K Number
K231394
Date Cleared
2023-08-09

(86 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intense Pulsed Light System is intended for medical use in the following conditions:

  1. Moderate inflammatory acne vulgaris;

  2. Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  3. Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations;

  4. Permanent hair reduction-long-term stable reduction in number of hairs re-growing after a treatment regimen.

Device Description

Intense Pulsed light (IPL) System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. There are six optical filters that can be using in this system listed in table. With these special properties, the IPL System has a wide application in non-ablative therapies based on theory of human skin tissue's selective absorption.

AI/ML Overview

The provided document is a 510(k) premarket notification for an Intense Pulsed Light System. It details the device's indications for use, technological characteristics, and comparison to a predicate device. However, the document does not contain information on acceptance criteria for a study demonstrating device performance against specific metrics, nor does it describe a study involving human subjects or AI performance.

The "Performance Data" section of the document primarily addresses:

  • Electrical safety and electromagnetic compatibility: Compliance with standards like IEC 60601-1, IEC 60601-1-2, IEC 62471, and IEC 60601-2-57.
  • Software Verification and Validation Testing: Stating that V&V testing was conducted, and the software was considered "moderate" level of concern.
  • Sterilization and Shelf-Life: Asserting the device is not sterile, is reusable, and has no restricted shelf-life.
  • Biocompatibility: Reporting results of cytotoxicity, sensitization, and irritation tests on the handpiece sapphire tip, concluding compliance with ISO 10993-1.
  • Bench testing: Verifying that the spatial variation of the light source equipment output over the treatment area does not deviate by more than ±20%, and fulfilling IEC 60601-2-57 requirements.

Therefore, I cannot fulfill your request for: a table of acceptance criteria and reported device performance (in the context of clinical or AI performance), sample size for test sets, data provenance, number and qualifications of experts, adjudication methods, MRMC study details, standalone AI performance, type of ground truth, training set sample size, or how training ground truth was established.

The document primarily focuses on demonstrating substantial equivalence to a predicate device based on similar technological characteristics and compliance with safety and bench testing standards, rather than proving efficacy or performance through clinical endpoint studies or AI-driven assessments.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

August 9, 2023

Smedtrum Medical Technology Co., Ltd. Crimson Wu Senior Regulatory Engineer 1F., No. 8, Ln. 97, Wugong Rd., New Taipei City, Xinzhuang Dist. 248016 Taiwan

Re: K231394

Trade/Device Name: Intense Pulsed Light System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF Dated: May 12, 2023 Received: May 15, 2023

Dear Crimson Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231394

Device Name Intense Pulsed Light System

The Intense Pulsed Light System is intended for medical use in the following conditions:

  1. Moderate inflammatory acne vulgaris;

  2. Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  3. Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations;

  4. Permanent hair reduction-long-term stable reduction in number of hairs re-growing after a treatment regimen.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo for a company called "Sme". The logo consists of a stylized letter "S" made up of blue triangles, along with the text "Sme" in a gray sans-serif font. Below the text "Sme" is the text "MEDICAL TECH" in a smaller font size.

Smedtrum Medical Technology Co., Ltd. 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist., New Taipei City 248016 , Taiwan TEL:+886 (02) 2298 9578 FAX:+886 (02) 2298 9426

Section 5 : 510(k) Summary Intense Pulsed Light System

SUBMITTER I.

Smedtrum Medical Technology Co., Ltd. 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist., New Taipei City 248016, Taiwan (R.O.C.) Contact Person

Crimson Wu Position: Senior Regulatory Engineer Tel: +886-2-2298-9578, Ext. 301 Fax: +886-2-2298-9426 E-mail: crimsonwu@smedtrum.com Date of preparation: May 12th, 2023

II. DEVICE

Trade Name:Intense Pulsed Light System
Common or Usual Name:Intense Pulsed Light System
Product code:ONF
Classification Name:Powered Light Based Non-Light Surgical Instrument With Thermal Effect

21 C.F.R. § 878.4810, Device Class II

PREDICATE DEVICE III.

ManufacturerShanghai Apolo Medical Technology Co., Ltd.
Trade Name:IPL Treatment Systems
Common or Usual Name:Laser surgical instrument for use in general and plasticsurgery and in dermatology
Classification Name:Powered Light Based Non-Light Surgical InstrumentWith Thermal Effect
21 C.F.R. § 878.4810, Device Class II
Premarket Notification:K200746 May 15th, 2020

IV. DEVICE DESCRIPTION

Intense Pulsed light (IPL) System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. There are six optical filters that can be using in this system listed in table. With these special properties, the IPL System has a wide application in non-ablative therapies based on theory of human

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains a logo for Smedtrum Medical Technology Co., Ltd. The logo features a stylized "S" shape formed by overlapping blue and teal triangles, with a gray shadow extending from the left side. To the right of the symbol is the company name, "Smedtrum," in a light gray sans-serif font, with the words "MEDICAL TECHNOLOGY CO., LTD." in smaller letters below. Above the company name is the company name in Chinese.

