K Number
K213226
Device Name
RCT700
Manufacturer
Date Cleared
2021-11-29

(61 days)

Product Code
Regulation Number
892.1750
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
RCT700 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatic patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals
Device Description
RCT700 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle. Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image, and Model Scan option for attaining dental model CBCT image are included.
More Information

Not Found

No
The summary describes a standard CBCT and panoramic x-ray imaging system and does not mention any AI or ML capabilities. The testing described focuses on imaging performance and bench testing, not the validation of an AI/ML algorithm.

No
The device is an imaging system intended for diagnostic support through radiographic examination, not for treatment.

Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "intended to radiographic examination...for diagnostic support."

No

The device description clearly details hardware components like a c-arm, high voltage generator, x-ray tube, and detector, indicating it is a physical imaging system, not software-only.

Based on the provided information, the RCT700 device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD devices are used to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
  • The RCT700 is an imaging system that uses X-rays to create images of anatomical structures within the body. It does not analyze biological samples.

The intended use and device description clearly state that the RCT700 is for radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway, ENT structure, and wrist. This is a form of in vivo diagnostic imaging, not in vitro diagnostic testing.

N/A

Intended Use / Indications for Use

RCT700 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatic patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals

Product codes

OAS

Device Description

System purpose RCT700 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.

Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image, and Model Scan option for attaining dental model CBCT image are included.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-Ray

Anatomical Site

dento-maxillofacial, sinus, TMJ, Airway and ENT structure, wrist (for carpus images)

Indicated Patient Age Range

adult and pediatric patients

We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.

Intended User / Care Setting

dentists or other legally qualified heath care professionals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

The clinical imaging samples are collected from the all detector on propose device at the 2 offices where the predicate device is installed on clinical consideration report for the clinical test images. These images were gathered from the all detector installed with RCT700 on any protocols with random patient age, gender, and size. A licensed practitioner reviewed the sample clinical images and found them to be of acceptable quality for the intended use.

Summary of Performance Studies

Bench testing was conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s.

Bench testing is used to assess whether or not the parameter measured required for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance.

Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4 and IEC 61223-3-5.

All test results were satisfactory.

Key Metrics

Not Found

Predicate Device(s)

K182614, K192737

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

November 29, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

RAY Co., Ltd % Ms. Youna Im RA Assistant Manager 332-7. Samsung 1-ro Hwaseong-si. 18380 REPUBLIC OF KOREA

Re: K213226

Trade/Device Name: RCT700 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: September 27, 2021 Received: September 29, 2021

Dear Ms. Im:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

, for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K213226

Device Name RCT700

Indications for Use (Describe)

RCT700 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatic patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

PSC Publishing Services (301) 443-6740 EF

3

5. 510(K) Summary

K213226

4

510(k) Summary

K213226

1. 510(k) Summary

The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

    1. Date: September 27, 2021

3. Administrative Information

APPLICANTRAY Co.,Ltd
ADDRESS#332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 18380, Korea
ManufacturerRAY Co.,Ltd
332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 18380, Korea
TEL : +82-31-605-1000
FAX : +82-2-6280-5534
Contact Persone-mail : youna.im@raymedical.co.kr

4. Device Information

Device Name

Trade/Proprietary Name: RCT700 Common Name: Dental panoramic/tomography and cephalometric x-ray system

Classification

Classification Name: Computed tomography x-ray system Regulation Number : 21 CFR 892.1750 Class : Il Product code : OAS Panel : Radiology

5. Predicate device

ParameterPredicate Device-1Predicate Device-1
Device NameRCT700RCT800
ManufacturerRAY Co., LtdRAY Co., Ltd

5

510(K) NumberK182614K192737
Classification nameComputed tomography x-ray systemComputed tomography x-ray system
Regulation number892.1750892.1750
Primary product
codeOASOAS

7. Device Description

System purpose RCT700 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.

Panoramic image scanning function for attaining image of whole teeth, cephalometric scanning option for attaining cephalic image, and Model Scan option for attaining dental model CBCT image are included.

8. Indication for use

RCT700 is CBCT and panoramic x-ray imaging system with cephalometric.

Which is intended to radiographic examination of the dento-maxillofacial, sinus, TMJ, Airway and ENT structure for diagnostic support for adult and pediatric patients. And a model scan is included as an option. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified heath care professionals

9. Patient population

The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure.

There is no restriction for ethnic group, Gender, weight, health, or condition.

We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.

