K Number
K231408
Date Cleared
2023-08-11

(88 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

Device Description

The subject device is a patient examination glove made of synthetic nitrile latex compound. It is Non-Sterile, Powder free and is Blue in color. The device is ambidextrous and can be worn on either the left or right hand. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per ISO 10993-11.

AI/ML Overview

The provided text describes Shandong Intco Medical Products Co, Ltd.'s Basic Synguard Nitrile Exam Gloves. This is a medical device submission, and the acceptance criteria and study described relate to the physical and biocompatibility performance of these gloves.

Here's the breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard)Reported Device Performance
DimensionsASTM D6319-19
Length230 mm minPass (240 mm min)
WidthXS: 70±10, S: 80±10, M: 95±10, L: 110±10, XL: 120±10, XXL: 130±10Pass (XS: avg 78.3mm, S: avg 86.5mm, M: avg 97.6mm, L: avg 108.7mm, XL: avg 115.4mm, XXL: avg 123.7mm)
ThicknessPalm: Minimum 0.05 mm, Finger: Minimum 0.05 mmPass (Palm 0.06mm min, Finger 0.082mm min)
Physical PropertiesASTM D6319-19
Before Aging
Tensile Strength14 MPa, minPass (22.6 MPa, min)
Elongation500%, minPass (531%, min)
After Aging
Tensile Strength14 MPa, minPass (21.9 MPa, min)
Elongation400%, minPass (437%, min)
Freedom from holesASTM D5151-19 (No leakage at sampling level of G-1, AQL 2.5)Pass (No leakage, 312 of 315 gloves passed)
Residual PowderASTM D6124-06 (Reapproved 2017) (< 2mg per glove)Pass (average 0.17 mg per glove)
Biocompatibility
Skin SensitizationISO 10993-10 (Not a sensitizer)Under the conditions of the study, not a sensitizer
Intracutaneous ReactivityISO 10993-10 (Not an irritant)Under the condition of study, not an irritant
Acute Systemic ToxicityISO 10993-11 (No evidence of systemic toxicity)Under the conditions of this study, there was no evidence of systemic toxicity. (also stated as "the device extracts do not pose a systemic toxicity" in the predicate comparison table)

2. Sample size used for the test set and the data provenance

  • Freedom from holes: 315 gloves were tested.
  • Other physical properties (Dimensions, Physical Properties, Residual Powder): The document does not explicitly state the sample size for each test, but it references ASTM D6319-19, ASTM D6124-06, and ASTM D5151-19, which would define the sampling plans. For "Freedom from holes," the sample size of 315 and AQL 2.5 is typical for ASTM D5151.
  • Biocompatibility: The sample sizes for the ISO 10993 tests are not explicitly stated in this document but are defined by the respective ISO standards.
  • Data Provenance: The studies were conducted by Shandong Intco Medical Products Co., Ltd. (China) for their device. The nature of these tests (physical and biocompatibility testing on manufactured products) categorizes them as prospective performance evaluations of the device itself.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the device is a medical glove, and the "ground truth" is established through standardized laboratory testing against objective criteria (e.g., tensile strength, dimensions, presence of holes, chemical residues, biological reactions in test models) as defined by ASTM and ISO standards, not expert interpretation of cases.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable as the device is a medical glove, and the performance is measured directly by physical and chemical tests, not through human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device is a medical glove, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is not applicable. The device is a medical glove, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the device's performance is established by:

  • Pre-defined physical and chemical specifications from ASTM standards (e.g., minimum tensile strength, maximum residual powder, dimensional tolerances).
  • Biological responses (e.g., no irritation, no sensitization, no systemic toxicity) in animal or in vitro test models, as defined by ISO 10993 standards.

8. The sample size for the training set

This section is not applicable. The device is a medical glove, and this is not a machine learning model requiring a training set.

