K Number
K230464
Date Cleared
2023-05-22

(90 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for the transfer and injection of drugs contained in a vial.

Device Description

The subject device, MixJect® Transfer Device (MXJ), is a single-use, gamma sterilized, nonpyrogenic device intended for the transfer and injection of drugs contained in a vial. The MXJ device is intended for use by Healthcare Professionals (HCPs) in a clinical, hospital, or other healthcare environment. The subject device is available by prescription use only and has no known contraindications. The MXJ device does not contain any medicinal substances and is only intended for use with drug vials having a neck diameter of 13mm. The subject device has a 3-year shelf life. The MXJ device configuration consists of four main components inclusive of a 30-gauge needle, a vial adapter including integral cannulated spike, MXJ Body, and MXJ Core. The subject device interfaces with a syringe (not supplied) that connects to the female Luer lock port located in the main body of the MXJ device. The MXJ Vial Adapter component connects to a drug vial having a neck diameter of 13mm. The Vial Adapter component contains a piercing spike and a female Luer lock connector. Puncturing of the drug vial stopper membrane is achieved by means of an integral spike located in the center of the MXJ Vial Adapter component. Once the drug vial stopper membrane is breached by the cannulated spike, fluid can travel from the drug vial into the MXJ device main body. A prefilled diluent syringe (not supplied) is then connected to MXJ female Luer lock port. The diluent is injected from the syringe into the drug vial. The reconstituted medicament is then aspirated back into the syringe. The MXJ device is then twisted in a counterclockwise direction, changing the fluid path from the syringe-vial to the syringe-needle. After reconstitution and aspiration, the drug is ready to be administered through the attached MXJ needle. The MXJ primary device package consists of a polyethylene terephthalate glycol (PETG) blister sealed with a Tyvek® lid sealed on top of the blister pack.

AI/ML Overview

The provided FDA 510(k) summary for the MixJect® Transfer Device (K230464) describes performance testing, but it does not specify explicit acceptance criteria in a quantitative manner or provide specific performance data points in a table as requested. Instead, it states that testing "confirm[s] the MixJect® Transfer Device meets all applicable design and performance requirements throughout its defined shelf life and verify conformity to the applicable external and internal standards and demonstrate substantial equivalence to the predicate device."

Therefore, I cannot generate a table of acceptance criteria and reported device performance with specific numerical values based on this document. However, I can infer the types of performance metrics and the general conclusion of the study.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

As mentioned, specific quantitative acceptance criteria and reported performance values are not provided in this document. The document primarily lists the types of performance tests conducted and generally states that the device meets requirements and standards.

Inferred Performance Metrics (from "Performance Testing" section):

Performance Metric(Inferred) Acceptance CriteriaReported Device Performance
Mechanical/Physical Performance
FragmentationConforms to ISO 8536-2:2010 section 6.2.2 requirementsMet applicable design and performance requirements.
Particulate MatterConforms to USP <788> requirementsMet applicable design and performance requirements.
Internal Diameter Upper SkirtConforms to ISO 8362-6:2010 Section 4.2 requirements and in-house methodMet applicable design and performance requirements.
Luer GaugingConforms to ISO 594-1:1986 and ISO 594-2:1998 requirementsMet applicable design and performance requirements.
Luer Stability & Compliance (ISO 80369-7)Conforms to ISO 80369-7:2021, ISO 80369-20:2015 Annex B, C, D, E, F, G requirementsMet applicable design and performance requirements.
Residual VolumeConforms to in-house test method requirementsMet applicable design and performance requirements.
Device LeakageConforms to in-house test method requirementsMet applicable design and performance requirements.
Device Total Penetration ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Vial Adapter Detachment ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Product Retention in BlisterConforms to in-house test method requirementsMet applicable design and performance requirements.
Flow RateConforms to in-house test method requirementsMet applicable design and performance requirements.
Device Removal Force from BlisterConforms to in-house test method requirementsMet applicable design and performance requirements.
Tyvek Total Peel Test ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Injection ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Aspiration ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Impact ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Needle Protective Cap Removal ForceConforms to in-house test method requirementsMet applicable design and performance requirements.
Torque TestConforms to in-house test method requirementsMet applicable design and performance requirements.
Functional Performance
Functionality according to IFUDevice operates as described in Instructions For Use (IFU)Met applicable design and performance requirements.
Label LegibilityLabels are clear and legibleMet applicable design and performance requirements.
Biocompatibility
CytotoxicityConforms to ISO 10993-5:2009 standardsBattery of tests conducted; results were acceptable.
Hemolysis (ASTM & ISO)Conforms to ISO 10993-4:2017 and ASTM F756 standardsBattery of tests conducted; results were acceptable.
Maximization and SensitizationConforms to ISO 10993-10:2010 standardsBattery of tests conducted; results were acceptable.
Intracutaneous ReactivityConforms to ISO 10993-10:2010 standardsBattery of tests conducted; results were acceptable.
Acute Systemic ToxicityConforms to ISO 10993-11:2017 standardsBattery of tests conducted; results were acceptable.
Material Mediated PyrogenicityConforms to ISO 10993-11:2017 standardsBattery of tests conducted; results were acceptable.
Sterilization
Sterility Assurance Level (SAL)Achieve SAL of 10-6Achieved SAL of 10-6.
Bacterial EndotoxinAcceptable levels per limulus amebocyte lysate (LAL) testingPassed with acceptable levels.

