(128 days)
Internal Locking Plate and Screw Fixation System is intended for temporary fixation, correction or bones in various anatomical regions.
3.5mm LCP Clavicle Plate is indicated for fixation of fractures, malunions, and osteotomies of the clavice.
LCP Proximal Humerus Plate is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
3.5mm LCP Distal Humerus plate is indicated for intraarticular fractures of the distal humerus, comminuted supracondylar fractures, osteotomies, and non-unions of the distal humerus.
2.4 mm LCP Volar Distal Radius Plate is indicated for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
5.0 LCP Distal Femur Plate is indicated for buttressing multifragmentary distal femur fractures including; supracondylar, intra-articular and extra-aticular condylar, periprosthetic fractures in normal or osteopenic bone, nonunions and malunions, and osteotomies of the femur.
3.5mm LCP Medial Proximal Tibia Plate is indicated to buttress metaphyseal fractures of the medial tibia plateau, split-type fractures of the medial tibia plateau, medial split fractures with associated depressions and split or depression fractures of the medial tibia plateau. Also, for use in the fixation of osteopenic bone and fixation of nonunions and malunions of the medial proximal tibia and tibia shaft.
3.5mm LCP Anterolateral Distal Tibia Plate is indicated for fractures, and non-unions of the distal tibia, especially in osteopenic bone.
2.7mm/3.5mm LCP Distal Fibula Plate is indicated for fractures, and non-unions of the metaphyseal and diaphyseal region of the distal fibula, especially in osteopenic bone.
2.0mm LCP condylar Plate is indicated for fixation of fractures, nonunions, replantations, and fusions of small bones and small bone fragments, particularly in osteopenic bone. Examples include, but are not limited to, the hand, wrist, foot, and ankle.
The 3.5mm Round Hole Reconstruction Locking Plates is indicated for fixation of fractures, osteotomies and non-unions of the pelvis, particularly in osteopenic bone for adult patients.
Internal Locking Plate and Screw Fixation System consists of plates in various designs and size. Plates are provided in straight designs and various geometric configurations that are commonly used in traumatic and reconstructive surgery. Plates are provided with screw holes to accommodate non-locking and locking screws. They are named according to both anatomical positions and biomechanical functions of the plates. The Internal Locking Bone Plate and Screw Fixation System is divided into four types: large, small, mini, and reconstruction. The length of the plate varies from 45 to 316mm, and hole number from 3 to 22 holes.
Geasure Internal Locking Bone Screw System is used either to fasten plates or similar devices onto bones or to hold bone fragments together. The screws are differentiated by the manner in which they are inserted into bone, their function, their size, and the type of bone they are intended for. There are two styles: locking screw, cortex or cancellous screw. The screw recess is hexagon or star shaped to allow screw removal and insertion. The thread diameter of screw varies from 2.0 to 6.5mm, total length from 6 to 120mm.
Internal Locking Plate and Screw Fixation System is a construct in which the screws are locked in the plate. The heads of the locking screws contain male threads while the holes in the plates contain female thread, this allows the screw head to be threaded into the plate. Various plate designs are available depending on specific anatomic locations, these may be larger or smaller, thicker or thinner as appropriate to various anatomic sites and the loads to which they will be subjected. The holes in the plate are designed for locking screws, or locking and non-locking screws combination to facilitate dynamic compression. By combining locking screw holes with compression screw slots in the shaft, the plate can be used as both a locking device and a fracture compression device. Internal Locking plate and screw Fixation systems do not disrupt the underlying cortical bone perfusion as much as conventional plates, which compress the plate to the cortical bone.
This document primarily focuses on the K232394 Internal Locking Plate and Screw Fixation System, a medical device for bone fixation, and its substantial equivalence to predicate devices, as reviewed by the FDA. It outlines the regulatory classification, intended use, and technical characteristics.
Crucially, this document is a 510(k) Premarket Notification and does NOT contain information about an AI/ML-based device or a clinical study proving performance against specific acceptance criteria for such a device.
Therefore, I cannot fulfill your request for descriptions of:
- A table of acceptance criteria and the reported device performance: This document refers to bench tests for mechanical properties (e.g., bend testing, torsional properties, pullout strength) and reprocessing and sterilization validation, which are standard for implantable devices. There are no acceptance criteria or performance metrics related to an AI/ML algorithm or its diagnostic/clinical performance.
