K Number
K230103
Manufacturer
Date Cleared
2023-04-24

(101 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Klockner Abutments are intended for use in dental implants to provide a support structure so that the edentulous or partially edentulous patients will regain their masticatory and aesthetic functions.

All digitally designed abutments for use with Klockner TiBase Abutments are intended to be sent to a SOADCO S.L. validated milling center for manufacture.

Device Description

The abutments consist of healing abutments, temporary abutments, straight abutments, and an overdenture attachment system that consists of abutments, denture housings, and retention inserts, and TiBase abutments with screws.

AI/ML Overview

This is a medical device 510(k) premarket notification for Klockner Abutments which are endosseous dental implant abutments. The submission aims to demonstrate substantial equivalence to previously cleared devices.

Here's the breakdown of the acceptance criteria and supporting studies, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in terms of specific performance targets (e.g., "must achieve X% accuracy"). Instead, it lists various non-clinical tests conducted to demonstrate substantial equivalence to predicate devices. The "reported device performance" is essentially that the device passed these tests in comparison to the predicate devices.

Acceptance Criteria (Implied by tests performed)Reported Device Performance (as stated in the document)
Sterilization validation (according to ISO 17665-1 and 17665-2)Leveraged from previously cleared submissions K151194 and K170022. (Implies compliance with these ISO standards)
Biocompatibility (for TiBase, straight, healing, temporary, denture housing)Leveraged from K170022 and K151194. (Implies compliance with biocompatibility standards for these components)
Biocompatibility for TiCN and ZrN coated abutments (according to ISO 10993-5, 10993-10, 10993-12)Passed tests for in vitro cytotoxicity, skin sensitization for irritation and sensitization, and sample preparation/reference materials. (Implies compliance with these ISO standards for the coated abutments)
Mechanical testing (according to ISO 14801 Dentistry Implants Dynamic loading test)Conducted. (Implies the device meets mechanical performance requirements for dynamic loading of endosseous dental implants)
Retention testingConducted to ensure retention is maintained throughout expected use. (Implies the device meets retention performance requirements)
Coating characterization (according to FDA Guidance Class II Special Controls Guidance Document)Conducted using methods outlined in the FDA Guidance. (Implies the coating properties were characterized and meet relevant guidance)
MRI compatibility/safetyNon-clinical worst-case MRI review to evaluate metallic devices in the MRI environment using scientific rationale and published literature. Rationale addressed parameters per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque. (Implies MRI safety has been addressed)
MaterialsTitanium alloy conforming to ASTM F136; Zirconia (for copings) conforming to ISO 13356. Same materials as previously cleared products.
Manufacturing ProcessIdentical manufacturing processes as used for SOADCO S.L. products cleared previously in K151194 and K170022.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state a "sample size" in terms of number of patients or clinical cases for a test set, nor does it specify data provenance (country of origin, retrospective/prospective).

The studies mentioned are primarily non-clinical bench testing (e.g., sterilization, biocompatibility, mechanical testing, retention testing, coating characterization, MRI review). For these types of tests, "sample size" would refer to the number of devices or components tested, which is not detailed in this summary.

The "leveraged" data refers to studies performed on previously cleared predicate devices (K151194, K170022) and is therefore retrospective in relation to this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is typically relevant for studies involving human assessment or interpretation (e.g., imaging studies, clinical trials with expert review). Since the provided text describes primarily non-clinical bench testing, there is no mention of experts establishing ground truth in this context. The "truth" for these tests is based on objective measurements and compliance with recognized standards.

4. Adjudication Method for the Test Set

Again, this is applicable to studies with subjective assessments. As the studies are non-clinical bench tests, there is no adjudication method mentioned or applicable. Compliance is determined by objective measurements against established standards.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was mentioned. This submission focuses on the substantial equivalence of a physical dental device (abutments) through non-clinical performance testing against established standards and predicates, not on AI-assisted diagnostic or interpretive tasks.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

No standalone algorithm performance study was mentioned. The device is a physical dental implant component, not a software algorithm.

