K Number
K223295
Date Cleared
2023-01-11

(77 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Endoscope EN-580T: This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine and rectum. Never use this product for any other purposes.

Over-tube TS-1314B: This product is used in combination with a FUJIFILM Double Balloon Endoscope to assist the insertion of the Endoscope under the management of physicians in medical facilities. This product is used to assist with the movement of the scopes inside the upper or lower digestive tract. Do not use this product for any other purposes. This product is not intended for use for any neonates, infants or children.

Balloon BS-4: This product is used in combination with FUJIFILM Double Balloon Endoscopes at medical facilities under the management of physicians. Being attached to the endoscope, this product is inserted into the mouth or anus to stabilize the distal end of the endoscope to the digestive tract's mucous membrane. Do not use this product for any other purposes. This product is not intended for use for any neonates, infants or children.

Tube Kit TY-500D: This product is the tube kit used in combination with the compatible balloon controller in medical facilities. Do not use this product for any other purpose.

Device Description

The endoscope EN-580T is inserted both perorally and transanally into the gastrointestinal tract during clinical use. The insertion of the device has a mechanism which bends the tip from right to left and up and down, and a flexible tube consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument.

The over-tube TS-1314B is introduced in the patient's anatomy with the pairing endoscope. TS-1314B is assembled over the outer diameter of endoscope. The endoscope and the over-tube are not advanced both at the same time, but alternatingly and successively. TS-1314B is provided sterile and single-patient use only.

The balloon BS-4 is attached to the balloon-compatible endoscope. The balloon air feed outlet should be contained inside BS-4. The accompanying fixing rubber is used to affix the scope balloon in place. The fixing rubbers are made specifically for BS-4 but should not be mixed with the fixing rubber of other scope balloons. BS-4 is provided sterile and single-patient use only.

The tube kit TY-500D consists of a set of two tubes connects the over-tube to the balloon controller by its air inlet. The other tube connects the same balloon controller to the endoscope by its balloon air feed inlet. TY-500D is only compatible with PB-30, cleared K153483. TY-500D is provided non-sterile. The tubes are not patient-contacting and reusable, but each filter is single-patient use only.

AI/ML Overview

This document does not contain information about a study that proves the device meets specific acceptance criteria in the manner typically seen for AI/ML-driven devices. Instead, it describes a 510(k) premarket notification for updated medical devices (Endoscope and accessories) that aim to demonstrate substantial equivalence to previously cleared predicate devices.

The "acceptance criteria" here are inherent to the regulatory process of demonstrating substantial equivalence for traditional medical devices, focusing on safety and effectiveness compared to existing devices rather than a performance benchmark against a clinical gold standard using a specific test set.

Here's an breakdown based on the document's content:

1. Table of Acceptance Criteria and Reported Device Performance

Instead of a table with specific numerical performance metrics (like sensitivity, specificity, AUC), the "acceptance criteria" are compliance with relevant safety and performance standards for medical devices, and the "reported device performance" is the successful completion of testing to demonstrate this compliance and substantial equivalence to predicate devices.

Acceptance Criteria (Implied)Reported Device Performance
Sterility (for TS-1314B and BS-4)Evaluated using ISO 11135:2014, ISO 10993-7:2008/AMD1:2019, ISO 11607-1:2019, ISO 11607-2:2019, ISO 11737-1:2018, and ISO 11138-1:2017 with acceptable results.
Biocompatibility (for new accessory materials)Evaluated using ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010, and ISO 10993-12:2012, in accordance with FDA guidance, with acceptable results.
Endoscope Specific Performance (EN-580T)Conducted according to ISO 8600-1:2015, ISO 8600-3:2019, and ISO 8600-4:2014 with acceptable results.
Endoscope Compatibility with New AccessoriesConducted with acceptable results.
Functional Equivalence (to predicate devices)Demonstrated through comparison of intended use, principle of operation, technical characteristics (Tables 1-4), and material construction (Tables 5-6). Differences were evaluated and found acceptable.
Product Expiration (for TS-1314B and BS-4)Product expiration of 3 years (predicate devices were 2 years). No specific acceptance criteria for this change are provided beyond the fact that it's a change.
Storage and Transport Environment (for TS-1314B and BS-4)Revised temperature and humidity ranges. No specific acceptance criteria for these changes are provided beyond the fact that they are changes.