- Sample sizes used for the test set and data provenance: No test sets in the context of AI/ML are mentioned. The "tests" performed are mechanical and reprocessing validations.
- Number of experts used to establish ground truth & qualifications: Not applicable as there's no diagnostic AI/ML component.
- Adjudication method: Not applicable.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not applicable. The document explicitly states "No clinical study data is submitted in this 510(k)."
- Standalone (algorithm only) performance: Not applicable.
- Type of ground truth used: Not applicable.
- Sample size for the training set: Not applicable.
- How ground truth for the training set was established: Not applicable.
What the document does provide is information critical for a mechanical implant device:
- Device Name: Internal Locking Plate and Screw Fixation System
- Intended Use: Temporary fixation, correction, or stabilization of bones in various anatomical regions, with specific indications for different plates (e.g., clavicle, humerus, radius, femur, tibia, fibula, small bones, pelvis).
- Regulatory Class: Class II
- Product Codes: HRS (Plate Fixation, Bone), HWC (Screw, Fixation, Bone)
- Materials: Titanium alloy (Ti-6Al-4V) ELI
- Sterilization: Provided non-sterile; manual cleaning and steam sterilization instructions are provided for the end-user.
- Non-clinical tests performed (bench tests):
- Single Cycle Bend Testing for bone plate (per ASTM F382-17)
- Fatigue Property Testing for bone plate (per ASTM F382-17)
- Torsional Properties Test for screws (per ASTM F543-17)
- Driving Torque Test for screws (per ASTM F543-17)
- Axial Pullout Strength Test for screws (per ASTM F543-17)
- Self-Tapping Performance Test for screws (per ASTM F543-17)
- Reprocessing and sterilization validation (following FDA guidance, AAMI TIRs, and ISO standards)
- Premarket Submission Type: 510(k) Premarket Notification – this means the device is seeking clearance based on substantial equivalence to predicate devices, not approval based on extensive clinical trials for a novel technology like AI.
- Lack of Clinical Data: The document explicitly states, "No clinical study data is submitted in this 510(k)." and "No animal study data is submitted in this 510(k)."
In summary, the provided text describes a conventional orthopedic implant, not an AI/ML diagnostic device, and therefore does not contain the information required to answer your specific questions related to AI acceptance criteria and study methodologies.
{0}------------------------------------------------
December 15, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Changzhou Geasure Medical Apparatus and Instruments Co., Ltd % Xiaoqing Xue Registration Engineer Sinow Medical AS Høyteknologisenteret, Thormøhlens Gate 55 Bergen, 5008 Norway
Re: K232394
Trade/Device Name: Internal Locking Plate and Screw Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 13, 2023 Received: November 13, 2023
Dear Xiaoqing Xue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
K232394 - Xiaoqing Xue
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K232394
Device Name
Internal Locking Plate and Screw Fixation System
Indications for Use (Describe)
Internal Locking Plate and Screw Fixation System is intended for temporary fixation, correction or bones in various anatomical regions.
3.5mm LCP Clavicle Plate
3.5mm LCP Clavicle Plate is indicated for fixation of fractures, malunions, and osteotomies of the clavice.
3.5mm LCP Proximal Humeral Plate
LCP Proximal Humerus Plate is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.
3.5mm LCP Distal Humerus plate
3.5mm LCP Distal Humerus plate is indicated for intraarticular fractures of the distal humerus, comminuted supracondylar fractures, osteotomies, and non-unions of the distal humerus.
2.4mm LCP Volar Distal Radius Plate
2.4 mm LCP Volar Distal Radius Plate is indicated for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
5.0mm LCP Distal Femur Plate
5.0 LCP Distal Femur Plate is indicated for buttressing multifragmentary distal femur fractures including; supracondylar, intra-articular and extra-aticular condylar, periprosthetic fractures in normal or osteopenic bone, nonunions and malunions, and osteotomies of the femur.
3.5mm LCP Medial Proximal Tibial Plate
3.5mm LCP Medial Proximal Tibia Plate is indicated to buttress metaphyseal fractures of the medial tibia plateau, splittype fractures of the medial tibia plateau, medial split fractures with associated depressions and split or depression fractures of the medial tibia plateau. Also, for use in the fixation of osteopenic bone and fixation of nonumions and malunions of the medial proximal tibia and tibia shaft.