7. The Type of Ground Truth Used

For the non-clinical tests, the "ground truth" is defined by:

  • Established standards: e.g., ISO 17665-1, ISO 17665-2, ISO 10993-5, ISO 10993-10, ISO 10993-12, ISO 14801, ASTM F136, ISO 13356.
  • FDA Guidance: e.g., "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments," and "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment."
  • Performance of predicate devices: The subject device's performance is compared to that of previously cleared predicate devices (K151194, K170022, K222288, K191123, K220612) to demonstrate substantial equivalence.

8. The Sample Size for the Training Set

There is no mention of a training set as this is not an AI/machine learning device. The studies described are non-clinical laboratory tests.

9. How the Ground Truth for the Training Set Was Established

Since there is no training set mentioned, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

SOADCO S.L. % Melissa Burbage Senior Regulatory Specialist PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K230103

Trade/Device Name: Klockner Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: Class II Product Code: NHA Dated: January 27, 2023 Received: January 27, 2023

Dear Melissa Burbage:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K230103

Device Name Klockner Abutments

Indications for Use (Describe)

Klockner Abutments are intended for use in dental implants to provide a support structure so that the edentulous or partially edentulous patients will regain their masticatory and aesthetic functions.

All digitally designed abutments for use with Klockner TiBase Abutments are intended to be sent to a SOADCO S.L. validated milling center for manufacture.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K230103

Klockner Abutments

SOADCO, S.L.

April 24, 2023

ADMINISTRATIVE INFORMATION

Manufacturer NameSOADCO S.L.Av. del Pessebre, 76-82AD700 Escaldes - Engordany, AndorraTelephone: +37 (6) 800-590
Official ContactMs. Mercedes Roldán, General Manager
Representative/ConsultantMelissa BurbageKevin A. Thomas, PhD; Floyd G. Larson, MS, MBAPaxMed International, LLC12264 El Camino Real, Suite 400San Diego, CA 92130Telephone: +1 858-792-1235Fax: +1 858-792-1236Email: mburbage@paxmed.comkthomas@paxmed.com; flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Device NameKlockner Abutments
Common NameDental implant abutment
Classification NameEndosseous dental implant abutment
Regulation Number21 CFR 872.3630
Regulation NameEndosseous dental implant abutment
Regulatory ClassClass II
Product CodeNHA
Classification PanelDental Products Panel
Reviewing BranchDental Devices Branch

PREDICATE DEVICE INFORMATION

Primary Predicate Device

K151194, Klockner Dental Implant Abutments (II), SOADCO, S.L.

Reference Devices

K170022, Klockner Vega TiBase for CEREC®, SOADCO, S.L. K222288, DESS Dental Smart Solutions, Terrats Medical SL K191123, Multi-unit Abutments, Medentika GmbH K220612, PrimeLOC Attachment System, Innovative Product Brands, Inc.

{4}------------------------------------------------

Klockner Abutments are intended for use in dental implants to provide a support structure so that the edentulous or partially edentulous patients will regain their masticatory and aesthetic functions.

All digitally designed abutments for use with Klockner TiBase Abutments are intended to be sent to a SOADCO S.L. validated milling center for manufacture.

SUBJECT DEVICE DESCRIPTION

The purpose of this submission is to obtain marketing clearance for abutments that are compatible with components from the Klockner Dental Implant Systems cleared in K153098, K082200, K080224. and K010132. The abutments consist of healing abutments, temporary abutments, straight abutments, and an overdenture attachment system that consists of abutments, denture housings, and retention inserts, and TiBase abutments with screws.

The subject device Vega Mimetic® Healing Abutments are designed to be connected to the implant during the healing period between implant placement and final abutment placement. Restorations are not placed on the healing components. The healing components are compatible with the Klockner Vega Implant System. They are offered in coronal diameter of 3.6 mm and with gingival heights of 2.0, 3.0, and 5.0 mm. The healing abutments attach to the dental implant with an integrated apical threaded portion. They are manufactured from titanium alloy and are color anodized to aid in identification.