2. Sample Size Used for the Test Set and Data Provenance

This document does not describe a "test set" in the context of AI/ML evaluation (e.g., a set of medical images or patient cases). The performance data cited refers to standard engineering and biological testing (sterility, biocompatibility, endoscope-specific tests). Therefore, information on sample size or data provenance for such a "test set" is not applicable or provided here.

3. Number of Experts Used to Establish the Ground Truth and Qualifications of Those Experts

Not applicable. Ground truth, in this context, would imply expert annotations or diagnoses on patient data. This document describes the substantial equivalence of physical medical devices and their components, not an AI/ML algorithm's diagnostic performance. Therefore, no experts were used to establish ground truth for a test set in this manner.

4. Adjudication Method for the Test Set

Not applicable, as there is no diagnostic "test set" to adjudicate.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with vs. without AI Assistance

No. This document does not pertain to AI or an AI-assisted device. Therefore, no MRMC study or AI-related effectiveness assessment was conducted or reported.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No. This document does not pertain to an algorithm or AI device.

7. The Type of Ground Truth Used

The "ground truth" for this submission is compliance with established international and FDA standards for medical device safety and performance. This is confirmed through laboratory testing for sterility, biocompatibility, and functional performance, rather than clinical outcomes or pathology data.

8. The Sample Size for the Training Set

Not applicable, as there is no AI/ML algorithm involved, and thus no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no AI/ML algorithm and training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2023

FUJIFILM Corporation % Kotei Aoki Senior Regulatory Affairs Specialist FUJIFILM Healthcare Americas Corporation 81 Hartwell Avenue, Suite 300 Lexington, MA 02421

K223295 Re:

Trade/Device Name: Endoscope Model EN-580T, Over-Tube TS-1314B, Scope Balloon BS-4, Tube Kit TY-500D Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Codes: FDA, FDF Dated: December 12, 2022 Received: December 12, 2022

Dear Kotei Aoki:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any

{1}------------------------------------------------

Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223295

Device Name Endoscope EN-580T, Over-tube TS-1314B, Balloon BS-4, Tube Kit TY-500D

Indications for Use (Describe)

Endoscope EN-580T

This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine and rectum. Never use this product for any other purposes.

Over-tube TS-1314B

This product is used in combination with a FUJIFILM Double Balloon Endoscope to assist the insertion of the Endoscope under the management of physicians in medical facilities.

This product is used to assist with the movement of the scopes inside the upper or lower digestive tract.

Do not use this product for any other purposes.

This product is not intended for use for any neonates, infants or children.

Balloon BS-4

This product is used in combination with FUJIFILM Double Balloon Endoscopes at medical facilities under the management of physicians.

Being attached to the endoscope, this product is inserted into the mouth or anus to stabilize the distal end of the endoscope to the digestive tract's mucous membrane.

Do not use this product for any other purposes.

This product is not intended for use for any neonates, infants or children.

Tube Kit TY-500D

This product is the tube kit used in combination with the compatible balloon controller in medical facilities. Do not use this product for any other purpose.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

FUJIFILM Corporation

Endoscope EN-580T, Over-tube TS-13140, Balloon BS-3, Tube Kit TY-06D

January 6, 2023

Submitter's Information:

FUJIFILM Corporation 798 MIYANODAI KAISEI-MACHI ASHIGARAKAMI-GUN, KANAGAWA 258-8538 JAPAN

Contact Person:

Kotei Aoki Senior Regulatory Affairs Specialist E-Mail: kotei.aoki@fujifilm.com Telephone: (765) 246-2931

Identification of the Proposed Device:

Device Names:Endoscope EN-580T,Over-tube TS-13140, Balloon BS-3, Tube Kit TY-06D
Common Names:Endoscope, Accessories
Product Codes:FDA, FDF
Device Class:Class II
Regulation Number:21 CFR 876.1500
Regulation Description:Endoscope and Accessories
Review Panel:Gastroenterology/Urology

Predicate Device(s):

  • FUJIFILM Double Balloon Endoscope EN-580T (K183683) ●
  • Over-tube TS-13140 (K143556) ●
  • . Balloon BS-3 (K213195)
  • . Tube Kit TY-06D (K143556)

Intended Use / Indications for Use:

Endoscope EN-580T

This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine and rectum.

Never use this product for any other purposes.

Over-tube TS-1314B

This product is used in combination with a FUJIFILM Double Balloon Endoscope to assist the insertion of the Endoscope under the management of physicians in medical facilities.

This product is used to assist with the movement of the scopes inside the upper or lower digestive tract. Do not use this product for any other purposes.

This product is not intended for use for any neonates, infants or children.

{4}------------------------------------------------

Balloon BS-4

This product is used in combination with FUJIFILM Double Balloon Endoscopes at medical facilities under the management of physicians.

Being attached to the endoscope, this product is inserted into the digestive tract from the mouth or anus to stabilize the distal end of the endoscope to the digestive tract`s mucous membrane.

Do not use this product for any other purposes.

This product is not intended for use for any neonates, infants or children.

Tube Kit TY-500D

This product is the tube kit used in combination with the compatible balloon controller in medical facilities. Do not use this product for any other purpose.

Device Description:

The endoscope EN-580T is inserted both perorally and transanally into the gastrointestinal tract during clinical use. The insertion of the device has a mechanism which bends the tip from right to left and up and down, and a flexible tube consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument.

The over-tube TS-1314B is introduced in the patient's anatomy with the pairing endoscope. TS-1314B is assembled over the outer diameter of endoscope. The endoscope and the over-tube are not advanced both at the same time, but alternatingly and successively. TS-1314B is provided sterile and single-patient use only.

The balloon BS-4 is attached to the balloon-compatible endoscope. The balloon air feed outlet should be contained inside BS-4. The accompanying fixing rubber is used to affix the scope balloon in place. The fixing rubbers are made specifically for BS-4 but should not be mixed with the fixing rubber of other scope balloons. BS-4 is provided sterile and single-patient use only.

The tube kit TY-500D consists of a set of two tubes connects the over-tube to the balloon controller by its air inlet. The other tube connects the same balloon controller to the endoscope by its balloon air feed inlet. TY-500D is only compatible with PB-30, cleared K153483. TY-500D is provided non-sterile. The tubes are not patient-contacting and reusable, but each filter is single-patient use only.

Comparison of Technological Characteristics:

The proposed EN-580T differs from the predicate EN-580T in terms of the compatible accessories. The modifications are also proposed in some technical characteristics and the material construction of the said accessories. The proposed device and the predicate device share intended use, principle of operation, and other technical characteristics of the compatible accessories. A summary of modifications is provided below.

A comparison of technological characteristics between the proposed device and the predicate device is provided in Table 1 through Table 6:

{5}------------------------------------------------

Proposed DeviceEndoscopeEN-580T (K223295)Predicate DeviceFUJIFILM Double Balloon EndoscopeEN-580T (K183683)
Intended UseThis device is intended for thevisualization of the upper and lowerdigestive tracts, specifically for theobservation, diagnosis, and endoscopictreatment of the esophagus, stomach,duodenum, small intestine, largeintestine and rectum.Never use this product for any otherpurposes.This device is intended for thevisualization of the upper and lowerdigestive tracts. Specifically, for theobservation, diagnosis, and endoscopictreatment of the esophagus, stomach,duodenum, small intestine, largeintestine, and rectum.Never use this product for any otherpurposes.
Over-tubeTS-13140TS-1314B (proposed)TS-13140
Scope BalloonBS-2, BS-3BS-4 (proposed)BS-2, BS-3
Balloon ControllerPB-30PB-20, PB-30
Tube KitTY-06D [1]TY-500D (proposed) [2]TY-06D [1]

Table 1 Comparison of endoscopes – compatible accessories

[1] TY-06D should be paired only with TS-13140.

