K Number
K210093
Device Name
AccuFFRangio
Date Cleared
2021-09-10

(239 days)

Product Code
Regulation Number
892.1600
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
AccuFFRangio is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease. When the quantified results provided by AccuFFRangio are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
Device Description
ArteryFlow® AccuFFRangio is designed as a stand-alone software package to run on a PC. This software can read traditional x-ray angiographic images with DICOM format from the local file directory. The AccuFFRangio is composed of the following analysis workflows: Image Loading, Frame Selection, Vessel Reconstruction, QCA Vessel Quantification, and AccuFFRangio Calculation for visualization of the target coronary segment, quantification of the stenosis and pressure drop of the coronary seqment. The AccuFFRangio parameter is only for quantitative imaging output but not for diagnosis and the AccuFFRanigo product has a moderate level of concern. The user can calculate the pressure drop and AccuFFRangio (FFR) value for the coronary vessel. To obtain these values for a specific lesion in a coronary vessel, the user has to start with Frame Selection using two angiographic images from different views. In each of these images, a classic 2D coronary contour detection is performed, after which a reconstruction of the coronary segment is obtained in 3D space. Based on the 3D reconstruction and user input of the aortic pressure, the pressure drop and AccuFFRangio value can be calculated. AccuFFRangio enables interventional cardiologists to obtain accurate anatomical quantifications of one or more lesions in the analyzed coronary segment, and to assess the best viewing angles which can be helpful for optimal visualization of the lesion during percutaneous coronary intervention (PCI) treatment. Results can be displayed and generated by the software, which contains patient information, imaging of actual and reference vessel boundaries, dimensions of the vessel sizing, pressure drop, and AccuFFRangio value. The results can be export in PDF format. This functionality is independent of the type of vendor acquisition equipment.
More Information

Not Found

No
The description details traditional image processing techniques (2D contour detection, 3D reconstruction) and calculations based on these reconstructions and user input. There is no mention of AI, ML, or related concepts in the device description or performance studies.

No

The device is indicated for quantify assessment of coronary vessels and pressure drop for diagnosis, but not for treatment.

No
Explanation: The "Intended Use / Indications for Use" states that the device provides quantified results to support the assessment of coronary vessels and that "the results are only intended for use by the responsible clinicians." Additionally, the "Device Description" explicitly states that "The AccuFFRangio parameter is only for quantitative imaging output but not for diagnosis".

Yes

The device is explicitly described as a "stand-alone software package to run on a PC" and its function is to process existing X-ray angiographic images. There is no mention of accompanying hardware components required for its operation or intended use beyond the standard PC.

Based on the provided information, it is unlikely that this device is an IVD (In Vitro Diagnostic). Here's why:

  • Intended Use: The intended use is to "support the assessment of coronary vessels in X-ray angiographic images" and provide "quantified results" for "individual patients with coronary artery disease." While these results are used in a clinical setting, they are derived from imaging data (X-ray angiograms), not from the analysis of biological specimens (like blood, urine, tissue, etc.).
  • Device Description: The device is described as a "stand-alone software package" that reads "traditional x-ray angiographic images." It performs "Image Loading, Frame Selection, Vessel Reconstruction, QCA Vessel Quantification, and AccuFFRangio Calculation." These are all processes applied to medical images.
  • Input Imaging Modality: The input is explicitly stated as "X-ray angiographic images."
  • Anatomical Site: The analysis is performed on "coronary vessels," which are structures within the body, visualized through imaging.
  • Lack of Biological Specimen Analysis: There is no mention of the device analyzing any biological samples or specimens from the patient. IVDs are defined by their use in examining specimens derived from the human body to provide information for diagnosis, monitoring, or screening.

In summary, AccuFFRangio is a software device that processes medical images to provide quantitative information about coronary vessels. This falls under the category of medical image analysis software, not In Vitro Diagnostics.