6

10. Comparison with predicate device

ParameterProposed DevicePredicate Device-1Predicate Device-2
ManufacturerRAY Co., Ltd.RAY Co., Ltd.RAY Co., Ltd.
Device nameRCT700RCT700RCT800
510(K) Number(Traditional 510K)K182614
(Special 510K)K192737
(Traditional 510K)
Common
NameDental panoramic/tomography and
cephalometric x-ray systemDental panoramic/tomography and
cephalometric x-ray systemDental panoramic/tomography and
cephalometric x-ray system
Indications
for useRCT700 is CBCT and panoramic x-ray
imaging system with cephalometric.
Which is intended to radiographic
examination of the dento-maxillofacial,
sinus, TMJ, Airway and ENT structure
for diagnostic support for adult and
pediatric patients. And a model scan is
included as an option. Cephalometric
image is also includes wrist to obtain
carpus images for growth and maturity
assessment for orthodontic treatment.
The device is to be operated and used
by dentists or other legally qualified
heath care professionalsRCT700, panoramic x-ray imaging
system with cephalostat, is an extraoral
source x-ray system, which is intended
for dental radiographic examination of
the teeth, jaw, and oral structures,
specifically for panoramic examinations
and implantology and for TMJ studies
and cephalometry, and it has the
capability, using the CBVT technique, to
generate dento-maxillo-facial 3D
images. The device uses cone shaped
x-ray beam projected on to a flat panel
detector, and the examined volume
image is reconstructed to be viewed in
3D viewing stations. 2D Images are
obtained using the standard narrow
beam technique.RCT800 is CBCT and panoramic x-ray
imaging system with cephalometric.
Which is intended to radiographic
examination of the dento-maxillofacial,
sinus, TMJ, Airway for diagnostic
support for adult and pediatric patients.
And a model scan is included as an
option.
Cephalometric image is also includes
wrist to obtain carpus images for growth
and maturity assessment for orthodontic
treatment.
The device is to be operated and used
by dentists or other legally qualified
heath care professionals.
Mode of
OperationSame as predicate device #1Continuous operation with
intermittent, stated permissible
loadingSame as predicate device #1
3D technologySame as predicate device #1CBCT Cone beam
Computed TomographySame as predicate device #1
Performance
SpecificationSame as predicate device #11) CBCT Computed tomography
  • Patient
  1. Panoramic
  2. Cephalometric(optional)
  • One shot type
  • Scan type | 1) CBCT Computed tomography
  • Patient
  • Dental Model Scan (Optional)
  1. Panoramic
  2. Cephalometric(optional)
  • One shot type

  • Scan type |
    | Functional
    Option | | Same as predicate device #1 | Base
    CT+PANO
    Option(CEPH)
    CT + PANO + SCAN CEPH
    CT + PANO + One shot(One shot,
    Standard Type)
    CT + PANO + One shot(One shot,
    Large Type). | Same as predicate device #1 |
    | Detector
    Type | CT | Same as predicate device #1 | FXDD-0606CA | Same as predicate device #1 |
    | | CT | Jupi0606X | SiX 650HD-E | FXDD-1012CH |
    | | PANO | Same as predicate device #1 | FXDD-0606CA | Same as predicate device #1 |
    | | PANO | Jupi0606X | SiX 650HD-E | FXDD-1012CH |
    | | Ceph (Scan) | Same as predicate device #1 | XID-C24DC | Same as predicate device #1 |
    | | Ceph (One shot) | FXDD-1012CA | PaxScan 2530C | Same as predicate device #1 |
    | | Ceph (One shot) | FXRD-1717VA | 1717SCC | Same as predicate device #1 |
    | Exposure switch
    Type | | Same as predicate device #1 | "Deadman" Button type | Same as predicate device #1 |
    | Main
    Components | | Same as predicate device #1 | Ceph Apparatus | Same as predicate device #1 |
    | Main
    Components | | Same as predicate device #1 | Vertical Carriage | Same as predicate device #1 |
    | Main
    Components | | Same as predicate device #1 | Rotator | Same as predicate device #1 |
    | | | | | |
    | | Same as predicate device #1 | X-RAY Generator | Same as predicate device #1 | |
    | | Same as predicate device #1 | X-ray tube | Same as predicate device #1 | |
    | | Same as predicate device #1 | High Frequency Generator | Same as predicate device #1 | |
    | | Same as predicate device #1 | Column | Same as predicate device #1 | |
    | | Same as predicate device #1 | Touch monitor (panel) | Same as predicate device #1 | |
    | | Detector

  • CT
    FXDD-0606CA
    Jupi0606X

  • PANO
    FXDD-0606CA
    Jupi0606X

  • Ceph
    XID-C24DC(Scan)
    FXDD-1012CA(One shot, Standard
    Size)
    FXRD-1717VA(One shot, Large Size) | Detector