[2] TY-500D should be paired only with TS-1314B.

Table 2 Comparison of over-tubes

Table 2 Comparison of over-tubes
Proposed DeviceOver-tube TS-1314B (K223295)Predicate DeviceOver-tube TS-13140 (K143556)
Intended UseThis product is used in combinationwith a FUJIFILM Double BalloonEndoscope to assist the insertion ofthe Endoscope under the managementof physicians in medical facilities.This product is used to assist with themovement of the scopes inside theupper or lower digestive tract.Do not use this product for any otherpurposes.This product is not intended for use forany neonates, infants or children.These sterile Over-tubes areintended to be used asaccessories with theFUJINON/FUJIFILM DoubleBalloon Endoscopes cleared foruse with Over-tubes. The Over-tubes are used to assist with themovement of the scopes insidethe upper or lower digestive tract.This product is not intended foruse for any neonates, infants orchildren.
Working length / Total length1350mm / 1450mm
Insertion portion (diameters)(inner/outer/outer maximum)10.8mm/13.2mm/16mm
Outer diameter of balloon40mm
Working length of balloon50mm
Provided sterileYes, EO Sterilized
ReuseNo, Single patient use only
Product expiration3 years2 years
Water inletLuer connector (Female)
Connector typeAir inletLuer connector (Male)Luer connector (Female)
Transport and storageEnvironmentTemperature: -10 to +60°CHumidity: 30 to 95% (nocondensation)Atmosphere: 70 to 106 kPaTemperature: +10 to +40°CHumidity: 30 to 85% (nocondensation)Atmosphere: 70 to 106 kPa

{6}------------------------------------------------

Table 3 Comparison of balloons

Proposed DeviceBalloon BS-4 (K223295)Predicate DeviceBalloon BS-3 (K213195)
Intended UseThis product is used in combinationwith FUJIFILM Double BalloonEndoscopes at medical facilities underthe management of physicians.Being attached to the endoscope, thisproduct is inserted into the digestivetract from the mouth or anus tostabilize the distal end of theendoscope to the digestive tract'smucous membrane.Do not use this product for any otherpurposes.This product is not intended for use forany neonates, infants or children.Balloon BS-3 is intended to beused in combination withFUJIFILM double balloonendoscopes to assist withinsertion inside the upper orlower digestive tract at medicalfacilities under the managementof physicians.Do not use this product for anyother purpose. It is not intendedfor use for any neonates, infantsor children.
Outer diameter35mm
Thickness0.25mm0.10mm
Working length / Total length50mm / 66mm40mm / 60mm
Provided sterileYes, EO Sterilized
ReuseNo, Single patient use only
Product expiration3 years2 years
Transport and storageEnvironmentTemperature: -10 to +60°CHumidity: 30 to 95% (nocondensation)Atmosphere: 70 to 106 kPaTemperature: +10 to +40°CHumidity: 30 to 85% (nocondensation)Atmosphere: 70 to 106 kPa

Table 4 Comparison of tube kits

Proposed DeviceTube Kit TY-500D (K223295)Predicate DeviceTube Kit TY-06D (K143556)
Intended UseThis product is the tube kit used incombination with the compatibleballoon controller in medical facilities.Do not use this product for any otherpurpose.This product is the tube kit used incombination with the ballooncontroller PB-20 in medical facilities.Do not use this product for any otherpurpose.
WorkinglengthOver-tube side2050mm
Scope side1300mm
ConnectortypeBallooncontroller sideLuer connector (Female)
Over-tube sideLuer connector (Female)Luer connector (Male)
Scope sideDedicated connector
Provided sterileNo
ReuseTube: YesFilter: No, Single patient use only
Transport and storageEnvironmentTemperature: -10 to +45°CHumidity: 30 to 95%RH (no dew condensation)Atmosphere: 70 to 106 kPa