The fact that it calculates parameters like "pressure drop" and "AccuFFRangio (FFR) value" based on the image analysis and user input (aortic pressure) further reinforces that it's an image-based analysis tool, not an IVD. The results are intended to aid clinicians in interpreting the imaging data and making decisions, but they are not derived from the direct analysis of biological material.

N/A

Intended Use / Indications for Use

AccuFFRangio is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.

When the quantified results provided by AccuFFRangio are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.

Product codes (comma separated list FDA assigned to the subject device)

QHA, LLZ

Device Description

ArteryFlow® AccuFFRangio is designed as a stand-alone software package to run on a PC. This software can read traditional x-ray angiographic images with DICOM format from the local file directory.

The AccuFFRangio is composed of the following analysis workflows: Image Loading, Frame Selection, Vessel Reconstruction, QCA Vessel Quantification, and AccuFFRangio Calculation for visualization of the target coronary segment, quantification of the stenosis and pressure drop of the coronary segment. The AccuFFRangio parameter is only for quantitative imaging output but not for diagnosis and the AccuFFRanigo product has a moderate level of concern.

The user can calculate the pressure drop and AccuFFRangio (FFR) value for the coronary vessel. To obtain these values for a specific lesion in a coronary vessel, the user has to start with Frame Selection using two angiographic images from different views. In each of these images, a classic 2D coronary contour detection is performed, after which a reconstruction of the coronary segment is obtained in 3D space. Based on the 3D reconstruction and user input of the aortic pressure, the pressure drop and AccuFFRangio value can be calculated.

AccuFFRangio enables interventional cardiologists to obtain accurate anatomical quantifications of one or more lesions in the analyzed coronary segment, and to assess the best viewing angles which can be helpful for optimal visualization of the lesion during percutaneous coronary intervention (PCI) treatment.

Results can be displayed and generated by the software, which contains patient information, imaging of actual and reference vessel boundaries, dimensions of the vessel sizing, pressure drop, and AccuFFRangio value. The results can be export in PDF format. This functionality is independent of the type of vendor acquisition equipment.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray angiographic images

Anatomical Site

Coronary vessels / coronary arteries

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Interventional cardiologists in clinical settings.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A phantom study was implemented using three different types of brass models with stenosis to evaluate the accuracy of the 3D quantitative coronary angiography (QCA) of AccuFFRangio software. The software's technical parameters validated include: lesion length accuracy, diameter stenosis accuracy, area stenosis accuracy, minimal lumen diameter accuracy, and reference diameter accuracy.

For AccuFFRangio calculation, a series of X-ray angiographic datasets with known pressure drops were analyzed to confirm the performance of pressure drop and AccuFFRangio (FFR) calculation in the coronary vessel.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Study Type: Phantom study and analysis of X-ray angiographic datasets.

Sample Size: Not specified.

Standalone Performance:

  • 3D vessel reconstruction: The phantom test demonstrated that the quantitative coronary angiography for coronary vessel has similar performance compared to the predicate device QAngio XA 3D (K182611).
  • AccuFFRangio calculation: The results demonstrated that the quantification of AccuFFRangio in coronary vessel has similar performance compared to the QFR by the predicate device QAngio XA 3D (K182611).

Key Results:

  • Accuracy was evaluated for 3D QCA.
  • Accuracy, sensitivity, specificity, positive predictive value, and negative predictive value for per-vessel were calculated for AccuFFRangio calculation.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Accuracy, sensitivity, specificity, positive predictive value, and negative predictive value for per-vessel.

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K182611

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.1600 Angiographic x-ray system.

(a)
Identification. An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

September 10, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

ArteryFlow Technology Co., Ltd. % Ashley Fu International Registeration Engineer 459 Oianmo Road, Suite C1-501, Binjiang District Hangzhou. Zhejiang 310051 CHINA

Re: K210093

Trade/Device Name: AccuFFFRangio Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: Class II Product Code: QHA, LLZ Dated: August 12, 2021 Received: August 16, 2021

Dear Ashley Fu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

, for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K210093

Device Name

AccuFFRangio

Indications for Use (Describe)

AccuFFRangio is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.