Smedtrum Medical Technology Co., Ltd. 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist., New Taipei City 248016 , Taiwan TEL:+886 (02) 2298 9578 FAX:+886 (02) 2298 9426

skin tissue's selective absorption.

SpectraApplication areasExpected Working Life
420-1200nmVascular, acne, pigmentation100,000 shots
510-1200nmVascular, pigmentation80,000 shots
560-1200nmVascular, pigmentation70,000 shots
610-1200nmHair removal60,000 shots
640-1200nmHair removal50,000 shots
610-980nm (SHR)Hair removal30,000 shots

V. INDICATIONS FOR USE

The Intense Pulsed Light System is intended for medical use in the treatment of the following conditions:

    1. Moderate inflammatory acne vulgaris;
  1. Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  2. Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations;

  3. Permanent hair reduction-long-term stable reduction in number of hairs re-growing after a treatment regimen.

COMPARISON OF TECHNOLOGICAL CHARATERISTICS WITH THE
PREDICATE DEVICE
FeatureProposed devicePredicate device(K200746)
Device NameIntense Pulsed Light SystemIPL Treatment Systems
Product CodeONFONF
Regulation No.21 CFR 878.481021 CFR 878.4810
Device ClassClass IIClass II
FeatureProposed devicePredicate device(K200746)
Indication forUseThe Intense Pulsed LightSystem is intended for medicaluse in the treatment of thefollowing conditions:1. Moderate inflammatory acnevulgaris;2. Benign pigmented epidermallesions including dyschromia,hyperpigmentation, melasma,ephelides (freckles);3. Benign cutaneous vascularlesions including port winestains, hemangiomas, facialtruncal and leg telangiectasias,erythema of rosacea, leg veins,spider angiomas and venousmalformations;4. Permanent hair reduction-long-term stable reduction innumber of hairs re-growingafter a treatment regimen.The IPL treatment systems isintended for medical use in thetreatment of the followingdermatologic conditions:- Permanent hair reduction- long-term stable reduction in numberof hairs re-growing after atreatment regimen;- Moderate inflammatory acnevulgaris;- Benign pigmented epidermallesions including dyschromia,hyperpigmentation, melasma,ephelides (freckles);- Cutaneous lesions includingscars;- Benign cutaneous vascular lesionsincluding port wine stains,hemangiomas, facial truncal andleg telangiectasias, erythema ofrosacea, leg veins, spiderangiomas and venousmalformations.
Light SourceIntense Pulsed light (XenonFlash Lamp)Intense Pulsed light (Xenon FlashLamp)
Wavelength420 – 1200 nm420 – 1200 nm
Light DeliverySystemHandpieceHandpiece
Filter420 -1200nm: Acne;510 -1200nm: Acne, vascular,pigment;560 -1200nm: Vascular,pigment;610-1200nm: Hair removal;640-1200nm: Hair removal;420 -1200nm: Acne;510 -1200nm: Acne, vascular,pigment;560 -1200nm: Acne, vascular,pigment;610-1200nm: Hair removal;640-1200nm: Hair removal;690-1200nm: Hair removal:
FeatureProposed devicePredicate device(K200746)
Fluence420 -1200nm: 4.1-34.8 J/cm2;510 -1200nm: 3.8-31.6 J/cm2;560 -1200nm: 3.9-30.2 J/cm2;610-1200nm: 3.9-27.8 J/cm2;640-1200nm: 3.3-24.9 J/cm2;610-980nm: 3.2-23.3 J/cm2;420 -1200nm: 4.1-50.8 J/cm2;510 -1200nm: 3.8-47 J/cm2560 -1200nm: 3.7-43.3 J/cm2610-1200nm: 3.5-38.7 J/cm2640-1200nm: 3.3-37.4 J/cm2690-1200nm: 3.1-33.4 J/cm2
Pulsed Energydensity4.1-35 J/cm2 (10~50 Level)4.1-50.8 J/cm2
Pulsed width5~20 ms5~20 ms
Pulsedduration5~50 ms5~50 ms
Spot Size12 mm × 35 mm &15 mm × 50 mm12 mm × 35 mm &15 mm × 50 mm
Coolingwater + air + TECwater + air + TEC

