(134 days)
The SCENARIA View 4.2 system is indicated to acquire axial volumes of the whole body including the head. Images can be acquired in axial, helical, or dynamic modes. The SCENARIA View system can also be used for interventional needle quidance. Volume datasets acquired by a SCENARIA View system can be post-processed in the SCENARIA View system to provide additional information. Post-processing capabilities of the SCENARIA View software include, multi-planar reconstruction (MPR), and volume rendering. Volume datasets acquired by a SCENARIA View system can be transferred to external devices via a DICOM standard interface. The Low Dose CT Lung Cancer Screening Option for the SCENARIA View system is indicated for using low dose CT for lung cancer screening. The screening must be conducted with the established program criteria and protocols that have been approved and published by a governmental body, a professional medical society, and/or FUJIFILM Healthcare Corporation.
The SCENARIA View 4.2 is a multi-slice computed tomography system that uses x-ray data to produce cross-sectional images of the body at various angles. The SCENARIA View 4.20system uses 128-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator, X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles. The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system. The SCENARIA View 4.2 system consists of a Gantry, Operator's Workstation, Patient Table, High- Frequency X-ray Generator, and accessories.
Here's a breakdown of the acceptance criteria and study information for the Scenaria View 4.2 CT system, based on the provided text:
Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. This type of document often relies on comparative analysis and verification that new features do not negatively impact existing performance, rather than extensive clinical efficacy studies for the entire device. For a CT Scanner, the primary acceptance criteria revolve around image quality, dose, and safety, often demonstrated through bench testing and compliance with recognized standards.
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Image Quality & Dose | Dose Profile | Evaluated, no change from the predicate device. Compliance with applicable requirements. |
| Image Noise | Evaluated, no change from the predicate device. Compliance with applicable requirements for Noise, Mean CT number and Uniformity. | |
| Modulation Transfer Function (MTF) / Spatial Resolution | Evaluated, no change from the predicate device. Compliance with applicable requirements for Spatial Resolution. | |
| Slice Thickness and Sensitivity Profile | Evaluated, no change from the predicate device. Compliance with applicable requirements for Tomographic Section Thickness and Sensitivity Profile. | |
| Slice Plane Location / Tomographic Plane Location | Evaluated, no change from the predicate device. Compliance with applicable requirements. | |
| CT Dose Index | Evaluated, no change from the predicate device. Compliance with applicable requirements. | |
| System Functionality | AutoPositioning Function | Verified working correctly through system testing. Various cases were tested to verify it shows up as an option. All test results performed as expected (installed, version info, enable, prohibit when not installed). |
| ExamSplit Function | Verified and validated to work as expected. A mock procedure confirmed the following: - Automatic analysis queue for main recon of original inspection registered and executed. - Automatic analysis queue for original inspection multi-recon registered and executed after multi-recon. - Automatic analysis queue for PES registered and executed after Exam Split. - Exam Split starts after scan, Multi Recon starts after recon completion. All testing passed. | |
| Safety & Standards | Compliance with: - AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012. C1:2009/(R)2012 and . A2:2010/(R)2012 (Medical electrical equipment - General requirements for basic safety and essential performance) | Conformance demonstrated. |
| - IEC 60601-1-2 Edition 4.0 (Electromagnetic compatibility) | Conformance demonstrated. | |
| - IEC 60601-1-3 Edition 2.1 (Radiation protection in diagnostic X-ray equipment) | Conformance demonstrated. | |
| - IEC 60601-2-44 Edition 3.2 (Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography) | Conformance demonstrated. | |
| - IEC 62304 Edition 1.1 (Medical device software - Software life cycle processes) | Conformance demonstrated. | |
| - NEMA XR 25 (Computed Tomography Dose Check) | Conformance demonstrated. | |
| Technological Equivalence | Features (Tilt for helical scan, improved Exam Split, AutoPositioning) do not impact safety and effectiveness compared to the predicate device, and the subject device is substantially equivalent to the predicate device (K213829). | Thorough analysis and comparison conducted. Technological characteristics (e.g., SSD for image storage, new features) do not impact safety and effectiveness. |
Study Details
-
Sample size used for the test set and the data provenance:
- The document does not specify a numerical "sample size" in terms of patient cases for testing.
