K Number
K230753
Device Name
RCT800
Manufacturer
Date Cleared
2023-04-11

(25 days)

Product Code
Regulation Number
892.1750
Panel
RA
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RCT800 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified health care professionals.

Device Description

The system's purpose is RCT800, a 3D computed tomography scanner for scanning hard tissues like bones and teeth. By rotating the c-arm, which is embedded with an all-in-one x-ray tube and a detector on each end, CBCT images of the dental maxillofacial area can be attained by recombining data from the same level that is scanned from different angles. Additionally, the system includes a panoramic image scanning function for attaining images of the whole teeth, a cephalometric scanning option for attaining a cephalic image, and a Model Scan option for attaining a dental model CBCT image.

AI/ML Overview

The provided text describes the 510(k) summary for the Ray Co., Ltd.'s RCT800 device. While it mentions various tests and compliance with standards, it does not explicitly provide a table of acceptance criteria and reported device performance in the typical format of a clinical study report with specific metrics and thresholds. Instead, it states that "All test results were satisfactory" for performance (imaging performance) testing conducted according to IEC 61223-3-4 and IEC 61223-3-7.

It also mentions "Clinical images were provided" and that "A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the intended use," but it doesn't detail a formal clinical study with specific acceptance criteria beyond subjective expert opinion.

Therefore, much of the requested information regarding the study that proves the device meets the acceptance criteria in a quantifiable manner (e.g., statistical significance, specific performance numbers) is not present in the provided document. The document primarily focuses on demonstrating substantial equivalence to a predicate device through technical similarities, bench testing, and compliance with general safety and performance standards.

However, based on the provided text, I can infer and extract some information:


Acceptance Criteria and Device Performance (Inferred from compliance statements)

Since the document states that "All test results were satisfactory" for imaging performance tests and that a licensed practitioner found clinical images to be of "acceptable quality for the intended use," the implied acceptance criteria were met. However, the specific quantitative criteria are not listed.

Table 1: Implied Acceptance Criteria and Reported Device Performance

Criterion CategoryImplied Acceptance CriterionReported Device Performance
Imaging PerformanceParameters required to describe functionalities related to imaging properties satisfy designated tolerance (as per IEC 61223-3-4 and IEC 61223-3-7)."All test results were satisfactory." (No specific quantitative results provided)
Clinical Image QualitySample clinical images are of acceptable quality for the intended use by a licensed practitioner."A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the intended use." (Qualitative assessment)
Electrical, Mechanical & Environmental SafetyConformity to IEC 60601-1:2005/AMD1:2012 (3.1 Edition), IEC 60601-1-3:2008/AMD1:2013 (Second Edition), IEC 60601-1-6:2010 (Third Edition), and IEC 60601-2-63:2012 (First Edition)."Electrical, mechanical and environmental safety testing... were performed." (Implied satisfactory outcome as part of substantial equivalence)
EMCConformity to IEC 60601-1-2:2014 (Edition 4.0)."EMC testing was conducted in accordance with the standard IEC 60601-1-2:2014 (Edition 4.0)." (Implied satisfactory outcome as part of substantial equivalence)
Software ValidationCompliance with FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" for moderate level of concern."That has been validated according to the FDA 'Guidance...'" and "...assure substantial equivalence." "Based on our risk analysis of software, the difference does not affect its safety and effectiveness." (Implied successful validation and no safety/effectiveness issues due to software)

Study Details (Based on the provided text)

  1. Sample sizes used for the test set and the data provenance:

    • Test Set (Clinical Images): The document states "Clinical imaging samples were collected from new detectors on the proposed device at the two offices where the predicate device was installed for the clinical test images. These images were gathered from all detectors installed with RCT800 using protocols with random patient age, gender, and size."
      • Specific Number: Not specified. It indicates "samples" and "images gathered from all detectors," implying a collection across multiple patient demographics, but no numerical count is provided.
      • Provenance: Clinical images obtained from "two offices where the predicate device was installed." The country of origin is not explicitly stated for these clinical images, but the applicant (Ray Co., Ltd.) is from SOUTH KOREA, so it is plausible these were collected within South Korea or countries where the predicate device was installed.
      • Retrospective/Prospective: Not explicitly stated, but the mention of collecting "new detectors" and "random patient age, gender, and size" suggests some form of prospective or concurrent collection for evaluation.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: "Two licensed practitioners/clinicians" clinically tested and approved the features. "A licensed practitioner reviewed the sample clinical images." It's unclear if these refer to the same "two" or one specific "licensed practitioner."
    • Qualifications: "Licensed practitioners/clinicians." No further details on their years of experience, subspecialty (e.g., specific type of radiologist/dentist), or formal board certifications are provided.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document states "A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the intended use." This suggests a qualitative assessment, but no formal adjudication method (like 2+1 consensus) is described. It implies that the single or (if two practitioners were involved) potentially un-adjudicated opinion of the expert(s) served as the "truth."
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned. The document focuses on demonstrating that the device itself produces images of acceptable quality for human interpretation, not on the improvement of human readers with AI assistance. This device is a source of images (an X-ray system), not an AI algorithm for image analysis.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device is a medical imaging system (CT/panoramic/cephalometric X-ray system), not an AI algorithm for image analysis. Therefore, the concept of "standalone performance" for an algorithm does not apply in the context of this device. The performance refers to the image acquisition capabilities of the system itself.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For clinical image quality, the ground truth was expert opinion/assessment by "licensed practitioners/clinicians" who deemed the images "of acceptable quality for the intended use."
    • For bench testing, the ground truth was based on compliance with standards (e.g., IEC 61223-3-4, IEC 61223-3-7) and "designated tolerance" parameters.
  7. The sample size for the training set:

    • This device is an X-ray imaging system. It takes images (presumably raw data) and reconstructs them into clinically useful images using reconstruction algorithms such as FBP (Filtered Back Projection) (as mentioned for the predicate device). The document does not describe the use of machine learning that would require a "training set" in the conventional AI sense. If any internal algorithms are adaptable or "learn," details are not provided. The phrase "training set" is typically applicable to AI/ML devices, which this is not identified as in the provided text.
  8. How the ground truth for the training set was established:

    • Not applicable, as no training set (for AI/ML) is mentioned or implied for this device.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.