Study Proving Device Meets Acceptance Criteria:

The study proving the device meets its requirements is a series of non-clinical bench performance tests and biocompatibility tests as detailed in the "Performance Data" section of the 510(k) submission.

2. Sample Size and Data Provenance:

  • Sample Size for Test Set: The document does not specify the sample size (number of devices) used for each individual performance test or biocompatibility test.
  • Data Provenance: The document does not explicitly state the country of origin of the data for the performance or biocompatibility tests. It indicates the manufacturer, West Pharma. Services IL, Ltd., is located in Ra'anana, Israel. The studies are retrospective in the sense that they are laboratory and bench studies conducted on the device, not ongoing clinical observations.

3. Number of Experts and Qualifications:

  • Not Applicable. For non-clinical bench testing and biocompatibility assessments, the ground truth is established by objective measurements against recognized international and internal standards, rather than expert consensus on medical images or diagnoses. These tests are typically performed by qualified laboratory technicians and overseen by engineers or scientists with expertise in material science, mechanical testing, and biocompatibility, who interpret the results against established regulatory requirements and standards.

4. Adjudication Method:

  • Not Applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where expert consensus is required for ground truth labeling, often for diagnostic accuracy studies. This document reports on bench and biocompatibility testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No. The document explicitly states: "Clinical trials were not performed for the MixJect® Transfer Device." Therefore, no MRMC comparative effectiveness study was conducted, and no effect size for human reader improvement with AI assistance is applicable or reported.

6. Standalone Performance Study:

  • Yes, in the context of a medical device. The entire "Performance Data" section describes standalone performance of the device itself through a variety of engineering, material, and biological safety tests (bench tests, biocompatibility tests). These tests establish the device's inherent properties and functionality without a human-in-the-loop performance assessment (which would be a clinical trial).

7. Type of Ground Truth Used:

  • The ground truth for the performance tests and biocompatibility tests is based on:
    • International Standards: e.g., ISO 8536-2, USP <788>, ISO 80369-7, ISO 10993 series, BS EN ISO 11137 series.
    • In-house test methods: Developed and validated by the manufacturer to assess specific device functions.
    • Established biological principles and safety thresholds: For biocompatibility and sterility.

8. Sample Size for Training Set:

  • Not Applicable. This device is a physical medical device (an intravascular administration set), not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The device design and manufacturing processes are likely informed by engineering principles, material science knowledge, and previous device iterations.

9. How Ground Truth for Training Set Was Established:

  • Not Applicable. As there is no AI/ML algorithm or training set, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

May 23, 2023

West Pharma Services IL, Ltd. % Fred Cowdery Director, Regulatory Affairs West Pharmaceutical Services, Inc. 530 Herman O. West Drive Exton, Pennsylvania 19341

Re: K230464

Trade/Device Name: MixJect® Transfer Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI

Dear Fred Cowdery:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 22, 2023. Specifically, FDA is updating this SE Letter to add the missing clearance date as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact David Wolloscheck, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, 301-796-1480, david.wolloscheck(@fda.hhs.gov.