- The performance evaluations for image quality and dose (Dose Profile, Noise, MTF, Slice Thickness, CT dose index, etc.) are described as "Performance Testing - Bench." This typically involves phantom studies or specified calibration procedures rather than patient data.
- For the new software features (AutoPositioning, ExamSplit), testing involved "various cases" and "mock procedure," which are likely internal verification and validation tests rather than tests on a diverse patient dataset.
- Data Provenance: Not explicitly stated as retrospective or prospective patient data, but the descriptions strongly suggest bench testing with phantoms/test objects or internal system validation using mock scenarios/simulated data. No country of origin for patient data is mentioned as no clinical studies with patient data were detailed for the new features.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the summary. For performance testing on a CT scanner, ground truth is usually established through physical measurements, calibration standards, and technical specifications, rather than expert interpretation of images for diagnosis. For the software features, the "ground truth" was whether the feature performed as designed based on the functional requirements.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- None is explicitly mentioned. The testing described (bench tests, functional verification) does not typically involve human adjudication in the way clinical studies for diagnostic accuracy would.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done or described. The device is a CT imaging system, and while it has a "Low Dose CT Lung Cancer Screening Option," the submission discusses the system's performance and new software features like AutoPositioning and ExamSplit, not a specific AI-assisted diagnostic tool. There is no mention of AI assistance for human readers or any effect size related to reader improvement.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The document describes the performance of the CT system and its software features. The "Low Dose CT Lung Cancer Screening Option" is an indication of use for the system, not a separate standalone algorithm whose performance is being evaluated independently. The new features (AutoPositioning, ExamSplit) are functionalities of the CT system itself, not standalone algorithms for diagnosis. The performance testing is focused on the system's technical specifications and the correct functioning of these features. Therefore, this question is not directly applicable in the context of the information provided for this CT device's 510(k).
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For image quality and dose metrics (Dose Profile, Noise, MTF, etc.), the ground truth is established by physical measurements, standardized phantoms, and adherence to technical specifications and industry standards.
- For the new software features (AutoPositioning, ExamSplit), the ground truth was the pre-defined functional requirements and expected system behavior.
-
The sample size for the training set:
- This information is not provided as the document describes a CT imaging system and its software features, not a machine learning or AI model that requires a dedicated training set for diagnostic purposes described in the output. The software features mentioned (AutoPositioning, ExamSplit) are likely rule-based or algorithmic improvements, not deep learning models.
-
How the ground truth for the training set was established:
- This information is not provided and is not applicable given the absence of a stated training set for a machine learning or AI algorithm.
{0}------------------------------------------------
October 12, 2023
Fujifilm Healthcare Americas Corporation % Chaitrali Kulkarni Sr. Regulatory Affairs Specialist 81 Hartwell Ave. Suite 300 LEXINGTON, MA 02421
Re: K231574
Trade/Device Name: Scenaria View 4.2 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: August 29, 2023 Received: August 29, 2023
Dear Chaitrali Kulkarni:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
Submission Number (if known)
Device Name
Scenaria View 4.2
Indications for Use (Describe)
The SCENARIA View 4.2 system is indicated to acquire axial volumes of the whole body including the head. Images can be acquired in axial, helical, or dynamic modes. The SCENARIA View system can also be used for interventional needle quidance. Volume datasets acquired by a SCENARIA View system can be post-processed in the SCENARIA View system to provide additional information. Post-processing capabilities of the SCENARIA View software include, multi-planar reconstruction
(MPR), and volume rendering. Volume datasets acquired by a SCENARIA View system can be transferred to external devices via a DICOM standard interface. The Low Dose CT Lung Cancer Screening Option for the SCENARIA View system is indicated for using low dose CT for lung cancer screening. The screening must be conducted with the established program criteria and protocols that have been approved and published by a governmental body, a professional medical society, and/or FUJIFILM Healthcare Corporation.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Value from Innovation
