(41 days)
RAIN SheathTM Tibial Pedal Introducer is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.
The RAIN Sheath™ Tibial Pedal Introducer device (hereafter referred as RAIN Sheath™ Pedal) consists of a sheath introducer, a vessel dilator (0.021" guidewire compatible), a stainless-steel entry needle, and a 0.021" guidewire (either stainless steel or nitinol). The sheath introducer has a lubricious hydrophilic coating, a smooth shoulder transition to the dilator and low-profile outer diameters (OD). To provide an atraumatic transition between the dilator tip and the 0.021" mini guide wire, the RAIN Sheath™ dilator is tapered at the distal end and the inner diameter (ID) of the dilator has been optimized. The hub is overmolded to the dilator and includes a locking mechanism between the hub of the vessel dilator and the hub of the sheath cannula to facilitate insertion of the product while preventing the vessel dilator from backing out of the cannula. The RAIN Sheath™ Pedal device is available in four (4) product configurations which differ based on the French sizes (4F and 5F) and specific wire included in the system. RAIN Sheath™ Pedal is a single-use sterile device, sterilized by ethylene oxide.
The provided text is a 510(k) summary for the RAIN Sheath™ Tibial Pedal Introducer, which is a medical device. This type of document is used to demonstrate substantial equivalence to a predicate device, not necessarily to prove efficacy through rigorous clinical studies. Therefore, it focuses on non-clinical tests and comparisons rather than comprehensive clinical trial data.
Based on the provided text, here's an analysis regarding acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance:
The document states that the RAIN Sheath™ Pedal device uses the same fundamental technology and operating principles as the predicate device. Therefore, the acceptance criteria are implicitly tied to demonstrating that the new device meets the established performance characteristics and safety profile of the predicate. The document doesn't explicitly list specific quantitative acceptance criteria in a table format, but rather discusses successful completion of various non-clinical tests.
| Test Category | Acceptance Criteria (Implicit) | Reported Device Performance and Conclusion |
|---|---|---|
| Kinking Resistance | Device should not kink; user should be able to handle device with reasonable force upon insertion. | Design Validation testing demonstrated no kinking and reasonable handling force. |
| Usability | Critical tasks should be comparable to the predicate device in simulated use. | All results met acceptance criteria; device demonstrated comparability to its predicate in simulated use conditions using a benchtop model. |
| Biocompatibility | Must be biocompatible for intended use (externally communicating device with limited blood contact ≤ 24 hours). | Leveraged tests from RAIN Sheath™ Transradial (K181592) showed compliance with ISO 10993-1. These included Chemical Characterization, Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Hemocompatibility (Hemolysis, PTT, Platelet & Leukocyte Counts, Complement Activation C3a & SC5b-9), and In-vivo Thrombogenicity. No additional biocompatibility testing was performed as both devices have identical materials. |
| Sterilization | Achieve a Sterility Assurance Level (SAL) of 10⁻⁶. | Adopted the validated ethylene oxide sterilization protocol for the reference device (RAIN Sheath™ Transradial, K181592), which was previously validated per ISO 11135:2014+A1:2018. Evaluations of ethylene oxide and ethylene chlorohydrin residuals were also leveraged from K181592. |
2. Sample Size Used for the Test Set and Data Provenance:
- Non-Clinical Performance (Kinking Resistance): The document does not specify a numerical sample size for the "Design Validation testing" which addressed kinking. It simply states "Design Validation testing was performed."
- Usability: The document states that "devices were tested in the simulated use condition using a clinically relevant bench top model." It does not specify the number of devices or number of users (sample size).
- Biocompatibility: The biocompatibility data was leveraged from the previously cleared RAIN Sheath™ Transradial (K181592). The sample size for these original tests is not provided in this document, nor is the provenance of the biological materials/samples used within those tests.
- Sterilization: The sterilization validation was leveraged from the RAIN Sheath™ Transradial (K181592) and followed ISO 11135. The sample size for the validation and provenance is not detailed here.
Data Provenance: For the RAIN Sheath™ Pedal, the primary "data" for equivalence comes from its identical design, materials, and manufacturing process to the RAIN Sheath™ Transradial (the reference device) and from non-clinical benchtop testing. The origin of the non-clinical testing is implied to be internal (Cordis US Corp). There is no mention of country of origin of data or whether it was retrospective or prospective in the context of clinical data, as no clinical data was used for this submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable / Not explicitly stated for non-clinical tests. For device performance tests like kinking or usability, the "ground truth" is typically objective measurements or observations by test engineers and subject matter experts. The document does not specify the number or qualifications of experts involved in conducting or evaluating these non-clinical tests.