The subject device Vega Mimetic® Temporary Abutments are straight, prepable abutments designed to retain a temporary cemented prosthesis. They are provided straight only and not intended to provide any angulation correction. They are compatible with the Klockner Vega Implant System. The abutment has a hex connection and attaches to the implant with a separate screw, previously cleared under K153098. The abutments are offered in a prosthetic platform diameter of 3.6 mm and a 7.0 mm post height, with gingival heights of 1.0, 2.0, and 3.0 mm. They are manufactured from titanium alloy and are anodized on a portion of the surface for identification purposes.

The subject device Vega Straight Permanent Abutments are designed for single-unit and multipleunit cement-retained restorations. The straight abutments are compatible with the Klockner Vega Implant System. They are available in an engaging and a non-engaging connection. The engaging design has a hex connection and attaches to the implant with a separate screw. The non-engaging design has a threaded apical portion. They are offered in a prosthetic platform diameter of 4.2 mm and gingival heights of 1.0, 2.0, and 3.0 mm, with no angulation. They are manufactured from titanium alloy and anodized on a portion of the surface for identification purposes.

The subject device K-LOCK Overdenture Attachment System is compatible with the Klockner Vega and Essential Cone Implant System and consists of abutments, retention inserts, and denture housings for attachment of prostheses to endosseous dental implants. All K-LOCK abutments are made of titanium alloy and have the same coronal ridge retention design that attaches to the overdenture component by an interference (snap) fit. The threaded apical end of the abutment connects to the implant and is specific to each compatible implant system and diameter. The

{5}------------------------------------------------

510(k) Summary Page 3 of 8

retention connection is designed to accommodate a path of insertion on implants that are divergent up to 40° (with 20°on each side). K-LOCK abutments are provided with a titanium carbonitride (TiCN) coating and are available in cuff (gingival) heights up to 6 mm, with a minimum implant interface diameter of 3.0 mm. A thin titanium carbonitride (TiCN) coating is applied to the abutments in the area that contacts gingival tissue and in the area that contacts the retention inserts.

Subject device TiBase Abutments are two-piece abutments composed of a bottom titanium component and are designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. The patient-specific CAD/CAM zirconia superstructure is the top-half of the two-piece abutment. They also may be used for support of a crown directly on the abutment. There are two types of TiBase Abutments; TiBases that attach directly to the implant, and TiBases that attach to previously-cleared abutments from the sponsor (similar to a prosthetic coping).

All patient-specific custom abutment fabrication for TiBase Abutments is by prescription on the order of the clinician. Zirconia superstructures for use with the subject device TiBase Abutments will be made at a SOADCO validated milling center under FDA quality system regulations and the material will conform to ISO 13356 Implants for surgery – Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP). The cement recommended in labeling for bonding of superstructures is Panavia F 5.0 cement by Kuraray Noritake Dental, cleared under K183537.

TiBase Abutments that attached directly to the implant are available in an engaging design and a non-engaging design. TiBase Abutments are made of titanium alloy. The TiBase Abutments are provided in four different designs: MedproTiBase, Mimetic TiBase, EC Essential TiBase, and SK2 Medpro TiBase. The EC Essential TiBase is a line extension from previously cleared TiBases for Cerec (K170022). The MedproTiBase abutments are anodized gold. The Mimetic TiBase abutments are coated with a thin layer of zirconium nitride (ZrN).

