(27 days)
Not Found
No
The summary describes a system for performing calculations on X-ray angiographic images, primarily focusing on hardware changes and leveraging testing from a predicate device. There is no mention of AI, ML, or related concepts in the provided text.
No
The device is a diagnostic tool that provides quantified results of coronary vessels from X-ray images; it does not directly treat or alleviate a medical condition.
Yes
The device supports the assessment of coronary vessels in X-ray angiographic images, providing quantified results intended for clinical use by responsible clinicians, which indicates its role in aiding in the diagnosis or prognosis of coronary artery disease.
No
The device description explicitly states that AccuFFRangio Plus includes both hardware and software, and the performance studies section details hardware testing.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: AccuFFRangio Plus processes and analyzes X-ray angiographic images of coronary vessels. This is an in vivo (within the living body) imaging modality, not an in vitro (in glass/outside the body) test.
- Intended Use: The intended use is to support the assessment of coronary vessels based on image analysis, not on the analysis of biological samples.
Therefore, AccuFFRangio Plus falls under the category of a medical device that performs image processing and analysis, but it is not an IVD.
N/A
Intended Use / Indications for Use
AccuFFRangio Plus is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by AccuFFRangio Plus are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
Product codes
QHA, LLZ
Device Description
AccuFFRangio Plus is a system that is used to perform calculations in X-ray angiographic images of the coronary arteries. It includes hardware and software (AccuFFRangio) and the hardware of the device which mainly has a display function and provide the software an operation environment. AccuFFRangio Plus is changed from our own legally marketed predicate device AccuFFRangio that is a stand-alone software package. Therefore, the significant change lies in equipping a computer system to the software on a particular mobile cart.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray angiographic images
Anatomical Site
coronary vessels / coronary arteries
Indicated Patient Age Range
Not Found
Intended User / Care Setting
clinical settings, responsible clinicians
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The hardware modifications to the AccuFFRangio Plus were implemented under the design controls that are compliant with 21 CFR 820.30. A risk analysis was conducted in accordance with ISO 14971:2019 Medical devices - Application of risk management to medical devices to assess the risks and risk mitigations for the device hardware modifications. Based on this risk assessment, the following verification tests were identified and conducted. All tests met the pre-defined acceptance criteria and were passed.
Electrical Safety Testing: The AccuFFRangio Plus was evaluated and found to be in compliance with the applicable requirements of IEC 60601-1:2012, "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance." All emissions and immunity tests were passed.
Electromagnetic Compatibility Testing: The AccuFFRangio Plus was tested and found to be in compliance with the applicable requirements of IEC 60601-1-2:2014 (4th edition), "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests."
Hardware Verification Testing: The AccuFFRangio Plus underwent testing per internal specifications through incoming inspections of raw materials and final inspections of finished devices. All hardware requirements of the system were evaluated/tested and found to meet the pre-defined acceptance criteria.
Transportation Testing: The AccuFFRangio Plus was shipped in a padded, wooden box. Transportation testing was conducted in accordance with ASTM D4169-16, "Standard Practice for Performance Testing of Shipping Containers and Systems." All tests were passed.
Human Factors Testing: Usability testing of the modified AccuFFRangio Plus and its operator manual was conducted in accordance with IEC 62366-1:2015, "Application of usability engineering to medical devices". Fifteen qualified participants performed all critical tasks necessary to validate the usability of AccuFFRangio Plus and user manual without any use errors. The conclusion of the testing was that the AccuFFRangio Plus can be used safely and effectively by the intended user population. No residual use-related risks were identified.
Accelerated Aging Testing: Accelerated aging testing was conducted in accordance with IEC 62506:2013, "Methods for product accelerated testing". All pre-defined acceptance criteria were met and the service life of the device is validated to be 5 years.
Labeling Inspection: Labeling inspection was conducted to in accordance with documentation of the company's quality management system. All inspections were passed.
No clinical testing was necessary to support the device modifications described in this Special 510(k).