  • CT
    FXDD-0606CA
    SIX 650HD-E

  • PANO
    FXDD-0606CA
    SIX 650HD-E

  • Ceph
    XID-C24DC(Scan)
    PaxScan 2530C(One shot, Standard
    Size)
    1717SCC(One shot, Large Size) | Detector

  • CT
    FXDD-0606CA
    FXDD-1012CH

  • PANO
    FXDD-0606CA
    FXDD-1012CHA

  • Ceph
    XID-C24DC(Scan)
    PaxScan 2530C(One shot, Standard
    Size)
    1717SCC(One shot, Large Size) | |
    | | Same as predicate device #1 | Chinrest | Same as predicate device #1 | |
    | | Same as predicate device #1 | Head rest | Same as predicate device #1 | |
    | | Same as predicate device #1 | Automatic Collimator | Same as predicate device #1 | |
    | | Same as predicate device #1 | Exposure switch | Same as predicate device #1 | |
    | | Same as predicate device #1 | Emergency stop switch | Same as predicate device #1 | |
    | | Same as predicate device #1 | Console PC set | Same as predicate device #1 | |
    | Automatic
    Collimator | Same as predicate device #1 | CT exams
    Panoramic exams
    Cephalometric exams | Same as predicate device #1 | |

The following table provides the summary of the technological characteristics of RCT800 compared to the predicate device.

7

8

9

| Display Type | | Same as predicate device #1 | TFT LCD type(Normally black)
*1280x800 pixel | Same as predicate device #1 |
|-------------------------|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Class | | Same as predicate device #1 | Class I with type B applied parts
according to IEC 60601-1 | Same as predicate device #1 |
| Focal size | | Same as predicate device #1 | 0.5 | Patient 0.5
Model scan 0.04 (Optional) |
| Field of
View(CT) | | Same as predicate device #1 | Max.160x100 mm | FXDD-0606CA : Max.160x100 mm
FXDD-1012CHA : Max. 200x200 mm |
| X-ray Voltage(Patient) | | Same as predicate device #2 | 6090kVp | 60100kVp |
| X-ray Current(Patient) | | 117mA | 417mA | Same as predicate device #1 |
| Total Filtration | | Same as predicate device #1 | Min. 2.8 mm Al equivalent | Same as predicate device #1 |
| Detector
Pixel size | CT | Same as predicate device #1 | FXDD-0606CA: 119μm | Same as predicate device #1 |
| | Jupi0606X : 100μm | | SiX 650HD-E : 150μm | FXDD-1012CHA: 124μm |
| | PANO | Same as predicate device #1 | FXDD-0606CA: 119μm | Same as predicate device #1 |
| | Jupi0606X : 100μm | | SiX 650HD-E : 150μm | FXDD-1012CHA: 124μm |
| | Ceph (Scan) | Same as predicate device #1 | XID-C24DC: 100μm | Same as predicate device #1 |
| | Ceph(One shot) | FXDD-1012CA : 124μm | PaxScan 2530C: 139μm | Same as predicate device #1 |
| | FXRD-1717VA : 140μm | | 1717SCC: 127μm | Same as predicate device #1 |
| | CT | Same as predicate device #1 | 1.44 | 1.44(Patient)
1.91(Model Scan) |
| | PANO | Same as predicate device #1 | 1.31 | Same as predicate device #1 |
| Magnifica
tion | Ceph (Scan) | Same as predicate device #1 | 1.11 | Same as predicate device #1 |
| | Ceph(One shot) | Same as predicate device #1 | Standard Size : 1.12 | Same as predicate device #1 |
| | | Same as predicate device #1 | Large Size : 1.13 | Same as predicate device #1 |
| Scan time | | Same as predicate device #1 | CT : below 14sec | CT : below 20sec |
| | | | | |
| | Same as predicate device #1 | Pano : below 14sec | Same as predicate device #1 | |
| | Same as predicate device #2 | Ceph[Scan type] : below 19sec | Ceph[Scan type] : below 20sec | |
| | Same as predicate device #1 | Ceph[One shot type]:
below 2sec | Same as predicate device #1 | |
| Format compatible | Same as predicate device #1 | DICOM 3.0 Format compatible | Same as predicate device #1 | |
| Image Viewing Software | Same as predicate device #1 | RayScan
(Cleared under K182614) | RayScan
(Cleared under K192737) | |
| Image acquisition | Same as predicate device #1 | Giga-Ethernet Network | Same as predicate device #1 | |
| Total Height | Same as predicate device #1 | Max 2,296mm | Same as predicate device #1 | |
| | | 1) Computed Tomography(CT) +
Panoramic(PANO)=185kg(407.9lb) ±
10% | 1) Computed Tomography(CT) +
Panoramic(PANO)=189kg(416.6lb) ±
10% | |
| | | 2) Computed Tomography(CT) +
Panoramic(PANO) + Ceph
(Scan type)= 212.5kg (468.5lb) ± 10% | 2) Computed Tomography(CT) +
Panoramic(PANO) + Ceph
(Scan type)= 219kg (482.8lb) ± 10% | |
| Weight | Same as predicate device #1 | 3) Computed Tomography(CT) +
Panoramic(PANO) + Ceph
(One shot type, installed in Standard
size)= 211kg (465.2lb) ± 10% | 3) Computed Tomography(CT) +
Panoramic(PANO) + Ceph
(One shot type, installed in Standard
size)= 217kg (478.4lb) ± 10% | |
| | | 4) Computed Tomography(CT) +
Panoramic(PANO) + Ceph
(One shot type, installed in Large size)
211kg (465.2lb) ± 10% | 4) Computed Tomography(CT) +
Panoramic(PANO) + Ceph
(One shot type, installed in Large size)
212kg (467.3lb) ± 10% | |
| Type of
installation | Same as predicate device #1 | Wall or floor mount | Same as predicate device #1 | |
| Patient position | Same as predicate device #1 | Standing / Wheelchair | Same as predicate device #1 | |
| Applicable
Standards | Same as predicate device #1 | IEC 60601-1
IEC 60601-1-3
IEC 60601-2-63
IEC 60601-1-2 | Same as predicate device #1 | |