The design parameters for the CAD/CAM zirconia superstructure to be used on MedproTiBase Abutments are as follows:

Minimum wall thickness – 0.5 mm Minimum post height - 4.0 mm1 Minimum gingival height - 0.5 mm Maximum gingival height – 6.0 mm All zirconia superstructures are for straight abutments only 1 For devices 84 01 36 must be a minimum post height of 5.5 mm

The design parameters for the CAD/CAM zirconia superstructure to be used on Mimetic TiBase Abutments are as follows:

Minimum wall thickness - 0.35 mm Minimum post height - 4.2 mm Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm All zirconia superstructures are for straight abutments only

{6}------------------------------------------------

510(k) Summary Page 4 of 8

The design parameters for the Sirona Dental CAD/CAM system to be used on EC Essential TiBase Abutments for Cerec® are as follows:

Minimum wall thickness - 0.8 mm Minimum post height - 5.2 mm Minimum gingival height - 0.5 mm Maximum gingival height – 6.0 mm Maximum angulation of the final abutment - 20°

TiBase Abutments that attached directly to previously-cleared angled abutments (like a coping) from SOADCO S.L. are available in an engaging design and a non-engaging design. These TiBase Abutments have no angulation, but some may be placed on abutments that are angled up to 30°. TiBase Abutments are made of titanium alloy and anodized gold. The TiBase Abutments are provided in three different designs: Permanent Vega Medpro, Essential Cone 25° Medpro, and Essential Cone Multi-Cone Medpro. The differences between these designs are primarily related to the compatible implant system or related to connection to a straight or angled abutment.

The design parameters for the CAD/CAM zirconia superstructure to be used on Medpro TiBase to Abutments are as follows:

Minimum wall thickness - 0.5 mm Minimum post height - 4.0 mm1 Minimum gingival height - 0.5 mm Maximum gingival height - 6.0 mm Not intended to provide angulation correction 1 For devices 84 01 26 and 84 01 27 must be a minimum post height of 5.5 mm

PERFORMANCE DATA

Non-clinical testing data submitted to demonstrate substantial equivalence included:

  • Sterilization validation according to ISO 17665-1 Sterilization of health care products Moist ● heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2 Sterilization of health care products — Moist heat — Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1 was leveraged from previously cleared submissions K151194 and K170022.
  • Biocompatibility testing for Klockner TiBase, straight abutments, healing abutments, temporary abutments, and denture housing subject devices leveraged from K170022 and K151194.
  • Biocompatibility testing for TiCN coated K-Lock Abutments and ZrN coated Mimetic . Abutments according to ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for skin sensitization for irritation and sensitization, and ISO 10993-12 Biological evaluation of medical devices – Part 12: Sample preparation and reference materials.
  • Mechanical testing according to ISO 14801 Dentistry Implants Dynamic loading test ● for endosseous dental implants.
  • . Retention testing was conducted to ensure that retention is maintained throughout the expected use of the product.
  • The coating characterization using the methods outlined in FDA Guidance Class II Special ● Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous

{7}------------------------------------------------

Dental Implant Abutments, issued May 12, 2002.

  • Non-clinical worst-case MRI review to evaluate the metallic devices in the MRI . environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the subject device components and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.

EQUIVALENCE TO MARKETED DEVICES

The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K151194 and reference device K170022. There is difference in terminology used, however the intended use is the same. All abutments are intended to be used with dental implants for support for prosthetic restoration for edentulous or partially edentulous patients. The digitally designed abutments are to be sent to a validated milling center for manufacture.

The subject device consists of various types of abutments; healing abutments, temporary abutments, straight abutments, overdenture attachment system, and TiBase abutments.

The subject device healing abutments, temporary abutments, and straight abutments are identical in design, materials, and technological characteristics to similar abutments cleared in the primary predicate K151194.

The subject device overdenture attachment system consists of abutments, retention inserts and denture housings for resilient attachment of prostheses to endosseous implants. The subject device abutment is substantially equivalent to the abutment of the reference device K220612 in its design, materials, and technological characteristics. It is similar to the reference device K220612 in coronal geometry in that the mechanism for overdenture retention is the same. The ability of the attachment system to accommodate abutment divergence is the same as that of the reference device K220612. The diameters and cuff heights offered are similar to those of the reference device K220612. The subject device is manufactured from titanium alloy and has the external surfaces above the implant/abutment interface modified by application of titanium carbonitride (TiCN) coating, similar to the titanium nitride (Ti) coating of the reference device K220612. As with the reference device K220612, the coating is non-porous, the surface roughness of the machined abutment surface is maintained, and the coating is not applied to enhance tissue attachment to the device.