3.5mm LCP Anterolateral Distal Tibia Plate
3.5mm LCP Anterolateral Distal Tibia Plate is indicated for fractures, and non-unions of the distal tibia, especially in osteopenic bone.
2.7mm/3.5mm LCP Distal Fibula Plate
2.7mm/3.5mm LCP Distal Fibula Plate is indicated for fractures, and non-unions of the metaphyseal and diaphyseal region of the distal fibula, especially in osteopenic bone.
2.0mm LCP Condylar Plate
2.0mm LCP condylar Plate is indicated for fixation of fractures, nonunions, replantations, and fusions of small bones and small bone fragments, particularly in osteopenic bone. Examples include, but are not limited to, the hand, wrist, foot, and ankle.
{3}------------------------------------------------
3.5mm Round Hole Reconstruction Locking Plate
The 3.5mm Round Hole Reconstruction Locking Plates is indicated for fixation of fractures, osteotomies and non-unions of the pelvis, particularly in osteopenic bone for adult patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(K) Summary
K232394
| Preparation Date: | Dec. 09, 2023 | |
|---|---|---|
| Submitter | Changzhou Geasure Medical Apparatus and Instruments Co., LtdNo. 12, Jinfeng Road, West Taihu Science and Technology IndustrialPark, Changzhou, Jiangsu, P.R. China | |
| Contact | Jing Huang, Management RepresentativeChangzhou Geasure Medical Apparatus and Instruments Co., LtdNo. 12, Jinfeng Road, West Taihu Science and Technology IndustrialPark, Changzhou, Jiangsu, P.R. ChinaPostcode: 213149Email: huangjing@geasure.comPhone:+86 13656146897 | |
| DesignatedSubmissionCorrespondent | Company: Sinow Medical ASAddress: Vestre Fantoftåsen 44, 5072, Bergen, NorwayContact Person: Xiaoqing XueTelephone: +86 15161196032Email: xue@bergemed.com | |
| Subject Device | Trade name | Internal Locking Plate and Screw Fixation System |
| Regulatory Class | II | |
| Regulation Number | 21 CFR 888.3030, 21 CFR 888.3040 | |
| Classification Panel | Orthopedic | |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories;Smooth or threaded metallic bone fixation fastener; | |
| Product Codes | HRS, HWC | |
| Common name forproduct codes | HRS:Plate Fixation, BoneHWC: Screw, Fixation, Bone | |
| Manufacturer | DePuy Synthes | |
| Primary PredicateDevice | Trade name | Synthes (USA) Modular Mini Fragment LCP System; |
| 510(K) number | K063049 | |
| Regulatory Class | II | |
| Regulation Number | 21 CFR 888.3030, 21 CFR 888.3040 | |
| Classification Panel | Orthopedic | |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories;Smooth or threaded metallic bone fixation fastener; | |
| Product Codes | HRS, HWC | |
| Common name for | HRS:Plate Fixation, Bone | |
| Additional PredicateDevices | product codes | HWC: Screw, Fixation, Bone |
| Manufacturer | DePuy Synthes | |
| K073186, Synthes 3.5mm LCP Clavicle PlateSystem; | ||
| K011815, Synthes LCP Proximal Humerus Plate; | ||
| K033995, 3.5mm LCP Distal Humerus System; | ||
| K083694, Synthes 2.4mm VA-LCPTwo-Column Volar Distal Radius Plate; | ||
| Trade name /510(K)number | K062564, Synthes LCP Distal Femur Plate; | |
| K032269, Synthes (USA) 3.5/4.5mm LCPMedial Proximal Tibia Plates; | ||
| K092812, Synthes (USA) 2.7mm /3.5mm LCPAnterolateral Distal Tibia Plates; | ||
| K083213, Synthes 2.7mm/3.5mm LCP DistalFibula Plates-Modification; | ||
| K082807, Synthes (USA) 3.5 mm and 4.5 mmLocking Compression Plate (LCP) System withExpanded Indications | ||
| Regulatory Class | II | |
| Regulation Number | 21 CFR 888.3030, 21 CFR 888.3040 | |
| Product Codes | HRS, HWC | |
| Classification Panel | Orthopedic | |
| Classification Name | Single/multiple component metallic bone fixationappliances and accessories;Smooth or threaded metallic bone fixationfastener; | |
| Reference device | Manufacturer | Changzhou Geasure Medical Apparatus andInstruments Co., Ltd |
| Trade name | Spinal Inner Fixation System | |
| 510(K) number | K222031 | |
| Regulatory Class | II | |
| Regulation Number | 21 CFR 888.3070 | |
| Product Codes | NKB | |
| Classification Panel | Orthopedic | |
| Classification Name | Thoracolumbosacral Pedicle Screw system | |