4. Adjudication Method for the Test Set:
- Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based AI studies where there are subjective interpretations that need consensus. Since the tests described are non-clinical and product-level, such adjudication methods are not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. The document explicitly states: "No clinical studies were deemed necessary to support substantial equivalence." Therefore, an MRMC comparative effectiveness study was not performed.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was done:
- Not applicable. This device is a physical medical device (an introducer sheath), not an AI algorithm. Therefore, "standalone algorithm" performance is not relevant.
7. The Type of Ground Truth Used:
- Non-clinical objective measurements and industry standards. For kinking resistance, the ground truth would be physical observation of kinking under simulated use conditions and evaluation against reasonable force. For usability, the ground truth would be successful completion of critical tasks by users on a benchtop model. For biocompatibility and sterilization, the ground truth is compliance with recognized international standards (ISO 10993-1, ISO 11135) and successful results from the leveraged tests.
8. The Sample Size for the Training Set:
- Not applicable. This submission is for a physical medical device, not an AI/ML algorithm. There is no concept of a "training set" in this context. The "design validation" and "usability testing" can be considered as verification and validation activities rather than training.
9. How the Ground Truth for the Training Set was Established:
- Not applicable. As explained above, there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
April 24, 2023
Cordis US Corp. Wai Morgan Project Manager, Regulatory Affairs 14201 North West 60th Avenue Miami Lakes, Florida 33014
Re: K230704
Trade/Device Name: RAIN Sheath™ Tibial Pedal Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: March 14, 2023 Received: March 15, 2023
Dear Wai Morgan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Misti L. Malone -S
Misti Malone, PhD Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230704
Device Name RAIN Sheath™ Tibial Pedal Introducer
Indications for Use (Describe)
RAIN SheathTM Tibial Pedal Introducer is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
I. SUBMITTER
Applicant: Cordis US Corp. 14201 North West 60th Avenue Miami Lakes, Florida 33014 USA Establishment Registration: 1016427
Contact: Wai Morgan wai.morgan@cordis.com
Date Prepared: March 10, 2023
II. DEVICE
Name of Device: RAIN Sheath™ Tibial Pedal Introducer Common Name: Introducer Sheath Classification Name: Introducer, Catheter Product Code: DYB
III. PREDICATE DEVICE
Glidesheath Slender Pedal Tibial Kit cleared on 08/03/2018 under K181237.
Reference Device: RAIN Sheath™ Transradial Introducer cleared on 08/15/2018 under K181592.
Predicate device and Reference device cited above has not been the subject of a recall.
IV. DEVICE DESCRIPTION
The RAIN Sheath™ Tibial Pedal Introducer is used to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.
The RAIN Sheath™ Tibial Pedal Introducer device (hereafter referred as RAIN Sheath™ Pedal) consists of a sheath introducer, a vessel dilator (0.021" guidewire compatible), a stainless-steel entry needle, and a 0.021" guidewire (either stainless steel or nitinol).
The sheath introducer has a lubricious hydrophilic coating, a smooth shoulder
{4}------------------------------------------------
transition to the dilator and low-profile outer diameters (OD). To provide an atraumatic transition between the dilator tip and the 0.021" mini guide wire, the RAIN Sheath™ dilator is tapered at the distal end and the inner diameter (ID) of the dilator has been optimized. The hub is overmolded to the dilator and includes a locking mechanism between the hub of the vessel dilator and the hub of the sheath cannula to facilitate insertion of the product while preventing the vessel dilator from backing out of the cannula.
The RAIN Sheath™ Pedal device is available in four (4) product configurations which differ based on the French sizes (4F and 5F) and specific wire included in the system. RAIN Sheath™ Pedal is a single-use sterile device, sterilized by ethylene oxide,
RAIN Sheath™ device is for professional use in a hospital, catheterization laboratory, or other suitable healthcare facility only.
V. INDICATIONS FOR USE
RAIN Sheath™ Tibial Pedal Introducer is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.
VI.COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE and REFERENCE DEVICE
RAIN Sheath™ Tibial Pedal Introducer device and the predicate device Glidesheath Slender Pedal Tibial Kit (also referred as "GSS Pedal") facilitate low profile access into the peripheral vasculature below the knee using the same fundamental mechanism of action. Both devices have introducer sheaths with hydrophilic coating which imparts lubricity and facilitates ease of insertion of the device. The dilator component of each device supports the sheath in which it is inserted and dilates the vessel during insertion. The introducer sheath with hydrophilic coating and the tapered dilator which provides a smooth transition between the miniquidewire and the introducer sheath facilitates atraumatic entry into the artery over a miniguidewire and minimizes tissue damage.
