, ATTEST Dip Card format

c. Clinical Studies

  • i. Clinical sensitivity: refer to accuracy table shown in section M.b. above
  • ii. Clinical specificity: refer to accuracy table show in section M.b. above
  • iii. Other clinical supportive data (when i and ii are not applicable): N/A

d. Read Time

To evaluate the reading time flexibility, the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card and were tested with drug standards at the concentration of 0, -50% and +50% cutoff level. The result was read and recorded at 4, 5, 10, 30, 60 and 75 minutes. Each solution was tested in duplicate. Data showed that the test results were stable for up to 75 minutes.

{18}------------------------------------------------

N. Proposed Labeling

The proposed labeling is deemed sufficient based on the outcomes of all intended user studies and meets the requirements of 21 CFR 809.10.

O. Conclusions

The data submitted in this premarket notification is complete and supports a substantial equivalence determination.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).