K Number
K223822

Validate with FDA (Live)

Device Name
ACTIVBRAID™
Manufacturer
Date Cleared
2023-06-02

(163 days)

Product Code
Regulation Number
878.5000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ACTIVBRAID™ is indicated for use in soft tissue approximation and/or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft tissues are used for repair.

Device Description

ACTIVBRAID is a partially absorbable surgical co-braid suture constructed of nonabsorbable Ultra High Molecular Weight Polyethylene (UHMWPE) and absorbable bovine-derived type I collagen. ACTIVBRAID suture ends are stiffened with cyanoacrylate.

ACTIVBRAID is available in several sizes (sutures meet U.S. Pharmacopeia standards for nonabsorbable sutures, except diameter). Sutures are oversized in diameter. ACTIVBRAID is provided in pre-cut lengths with and without needles. Suture strands that are dyed (Chromium Cobalt-Aluminum Oxide or D&C Black No. 4) are made of UHMWPE.

AI/ML Overview

The provided text is a 510(k) Summary for the ACTIVBRAID™ surgical suture. It details the device, its intended use, comparison to a predicate device, and performance data to support its substantial equivalence. However, the document does not describe a study involving a medical device software with AI/ML components that would have acceptance criteria related to accuracy, sensitivity, specificity, or human-in-the-loop performance.

The "performance data" section in the document refers to:

  • Biocompatibility Testing: Evaluating the biological response of the material.
  • Non-Clinical Performance Testing: Assessing physical properties of the suture (diameter, needle attachment, tensile strength) against USP standards and comparing other mechanical properties (abrasion, knot profile/security, physiological fluid absorption) to the predicate device.
  • Resorption Profile: Characterizing the device's absorption rate and tensile strength retention over time in an animal model.

These are standard performance tests for physical medical devices like sutures, not for AI/ML-driven diagnostic or assistive software. Therefore, the requested information about acceptance criteria for AI/ML performance, sample sizes for test/training sets, expert adjudication methods, MRMC studies, or standalone algorithm performance, cannot be extracted from this document.

The document explicitly states: "Biocompatibility and nonclinical performance testing data demonstrate that ACTIVBRAID is as safe, as effective, and performs as well as or better than the predicate device, K112899." This indicates that the "acceptance criteria" are related to meeting or exceeding the performance of the predicate device and established standards for surgical sutures.

In summary, the provided text does not contain the information requested about AI/ML device acceptance criteria and studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

June 2, 2023

Embody, Inc. Brianna Schehr Vice President, Quality, Regulatory & Clinical Affairs 4211 Monarch Way, Suite 500 Norfolk, Virginia 23508

Re: K223822

Trade/Device Name: ActivbraidTM Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable polyethylene terephthalate surgical suture Regulatory Class: Class II Product Code: GAT Dated: December 21, 2022 Received: December 21, 2022

Dear Brianna Schehr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cynthia Chang -S

Cynthia J. Chang, Ph.D. Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223822

Device Name ACTIVBRAID™

Indications for Use (Describe)

ACTIVBRAID™ is indicated for use in soft tissue approximation and/or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft tissues are used for repair.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "emBODY". The word "em" and "DY" are in a dark blue font. The "BODY" portion of the logo has an orange square with diagonal lines inside of it.

510(k) Summary - K223822

SUBMITTER l.

Owner/Submitter:Embody, Inc.4211 Monarch WaySuite 500Norfolk, VA 23508
Phone:236-994-7912
Contact Person:Brianna SchehrVice President, Quality, Regulatory & Clinical Affairs
Date Prepared:May 24, 2023
II. DEVICE
Name of Device:ACTIVBRAID™
Common or Usual Name:Suture
Classification Name:Suture, Nonabsorbable, Synthetic, Polyethylene
Regulation Number:21 CFR 878.5000
Regulatory Class:Class II
Product Code:GAT

lll. PREDICATE DEVICE

Medical Specialty:

Arthrex Bio-Suture, K112899. This predicate has not been subject to a design-related recall.

