K Number
K112899

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-04-06

(186 days)

Product Code
Regulation Number
878.5000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex Bio-Suture is intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair.

Device Description

The Arthrex Bio-Suture is a dyed or non-dyed braided polyester suture construct coated with type 1 bovine collagen. The suture construct is made of UHMWPE and polyester braided over a UHMWPE core. The suture tape construct is a flat suture construct composed of UHMWPE and polyester yarns braided over a FiberWire suture core and UHMWPE yarns. Arthrex Bio-Suture strands that are dyed black are made of nylon. The suture ends are stiffened with cyanoacrylate. The Arthrex Bio-Suture will be supplied in pre-cut lengths with or without various swaged needles. The Arthrex Bio-Suture constructs meet USP standards for suture, except for diameter.

AI/ML Overview

Here's an analysis of the provided text regarding the Arthrex Bio-Suture, focusing on the acceptance criteria and supporting studies.

Important Note: The provided document is a 510(k) Summary for a medical device (suture). It primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed clinical study for novel performance claims. As such, many of the requested categories (e.g., sample size for test set, number of experts for ground truth, MRMC study, training set details) are not applicable or not present in this type of regulatory submission. The assessment relies on a comparison to established standards and predicate devices.


Acceptance Criteria and Study Details for Arthrex Bio-Suture

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied by Regulatory Pathway & Standards)Reported Device Performance (Arthrex Bio-Suture)
Biocompatibility: No adverse biological reactionsBiocompatibility testing performed and determined to be substantially equivalent to predicate devices.
Tensile Strength: Meets or exceeds performance of predicate devices and relevant USP standards (where applicable).Tensile testing performed and determined to be substantially equivalent to predicate devices. The device meets USP standards for suture, "except for diameter." (This exception is noted with the understanding that the overall performance is still equivalent or acceptable based on testing).
Material Composition: Consistent with known safe and effective materials for sutures (UHMWPE, polyester, bovine collagen coating).Device is made of UHMWPE and polyester braided over a UHMWPE core, coated with type 1 bovine collagen. Dyed strands use nylon.
Physical Characteristics: Consistent with intended function (braided, stiffened ends).Described as a dyed or non-dyed braided polyester suture construct. Suture ends are stiffened with cyanoacrylate.
Sterility: (Implied, standard for medical devices)Not explicitly detailed in the summary, but presumed to meet regulatory requirements for sterile medical devices.
Freedom from Pyrogens: (Implied, standard for medical devices)Not explicitly detailed in the summary, but presumed to meet regulatory requirements.
Toxicity: (Implied by biocompatibility)Addressed via biocompatibility testing.
Intended Use: Safe and effective for soft tissue approximation and/or ligation, and as components in constructs with allograft or autograft tissues.Stated intended use matches the device description and is considered substantially equivalent to predicate devices with identical intended uses.
USP Standards for Suture: Meets general USP standards."The Arthrex Bio-Suture constructs meet USP standards for suture, except for diameter."

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated as this is a 510(k) summary focusing on substantial equivalence. The document indicates that "biocompatibility and tensile testing" were performed. The specific number of samples for these tests is not provided in this summary.
  • Data Provenance: Not specified, but generally, such testing would be conducted in a controlled laboratory setting (e.g., in-house or by a contract lab). The document does not indicate human trial data (retrospective or prospective) for this submission given its nature as a device comparison.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. For this type of device (suture) and regulatory submission (510(k) for substantial equivalence), the "ground truth" is established through standardized laboratory testing (e.g., mechanical tensile tests, biocompatibility assays) and comparison to predicate devices, rather than expert interpretation of clinical data in a test set. There isn't "ground truth" in the sense of expert consensus on qualitative observations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. See point 3. Adjudication methods are typically associated with human-interpreted clinical or image-based studies.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a physical medical device (suture), not an AI-powered diagnostic or assistive tool. MRMC studies are not relevant here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For Biocompatibility: The ground truth is established by standardized biological assays and tests (e.g., ISO 10993 series for medical devices), which determine if the material elicits an acceptable biological response.
  • For Tensile Strength: The ground truth is established by objective, quantitative measurements of mechanical properties according to recognized standards (e.g., ASTM, USP, or manufacturer's internal validated methods).
  • For Substantial Equivalence: The ultimate "ground truth" for regulatory clearance is that the device is shown to be as safe and effective as a legally marketed predicate device, as determined by the FDA.