510(k) Sum
Submitter Information
| Submitter: | FUJIFILM Healthcare Corporation2-1, Shintoyofuta,Kashiwa-Shi, Chiba, JP 277-0804 |
|---|---|
| Contact: | Chaitrali Kulkarni, Senior Regulatory Affairs Specialist |
| Telephone number: | (704)-517-4886 |
| E-mail: | chaitrali.kulkarni@fujifilm.com |
| Date: | July 25, 2023 |
Subject Device Name
| Trade/Proprietary Name: | SCENARIA View 4.2 |
|---|---|
| Regulation Number: | 21 CFR 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Product Code | JAK, System, X-Ray, Tomography, Computed |
| Class | II |
| Panel | Radiology |
Predicate Device Name
| Predicate Device(s): | SCENARIA View 4.0 (K213829) |
|---|---|
| Regulation Number: | 21 CFR 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Product Code | JAK, System, X-Ray, Tomography, Computed |
| Class | II |
| Panel | Radiology |
Indications for Use
The SCENARIA View 4.2 system is indicated to acquire axial volumes of the whole body including the head. Images can be acquired in axial, helical, or dynamic modes. The SCENARIA View system can also be used for interventional needle guidance. Volume datasets acquired by a SCENARIA View system can be post-processed in the SCENARIA View system to provide additional information. Post-processing capabilities of the SCENARIA View software include, multi-planar reconstruction (MPR), and volume rendering. Volume datasets acquired by a SCENARIA View system can be transferred to external devices via a DICOM standard interface. The Low Dose CT Lung Cancer Screening Option for the SCENARIA View system is indicated for using low dose CT for lung cancer screening. The screening must be conducted with the established program criteria and protocols that have been approved and published by a governmental body, a professional medical society, and/or FUJIFILM Healthcare Corporation.
Device Description
Function
The SCENARIA View 4.2 is a multi-slice computed tomography system that uses x-ray data to produce cross-sectional images of the body at various angles.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Fujifilm logo. The logo is in black, except for a small red square above the "I" in Fujifilm. The logo is simple and modern.
Value from Innovation
Scientific Concepts
The SCENARIA View 4.20system uses 128-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator, X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.
The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.
Physical and Performance Characteristics
The SCENARIA View 4.2 system consists of a Gantry, Operator's Workstation, Patient Table, High- Frequency X-ray Generator, and accessories. The system performance is similar to the predicate device.
Performance Comparison
Evaluations were conducted for dose profile, image noise, Modulation Transfer Function (MTF), slice thickness and sensitivity profile, slice plane location, and CT dose index. The feature of having tilt for helical scan is a combination of features that are included in the predicate device and do not affect the diagnosis. Prospective ExamSplit is a modification of the image information and also does not affect the diagnosis. The AutoPositioning feature supports the operator's work prior to the CT examination, and does not affect the diagnosis.
The AutoPositioning function was verified to be working correctly through system testing. To ensure the AutoPositioning option was working correctly, various cases were tested to verify if the AutoPositioning function will show up as an option. The following case testing was performed:
- . Install the software as an application added to the existing system that does not have AutoPositioning installed.
- . Indicate the name of this software and the version number as the version information in an environment where AutoPositioning option is installed.
- Enable AutoPositioning function in an environment where AutoPositioning is already installed.
- . Prohibit the AutoPositioning function when this software is not installed in an environment where the AutoPositioning option is not installed.
All test results performed as expected.
Verification and validation testing was performed on ExamSplit function as well to ensure it works as expected. A mock procedure was carried out during which ExamSplit function was turned ON and the scan was carried out as listed in the operation manual. The following criteria had to meet in order for the ExamSplit testing to pass:
- After scan, an automatic analysis queue for the main recon of the original inspection is . registered and executed.
- . After the multi-recon is completed, the automatic analysis queue for the original inspection multi-recon is registered and executed.
- . After Exam Split execution is completed, an automatic analysis queue for PES is registered and executed.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square above the letter 'I'. The text 'FUJIFILM' is written in a bold, sans-serif font.
