(265 days)
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
The monitors are only for use on one patient at a time.
The monitors are not therapeutic devices.
The monitors are for prescription use only.
The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC11).
ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.
BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.
The Intelli Vue NMT Module is intended to be used as an objective neuromuscular transmission monitor, using accelerometry for measuring the muscle contraction following an electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.
The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.
The noninvasive Masimo O3 Regional Oximeter System and accessories are intended for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and
The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and non-invasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, EEG, BIS, NMT, and gas analysis. The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals. The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display. The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.
This document is a 510(k) Summary for the Philips IntelliVue Patient Monitor MX400, MX450, MX500, and MX550. It does not describe a study involving an AI/algorithm that requires detailed acceptance criteria, ground truth establishment, or specific human reader studies as typically seen for AI-enabled medical devices. Instead, this submission focuses on demonstrating substantial equivalence to a predicate device, primarily due to software and hardware modifications allowing integration with external measurement devices like the Masimo SedLine Sedation Monitor.
Therefore, many of the requested points, particularly those related to AI/algorithm performance studies (e.g., sample size for test set/training set, expert adjudication, MRMC studies, standalone performance, specific effect sizes), are not applicable to this type of 510(k) submission.
However, I can extract information related to the overall device performance and testing as described in the document.
Here's an analysis of the provided text in relation to your request, with an emphasis on what is and is not present:
Device: IntelliVue Patient Monitor MX400, MX450, MX500, MX550
Study Type: This is a 510(k) premarket notification primarily demonstrating substantial equivalence (predicate pathway) for modifications to an existing patient monitor, specifically related to software updates and integration of an external monitoring module (Masimo SedLine Sedation Monitor). It is not a de novo submission for a new AI/algorithm.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a table of specific quantitative acceptance criteria for an AI/algorithm's performance (e.g., accuracy, sensitivity, specificity metrics) nor does it provide a "reported device performance" table in that context.
Instead, the "acceptance criteria" can be inferred from the non-clinical tests performed to demonstrate compliance with recognized standards for medical electrical equipment. The "reported device performance" is that the device "passed all safety tests" and "perform as intended."
Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|
Compliance with IEC 62304 Ed 1.1 (Software life cycle processes) | Passed all safety tests for demonstrated compliance with the standard |
Compliance with ANSI AAMI ES60601-1 (Basic safety & essential performance) | Passed all safety tests for demonstrated compliance with the standard |
Compliance with IEC 60601-1-2 Ed 4.0 (Electromagnetic disturbances) | Passed all safety tests for demonstrated compliance with the standard |
Compliance with IEC 60601-1-8 Ed 2.2 (Alarm systems) | Passed all safety tests for demonstrated compliance with the standard |
Compliance with IEC 60601-1-6 Ed 3.1 (Usability) | Passed all safety tests for demonstrated compliance with the standard |
No new issues of substantial equivalence introduced | Substantially equivalent to predicate devices; performs as intended |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not applicable/not specified for an AI/algorithm test set. The testing described is primarily for system-level safety, performance, and software validation.
- Data Provenance: Not applicable in the context of an AI/algorithm dataset. The document discusses compliance with various electrical, safety, software, and usability standards for the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth establishment by experts is typically for diagnostic AI/imaging algorithms. This submission is for a patient monitor system.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are relevant for expert review of images or clinical data for AI algorithm validation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a patient monitor, not an AI-assisted diagnostic tool that would typically undergo MRMC studies to evaluate human reader performance with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/algorithm in the sense of a diagnostic or predictive model being evaluated in standalone mode. It's a medical device system. While it has "software," the performance evaluation is centered on its functionality as a patient monitor and its compliance with relevant standards.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable. The "ground truth" for this device's performance validation is its ability to correctly acquire, display, and alert on physiological parameters according to established benchmarks and safety standards (e.g., AAMI EC11 for ECG, IEC standards for electrical safety). It's not a diagnostic AI's ground truth based on clinical diagnoses or pathological findings.
8. The sample size for the training set
Not applicable. This is not an AI/ML model generated from a training set in the typical sense of deep learning or machine learning algorithms.
9. How the ground truth for the training set was established
Not applicable. (See #8)
Summary of what the document does discuss regarding validation and performance:
The document emphasizes non-clinical bench tests and compliance with harmonized standards (e.g., IEC 62304 for software, ANSI AAMI ES60601-1 for basic safety, IEC 60601-1-8 for alarms, IEC 60601-1-6 for usability). The "performance data" section primarily lists these standards and states that the devices "passed all safety tests."
Key statements from the document regarding performance:
- "The subject devices have passed all safety tests for demonstrated compliance with the recognized standards below." (See table in section 1.8 Performance Data)
- "No new issues of substantial equivalence are introduced because of using this device." (Section "Non-clinical Bench Tests")
- "The subject devices, like the primary predicate devices, did not require clinical trials. Any clinical studies performed for the Masimo SedLine parameter are still valid as the measurements are not modified; they are only being connected to an additional host patient monitor." (Section "Clinical Studies")
- "Compliance to the FDA Quality System Regulations, FDA recognized standards, FDA guidance documents, harmonized standards, verification and validation, software validation, usability validation, and risk management activities have taken place for the subject devices." (Section "Clinical Studies")
- "The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety and electromagnetic compatibility, software verification and validation, human factors, usability and interoperability testing, demonstrate that the modified devices do not raise different questions of safety and effectiveness when compared to the predicate, perform as intended, and have performance characteristics that are substantially equivalent to the predicate devices." (Section 1.9 CONCLUSIONS)
In conclusion, this 510(k) submission for the Philips IntelliVue Patient Monitor MX series primarily relies on demonstrating compliance with recognized electrical safety, performance, and software development standards, along with a comparison to predicate devices, rather than an AI/algorithm performance study with specific metrics, expert adjudication, or MRMC studies.
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.