8. The sample size for the training set

  • Not Applicable. There is no "training set" in the context of this device and submission, as it does not involve machine learning or AI.

9. How the ground truth for the training set was established

  • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

। ତା ।

Arthrex SPECIAL 510(K): Arthrex Bio-Suture

4 510(k) Summary of Safety and Effectiveness

Date Summary PreparedSeptember 30, 2011
Purpose of SubmissionTo obtain clearance to distribute new suture devices, Arthrex Bio-Suture
Manufacturer/Distributor/SponsorArthrex, Inc.1370 Creekside BoulevardNaples, FL 34108-1945 USA
510(k) ContactSally Foust, RACRegulatory Affairs Project ManagerArthrex, Inc.1370 Creekside BoulevardNaples, FL 34108-1945 USATelephone: 239/643.5553, ext. 1251, and Fax: 239/598-5508Email: sfoust@arthrex.com
Trade NameArthrex Bio-Suture: Bio-FiberWire, Bio-TigerWire, Bio-FiberLoop, Bio-TigerLoop, Bio-FiberTape, and Bio-TigerTape
Common NameSuture
Product Code -Classification Name878.5000 - GAT - Nonabsorbable poly(ethylene terephthalate) surgicalsuture
Predicate DevicesArthrex Graft Suturing Kit (K041553)Arthrex BioWire (K091018)
Device Description andIntended UseThe Arthrex Bio-Suture is a dyed or non-dyed braided polyester sutureconstruct coated with type 1 bovine collagen. The suture construct ismade of UHMWPE and polyester braided over a UHMWPE core. Thesuture tape construct is a flat suture construct composed of UHMWPE andpolyester yarns braided over a FiberWire suture core and UHMWPEyarns. Arthrex Bio-Suture strands that are dyed black are made of nylon.The suture ends are stiffened with cyanoacrylate. The Arthrex Bio-Suturewill be supplied in pre-cut lengths with or without various swaged needles.The Arthrex Bio-Suture constructs meet USP standards for suture, exceptfor diameter.The Arthrex Bio-Suture is intended for use in soft tissue approximationand or ligation. These sutures may be incorporated, as components, intosurgeries where constructs including those with allograft or autografttissues are used for repair.
Substantial EquivalenceSummaryThe Arthrex Bio-Suture is substantially equivalent to the predicatedevices in which the material, basic features and intended uses areidentical. Any differences between the Arthrex Bio-Suture and thepredicate devices are considered minor and do not raise questionsconcerning safety and effectiveness. Based on the biocompatibility andtensile testing performed, Arthrex, Inc. has determined that the ArthrexBio-Suture is substantially equivalent to the currently marketed predicatedevices.

.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR - 6 2012

Arthrex, Inc. . % Courtney Smith, RAC Regulatory Affairs Manager 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K112899

Trade/Device Name: Arthrex Bio-Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: II Product Code: GAT Dated: March 22, 2012 Received: March 28, 2012

Dear Courtney Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

Page 2 - Courtney Smith, RAC

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions.regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Eunel Keith

-Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

KI12899

Arthrex SPECIAL 510(k): Arthrex Bio-Suture

3 Indications for Use Form

Indications for Use

510(k) Number:

Device Name:

Arthrex Bio-Suture

Indications For Use:

The Arthrex Bio-Suture is intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair.

Prescription Use _X_AND/OR Over-The-Counter Use ___

(Per 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kime ber MXM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112899

PAGE 1 of 1

§ 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture.

(a)
Identification. Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.