{5}------------------------------------------------

巨興醫學科技股份有限公司

Smedtrum Medical Technology Co., Ltd. 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist., New Taipei City 248016 , Taiwan
TEL:+886 (02) 2298 9578 FAX:+886 (02) 2298 9426

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains a logo for Smedtrum Medical Technology Co., Ltd. The logo features an abstract geometric shape in shades of teal and gray on the left side. To the right of the shape, the text "Smedtrum" is displayed in a stylized font, with the "S" in gray and the rest of the letters in teal. Below "Smedtrum", the words "MEDICAL TECHNOLOGY CO., LTD." are written in smaller, gray font.

Smedtrum Medical Technology Co., Ltd. 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist., New Taipei City 248016 , Taiwan TEL:+886 (02) 2298 9578 FAX:+886 (02) 2298 9426

VI. PERFORMANCE DATA

The Intense Pulsed Light System has been determined through engineering testing to verify optical energy output and electrical safety.

Electrical safety and electromagnetic compatibility

The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2020 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems

IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains a logo for Smedtrum Medical Technology Co., LTD. The logo features a geometric design in shades of teal and gray on the left side. To the right of the design is the company name, "Smedtrum" in a stylized font, with "MEDICAL TECHNOLOGY CO., LTD." printed in smaller letters below.

for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since a failure of the software could result in minor injury to a patient or to a user of the device.

Sterilization and Shelf-Life

The proposed device is not provided sterile and does not need to be sterilized. The handpiece and the body are cleaned with a soft cloth moistened with ethanol of 70% strength or higher. The proposed device is reusable and does not have a restricted shelflife.

Biocompatibility

The handpiece sapphire tip may be contact with the intact skin of patients. According to FDA guidance document "Use of International Standard ISO 10993-1, " Biological evaluation of medical device - Part 1: Evaluation and testing within a risk management process "" three biological effects were determined in following three test: Cytotoxicity, Sensitization and Irritation.

In Cytotoxicity test, to determine the potential cytotoxicity of finished components, L-929 cell cultured in 96-well plates were treated with extract of 2 test article group:

Test article group 1: Test article extract, original extract

Test article group 2: 50% extract of the test article

The test results showed that the cell morphological grading of test group is the same as negative control group. The cell viability represented 100%, 93%, 11%, 74% and 96%; the mortality showed 0%, 7%, 89%, 26% and 4%. Under the conditions of this test, the extract of test article passed test requirement and did not show potential cytotoxicity to mouse fibroblast L-929 cells.

In Skin sensitization test. Guinea Pig Maximization Test (GPMT) were performed for determination of the skin sensitizing potential of test article extraction. The appearance of the challenge skin sites of the test and control animals 24 h and 48 h after removal of the patch was scored by grade of the skin reactions for erythema and oedema according to the Magnusson and Kligman grading scale. Under GPMT method, the results shows the test article did not cause delayed dermal contact sensitization in the guinea pig.

In Skin irritation test. The results showed that there were no erythema and edema findings in either the control or treatment group, and there were no mortalities. Furthermore, the PII values were 0 (Non-irritant). Therefore, a single topical

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains a logo with a geometric design on the left and text on the right. The geometric design consists of overlapping triangles in shades of blue and gray. To the right of the logo, there is text in Chinese characters, followed by the word "Smedtru" in a combination of gray and blue letters. Below "Smedtru", the words "MEDICAL TECHNOLOGY CO." are written in gray.

application of 0.5 ml of the test article extracts (Polar and non-polar groups) did not cause skin irritation.

Three biological effects, cytotoxicity, sensitization and irritation testing, were performed and test results did not identify any biological response or risk. The Intense Pulsed Light System meets the ISO 10993-1 standard requirements for biocompatibility and no further characterization testing is required.

Bench testing

Intense Pulsed Light system is an Light source equipment. The spatial variation of the LS equipment output over the treatment area shall not deviate from the average irradiance or radiant exposure by more than ±20%

The product fulfills the requirements of IEC 60601-2-57.

VII. CONCLUSION

The Intense Pulsed Light System has the same intended use, similar indications for use, the same technological characteristics, the same energy used, and the same operating principles as its predicates. The non-clinical data and performance testing reports in this submission demonstrate that Intense Pulsed Light System meets the expected performance requirements. Any difference between the subject and predicate device do not raise new issues of safety or effectiveness. Based on above analysis, the Intense Pulsed Light System is as safe, as effective, and performs as well as the cited predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.