When the quantified results provided by AccuFFRangio are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Tab #06 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

1. Submitter's Information

Submitter: ArteryFlow Technology Co., Ltd.

Address: 459 Qianmo Road, Suite C1-501, Binjiang District, 310051 Hangzhou City, Zhejiang Province, China

Phone Number: +86-571-86772567

Primary correspondent: XIANG Jianping, PhD, General Manager

Email: jianping.xiang@arteryflow.com

Secondary correspondent: Ashley Fu, Registration engineer

Email: fang.fu@arteryflow.com

Date of preparation: January 05, 2021

2. Device Information

Trade/ Device Name: AccuFFRangio

Common Name: Radiological Image Processing Software

Regulatory Class: Class II

Regulation Description: Angiographic X-ray system

Regulation number: 892.1600

Classification Product Code: QHA

Subsequent Product Code: LLZ

3. Predicate Device Information

Manufacturer: Medis medical imaging systems bv

Device Name: QANGIO XA 3D

Regulatory Class: Class II

Regulation Number: 892.1600

Classification Product Code: QHA

Subsequent Product Code: LLZ

510(k) number: K182611

4. Device Description

4

ArteryFlow® AccuFFRangio is designed as a stand-alone software package to run on a PC. This software can read traditional x-ray angiographic images with DICOM format from the local file directory.

The AccuFFRangio is composed of the following analysis workflows: Image Loading, Frame Selection, Vessel Reconstruction, QCA Vessel Quantification, and AccuFFRangio Calculation for visualization of the target coronary segment, quantification of the stenosis and pressure drop of the coronary seqment. The AccuFFRangio parameter is only for quantitative imaging output but not for diagnosis and the AccuFFRanigo product has a moderate level of concern.

The user can calculate the pressure drop and AccuFFRangio (FFR) value for the coronary vessel. To obtain these values for a specific lesion in a coronary vessel, the user has to start with Frame Selection using two angiographic images from different views. In each of these images, a classic 2D coronary contour detection is performed, after which a reconstruction of the coronary segment is obtained in 3D space. Based on the 3D reconstruction and user input of the aortic pressure, the pressure drop and AccuFFRangio value can be calculated.

AccuFFRangio enables interventional cardiologists to obtain accurate anatomical quantifications of one or more lesions in the analyzed coronary segment, and to assess the best viewing angles which can be helpful for optimal visualization of the lesion during percutaneous coronary intervention (PCI) treatment.

Results can be displayed and generated by the software, which contains patient information, imaging of actual and reference vessel boundaries, dimensions of the vessel sizing, pressure drop, and AccuFFRangio value. The results can be export in PDF format. This functionality is independent of the type of vendor acquisition equipment.

5. Intended Use

AccuFFRangio is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. AccuFFRangio enables interventional cardiologists to obtain quantifications of one or more lesions in the analyzed coronary vessel segment. In particular, AccuFFRangio provides:

  • Quantitative results of coronary vessel segments based on a 3D reconstructed model;
  • Dimensions of the cardiovascular vessels and lesions;
  • Quantification of the pressure drop in coronary vessels.

6. Indication for Use

AccuFFRangio is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.

When the quantified results provided by AccuFFRangio are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.

5

7. Technological Characteristic Comparison

A comparison of the technological characteristics of the predicate and subject device is given in the table below.