RAIN Sheath™ Pedal has the following similarities to the GSS Pedal predicate device:
- Same intended use ●
- Same principle of operation ●
- Same mechanism of action
- Same method of sterilization and sterility assurance level ●
- Same biocompatibility classification ●
- Biocompatible for intended use ●
- Labeled non-pyrogenic
- Similar materials ●
- Similar components ●
- . Similar device dimensions
- Similar packaging configuration ●
- Similar compatibility with other devices used in the below the knee procedures ●
{5}------------------------------------------------
Furthermore, RAIN Sheath™ Tibial Pedal Introducer shares significant similarities with the Cordis RAIN Sheath™ Transradial (reference device). The device components, material composition and the composition of the packaging materials remain the same as that of the RAIN Sheath™ Transradial. The predicate device (GSS Pedal) is offered in 5F size only, whereas the proposed RAIN Sheath™ Pedal will be offered in both 4F and 5F sizes similar to RAIN Transradial.
Based on a thorough analysis of technological characteristics, including design, materials, dimensions, mechanism of action, and clinical use, the RAIN Sheath™ Pedal is substantially equivalent to the predicate device.
VII. NON-CLINICAL TESTS
Performance
The RAIN Sheath™ Pedal device has same intended use, fundamental technology (operating principle), mode of operation, anatomical site of use, clinical utility, as that of the Predicate device, Glidesheath Slender Tibial/Pedal. All components of the RAIN Sheath™ Pedal device come from its reference device, RAIN Sheath Transradial. Both the devices are identical in design, materials, formulation, geometry, manufacturing process, packaging and sterility. Therefore, no design verification was performed on RAIN Sheath™ Pedal to demonstrate the substantial equivalence of the proposed RAIN Sheath™ Tibial/Pedal Introducer.
Design Validation testing was performed to validate that the RAIN Sheath™ Pedal device did not observe any kinking and allowed user to handle the device with reasonable force upon insertion. Additionally, design validation simulated use conditions testing was leveraged from the reference device, RAIN Sheath™ Transradial. The results of the design validation demonstrated that the subject device, RAIN Sheath™ Pedal, is appropriate for its use and is considered substantially equivalent to the predicate device.
Usabilitv
To evaluate critical tasks of the RAIN Sheath™ Pedal device to that of the predicate device, devices were tested in the simulated use condition using a clinically relevant bench top model. All results met the acceptance criteria and demonstrated that the RAIN Sheath™ Pedal device is comparable to its predicate device. The usability testing performed supports the substantial equivalence of the proposed RAIN Sheath™ Tibial Pedal Introducer device.
Biocompatibility Testing
RAIN Sheath™ Pedal, like the predicate, is an externally communicating device with limited contact duration (≤ 24 hours) with circulating blood. Biocompatibility testing was performed on the currently cleared RAIN Sheath™ Transradial (K181592) in accordance with FDA Guidance, Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (September 2020) and ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. On that basis, the following tests were leveraged from K181592:
- . Chemical Characterization
- . In vitro Cytotoxicity- MEM Elution
{6}------------------------------------------------
- Sensitization - Guinea Pig Maximization
- Irritation or Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- . Material Mediated Pvrogenicitv
- . Hemocompatibility
- . In vitro Hemolvsis -Direct/Extract
- Partial Thromboplastin Time (PTT)
- Platelet and Leukocyte Counts ●
- Complement Activation C3a & SC5b-9 Assay
- . In-vivo Thrombogenicity
Based on the biocompatibility testing performed on RAIN Sheath™ Transradial (cleared under K181592), no additional biocompatibility testing for the subject device was performed as both devices have identical materials.
Sterilization
The RAIN Sheath™ Tibial Pedal Introducer was adopted into the same validated ethylene oxide sterilization protocol for the reference device (RAIN Sheath™ Transradial, cleared under K181592) which was previously validated per ISO 11135:2014+A1:2018 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices to provide a sterility assurance level (SAL) of 10°. Previous evaluations of the ethylene oxide and ethylene chlorohydrin residuals was leveraged from K181592.
Clinical Studies
No clinical studies were deemed necessary to support substantial equivalence.
CONCLUSIONS VIII.
In summary, the RAIN Sheath™ Tibial/Pedal Introducer is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the predicate device, Glidesheath Slender Tibial Pedal Kit.
§ 870.1340 Catheter introducer.
(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).