Sincerely,

Porsche Bennett

Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo has a blue square with the letters FDA in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

West Pharma Services IL, Ltd. % Fred Cowdery Director, Regulatory Affairs West Pharmaceutical Services, Inc. 530 Herman O. West Drive Exton, Pennsylvania 19341

Re: K230464

Trade/Device Name: MixJect® Transfer Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: LHI Dated: February 17, 2023 Received: February 21, 2023

Dear Fred Cowdery:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Porsche Bennett

Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K230464

Device Name MixJect® Transfer Device

Indications for Use (Describe)
The device is intended for the transfer and injection of drugs contained in a vial.
Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K230464 - 510(k) SUMMARY

Date Prepared: May 22, 2023

Applicant:

West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297

Manufacturer:

West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297

Contact Person:

Fred Cowdery Director, Regulatory Affairs, Medical Devices Phone: 267-205-1273 Fax: 610-717-0668 E-mail: fred.cowdery@westpharma.com

SUBJECT DEVICE

Trade Name:MixJect® Transfer Device
Common/Usual Name:I.V. Fluid Transfer Set
Regulation Name:Intravascular Administration Set
Product Code:LHI
Regulation No.:880.5440
Class:II
Panel Identification:General Hospital Panel

PREDICATE DEVICE

Predicate DeviceMixJect Dispensing Pin / with Detachable Vial Holder / with Detachable VialHolder and Preattached Needle
510(k) NumberK001293
Common/Usual NameI.V. Fluid Transfer Set
Product CodeLHI

{5}------------------------------------------------

Regulation Name / Number / ClassIntravascular Administration Set / 21CFR880.5440 / Class II
PanelGeneral Hospital

INDICATION FOR USE

The device is intended for the transfer and injection of drugs contained in a vial.

DEVICE DESCRIPTION

Device Overview

The subject device, MixJect® Transfer Device (MXJ), is a single-use, gamma sterilized, nonpyrogenic device intended for the transfer and injection of drugs contained in a vial.

The MXJ device is intended for use by Healthcare Professionals (HCPs) in a clinical, hospital, or other healthcare environment. The subject device is available by prescription use only and has no known contraindications.

The MXJ device does not contain any medicinal substances and is only intended for use with drug vials having a neck diameter of 13mm. The subject device has a 3-year shelf life.

The MXJ device configuration consists of four main components inclusive of a 30-gauge needle, a vial adapter including integral cannulated spike, MXJ Body, and MXJ Core.

The subject device interfaces with a syringe (not supplied) that connects to the female Luer lock port located in the main body of the MXJ device.

The MXJ Vial Adapter component connects to a drug vial having a neck diameter of 13mm. The Vial Adapter component contains a piercing spike and a female Luer lock connector. Puncturing of the drug vial stopper membrane is achieved by means of an integral spike located in the center of the MXJ Vial Adapter component. Once the drug vial stopper membrane is breached by the cannulated spike, fluid can travel from the drug vial into the MXJ device main body.

A prefilled diluent syringe (not supplied) is then connected to MXJ female Luer lock port. The diluent is injected from the syringe into the drug vial. The reconstituted medicament is then aspirated back into the syringe. The MXJ device is then twisted in a counterclockwise direction, changing the fluid path from the syringe-vial to the syringe-needle. After reconstitution and aspiration, the drug is ready to be administered through the attached MXJ needle.

The MXJ primary device package consists of a polyethylene terephthalate glycol (PETG) blister sealed with a Tyvek® lid sealed on top of the blister pack.

In accordance with ISO 10993-1:2018, the subject device MixJect® Transfer Device is classified as an externally communicating device, having limited indirect contact with the patient blood path (< 24 hour). The subject device needle has transient contact (<60 minutes) with the patient body.

{6}------------------------------------------------

Principle of Operation

The MXJ is operated by a manual process. The subject device connects to a drug vial (supplied by the Drug Manufacturer) having a neck diameter of 13mm via the Vial Adapter portion of the MXJ device.

The vial adapter contains an integral plastic cannulated spike located in the center of the MXJ Vial Adapter component, intended to puncture the vial stopper membrane allowing access to the vial contents. This piercing cannulated spike then facilitates the transfer of the drug between the vial and the syringe.

A prefilled diluent syringe is then connected to the MXJ, interfacing to a female Luer lock port located on the MXJ body.

The diluent is then injected into the drug vial. The reconstituted medicament is then aspirated back into the syringe.

The MXJ body is then manually rotated in a counterclockwise direction, changing the fluid path from the syringe-vial to the syringe-needle.

After reconstitution and aspiration, the drug can be administered through the attached syringe / needle.

{7}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The MixJect® Transfer Device is substantially equivalent in its intended use, design/construction, technology/principle of operation, materials, and performance to the predicate device, MixJect Dispensing Pin / with Detachable Vial Holder / with Detachable Vial Holder and Preattached Needle, cleared under K001293.