General & Plastic Surgery

Reference Devices: Arthrex BioWire, K091018 ORTHOCORD® Braided Composite Suture, K040004

IV. DEVICE DESCRIPTION

ACTIVBRAID is a partially absorbable surgical co-braid suture constructed of nonabsorbable Ultra High Molecular Weight Polyethylene (UHMWPE) and absorbable bovine-derived type I collagen. ACTIVBRAID suture ends are stiffened with cyanoacrylate.

ACTIVBRAID is available in several sizes (sutures meet U.S. Pharmacopeia standards for nonabsorbable sutures, except diameter). Sutures are oversized in diameter. ACTIVBRAID is provided in pre-cut lengths with and without needles. Suture strands that are dyed (Chromium Cobalt-Aluminum Oxide or D&C Black No. 4) are made of UHMWPE.

V. INDICATION FOR USE

ACTIVBRAID is indicated for use in soft tissue approximation and/or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "embody" in a stylized font. The letters are a dark teal color, except for the "o", which is an orange square with diagonal lines inside. The word is written in all lowercase letters.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

ACTIVBRAID is substantially equivalent in terms of indication for use, technological characteristics, and performance characteristics to the predicate device. Both the subject and predicate devices meet the USP Monograph for nonabsorbable surgical sutures, except diameter.

The following technological differences exist between the subject and predicate devices:

  • The subject device is composed of nonabsorbable UHMWPE fiber braided with . absorbable bovine type I collagen fiber, whereas the predicate device is composed of nonabsorbable UHMWPE fiber braided with polyester fibers coated with absorbable type I bovine collagen, and
  • The subject device is available in sizes 2-0, 2, 1.5mm tape, and 2.5mm tape, whereas . the predicate device is available in sizes 2 and 2mm tape.

The differences between the subject and predicate device are considered minor and do not raise questions concerning safety and effectiveness. Biocompatibility and nonclinical performance testing data demonstrate that ACTIVBRAID is as safe, as effective, and performs as well as or better than the predicate device, K112899.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing

The biocompatibility evaluation for the ACTIVBRAID device was conducted in accordance with the FDA guidance document "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" dated September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: cytotoxicity, sensitization, systemic toxicity, pyrogenicity, genotoxicity, and implantation.

Non-Clinical Performance Testing

Nonclinical performance testing was performed to demonstrate that ACTIVBRAID meets the current USP Monograph for nonabsorbable surgical sutures, except diameter. Testing was performed in accordance with FDA's guidance document, "Surgical Sutures – Class II Special Controls Guidance Document for Industry and FDA Staff" and included the following tests:

  • Nonabsorbable Suture Diameter USP <861> Sutures Diameter .
  • Nonabsorbable Suture Needle Attachment USP <871> Sutures Needle Attachment .
  • . Nonabsorbable Suture Tensile Strength USP <881> Sutures – Tensile Strength

ACTIVBRAID meets or exceeds USP performance standards for nonabsorbable surgical sutures, except for an oversize in diameter. Bacterial Endotoxin Test (BET) was performed to demonstrate that the subject device meets pyrogen limit specifications.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "emBODY" in a stylized font. The letters "em" and "DY" are in a dark blue color, while the "BOD" is represented by an orange square with diagonal lines inside. The logo appears to be for a company or product named "emBODY."

Additional performance testing was conducted in comparison to the predicate device and included the following tests:

  • Abraded Tensile Strength .
  • . Knot Profile
  • Knot Security .
  • Abrasiveness to Tissue .
  • Physiological Fluid Absorption .

ACTIVBRAID performs as well as or better than the predicate device for all performance tests listed. Both the subject and predicate devices have suitable characteristics for their indication for use.

