K Number
K232435

Validate with FDA (Live)

Date Cleared
2024-04-26

(256 days)

Product Code
Regulation Number
874.4760
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The endoscope is a sterile, single endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via an Image Processor. The endoscope is intended for use in a hospital environment. It is designed for use in adults.

Device Description

The Rhinolaryngoscope System consists of Single-Use Rhinolaryngoscope (six models shown in below) to be introduced within the nasal lumens and upper airway anatomy and Image Processor (model: VLM-02) for clinical image processing. The Image Processor provides power and processes the images for medical electronic endoscope. The Single-Use Rhinolaryngoscope is a sterile single used flexible Rhinolaryngoscope. The Image Processor is a reusable monitor.

AI/ML Overview

The document provided is a 510(k) premarket notification for a medical device, the Rhinolaryngoscope System. It details the device's characteristics, intended use, and comparison to predicate devices, along with performance data. However, this document does not contain information about an AI/ML-based device or a comparative effectiveness study involving human readers and AI assistance.

The device described is a Rhinolaryngoscope system, which is an endoscope for visualization. The performance data provided are primarily related to the physical and electrical aspects of the device, such as biocompatibility, sterilization, electrical safety, EMC, and bench performance (optical performance, mechanical bending, photobiological safety, thermal safety). There is no mention of an algorithm or AI component that would require an acceptance criteria table for AI performance metrics, a test set with ground truth from experts, or multi-reader multi-case studies.

Therefore, I cannot fulfill the request for information on acceptance criteria and study proving an AI/ML device meets them based on the provided text. The questions regarding sample size, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, type of ground truth, training set size, and how training set ground truth was established are irrelevant to the information present in this 510(k) submission, as it does not pertain to an AI/ML component.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

April 26, 2024

Shenzhen HugeMed Medical Technical Development Co., Ltd. % Yang Jie Consultant Shenzhen Chonconn Medical Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China

Re: K232435

Trade/Device Name: Rhinolaryngoscope system Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: March 27, 2024 Received: March 27, 2024

Dear Yang Jie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232435

Device Name Rhinolaryngoscope System

Indications for Use (Describe)

The endoscope is a sterile, single endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via an Image Processor. The endoscope is intended for use in a hospital environment. It is designed for use in adults.

Type of Use ( Select one or both, as applicable )
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

The assigned 510(k) number is K232435.

Prepared Date: 2024/04/25

1. Submission sponsor

Name: Shenzhen HugeMed Medical Technical Development Co., Ltd.

Address: 415, 416-1,516-1, Building 2, No. 1, Mawu Road, Baoan Community, Yuanshan Street, Longgang District, Shenzhen, Guangdong, 518115, China Contact person: Cathy Title: Regulatory engineer E-mail: cathy.shi@hugemed.cn Tel: +86-755-22275833

2. Submission correspondent

Name: Shenzhen Chonconn Medical Device Consulting Co., Ltd. Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China Contact person: Yang Jie E-mail: yangjie@chonconn.com Tel: +86-755 33941160

Trade/Device NameRhinolaryngoscope System
ModelSingle-use Rhinolaryngoscope: RH-M58, RH-M52, RH-M50,RH-M40, RH-M32, RH-M22Image Processor: VLM-02
Common NameRhinolaryngoscope System
Regulatory ClassClass II
Classification21CFR 874.4760 / Nasopharyngoscope (flexible or rigid) andaccessories / EOB
Submission typeTraditional 510(K)

3. Subject Device Information

4. Predicate Device

    1. Predicate device 510(k) number: K191080 Product name: Ambu aScope 4 RhinoLaryngo Slim Submitter: Ambu Inc.
  • Reference device 2) 510(k) number: K190972 Product name: Ambu AScope 4 RhinoLaryngo Intervention

{4}------------------------------------------------

Submitter: Ambu A/S

Reference device 3)

510(k) number: K221581 Product name: Single-Use Flexible Rhinolaryngoscope Submitter: Hunan Vathin Medical Instrument Co., Ltd.

ട. Device Description

The Rhinolaryngoscope System consists of Single-Use Rhinolaryngoscope (six models shown in below) to be introduced within the nasal lumens and upper airway anatomy and Image Processor (model: VLM-02) for clinical image processing. The Image Processor provides power and processes the images for medical electronic endoscope.

System nameComponent nameModel
RhinolaryngoscopeSystemSingle-UseRhinolaryngoscopeRH-M58, RH-M52, RH-M50, RH-M40, RH-M32, RH-M22
Image ProcessorVLM-02

The Single-Use Rhinolaryngoscope is a sterile single used flexible Rhinolaryngoscope. The Image Processor is a reusable monitor.

The differences between the Single-use Rhinolaryngoscope models are as follow:

  • Have or haven't working channel
  • Working channel inner diameter
  • Insertion tube outer diameter

6. Intended use & Indication for use

The endoscope is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the nasal lumens and upper airway anatomy. The endoscope is intended to provide visualization via an Image Processor.

The endoscope is intended for use in a hospital environment. It is designed for use in adults.

7. Comparison to the Predicate Device

The Rhinolaryngoscope System consists of Single-Use Rhinolaryngoscope (six models shown in below) and Image Processor (model: VLM-02). However, the Image Processor (model: VLM-02) has been cleared by K222910. The following lists the comparison with the predicate device for the Single-Use Rhinolaryngoscope.