10

11

The product is principally just the same as in the previous 510(k) #K182614 and K192737. The complete of differences of the subject device to the predicate device is as follows

  • The minimum X-ray Current of the tube has been changed from 4mA to 1mA.
  • Added new CT, Pano, Scan Ceph and One shot Ceph detector.

12

11. Safety and Effectiveness Information

RCT700 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.

The following information further substantial equivalence between the subject device and predicate device.

The fundamental technological characteristics of the subject and predicate device are similar.

The imaging modes are similar; PANO, CEPH (Optional), CBCT All viewing software programs have been cleared with previous 510k submissions; RAYSCAN(K182614).

The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and the predicate device for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.

The differences are as follows.

  • -The minimum X-ray Current of the tube has been changed from 4mA to 1mA.
  • The irradiation time of One shot Ceph has been changed.

Electrical, mechanical and environmental safety testing according to standard of IEC 60601-1: 2005/AMD1:2012(3.1 Edition), IEC 60601-1-3: 2008/AMD1:2013(Second Edition), IEC 60601-1-6:2010(Third Edition) and IEC 60601-2-63: 2012/AMD1:2017(First Edition) were performed. EMC testing was conducted in accordance with the standard IEC 60601-1-2: 2014(Edition 4.0).

The device software, named Rayscan, saves the patient and image data and offers an inquiry function, in addition, supports the image generate function intended to obtain images using the RCT700 equipment and various sensors for diagnosis. And that has been validated according to FDA "Guidance for the Content d Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" to assure substantial equivalence.

RCT700 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics, and X-ray source.

The following information further substantial equivalence between the subject device and predicate device.

The fundamental technological characteristics of the subject and predicate device are similar.

The imaging modes are similar; PANO, CEPH (Optional), CBCT All viewing software programs have been cleared with previous 510k submissions; RAYSCAN(K182614).

The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.

As a result, we identified the level of concern associated with new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.

Bench testing was conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s.

Bench testing is used to assess whether or not the parameter measured required for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance.

Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4 and IEC 61223-3-5.

All test results were satisfactory.

Non-clinical considerations were conducted in accordance with FDA Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". Because the subject device used the

13

same detector as the predicate device, there is no significant difference between the two devices as a result of non-clinical testing.

Clinical considerations were conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s.

Clinical images were provided these images were not necessary to establish substantial equivalence based on the modifications to the device but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended.

The features of RCT700 have been clinically tested and approved by two licensed practitioners/ clinicians.

The clinical imaging samples are collected from the all detector on propose device at the 2 offices where the predicate device is installed on clinical consideration report for the clinical test images. These images were gathered from the all detector installed with RCT700 on any protocols with random patient age, gender, and size. A licensed practitioner reviewed the sample clinical images and found them to be of acceptable quality for the intended use.

12. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. RAY Co., Ltd. concludes that the newly RCT700 is safe and effective and substantially equivalent to predicate device as described herein.