Resorption Profile

ACTIVBRAID is partially absorbable; therefore, resorption profile testing was conducted in accordance with FDA's special controls guidance, "Surgical Sutures – Class II Special Controls Guidance Document for Industry and FDA Staff" to characterize the device's rate of absorption, time to complete absorption, and residual tensile strength over time. The results of implantation studies of ACTIVBRAID in animals indicate that approximately 85% of its original strength remains 16 weeks after implantation and absorption of the collagen component is complete between 8 and 16 weeks post-implantation. The ACTIVBRAID degradation profile combined with the tensile strength retention over time demonstrates the resorption profile is consistent with the intended use of the subject device, for the long-term approximation of soft tissue.

CharacteristicSubject DevicePredicate DeviceArthrex Bio-Suture(K112899)Comparison
Product CodeGATGATSame
Regulation21 CFR 878.500021 CFR 878.5000Same
ClassificationNameSuture, Nonabsorbable,Synthetic, PolyethyleneSuture, Nonabsorbable,Synthetic, PolyethyleneSame
Intended UseLong-term approximation of softtissueLong-term approximation of softtissueSame
Indications forUseFor use in soft tissueapproximation and/or ligation.These sutures may beincorporated, as components,into surgeries where constructsincluding those with allograft orautograft tissues are used forrepair.For use in soft tissueapproximation and/or ligation.These sutures may beincorporated, as components,into surgeries where constructsincluding those with allograft orautograft tissues are used forrepair.Same
Prescription UseYesYesSame
Material(s) ofCompositionUHMWPE, bovine-derived type IcollagenUHMWPE, polyester, bovine-derived type I collagenSimilar, nonew issues ofsafety orefficacy
ResorbablePartially absorbable (i.e.,nonabsorbable UHMWPE fiberand absorbable collagen fiberco-braid)Nonabsorbable (i.e.,nonabsorbable braidedUHMWPE and polyester fiberscoated with absorbable type Ibovine collagen)Similar, nonew issues ofsafety orefficacy

Substantial Equivalence

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "EMBODY" in a stylized font. The letters are primarily dark blue, except for the "B", which is enclosed in an orange square with diagonal lines inside. The word appears to be a logo or brand name, with a modern and clean design.

CharacteristicSubject DevicePredicate DeviceArthrex Bio-Suture(K112899)Comparison
SutureConfiguration(s)Configured as multifilamentbraided strands of multiple sizes(per USP size system)ACTIVBRAID suture:• Size 2-0 (blue) w/ taperedneedle• Size 2 (blue) w/ taperedneedle• Size 2 (white/black) w/ taperedneedleACTIVBRAID suture tape:• Flat, tape-style suture, 1.5mm(white/black) w/ taperedneedle• Flat, tape-style suture, 1.5mm(blue) w/ tapered needle• Flat, tape-style suture, 2.5mm(white/black)• Flat, tape-style suture, 2.5mm(blue)Configured as multifilamentbraided strands of multiple sizes(per USP size system)Bio-FiberWire:• Size 2 (blue) w/ taperedneedleBio-FiberTape:• Flat, tape-style suture, 2mm(blue)Similar, nonew issues ofsafety orefficacy
Residual TensileStrength OverTime85% of original strength remains16 weeks after implantationNot known to have significantchange in tensile strength in vivoSimilar, nonew issues ofsafety orefficacy
TechnologicalPrincipleLong-term approximation of softtissueLong-term approximation of softtissueSame
OperatingPrinciplesStandard surgical techniqueStandard surgical techniqueSame
How SuppliedSterile, single use only, with orwithout needles attachedSterile, single use only, with orwithout needles attachedSame
SterilitySterile, SAL 10-6Sterile, SAL 10-6Same
SterilizationMethodEthylene OxideEthylene OxideSame
Single UseSingle UseSingle UseSame

VIII. CONCLUSIONS

Results of the biocompatibility testing and nonclinical performance testing demonstrate that ACTIVBRAID is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K112899.

§ 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture.

(a)
Identification. Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.