ItemNew devicePredicate device
Device nameAccuFFRangioQANGIO XA 3D
ManufacturerArteryFlow Technology Co., Ltd.Medis medical imaging systems bv
510(k) No.-K182611
Product CodeQHA, LLZQHA, LLZ
Regulation No.892.1600892.1600
ClassIIII
Level of Concern
of the softwareModerateModerate
Intended Use and Indications for Use
Intended UseAccuFFRangio is software
intended to be used for performing
calculations in X-ray angiographic
images of the coronary arteries.
AccuFFRangio enables
interventional cardiologists to
obtain quantifications of one or
more lesions in the analyzed
coronary vessel segment. In
particular, AccuFFRangio
provides:
Quantitative results of
coronary vessel segments
based on a 3D reconstructed
model;Dimensions of the
cardiovascular vessels and
lesions;Quantification of the pressure
drop in coronary vessels.QANGIO XA 3D is software
intended to be used for performing
calculations in X-ray angiographic
images of the coronary arteries.
QANGIO XA 3D enables
interventional cardiologists and
researchers to obtain
quantifications of one or more
lesions in the analyzed coronary
vessel segment. In particular,
QANGIO XA 3D provides:
Quantitative results of
coronary vessel segments
based on a 3D reconstructed
model;Dimensions of the
cardiovascular vessels and
lesions;Quantification of the pressure
drop in coronary vessels.
Indication for useAccuFFRangio is indicated for use
in clinical settings where validated
and reproducible quantified results
are needed to support the
assessment of coronary vessels in
X-ray angiographic images, for use
on individual patients with
coronary artery diseaseQANGIO XA 3D is indicated for
use in clinical settings where
validated and reproducible
quantified results are needed to
support the assessment of
coronary vessels in X-ray
angiographic images, for use on
individual patients with coronary
When the quantified results
provided by AccuFFRangio are
used in a clinical setting on X-ray
images of an individual patient, the
results are only intended for use by
the responsible clinicians.artery disease.
When the quantified results
provided by QANGIO XA 3D are
used in a clinical setting on X-ray
images of an individual patient, the
results are only intended for use by
the responsible clinicians.
Technological Characteristics
Data type● X-ray angiographic data in
DICOM format
(vendor-independent)● X-ray angiographic data in
DICOM format
(vendor-independent)
Import of patient
data● Automatically load information
from patient's DICOM file;
● Manual through keyboard● Automatically load information
from patient's DICOM file;
● Manual through keyboard
Image display● 2D X-ray image visualization
with embedded ECG signal;
● 3D reconstruction of the
vessel segment based on 2
X-ray images;
● Graph for actual and
reference vessel diameter
with vessel longitudinal
position;
● Graph for AccuFFRangio
(FFR) value with vessel
longitudinal position;
● AccuFFRangio (FFR) value is
colored on the 3D
reconstructed vessel segment
model.● 2D X-ray image visualization
with embedded ECG signal;
● 3D reconstruction of the
vessel segment based on 2
X-ray images;
● Graph for actual and
reference vessel diameter
with vessel longitudinal
position;
● Graph for QFR (FFR) value
with vessel longitudinal
position;
● QFR (FFR) value is colored
on the 3D reconstructed
vessel segment model.
Centerline and
contour definition● Manual and semi-automatic
centerline definition based
contour detection of coronary
vessel;
● Contour correction and
restriction● Manual and semi-automatic
centerline definition based
contour detection of coronary
and peripheral vessel;
● Contour correction and
restriction
Image assessment● Manual and automatic
calibration;
● Vessel dimensions
(diameters, areas, lengths);
● Automatic stenosis
assessment;
● Pressure drop calculation in
coronary vessel● Manual and automatic
calibration;
● Vessel dimensions
(diameters, areas, lengths);
● Automatic stenosis
assessment;
● Pressure drop calculation in
coronary vessel
FFR calculationAccuFFRangio provides a
quantification parameter called
AccuFFRangio.QANGIO XA 3D provides a
quantification parameter called
QFR.
Storage of results● Printout
● Images
● PDF● Printout
● Images
● PDF
Different technological Characteristics
Gap 1With ECG display in the image
paneECG display in a separate pane
Gap 2The vessel is colored by stenosis
percentage in red to blueThe vessel is colored by stenosis
percentage in red to white
Gap 3When visualizing the reconstructed
vessel segment, there is a body
icon to show the relative position
for the vessel segment.In QAngio XA 3D, there is a box
icon to show the relative position
for the vessel segment.
Image: vessel segmentImage: vessel segment
Gap 4In the calculated vessel FFR
distribution, users can easily point
the position by the curser and the
specific AccuFFRangio value will
show up.In QAngio XA 3D, the user needs
to move four "checking disk" and
look at the FFR value in the
left-bottom window.
Gap 5Software operation environment:
● Microsoft Windows 7, SP1,
32 and 64-bit version
● Microsoft Windows 8.1, 32
and 64-bit version
● Microsoft Windows 10, 32
and 64-bit version
● Linux Ubuntu16.04 and
18.04, 32 and 64-bit versionSoftware operation environment:
● Microsoft Windows 7, SP1,
32 and 64-bit version
● Microsoft Windows 8.1, 32
and 64-bit version
● Microsoft Windows 10, 32
and 64-bit version
● Microsoft Windows Server
2008 R2, SP1, 64-bit version
● Microsoft Windows Server
2012 R2, 64-bit version
● Microsoft Windows Server
2016, 64-bit version