{7}------------------------------------------------

Proposed DeviceOver-tube TS-1314B (K223295)Predicate DeviceOver-tube TS-13140 (K143556)Contacttype
BalloonSilicone rubber(LSR 2030)Natural rubber(Fujilatex DPNR)ABSORBO HPDirect
Adhesive potionBond: Silicon series adhesive agentAdhesive coat: Silicon series adhesiveagent(Compound ratio A:B = 5:1)A) Silicon Rubber (YE3085)B) Silicon Sealing Agent (KE-42-T)Bond: Silicone compound(KE445B)Adhesive coat: N/A[1]
TubeTube: Silicon rubber(ELASTOSIL WS 9802 C)Coating: Polyvinylpyrrolidone(K Value:90)Tubing: Polyurethane(E380MNAT)Coating: Polyvinylpyrrolidone(Kollidon 30)Direct
Tip RingSilicon rubber(Compound ratio C:D = 1:1)C) Silicon Rubber(ELASTOSIL R 401/90 OH)D) Silicon Rubber w/ Barium Sulfate(ELASTOSIL EL 7704)Direct
HandleHandle: Silicon rubber(ELASTOSIL WS 9802 C)Coating: Polyvinylpyrrolidone(K Value:90)Indirect
Check ValveSilicon rubber(ELASTOSIL WS 9402 C)Indirect
Water feed inletSilicon rubber(ELASTOSIL WS 9802 C)Indirect
EndoscopeInsertion inletSilicon rubber(ELASTOSIL WS 9402 C)Polyvinylpyrrolidone(K Value:90)Indirect

Table 5 Material construction comparison of the over-tubes

[1] In the adhesive portion of the predicate device, the bonded portion is not coated and considered "Direct" contact type. On the other hand, the adhesive portion of the proposed device is coated. Consequently, the bonded portion is not patient contacting while the coating over the bonded portion is considered "Direct" contact type.

Table 6 Material construction comparison of the balloons

Proposed DeviceBalloon BS-4 (K223295)Predicate DeviceBalloon BS-3 (K213195)Contacttype
BalloonSilicone rubber(LSR 2030)Natural rubber(Fujilatex DPNR)Direct
ABSORBO HPDirect
Fixing rubberSilicone rubber(ELASTOSIL WS 9502 C)Natural rubber(Fujilatex DPNR)Direct
ABSORBO HPDirect

{8}------------------------------------------------

Performance Data:

Sterility of TS-1314B and BS-4 were evaluated using the following consensus standards: ISO 11135:2014. ISO 10993-7:2008/AMD1:2019, ISO 11607-1:2019, ISO 11607-2:2019, ISO 11737-1:2018, and ISO 11138-1:2017.

The new accessories are made of different materials compared to the respective predicate devices. Biocompatibility of each new accessory was evaluated using the following consensus standards: ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010, and ISO 10993-12:2012. Biocompatibility testing was performed in accordance with FDA's guidance, Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', issued September 4, 2020.

Endoscope specific testing was conducted according to ISO 8600-1:2015. ISO 8600-3:2019, and ISO 8600-4:2014. Endoscope compatibility with the new accessories was conducted with acceptable results.

Conclusions:

The proposed devices Endoscope EN-580T, Over-tube TS-13140, Balloon BS-3, and Tube Kit TY-06D shares the same intended use as the respective predicate devices. The endoscope's compatibility with the new accessories was evaluated. The differences among the accessories and their respective predicate devices have been evaluated for the biocompatibility and bench testing with acceptable results. Thus, the proposed device Endoscope EN-580T, Over-tube TS-13140, Balloon BS-3, and Tube Kit TY-06D are substantially equivalent to the respective predicate devices, FUJIFILM Double Balloon Endoscope EN-580T (K183683), Over-tube TS-13140 (K143556), Balloon BS-3 (K213195), and Tube Kit TY-06D (K143556).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.