K Number
K200746
Date Cleared
2020-05-15

(53 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IPL treatment system is intended for medical use in the treatment of the following dermatologic conditions:

  • Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen;

  • Moderate inflammatory acne vulgaris;

  • Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  • Cutaneous lesions including scars;

  • Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.

Device Description

The IPL Treatment Systems utilizes an intense, visible, broad-spectrum pulse of light that when used with interchangeable filters, allowing it to selectively heat the target for different indications. The optical component is intended to drive the IPL energy to concentrate where the optical energy has selectively heated the target.

The Intense pulsed light (IPL) is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. The light-based component delivers optical energy that is absorbed by specific chromophores in the skin, which is converted to heat, according to the principle of selective photothermolysis.

The proposed device consists of the main unit, control panel and the treatment handpiece(s).

AI/ML Overview

This document is a 510(k) premarket notification for an IPL Treatment System (K200746). It describes the device and claims substantial equivalence to a predicate device (iPulseLight IPL System, K093627). However, the document explicitly states:

VIII Clinical Testing: It is not applicable.

Therefore, there is no study conducted to prove the device meets acceptance criteria through clinical testing. The submission relies solely on non-clinical testing and comparison to the predicate device to establish substantial equivalence.

Given this, I cannot provide the requested information regarding acceptance criteria and a study that proves the device meets those criteria, as no such clinical study was performed or documented in the provided text.

Here is what can be inferred about acceptance criteria and device performance based on the non-clinical information provided:

1. A table of acceptance criteria and the reported device performance:

Since no clinical study was performed for performance, the acceptance criteria are implicitly met by the device demonstrating similar specifications and safety profiles to the predicate device through non-clinical testing.

Acceptance Criteria (Implied from predicate device comparison and non-clinical testing)Reported Device Performance (as presented in comparison table)
Indications for Use: Must be identical or covered by the predicate device's indications.Proposed device indications: Permanent hair reduction, moderate inflammatory acne vulgaris, benign pigmented epidermal lesions (dyschromia, hyperpigmentation, melasma, ephelides), cutaneous lesions (scars), benign cutaneous vascular lesions (port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations).
Product Code: ONFONF
Regulation No.: 21 CFR 878.481021 CFR 878.4810
Class: Class IIClass II
Light source: Intense pulsed light (Xenon Flash Lamp)Intense pulsed light (Xenon Flash Lamp)
Wavelength range: 420 – 1200 nm420 – 1200 nm
Energy output: Equivalent to predicate (10-50 J/cm²)4.1-50.8 J/cm² (Equivalent to predicate given similar energy levels for each filter)
Pulse width: Equivalent to predicate (2-20 ms)5-20 ms (Equivalent to predicate)
Pulse duration: Identical to predicate (5-50 ms)5-50 ms
Spot size: Identical to predicate (1235mm, 1550mm)1235mm, 1550mm
Filters: Identical to predicateIdentical filter ranges available
Fluences (per filter): Equivalent to predicateFluences for each filter are at similar energy level, considered equivalent to predicate.
Output mode: Pulse modePulse mode
Deliver materials: Direct sapphire CouplingDirect sapphire Coupling
Cooling method: Equivalent to predicate (TEC and circulating water & air)Water cooling, forced-air cooling, copper and TEC (for vertical models) and Water cooling and forced-air cooling (for table top models); considered equivalent.
Electrical safety and essential performance: Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57.Device complies with specified IEC standards.
Biocompatibility: Demonstrated through Cytotoxicity, Skin irritation, Skin Sensitization tests for contact materials.Tests performed show compliance with ISO 10993-1.

Regarding the other points, as no clinical study was performed:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
N/A - No clinical test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
N/A - No clinical test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
N/A - No clinical test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
N/A - This device is not an AI-assisted diagnostic tool. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
N/A - This is a physical IPL treatment system, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
N/A - No clinical ground truth was established for performance claims.

8. The sample size for the training set:
N/A - This is a physical device, not a machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established:
N/A - Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a caduceus, which is a traditional symbol of medicine. The text on the right is in blue and is arranged in three lines.