| Intendeduse/indications foruse | Internal Locking Plate and Screw Fixation System is intended fortemporary fixation, correction or stabilization of bones in variousanatomical regions. | |
| 3.5mm LCP Clavicle Plate | ||
| 3.5mm LCP Clavicle Plate is indicated for fixation of fractures,malunions, non-unions, and osteotomies of the clavicle. | ||
| 3.5mm LCP Proximal Humeral Plate | ||
| LCP Proximal Humerus Plate is indicated for fractures and fracture | ||
| 3.5mm LCP Distal Humerus plate3.5mm LCP Distal Humerus plate is indicated for intraarticular fracturesof the distal humerus, comminuted supracondylar fractures,osteotomies, and non-unions of the distal humerus. | ||
| 2.4mm LCP Volar Distal Radius Plate2.4 mm LCP Volar Distal Radius Plate is indicated for fixation ofcomplex intra- and extra-articular fractures and osteotomies of the distalradius and other small bones. | ||
| 5.0mm LCP Distal Femur Plate5.0 LCP Distal Femur Plate is indicated for buttressingmultifragmentary distal femur fractures including: supracondylar,intra-articular and extra-articular condylar, periprosthetic fractures andfractures in normal or osteopenic bone, nonunions and malunions, andosteotomies of the femur. | ||
| 3.5mm LCP Medial Proximal Tibial Plate3.5mm LCP Medial Proximal Tibia Plate is indicated to buttressmetaphyseal fractures of the medial tibia plateau, split-type fractures ofthe medial tibia plateau, medial split fractures with associateddepressions and split or depression fractures of the medial tibia plateau.Also, for use in the fixation of osteopenic bone and fixation ofnonunions and malunions of the medial proximal tibia and tibia shaft. | ||
| 3.5mm LCP Anterolateral Distal Tibia Plate3.5mm LCP Anterolateral Distal Tibia Plate is indicated for fractures,osteotomies, and non-unions of the distal tibia, especially in osteopenicbone. | ||
| 2.7mm/3.5mm LCP Distal Fibula Plate2.7mm/3.5mm LCP Distal Fibula Plate is indicated for fractures,osteotomies, and non-unions of the metaphyseal and diaphyseal regionof the distal fibula, especially in osteopenic bone. | ||
| 2.0mm LCP Condylar Plate2.0mm LCP condylar Plate is indicated for fixation of fractures,osteotomies, nonunions, replantations, and fusions of small bones andsmall bone fragments, particularly in osteopenic bone. Examplesinclude, but are not limited to, the hand, wrist, foot, and ankle. | ||
| 3.5mm Round Hole Reconstruction Locking PlateThe 3.5mm Round Hole Reconstruction Locking Plate is indicated forfixation of fractures, osteotomies and non-unions of the pelvis,particularly in osteopenic bone for adult patients. | ||
| Device Description | Internal Locking Plate and Screw Fixation System consists of plates invarious designs and size. Plates are provided in straight designs and | |
| Mechanism ofAction | various geometric configurations that are commonly used in traumaticand reconstructive surgery. Plates are provided with screw holes toaccommodate non-locking and locking screws. They are namedaccording to both anatomical positions and biomechanical functions ofthe plates. The Internal Locking Bone Plate and Screw Fixation Systemis divided into four types: large, small, mini, and reconstruction. Thelength of the plate varies from 45 to 316mm, and hole number from 3 to22 holes.Geasure Internal Locking Bone Screw System is used either to fastenplates or similar devices onto bones or to hold bone fragments together.The screws are differentiated by the manner in which they are insertedinto bone, their function, their size, and the type of bone they areintended for. There are two styles: locking screw, cortex or cancellousscrew. The screw recess is hexagon or star shaped to allow screwremoval and insertion. The thread diameter of screw varies from 2.0 to6.5mm, total length from 6 to 120mm.Internal Locking Plate and Screw Fixation System is a construct inwhich the screws are locked in the plate. The heads of the lockingscrews contain male threads while the holes in the plates contain femalethread, this allows the screw head to be threaded into the plate. Variousplate designs are available depending on specific anatomic locations,these may be larger or smaller, thicker or thinner as appropriate tovarious anatomic sites and the loads to which they will be subjected.The holes in the plate are designed for locking screws, or locking andnon-locking screws combination to facilitate dynamic compression. Bycombining locking screw holes with compression screw slots in theshaft, the plate can be used as both a locking device and a fracturecompression device. Internal Locking plate and screw Fixation systemsdo not disrupt the underlying cortical bone perfusion as much asconventional plates, which compress the plate to the cortical bone. | |