FeaturesSubject DevicePredicateDeviceComparison
K numberK232435K191080/
ManufacturerShenzhen HugeMedMedical TechnicalAmbu Inc./
FeaturesSubject DevicePredicate DeviceComparison
ModelDevelopment Co.,Ltd.Ambu aScope 4RhinoLaryngo Slim/
Single-useRhinolaryngoscope:RH-M58, RH-M52,RH-M50, RH-M40,RH-M32, RH-M22
Product code&Classification21 CFR 874.4760,EOB Class II21 CFR874.4760,EOB Class IISame
Intended useThe endoscope is asterile, single-use,flexible endoscopeintended forendoscopicprocedures andexamination withinthe nasal lumens andupper airwayanatomy. Theendoscope isintended to providevisualization via anImage Processor.The endoscope isintended for use in ahospitalenvironment. It isdesigned for use inadults.The endoscope isa sterile, single-use, flexibleendoscopeintended forendoscopicprocedures andexaminationwithin the nasallumens andupper airwayanatomy. Theendoscope isintended toprovidevisualization viaAmbu displayingunit.The endoscope isintended for usein a hospitalenvironment. Itis designed foruse in adults.Same
Scope typeFlexibleFlexibleSame
Field of view(degree)120°85°SimilarAlthough the field of viewis slightly different from thepredicate device, the area
FeaturesSubject DevicePredicate DeviceComparison
within the view can beadjusted by the advance orretreat of the endoscope.Therefore, the difference inFOV does not affect thesafety and effectiveness ofthe subject device.
Direction ofview(degree)Same
Depth ofField3-50mm6-50mmSimilarThe subject device has awider range of depth of viewthan the predicate device togive physicians moreoptions for diagnosis andtreatment based on thepatient's condition.Therefore, this differencewill not raise new questionon safety and effectivenessof the proposed device.
Bendingangle(degree)210° (up and down)130° (up anddown)Different, but the K221581reference device has thesame 210° (up and down)Bending angle.Because the Bending angleis larger, it means moreflexibility in clinical use.Therefore, this differencewill not raise new questionon safety and effectivenessof the proposed device.
Maximuminsertionportionwidth(mm)RH-M22:2.2±10%mmRH-M32:3.2±10%mmRH-M40:4.4±10%mm3.5 mmSimilar, and the K190972reference device also has asimilar 5.5 mm Maximuminsertion portion width.The difference in Maximuminsertion portion width for
FeaturesSubject DevicePredicate DeviceComparison
RH-M50:$5.0\pm10%mm$the subject device does notalter or change theindications for use or resultin a new intended use.
RH-M52:$5.2\pm10%mm$
RH-M58:$5.8\pm10%mm$
Minimuminsertionchannelwidth(mm)RH-M22: /RH-M32: $1.2\pm10%$mmRH-M40: Φ≥1.1 mmRH-M50: $2.8\pm10%$mmRH-M52: Φ≥2.1 mmRH-M58: Φ≥2.50mm3.0mmSimilarThe difference in Minimuminsertion channel width forthe subject device does notalter or change theindications for use or resultin a new intended use.
Workinglength(mm)350 mm300 mmSimilar, and the K190972reference device has thesame 350 mm Workinglength.The length of 300mm meetsthe needs of clinical use.Therefore, this differencewill not raise new questionon safety and effectivenessof the proposed device.
IlluminationSourceLEDLEDSame
LED LightsourceAt distal tipAt distal tipSame
Single-useYesYesSame
EMCIEC 60601-1-2IEC 60601-1-2Same
SafetyIEC 60601-1IEC 60601-2-18IEC 60601-1IEC 60601-2-18Same
Bench testISO 8600-1ISO 8600-3ISO 8600-4ISO 8600-1ISO 8600-3ISO 8600-4Same
SterilizationEO SterilizationEO SterilizationSame

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

8. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

Biocompatibility of the Single-Use Rhinolaryngoscope was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following tests were performed, as recommended:

  • . Cytotoxicity
  • . Sensitization
  • Oral Mucosa Irritation

The Single-Use Rhinolaryngoscope is considered surface - mucosal membrane contacting for a duration of less than 24 hours.

Sterilization and shelf life testing

The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters. The shelf life of the Single-use Rhinolaryngoscope is validated.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Rhinolaryngoscope System. The system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.

Bench performance testing

The following bench tests were performed:

  1. Optical performance testing according to ISO 8600 series.

  2. Color performance (color reproduction), geometric distortion, optical performance (resolution, depth of field and image intensity), SNR and dynamic range, image frame frequency and system delay test compared with the predicate device.

    1. Mechanical bending testing.
    1. Photobiological safety test verified compliance to IEC 62471:2006
  1. Thermal safety performance test verified compliance to Protection against excessive temperature and other safety hazards of IEC 60601-2-18:2009

9. Clinical study

No clinical study is included in this submission

10. Conclusion

Substantial equivalence comparisons, performance testing and compliance with voluntary

{9}------------------------------------------------

standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.

§ 874.4760 Nasopharyngoscope (flexible or rigid) and accessories.

(a)
Identification. A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.