Shanghai Apolo Medical Technology Co., Ltd. Felix Li RA Supervisor 4F, Building A, No.388, Yindu Road, Xuhui District Shanghai, Shanghai 200231 China

Re: K200746

Trade/Device Name: IPL Treatment Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: March 9, 2020 Received: March 23, 2020

Dear Felix Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K200746

Device Name IPL Treatment System

Indications for Use (Describe)

The IPL treatment system is intended for medical use in the treatment of the following dermatologic conditions:

  • Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen;

  • Moderate inflammatory acne vulgaris;

  • Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  • Cutaneous lesions including scars;

  • Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 3-510(k) summary

l Submitter

Shanghai Apolo Medical Technology Co., Ltd. 4F, Building A, No.388, Yindu Road, Xuhui District, Shanghai 200231, China

Establishment Registration Number: 3007120647

Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn

Date of preparation: Apr 28th, 2020

II Proposed Device

Trade Name of Device:IPL Treatment Systems
Common name:Laser surgical instrument for use in general and plastic surgeryand in dermatology
Regulation Number:21 CFR 878.4810
Regulatory Class:Class II
Product code:ONF
Review PanelGeneral & Plastic Surgery

III Predicate Devices

510(k) Number:K093627
Trade name:iPulseLight IPL System (HS 300C, HS 650)
Common name:Laser surgical instrument for use in general and plasticsurgery and in dermatology
Classification:Class II
Product Code:ONF
ManufacturerShanghai APOLO Medical Technology Co., Ltd.

IV Device description

The IPL Treatment Systems utilizes an intense, visible, broad-spectrum pulse of light that when used with interchangeable filters, allowing it to selectively heat the target

{4}------------------------------------------------

for different indications. The optical component is intended to drive the IPL energy to concentrate where the optical energy has selectively heated the target.

The Intense pulsed light (IPL) is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. The light-based component delivers optical energy that is absorbed by specific chromophores in the skin, which is converted to heat, according to the principle of selective photothermolysis.

The proposed device consists of the main unit, control panel and the treatment handpiece(s).

V Indication for use

The IPL treatment systems is intended for medical use in the treatment of the following dermatologic conditions:

  • Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen;
  • -Moderate inflammatory acne vulgaris;
  • Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);
  • -Cutaneous lesions including scars;
  • Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leq telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.
ItemProposed devicePredicate device(K093627)Discussion
ProductCodeONFONFIdentical
RegulationNo.21 CFR 878.481021 CFR 878.4810Identical
ClassClass IIClass IIIdentical
Indicationfor useThe IPL treatmentsystems is intended formedical use in thetreatment of the followingdermatologic conditions:- Permanent hairreduction- long-termiPulseLight IPL System (HS300C, HS 650) are identicalwith regard to indicationsfor use includingrecommended filters to beused with Fitzpatrick skintype. Both models areTheindicationsoftheproposeddevicecovered bythe
stable reduction innumber of hairsre-growing after atreatment regimen;intended for medical use inthe treatment of thefollowing dermatologicconditions:predicate.The coolingsystem isan integralpart of thedevice.
-Moderateinflammatory acnevulgaris;Permanent hairreduction- long-termstable reduction innumber of hairsre-growing after atreatment regimen.
-Benign pigmentedepidermal lesionsincluding dyschromia,hyperpigmentation,melasma, ephelides(freckles);Treatment of:
-Cutaneous lesionsincluding scars;• Moderateinflammatory acnevulgaris
-Benign cutaneousvascular lesionsincluding port winestains, hemangiomas,facial truncal and legtelangiectasias,erythema of rosacea,leg veins, spiderangiomas and venousmalformations.• Benign pigmentedepidermal lesionsincludingdyschromia,hyperpigmentation,melasma,ephelides(freckles).
• Cutaneous lesionsincluding scars
• Benign cutaneousvascular lesionsincluding port winestains,hemangiomas,facial, truncal andleg telangiectasia,erythema ofrosacea, leg veins,spider angiomasand venousmalformations.
The integrated thermalcooling is indicated for usein cooling the epidermis at
the treatment site prior to,
during, and after light
treatment in general
aesthetic dermatologic and
plastic surgery procedures.
- Reduce pain during light
treatment (via partial
anesthesia from
cooling)
- Reduce discomfort
during and/or
associated with light
treatment
- Minimize thermal injury,
including thermal
necrosis, to non-target
skin and skin structures
during and/or
associated with light
treatment, thus reducing
possible complications
such as scabbing,
scarring, hyper - and/or
hypo pigmentation
- Allow the use of higher
light or laser fluencies
for light treatments
(such as for hair
removal and the
treatment of vascular or
pigmented lesions)
- Reduce potential side
effects of light
treatments (such as for
hair removal and the
treatment of lesions)
Light sourceIntense pulsed light(Xenon Flash Lamp)Intense pulsed light (XenonFlash Lamp)Identical
Wavelength420 – 1200 nm420 – 1200 nmIdentical
range
Hanpiece portsHS-650K&HS-660K: 2HS620K, HS-300CK &HS-310K: 1HS 650: 2HS 300C: 1Identical
StructureHS-650K&HS-660K:VerticalHS620K, HS-300CK &HS-310K: Table topHS 650: VerticalHS 300C: Table topIdentical
Energy output4.1-50.8 J/cm²10-50 J/cm²Equivalent
Pulse width5-20 ms2-20 msEquivalent
Pulse duration5-50 ms5-50 msIdentical
Spot size1235mm, 1550mm1235mm, 1550mmIdentical
Filters420 -1200nm: Acne;510 -1200nm: Acne,vascular, pigment;560 -1200nm: Acne,vascular, pigment;610-1200nm: Hairremoval;640-1200nm: Hairremoval;690-1200nm: Hairremoval;420 -1200nm: Acne;510 -1200nm: Acne,vascular, pigment;560 -1200nm: Acne,vascular, pigment;610-1200nm: Hair removal;640-1200nm: Hair removal;690-1200nm: Hair removal;Identical
Fluences420 -1200nm:4.1-50.8J/cm²;510 -1200nm: 3.8-47J/cm²;560 -1200nm: 3.7-43.3J/cm²;610-1200nm: 3.5-38.7J/cm²;640-1200nm:3.3-37.4J/cm²;690-1200nm:3.1-33.4J/cm²;420 -1200nm: 10-50J/cm²;510 -1200nm: 9.2-46.35J/cm²;560 -1200nm: 9-42.6 J/cm²;610-1200nm: 8.5-38.1J/cm²;640-1200nm: 8-36.8J/cm²;690-1200nm:7.6-32.9J/cm²;Thefluences foreach filterare atsimilarenergylevel,consideredequivalent.
Output modePulse modePulse modeIdentical
DelivermaterialsDirect sapphire CouplingDirect sapphire CouplingIdentical
CoolingmethodHS-650K&HS-660K:Water cooling, forced-aircooling, copper and TEC;HS620K,HS-300CK&HS-310K:Water cooling andforced-air cooling,TEC and circulating water &airEquivalent

VI Comparison of technological characteristics with the predicate devices

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

VII Non-Clinical Testing

A series of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:

Electrical safety and electromagnetic compatibility

  • IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General ● requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC 60601-2-57: 2011 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

Biocompatibility Evaluation:

The contact material of the proposed device is identical to that of the iPulseLight IPL System cleared in K093627.

Per FDA's Biocompatibility Guidance issued on June 16, 2016 and with regard to Table A.1 Evaluation Tests for consideration in ISO, "Use of international Standard ISO 10993-1, Biological evaluation of medical - Part 1: Evaluation and testing within a risk management process," the following tests performed on the material which contacts with human for Biocompatibility:

  • Cytotoxicity;
  • Skin irritation; ●
  • Skin Sensitization .; ●

{9}------------------------------------------------

VIII Clinical Testing

It is not applicable.

IX Conclusion substantially equivalent to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.