| Materials | Titanium alloy (Ti-6Al-4V) ELI | |
| Patient Contact | Bone and surrounding tissue | |
| Contact Duration | Permanent (>30 days) | |
| Sterilization Method | The devices are provided non-sterile; validated manual cleaning andsteam sterilization instructions are provided for the end user beforeimplantation. | |
| Environment of Use | Healthcare facility/Hospital | |
| Single Use | Yes | |
| Summary ofintended use andtechnologicalcharacteristics | The Internal Locking plate and Screw Fixation System is substantiallyequivalent to the predicate device when evaluating intended use andtechnological characteristics.The subject device has the identical intended use as the predicate devicewith only minor differences for the 3.5mm Round Hole Reconstruction | |
| Locking Plate indication, which is included in the predicate device'sindication scope. | ||
| The subject device and predicate devices are substantially equivalentwith only minor differences in technological characteristics regarding:• Sterilization method• device material• Screw size | ||
| These differences do not raise new questions of safety and effectiveness. | ||
| • Performance-bench test including:Single Cycle Bend Testing for bone plate per ASTM F382-17Fatigue Property Testing for bone plate per ASTM F382-17Torsional Properties Test for the screws per ASTM F543-17Driving Torque Test for the screws per ASTM F543-17Axial Pullout Strength Test for the screws per ASTM F543-17Self-Tapping Performance Test for the screws per ASTM F543-17 | ||
| •Reprocessing and sterilizationGensure completed manual cleaning and steam sterilization validationtesting per the following guidance and standard document:FDA guidance document, Reprocessing Medical Devices in HealthCare Settings: Validation Methods and Labeling, issued March 17,2015.AAMI TIR12: 2010 Designing, testing, and labeling reusable medicaldevices for reprocessing in health care facilities: A guide for medicaldevice manufacturers.AAMI TIR30: 2011/(R)2016 A compendium of processes, materials,test methods, and acceptance criteria for cleaning reusable medicaldevices.ISO 17665- 1:2006 Sterilization of health care products - Moist heat -Requirements for the development, validation, and routine control ofsterilization process for medical devices.ISO 17665-2:2009 Sterilization of health care products - Moist heat -Part 2: Guidance of the application of ISO 17665-1.ISO 19227:2018 Implants for surgery - Cleanliness of orthopedicimplants - General requirements. | ||
| Non-clinical test | ||
| Performance -Animal | No animal study data is submitted in this 510(k). | |
| Performance -Clinical | No clinical study data is submitted in this 510(k). | |
| SubstantialEquivalence | The Internal Locking Plate and Screw Fixation System is substantiallyequivalent to the predicate device when evaluating intended use andtechnological characteristics.The subject device has the identical intended use as the predicatedevice. The subject device and predicate devices are substantially | |
| equivalent with only minor differences. These differences do not raisenew questions of safety and effectiveness. | ||
| The non-clinical data demonstrates the Internal Locking Plate and | ||
| Conclusion | Screw Fixation System is substantially equivalent to the predicate | |
| device. |
{5}------------------------------------------------
{6}------------------------------------------------
dislocations, osteotomies, and non-unions of the proximal humerus,
particularly in osteopenic bone.
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.