9. How the ground truth for the training set was established

This section is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 11, 2023

Shandong Intco Medical Products Co, Ltd. % Deze Wang Official Correspondent Intco Medical Industries, Inc 805 Barrington Ave. Ontario, California 91764

Re: K231408

Trade/Device Name: Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 11, 2023 Received: May 15, 2023

Dear Deze Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

BiFeng Qian, M.D., Ph.D Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231408

Device Name

Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for INTCO Medical. The logo features the company name in blue, with the letters "INTCO" in a bold, sans-serif font. Below the company name are the Chinese characters "英科 医疗", also in blue. Above the company name is a stylized blue graphic that resembles a flower or sunburst, with several petal-like shapes radiating from a central point.

510(K) Summary

(As requirement by 21 CFR 807.92)

K231408

Date : A. Applicant: Name: Shandong Intco Medical Products Co., Ltd. Address: Qiwang Road, Naoshan Industrial Park, Qingzhou, Shandong, China Contact: Ms. Min Gao Title: QA Tel: +86-536-6136888 Email: gaomin@intco.com

Submission Correspondent: Primary contact: Max Li Shandong Intco Medical Products Co., Ltd. Qiwang Road, Naoshan Industrial Park, Qingzhou, Shandong, China Tel: +86-536-6136888 Email: maxli@intco.com Secondary contact: Min Gao Qiwang Road, Naoshan Industrial Park, Qingzhou, Shandong, China Tel: +86-536-6136888 Email: gaomin@intco.com

B. Device:

Trade Name: Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile. Common Name: Non-Powdered Patient Examination Glove

Regulatory Information Classification Name: Non-Powdered Patient Examination Glove Classification: Class I Product code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital and Personal Use

C. Predicate device:

K221169 Name of the device: Nitrile Examination Gloves Powder Free-Black Owner's Name: JR Engineering & Medical Technologies (M) Sdn. Bhd. Classification: Class I Product code: LZA Regulation Number: 21 CFR 880.6250

{4}------------------------------------------------

D. Indications for use of the device:

The Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiners' hand or fingers to prevent contamination between patient and examiner.

E. Device Description:

The subject device is a patient examination glove made of synthetic nitrile latex compound. It is Non-Sterile, Powder free and is Blue in color. The device is ambidextrous and can be worn on either the left or right hand. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per ISO 10993-11.

F. Summary of Technological Characteristics

The Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile are compared with the predicate device in terms of intended use, design, material, specifications, and performance. The following table shows similarities and differences of use, design, and material between the proposed device and the predicate device.

DeviceProposed DevicePredicate DeviceResult
510K #K231408K221169-
Product NameBasic Synguard Nitrile Exam Gloves,Powder Free, Blue Colored, Non-SterileJR MEDIC Nitrile Examination GlovesPowder Free- Black-
Manufacturer(s)Shandong Intco Medical Products Co.,Ltd.JR Engineering & MedicalTechnologies (M) SDN.BHD.-
Product CodeLZALZASame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
Indications foruseThe Basic Synguard Nitrile ExamGloves, Powder Free, Blue Colored,Non-Sterile is a disposable deviceintended for medical purposes that isworn on the examiners' hand or fingerto prevent contamination betweenpatient and examiner.Nitrile Examination Gloves Powderfree is a disposable device intendedfor medical purpose that is worn onthe examiner's hand to preventcontamination between patient andexaminer.Same
Powder freeYesYesSame
Design featureAmbidextrous.Ambidextrous.Same
MaterialNitrileNitrileSame
ColorBlueBlackDifferent
SizeXS, S, M, L, XL, XXLXS, S, M, L, XLSimilar
SterileNon-SterileNon-SterileSame
UseSinge useSingle useSame
CHARACTERSTICSSTANDARDSDEVICE PERFORMANCERemarks
SUBJECTPREDICATE
510(k) NumberK231408K221169----
Name of deviceBasic Synguard NitrileExam Gloves, PowderFree, Blue Colored, Non-SterileNitrile ExaminationGloves Powder Free-Black----
DimensionsASTMD6319- 19Length Min 230 mWidth Min 95+/- 10 mm( for medium size)Length Min 230 mmWidth Min 95+/- 10mm(for medium size)Same
Physical PropertiesASTMD6319- 19Before AgingTensile Strengthmin 14 MPaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MPaUltimate ElongationMin 400%Before AgingTensile Strengthmin 14 MPaUltimate ElongationMin 500%After AgingTensile Strengthmin 14 MPaUltimate ElongationMin 400%Same
ThicknessASTMD6319- 19Palm min 0.05 mmFinger min 0.05 mmPalm min 0.05 mmFinger min 0.05 mmSame
Powder FreeASTMD6319- 19≤2 mg/glove≤2 mg/gloveSimilar
Water leakASTM D5151-193/315 AQL2.5Meets ASTMD6319 requirementsSimilar
BiocompatibilityPrimary SkinIrritation-ISO10993-10:2010DermalSensitization- ISO10993- 10:2010Under the condition ofstudy not an irritantUnder the conditions ofthe study not a sensitizerUnder the conditionof study not anirritantUnder the conditions ofthe study not a sensitizerSame
In vitro cytotoxicityISO10993-5 :2009Data Not availableUnder the conditionsof the study, cytotoxic.--------
Acute SystemicToxicity Test ISO10993- 11:2017Under the conditions of thisstudy, there was no evidenceof systemic toxicity.Under the condition ofstudy the device extractsdo not pose a systemictoxicity.Same