A summary of equivalence and differences between the subject device and the predicate device are provided in the table below.

Areas forComparisonSubject DeviceMixJect® Transfer DevicePredicate Device(K001293):Comparison
General Information
ManufacturerWest Pharma. Services IL, Ltd.Medimop Medical Projects, Ltd./West Pharma. Services IL., Ltd.Different; Medimop acquired by West Pharmaceutical Services, Inc.
Device Trade NameMixJect® Transfer DeviceMixJect Dispensing Pin / with Detachable Vial Holder / with Detachable Vial Holder and Preattached NeedleSimilar: Changed for marketing purposes only. Change does not impact device intended use.
Catalog Number970133800405Similar: Change does not impact device intended use.
Indications for UseThe MixJect® Transfer Device is intended for the transfer and injection of drugs contained in a vial.Transfer, mixing, and injection of drugs contained in a vial.Similar – Differences do not raise new or additional concerns regarding device safety profile or clinical effectiveness.
ContraindicationsNone knownNone knownIdentical
Intended UserPopulationIntended for use by Healthcare Professionals (HCPs)Intended for use by Healthcare Professionals (HCPs)Identical
Intended UseEnvironmentIntended for use in hospitals, outpatient nursing units and other suitable clinical environmentsIntended for use in hospitals, outpatient nursing units and other suitable clinical environmentsIdentical
Device Class &Classification NameClass II, I.V. Fluid Transfer SetClass II, I.V. Fluid Transfer SetIdentical
Regulation Number / Name21CFR 880.5440Intravascular Administration Set21CFR 880.5440Intravascular Administration SetIdentical
Areas forComparisonSubject DeviceMixJect® Transfer DevicePredicate Device(K001293):MixJect Dispensing Pin /with Detachable Vial Holder/ with Detachable VialHolder and PreattachedNeedleComparison
Product CodeLHIrhiIdentical
Prescription UseYesYesIdentical
Single UseYesYesIdentical
Shelf life3 years3 yearsIdentical
Design
Operation PrincipleManualManualIdentical
Design/constructionFeaturing a 13mm Vial Adaptorbody with tight grip hold("wings"), intended to beattached to a standard drug vialwith a neck diameter of 13mm.The device contains a piercingspike, a 30Ga Needle and afemale Luer lock fittingcompliant with BS EN ISO80369-7:2016, for attachment toa standard accessory such as asyringe (not supplied).Featuring a 13mm VialAdaptor body with tight griphold ("wings"), intended tobe attached to a standard drugvial with a neck diameter of13mm. The device containsa piercing spike, a 30GaNeedle and a female Luerlock fitting compliant withISO 594-1 and ISO 594-2,for attachment to a standardaccessory such as a syringe(not supplied).Similar - Differencesnoted with ISO LuerStandard compliance donot alter device intendeduse, clinical effectiveness,or safety profile.
Female Luer LockConnectorCompliant withBS EN ISO 80369-7:2016Compliant withISO 594-1 and ISO 594-2Similar – Subject deviceupdated for compliancewith latest ISO Luerstandard. No change inintended use.
Compatible VialSize13mm13mmIdentical
Body Diameter18.5mm to accommodate 13mmstandard vials.18.5mm to accommodate13mm standard vials.Identical
Piercing SpikeSingle lumenSingle lumenIdentical
Vial Adapter FitVial first, snap fit to vialVial first, snap fit to vialIdentical
MaterialMixJect Body: PolycarbonateMixJect Core: High DensityPolyethyleneMixJect Vial Adapter Body:PolycarbonateMixJect Body: PolycarbonateMixJect Core: High DensityPolyethyleneMixJect Vial Adapter Body:PolycarbonateIdentical
Areas forComparisonSubject DeviceMixJect® Transfer DevicePredicate Device(K001293):Comparison
MixJect Dispensing Pin /with Detachable Vial Holder/ with Detachable VialHolder and PreattachedNeedle
BiocompatibilityISO 10993-1:2018External Communicating,Transient Contact (<60 min)ISO 10993-1 and USPBiological ReactivityExternal Communicating,Transient Contact (<60 min)Identical
Non-pyrogenicYesYesIdentical
Sterilization
SterilitySterileSterileIdentical
Sterilization MethodGammaGammaIdentical
Sterility AssuranceLevelSAL of 10-6SAL of 10-6Identical
Packaging
PackagingSterile Barrier packagematerials: PETG blister withTyvek® sealSterile Barrier packagematerials: PETG blister withTyvek® sealIdentical
Sterile Barrier packageorientation: Devices are suppliedin an individual blister, vial firstorientation.Sterile Barrier packageorientation: Devices aresupplied in an individualblister, vial first orientation.