Value from Innovation
- Exam Split starts running after scan, and Multi Recon starts running Exam Split after recon . is completed
All testing passed the above requirements.
The evaluation results confirm the performance characteristics of the SCENARIA View 4.2 are comparable to the predicate device and support our conclusion that the subject system is substantially equivalent.
Device Technological Characteristics
A summary of the differences is listed in the following table.
| Systems | SCENARIA View | SCENARIA View (K213829) |
|---|---|---|
| Subject Device | Predicate Device | |
| Physical characteristics | Image: SCENARIA View Subject Device | Image: SCENARIA View (K213829) Predicate Device |
| Gantry | There are no differences between the two systems. | There are no differences between the two systems. |
| Detector | There are no differences between the two systems. | There are no differences between the two systems. |
| X-ray Tube | There are no differences between the two systems. | There are no differences between the two systems. |
| X-ray Generator | There are no differences between the two systems. | There are no differences between the two systems. |
| Patient Table | There are no differences between the two systems. | There are no differences between the two systems. |
| Display | There are no differences between the two systems. | There are no differences between the two systems. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Fujifilm logo. The logo is in black text, with a red square above the "I" in Fujifilm. Below the logo is the text "Value from Innovation".
Value from Innovation
| Systems | SCENARIA ViewSubject Device | SCENARIA View (K200498)Predicate Device |
|---|---|---|
| Image Storage | Storage type changed from hard disk to SSD. Size of raw data storage is reduced, max of raw data is not changed | |
| Scanning, Reconstruction | There are no differences between the two systems. | |
| Performance | There are no differences between the two systems. | |
| Dose Controls | There are no differences between the two systems. | |
| Dose Displays | There are no differences between the two systems. | |
| Features | The option to add tilt for helical scan was added. Exam split was improved by adding the ability to split time before and after the scan. New feature of AutoPositioning was added to the subject device. |
Therefore, based on a thorough analysis and comparison of subject device (SCENARIA View 4.2) and the predicate device, the technological characteristics do not impact safety and effectiveness.
Substantial Equivalence
A summary decision was based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics.
| ITEM | OVERALL RATIONALE ANALYSIS |
|---|---|
| Gantry | No changes |
| Detector | No changes |
| X-ray Tube | No changes |
| X-ray Generator | No changes |
| Patient Table | No changes |
| Display | No changes |
| Image Storage | Storage type changed from hard disk to SSD. However, max number of RawData (6000 scan) is not changed. |
| Scanning,Reconstruction | No changes |
| Performance | No changes |
| Dose Controls | No changes |
| Dose Displays | No changes |
| Features | The option to add tilt for helical scan was added. Exam split was improved byadding the ability to split time before and after the scan. New feature ofAutoPositioning was added to the subject device. |
Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed SCENARIA View 4.2 is considered substantially equivalent to the currently marketed predicate device (SCENARIA View 4.0 (K213829)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red square in the middle of the "I". The text "FUJIFILM" is in a bold, sans-serif font. The logo is simple and modern.
Value from Innovation
Summary of Non-Clinical Testing
This device complies with all applicable requirements for Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index.
In addition, the SCENARIA View system is in conformance with the applicable parts of the following standards:
-
AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012. C1:2009/(R)2012 and . A2:2010/(R)2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) -
. IEC 60601-1-2 Edition 4.0
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests -
IEC 60601-1-3 Edition 2.1 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
-
. IEC 60601-2-44 Edition 3.2
Medical electrical equipment Part 2-44: particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography.
- . IEC 62304 Edition 1.1 Medical device software - Software life cycle processes
- NEMA XR 25 Computed Tomography Dose Check .
Summary of Performance Testing
| Testing Type | Rationale Analysis |
|---|---|
| Performance Testing - Bench | There is no change is the following performance issues:Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, TomographicSection Thickness and Sensitivity Profile, Tomographic Plane Location, CT Dose Index. |
| Performance Testing - Clinical | N/A |
Conclusions
FUJIFILM believes that, based on the information included in the submission, SCENARIA View 4.2 subject device is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the SCENARIA View 4.0 (K213829).
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.