The subject device TiBase (to implant) abutment is substantially equivalent to the predicate device K170022 and the reference device K222288 in terms of design, materials, and technological characteristics. The subject device Mimetic TiBase abutments have external surfaces above the implant/abutment interface modified by application of zirconium nitride (Zr) coating, similar to the zirconium nitride (ZrN) coating of the reference device K222288. The superstructure design parameters, including angulation, are similar to those of reference device K22288.

{8}------------------------------------------------

510(k) Summary Page 6 of 8

The subject device TiBase to abutment (coping) is substantially equivalent to the reference device K191123 in terms of design, materials, and technological characteristics. The TiBase to abutment does not allow for any additional angulation but can be used on abutments with angulation up to 30°.

The subject device abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-4 Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications (UNS R56401). The titanium alloy subject device components are manufactured from identical materials, in the identical facilities using the identical manufacturing processes as used for SOADCO S.L. products cleared previously in K151194 and K170022.

All zirconia superstructures for use with the subject device TiBase Abutments will be made at a SOADCO S.L. validated milling center under FDA quality system regulations. The zirconia is inCoris ZI, cleared under K123664, and conforming to ISO 13356. Implants for surgerv -Ceramic materials based on vttria-stabilized tetragonal zirconia (Y-TZP). The cement recommended in labeling for bonding of superstructures is Panavia F 5.0 cement by Kuraray Noritake Dental, cleared under K183537. The zirconia and cement are the same material that was used in previously cleared products K170022.

Provided at the end of this summary is a table comparing the Indications for Use Statements (IFUS) and the technological characteristics of the subject device, the primary predicate device, and the reference devices.

CONCLUSION

The subject device, the primary predicate device, and reference devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