The following testing was leveraged from the predicate device. The results from the predicate were used to support the subject device the software contained in the AccuFFRangio Plus is the same as the predicate device.
Software verification and validation: Software verification testing in accordance with the design requirements to ensure that the software requirements were met. Software validation test to ensure the device meet user needs and perform as intended. Applicable standard(s): FDA Guidance, "Guidance for the Content of Premarket Submissions for software Contained in Medical Device".
Cybersecurity: Testing to verify Cybersecurity control and management. Applicable standard(s): Cybersecurity as recommended in FDA guidance, “Content of Premarket Submission for Management of Cybersecurity in Medical Device”.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1600 Angiographic x-ray system.
(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
March 2, 2023
ArteryFlow Technology Co., Ltd. % Ashley Fu RA Specialist 459 Oianmo Road, Suite C1-501, Binjiang District Hangzhou, ZHEJIANG 310051 CHINA
Re: K230303
Trade/Device Name: AccuFFRangio Plus Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: Class II Product Code: QHA, LLZ Dated: February 3, 2023 Received: February 3, 2023
Dear Ashley Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K230303
Device Name AccuFFRangio Plus
Indications for Use (Describe)
AccuFFRangio Plus is indicated for use in clinical settings where validated and repreproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by AccuFFRangio Plus are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Tab #06 Special 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
1. Submitter's Information
Submitter: ArteryFlow Technology Co., Ltd.
Address: 459 Qianmo Road, Suite C1-501, Binjiang District, 310051 Hangzhou City, Zhejiang Province, China
Phone Number: +86-571-86772567
Primary correspondent: XIANG Jianping, PhD, General Manager
Email: jianping.xiang@arteryflow.com
Secondary correspondent: Ashley Fu, RA Specialist
Email: fang.fu@arteryflow.com
Date of preparation: February 03, 2023
2. Device Information
Trade/ Device Name: AccuFFRangio Plus
Common Name: Radiological Image Processing System
Regulatory Class: Class II
Regulation Description: Angiographic X-ray system
Regulation number: 892.1600
Classification Product Code: QHA
Subsequent Product Code: LLZ
3. Predicate Device Information
Manufacturer: ArteryFlow Technology Co., Ltd. Device Name: AccuFFRangio Common Name: Radiological Image Processing Software Regulatory Class: Class II Regulation Number: 892.1600 Classification Product Code: QHA
Subsequent Product Code: LLZ
510(k) number: K210093
4
4. Device Description
AccuFFRangio Plus is a system that is used to perform calculations in X-ray angiographic images of the coronary arteries. It includes hardware and software (AccuFFRangio) and the hardware of the device which mainly has a display function and provide the software an operation environment. AccuFFRangio Plus is changed from our own legally marketed predicate device AccuFFRangio that is a stand-alone software package. Therefore, the significant change lies in equipping a computer system to the software on a particular mobile cart.
5. Intended Use
AccuFFRangio Plus is a system intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. AccuFFRangio Plus enables interventional cardiologists and researchers to obtain quantifications of one or more lesions in the analyzed coronary vessel seqment. In particular, AccuFFRanqio Plus provides:
- Quantitative results of coronary vessel segments based on a 3D reconstructed model:
- Dimensions of the cardiovascular vessels and lesions;
- Quantification of the pressure drop in coronary vessels.
6. Indications for Use
AccuFFRangio Plus is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by AccuFFRangio Plus are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
7. Technological Characteristic Comparison
Compared to the predicate device AccuFFRangio, intended use and indications for use of subject device remain unchanged. The software installed in the subject device is the same as the predicate device. The features and technology of the software part are the same as the predicate device.