Table 1 General Comparison

6

7

AccuFFRangio has the same intended use and indications for use as the predicate devices QANGIO XA 3D (K182611). The basic features in technological characteristics are the same, such as same types of data format, same method for importing patient data, similar

8

display of image, same definition for centerline and contour, same method for Image assessment, same output for 3D reconstruction and the calculation of the pressure drop, etc.

AccuFFRanqio is a quantitative imaging output device, like the QANGIO XA 3D. Though above five differences concerning the display pattern (gap 1), display interface (gap 2-3), specific operate method on showing result (gap 4) and software operating system (gap 5) do exist, they do not affect the function and safety of AccuFFRangio.

8. Performance Data

System requirements – derived from the intended use and indications for use – as well as risk control measures are verified by system testing. All requirements are tested and all results of the tests performed are summarized in the software test statement and especially the requirements coverage matrix of AccuFFRangio providing traceability between requirements and the tests successfully executed.

The main functionality (including 3D vessel reconstruction, the calculation of the pressure drop and AccuFFRangio) have all been validated:

  • 3D vessel reconstruction: A phantom study had been implemented by using three different types stenosis of brass model. This phantom test was to evaluate the accuracy of the 3D quantitative coronary angiography (QCA) of AccuFFRangio software. The following software technical parameters were validated, namely, lesion length accuracy, diameter stenosis accuracy, area stenosis accuracy, minimal lumen diameter accuracy and reference diameter accuracy. This demonstrated that the quantitative coronary angiography for coronary vessel have similar performance compare to the predicate device QAngio XA 3D (K182611).
  • AccuFFRanqio calculation: To confirm the performance of the pressure drop and AccuFFRangio (FFR) calculation in the coronary vessel, a series of X-ray angiographic dataset with known pressure drops were analyzed. The functionality of pressure drop and AccuFFRangio calculation for the new device is based on 3D reconstruction out of two angiographic images. The results, including accuracy, sensitivity, specificity, positive predictive value and negative predictive value for per-vessel were calculated. The results demonstrated that the quantification of AccuFFRangio in coronary vessel have similar performance compare to the QFR by the predicate device QAngio XA 3D (K182611).

9. Conclusion

AccuFFRangio product manufactured by our company shares the same intended use, and indications for use with QAnqio XA 3D by Medis.

Testing, validation and verification have produced results consistent with design input requirements. AccuFFRangio is a software-only device for which there are no applicable mandatory performance standards.

During the development, potential hazards were controlled by a risk management report, including risk analysis, risk mitigation, verification and validation.

ArteryFlow technology concludes that the AccuFFRangio is only for quantitative imaging

9

output but not for diagnosis, and it is a safe and effective medical device as its predicate device. The use of AccuFFRangio does not change the intended use of X-ray angiography acquisition, nor does the use of this software results in any new potential hazards.