{9}------------------------------------------------

Subject DevicePrimary Predicate DeviceReference DeviceReference DeviceReference DeviceReference Device
Klockner AbutmentsSOADCO S.L.K151194Klockner Dental ImplantAbutment (II)SOADCO S.L.K170022Klockner Vega TiBase forCEREC®SOADCO S.L.K222288DESS Dental SmartSolutionsTerrats Medical SLK191123Multi-unit AbutmentsMedentika GmbHK220612PrimeLOC AttachmentSystemInnovative ProductBrands, Inc.
Klockner Abutments areintended for use in dentalimplants to provide a supportstructure so that theedentulous or partiallyedentulous patients willregain their masticatory andaesthetic functions.All digitally designedabutments for use withKlockner TiBase Abutmentsare intended to be sent to aSOADCO S.L. validatedmilling center formanufacture.The Klockner Dental ImplantAbutments are intended to beplaced into dental implants toprovide support forprosthetic reconstructionssuch as crowns, bridges.The Klockner Dental ImplantAbutments include protectivecap, temporary abutment,angled abutments, straightabutments, cast abutments.All abutments are intended tobe used with the KlocknerDental Implant Systems,models: Essential EC,Essential EC 1.5.The Klockner Vega TiBasefor CEREC® abutments aretitanium alloy abutmentsplaced onto Klockner VegaImplants to provide supportfor customized prostheticrestorations. The KlocknerVega TiBase for CEREC®abutments are indicated forscrew-retained single toothor cement-retained singletooth and bridge restorations.All digitally designedcopings and/or crowns foruse with the Klockner VegaTiBase for CEREC®abutments are to be designedand milled using SironaCEREC Premium SW 4.2software and manufacturedusing a Sirona CEREC orinLab MC X or MC XLmilling unit.DESS Dental SmartSolutions abutments areintended to be used inconjunction with endosseousdental implants in themaxillary or mandibular archto provide support forprosthetic restorations.All digitally designed customabutments for use with DESSBases or Blanks are to besent to a Terrats Medicalvalidated milling center formanufacture.For complete Indications forUse statement on OEMCompatibility see 510(k)Summary for K222288 inSection 12Multi-unit abutments areindicated for use with dentalimplants as a support formulti-unit screw retainedbridges and bars in themaxilla or mandible of apartially or fully edentulouspatient.Multi-unit Abutments areused for the restoration of thefollowing dental implantsystems:For complete Indications forUse statement on OEMCompatibility see 510(k)Summary for K191123 inSection 12.The PrimeLOC AttachmentSystem is designed tofacilitate patient removal of adental prosthesis for use withfull arch overdentures orpartial dentures retained inwhole or in part byendosseous implants in themandible or maxilla.
Reason forPredicate DeviceNot applicableHealing abutment, temporaryabutment, straight abutmentTi Base, validated millingcenterTiBase to implant, 30degreesTiBase to abutment (coping)Attachment system design
Product CodesNHANHANHANHANHANHA
Designs/Features
Abutment DesignHealing, Temporary,Straight, Attachment System,TiBase, TiBase to abutment(coping)Conventional one-pieceabutmentsTitanium Base AbutmentsHealing, Temporary,Straight, TiBase to implantTiBase to abutment (coping)Overdenture Attachment
ProsthesisAttachmentCement-retainedScrew-retainedCement-retainedScrew-retainedCement-retainedScrew-retainedCement-retainedScrew-retainedCement-retainedScrew-retainedInternal Thread
RestorationSingle-unit, Multi-unitSingle-unit, Multi-unitSingle-unit, Multi-unitSingle-unit, Multi-unitMulti-unitMulti-unit
Subject DevicePrimary Predicate DeviceReference DeviceReference DeviceReference DeviceReference Device
Klockner AbutmentsSOADCO S.L.K151194Klockner Dental ImplantAbutment (II)SOADCO S.L.K170022Klockner Vega TiBase forCERECⓇSOADCO S.L.K222288DESS Dental SmartSolutionsTerrats Medical SLK191123Multi-unit AbutmentsMedentika GmbHK220612PrimeLOC AttachmentSystemInnovative ProductBrands, Inc.
Abutment-ImplantPlatform Diameter(mm)3.0 - 4.83.0 - 4.83.0 - 4.82.52 - 6.53.3 - 5.03.5 - 4.1
Prosthetic PlatformDiameter (mm)3.5 - 6.63.5 - 6.64.5 - 74.5 - 6.53.3 - 6.53.5
AngulationUp to 30°0°, 17°, 30°Up to 20°Up to 30°0°, 17°, 30°
Abutment-ImplantInterfaceInternal, ExternalInternalInternalInternal, ExternalInternalInternal
OverdentureDivergenceAllowance20°/40°n/an/an/an/a20°/40°
OverdentureAttachment TypeRetention Insert retained inDenture Attachment Housingn/an/an/an/aRetention Insert retained inDenture Attachment Housing
Materials
Abutment MaterialsTitanium alloy, ASTM F136Zirconia, ISO 13356(copings)Titanium alloy, ASTM F136Titanium alloy, ASTM F136Compatible with the SironaDental System inCoris ZIMeso L (K100152 /K111421)Titanium alloy, ASTM F136Cobalt-chromium alloy,ASTM F1537Zirconia, ISO 13356(copings)Ti 6Al-4V ELITi 6Al-4V ELI
Surface Coating(overdenture &Mimetic TiBase)Titanium Carbonitride,Zirconium nitriden/an/aZirconium nitriden/aTitanium Nitride
RetentionAttachmentComponentNylonn/an/an/an/aNylon

{10}------------------------------------------------

510(k) Summary

Page 8 of 8

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)