The difference lies in the addition of the hardware part. AccuFFRangio Plus has changed to an integrity with software and hardware. The hardware mainly equips a computer system on a mobile cart. The detailed comparisons of differences are provided in the following table:
Subject Device | Predicate Device | |
---|---|---|
Feature | AccuFFRangio Plus | |
(Hardware with software) | AccuFFRangio | |
(Stand-alone | ||
software) |
5
| Hardware | Monitor | The monitor converts signal
outputted from host
computer to visible image,
and user can operate based
on the information
displayed. | N/A |
|----------|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| | Mouse | Mouse is the input
component, user can locate
region of interest with the
mouse, buttons, and roller. | N/A |
| | Keyboard | User can input data through
keyboard. | N/A |
| | Host computer | The predicate software is
installed within it. | N/A |
| | Monitor mount | Adjust the height of the
monitor when necessary. | N/A |
| | Equipment shell | It is equipped with power
switch, USB port, cable
post, heat vent,
equipotential terminal and
power socket and switch. | N/A |
| | Casters | Four casters with brake
make the product mobile,
improving the flexibility and
safety. | N/A |
8. Performance Data to Support Substantial Equivalence
The hardware modifications to the AccuFFRangio Plus was implemented under the design controls that are compliant with 21 CFR 820.30. A risk analysis was conducted in accordance with ISO 14971:2019 Medical devices - Application of risk management to medical devices to assess the risks and risk mitigations for the device hardware modifications. Based on this risk assessment, the following verification tests were identified and conducted. All tests met the pre-defined acceptance criteria and were passed.
Electrical Safety Testing: The AccuFFRangio Plus was evaluated and found to be in compliance with the applicable requirements of IEC 60601-1:2012, "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance." All emissions and immunity tests were passed.
6
Electromagnetic Compatibility Testing: The AccuFFRangio Plus was tested and found to be in compliance with the applicable requirements of IEC 60601-1-2:2014 (4th edition), "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests."
Hardware Verification Testing: The AccuFFRangio Plus underwent testing per internal specifications through incoming inspections of raw materials and final inspections of finished devices. All hardware requirements of the system were evaluated/tested and found to meet the pre-defined acceptance criteria.
Transportation Testing: The AccuFFRangio Plus was shipped in a padded, wooden box. Transportation testing was conducted in accordance with ASTM D4169-16, "Standard Practice for Performance Testing of Shipping Containers and Systems." All tests were passed.
Human Factors Testing: Usability testing of the modified AccuFFRangio Plus and its operator manual was conducted in accordance with IEC 62366-1:2015, "Application of usability engineering to medical devices". Fifteen qualified participants performed all critical tasks necessary to validate the usability of AccuFFRangio Plus and user manual without any use errors. The conclusion of the testing was that the AccuFFRangio Plus can be used safely and effectively by the intended user population. No residual use-related risks were identified.
Accelerated Aging Testing: Accelerated aging testing was conducted in accordance with IEC 62506:2013, "Methods for product accelerated testing". All pre-defined acceptance criteria were met and the service life of the device is validated to be 5 years.
Labeling Inspection: Labeling inspection was conducted to in accordance with documentation of the company's quality management system. All inspections were passed.
No clinical testing was necessary to support the device modifications described in this Special 510(k).
The following testing was leveraged from the predicate device. The results from the predicate were used to support the subject device the software contained in the AccuFFRangio Plus is the same as the predicate device.
Test | Test description | Applicable standard(s) |
---|---|---|
Software verification | ||
and validation | Software verification testing in | |
accordance with the design | ||
requirements to ensure that the | ||
software requirements were met. | ||
Software validation test to ensure | ||
the device meet user needs and | ||
perform as intended. | FDA Guidance, "Guidance for | |
the Content of Premarket | ||
Submissions for software | ||
Contained in Medical Device" | ||
Cybersecurity | Testing to verify Cybersecurity | |
control and management. | Cybersecurity as | |
recommended in FDA |
7
| | | guidance, “Content of
Premarket Submission for
Management of
Cybersecurity in Medical
Device” |
-- | -- | ----------------------------------------------------------------------------------------------------------- |
---|
9. Conclusion
AccuFFRangio Plus shares the same intended use, and indications for use with AccuFFRangio. Besides, the software application of the device is totally the same as the predicate device. The information and testing presented demonstrate that AccuFFRangio Plus is clinically feasible, and the performance is substantially equivalent to the original ArteryFlow AccuFFRangio cleared device under K210093.