Substantial Equivalence Comparison Table

{8}------------------------------------------------

{9}------------------------------------------------

PERFORMANCE DATA

The following non-clinical performance data were provided in support of the substantial equivalence determination.

Performance Testing

Performance testing was conducted to confirm the MixJect® Transfer Device meets all applicable design and performance requirements throughout its defined shelf life and verify conformity to the applicable external and internal standards and demonstrate substantial equivalence to the predicate device. The table below provides a list of non-clinical bench performance tests that were completed on the device and provided within this submission.

{10}------------------------------------------------

TestTest Method/ Standard
Fragmentation TestISO 8536-2:2010 section 6.2.2
Particulate TestingUSP <788>
Internal Diameter Upper SkirtISO 8362-6:2010 Section 4.2
Luer Gauging TestISO 594-1:1986 and ISO 594-2:1998
Luer Stability and compliance to ISO 80369-7ISO 80369-7:2021
Luer Stability and compliance to ISO 80369-7ISO 80369-20:2015, Annex B & Annex C forthe leakage reference connector (fluidleakage)
Luer Stability and compliance to ISO 80369-7ISO 80369-20:2015, Annex D & Annex C forthe leakage reference connector (air leakage)
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex E & Annex C forthe stress cracking reference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex F & Annex C forthe axial load reference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex G & Annex C forthe resistance separation from unscrewingreference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex G & Annex C forthe overriding reference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-7:2021 Table B.2 and B.5(compliance to dimensions)
Residual VolumeIn-house test method
Device LeakageIn-house test method
Device Total Penetration ForceIn-house test method
Vial Adapter Detachment ForceIn-house test method
Product Retention in BlisterIn-house test method
Flow RateIn-house test method
Device Removal Force from BlisterIn-house test method
Tyvek Total Peel Test ForceIn-house test method
TestTest Method/ Standard
Internal Diameter Dimensional Measurements Upper SkirtIn-house test method
Functionality according to IFUIn-house test method
Injection ForceIn-house test method
Aspiration ForceIn-house test method
Label LegibilityIn-house test method
Impact ForceIn-house test method
Needle Protective Cap Removal ForceIn-house test method
Torque TestIn-house test method

Summary of Performance Testing

{11}------------------------------------------------

Performance testing and risk management review indicate all product design requirements are verified and the residual risk level is acceptable based on the test results. Together, objective evidence satisfies the product requirements for performance, safety and effectiveness and the results support a determination of substantial equivalence to the predicate device.

Biocompatibility Testing

The biocompatibility evaluation for the MixJect® Transfer Device was conducted in accordance with, 2016 FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process. and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

ISO 10993-5:2009: Cytotoxicity ISO 10993-4: 2017: ASTM Hemolysis ASTM F756: Hemolysis Study ISO 10993-10:2010: Maximization and Sensitization ISO 10993-10:2010: Intracutaneous Reactivity ISO 10993-11:2017: Acute Systemic Toxicity ISO 10993-11:2017: Material Mediated Pyrogenicity

{12}------------------------------------------------

Sterilization

The subject device is terminally sterilized using a Gamma irradiation sterilization method, validated in accordance with standard BS EN ISO 11137-1:2015 & A2:2019 Sterilization of health care products – Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and BS EN ISO 11137-2:2015 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, and ISO 13004 -Sterilization of health care products – Radiation – Substantiation of a selected sterilization dose: Method VDmaxD. The sterilization method of Gamma irradiation provides a sterility assurance level (SAL) of 10-6.

Bacterial Endotoxin Testing by limulus amebocyte lysate (LAL) was also performed on the same batch of product used for sterility dose verification, which passed with acceptable levels, further ensuring the safety of the device. The Sterility Validation and Bacterial Endotoxin Testing are provided within this submission.

CLINICAL DATA

Clinical trials were not performed for the MixJect® Transfer Device.

CONCLUSION

The differences between the predicate and the subject devices do not raise any new or different questions of safety or effectiveness. The subject device, MixJect® Transfer Device is substantially equivalent to the predicate device, MixJect Dispensing Pin / with Detachable Vial Holder / with Detachable Vial Holder and Preattached Needle.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.