Table 1 General Comparison of Proposed and Predicate Devices

{5}------------------------------------------------

Analysis: The proposed device has product size XS, S, M, L, XL, XXL . while the predicate device has product size XS, S, M, L, XL. The proposed device is Blue color, while the predicate device is black color. But safety and performance testing has been done to the proposed device and the results showed that the device meets the requirements of standard ASTM D6319-19. Therefore, this difference do not raise any new safety or performance questions.

Table 2 Comparison of Physical, Biocompatibility and Performance Testing

{6}------------------------------------------------

Analysis: The proposed device has the Skin Sensitization test, Primary Skin Irritation test and the Acute Systemic Toxicity test. The result is the same as the PREDICATE device's.

G. Summary of Non-Clinical Testing

A Biocompatibility

The following tests for the subject device were conducted to evaluate the biocompatibility of Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile:

  • ISO 10993-10: Skin Sensitization test
  • ISO 10993-10: Intracutaneous Reactivity test
  • ISO 10993-11: Acute Systemic Toxicity test

A Performance Testing

Physical performance testing of the proposed device was conducted as per ASTM D6319-19 Standard Specification for Nitrile Examination Gloves.

To summarize, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

• ASTM D6319-19 Standard Specification for Nitrile Examination Gloves.

  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
  • · ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
Test MethodPurposeAcceptance CriteriaResults
Dimensions (length)(width)(thickness)The purpose of thetest is to evaluatethe physicaldimension of thegloveLength230 mm minPass240 mm min
Width(mm)Pass
XS: 70±10XS: average 78.3mm
S: 80±10S: average 86.5mm
M: 95±10M: average 97.6mm
L: 110±10L: average 108.7mm
XL: 120±10XL: average 115.4mm
XXL: 130±10XXL: average 123.7mm
Thickness(mm):Palm: Minimum 0.05PassPalm 0.06mm min.

{7}------------------------------------------------

Finger: Minimum 0.05Finger - 0.082mm min
Physical propertiesThe purpose of thetest is to evaluatethe tensile strengthand ultimateelongation beforeand after agingBefore Aging:Tensile Strength: 14 MPa, min.Elongation: 500%, min.After Aging:Tensile Strength: 14 MPa, min.Elongation: 400%, min.PassBefore Aging:Tensile Strength: 22.6MPa, min.Elongation: 531%, min.After Aging:Tensile Strength: 21.9MPa, min.Elongation: 437%, min.
Freedom from holesThe purpose of thetest is to detectholes in the glovesNo leakage at sampling level ofG-1, AQL 2.5PassNo leakage, 312 of 315 passed
Residual PowderThe purpose of thetest is to detect thepowder residue inthe glove<2mg per glovePassaverage 0.17 mg per glove

H. Clinical Test Conclusion

No clinical study is included in this submission.

l. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Basic Synguard Nitrile Exam Gloves, Powder Free, Blue Colored, Non-Sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K221169.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.