K Number
K183683

Validate with FDA (Live)

Date Cleared
2019-02-27

(61 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EN-580T, EN-450P5/20 and EN-450T5 are intended for the upper and lower digestive tracts. Specifically, for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine, and rectum.

EC-450B15 is intended for the optical visualization of the gastrointestinal tract. This includes the rectum, large and small intestines. It is intended for observation, diagnosis, and endoscopic treatment.

Device Description

FUJIFILM Double Balloon Endoscopes EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5 are comprised of three general sections: a control portion, an insertion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary Charge-Coupled Device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when pluqged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.

AI/ML Overview

The document describes the regulatory submission for FUJIFILM Double Balloon Endoscopes, focusing on modifications made to existing devices. The acceptance criteria and the study proving the device meets these criteria are outlined in the "Performance Data" section (pages 4-5).

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states that a series of performance tests were conducted to ensure the modified devices perform equivalently to the predicate devices. The acceptance criteria were "pre-defined," and in all cases, the devices met these criteria. However, the specific quantitative acceptance criteria for each test and the precise reported device performance values are not explicitly detailed in the provided text. Instead, a list of tested parameters is given.

Acceptance Criteria (Stated)Reported Device Performance
Field of viewMet pre-defined criteria
Bending capabilityMet pre-defined criteria
Rate of air supplyMet pre-defined criteria
Rate of water supplyMet pre-defined criteria
Suction rateMet pre-defined criteria
Working lengthMet pre-defined criteria
Forceps channel diameterMet pre-defined criteria
Viewing directionMet pre-defined criteria
ResolutionMet pre-defined criteria
LG outputMet pre-defined criteria
Electrical SafetyMet ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009 standards
BiocompatibilityMet ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010 standards
Reprocessing ValidationMet AAMI TIR12:2010, AAMI TIR30:2011, and FDA's guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (March 17, 2015)

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample size (number of devices or tests performed) for the performance testing. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) for these performance studies.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not applicable and not provided in the document. The performance tests described (e.g., field of view, bending capability, electrical safety, biocompatibility) are objective engineering and material science evaluations, not assessments requiring expert interpretation of clinical data to establish ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable and not provided. The performance tests described are objective and do not involve adjudication by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence of modified endoscopes to predicate devices through technical and physical performance metrics, not on evaluating human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

No, a standalone (algorithm-only) performance study was not done. This device is an endoscope, a physical instrument, and does not involve AI algorithms for diagnosis or analysis.

7. The Type of Ground Truth Used:

The "ground truth" for the performance tests was based on established engineering specifications, consensus standards (e.g., ISO, AAMI, IEC), and regulatory guidance documents. For instance, electrical safety was measured against specific IEC standards, and biocompatibility against ISO standards. The validation of reprocessing instructions was performed in accordance with FDA guidance.

8. The Sample Size for the Training Set:

This information is not applicable and not provided. The studies described are performance tests of physical devices against pre-defined specifications, not machine learning model training.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable and not provided for the reason mentioned above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name below it.

February 27, 2019

FUJIFILM Corporation % Jeffrey Wan Specialist. Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc 81 Hartwell Avenue, Suite 300 Lexington, MA 02421

Re: K183683

Trade/Device Name: FUJIFILM Double Balloon Endoscopes EN-580T, EC-450B15, EN-450P5/20, and EN-450T5 Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDA, FDF Dated: December 27, 2018 Received: December 28, 2018

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183683

Device Name

FUJIFILM Double Balloon Endoscopes EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5

Indications for Use (Describe)

EN-580T, EN-450P5/20 and EN-450T5 are intended for the upper and lower digestive tracts. Specifically, for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine, and rectum.

EC-450B15 is intended for the optical visualization of the gastrointestinal tract. This includes the rectum, large and small intestines. It is intended for observation, diagnosis, and endoscopic treatment.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

FUJIFILM Double Balloon Endoscopes EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5

Date: December 27, 2018

Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

Contact Person:

Jeffrey Wan Specialist, Regulatory Affairs Telephone: (201) 675-8947 E-Mail: jeffrey.wan@fujifilm.com

Identification of the Proposed Device:

Proprietary/Trade Name:FUJIFILM Double Balloon Endoscopes EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5
Common Name:Enteroscope (EN-580T, EN-450P5/20, EN-450T5)Colonoscope (EC-450BI5)
Device Class:Class II
Review Panel:Gastroenterology/Urology
Classification:Endoscope and accessories, 21 C.F.R. § 876.1500
Product Code:FDA, FDF

Predicate Devices:

FUJIFILM Double Balloon Endoscope Model EN-580T (K143556) Fujinon EC-450BI5 (K090116) Fujinon EN-450P5/20 (K040048)

Intended Use / Indications for Use:

EN-580T, EN-450P5/20 and EN-450T5 are intended for the visualization of the upper and lower digestive tracts. Specifically, for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine, and rectum.

EC-450Bl5 is intended for the optical visualization of the gastrointestinal tract. This includes the rectum, large and small intestines. It is intended for observation, diagnosis, and endoscopic treatment.

{4}------------------------------------------------

Device Description:

FUJIFILM Double Balloon Endoscopes EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5 are comprised of three general sections: a control portion, an insertion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary Charge-Coupled Device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when pluqged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.

Technological Characteristics:

The proposed devices EN-580T, EC-450Bl5, EN-450P5/20, and EN-450T5 differ from the predicate devices in the following minor modifications:

  • . Compatibility with different components and accessories
  • . Expansion of transport and storage conditions to a temperature range of -20°C to 60°C and a humidity range of 10 to 85% RH.
  • . Use of a new epoxy resin in the manufacturing process. Both the new and original epoxy resins can be used for repair.
  • . Removal of the option to inflate a balloon by manual syringe.
  • . Design specification changes (EN-450T5 only)

Performance Data:

Electrical safety of the proposed devices were evaluated using the following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.

The proposed devices EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5 were adopted into biocompatibility testing conducted on the predicate devices EN-580T, EC-450B15, EN-450P5/20, and EN-450T5 using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.

The proposed devices EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5 were adopted into the reprocessing validation of the predicate devices using the following consensus standards: AAMI TIR12:2010, AAMI TIR30:2011. Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17, 2015.

Fuilfilm conducted the following performance testing on the proposed devices EN-580T. EC-450BI5. EN-450P5/20, and EN-450T5 to ensure that the modified devices perform equivalently to the predicate devices:

{5}------------------------------------------------

  • Field of view ●
  • Bending capability
  • Rate of air supply ●
  • . Rate of water supply
  • Suction rate
  • . Working length
  • . Forceps channel diameter
  • Viewing direction ●
  • . Resolution
  • . LG output

In all cases, the devices met the pre-defined acceptance criteria for the test.

Substantial Equivalence:

The company's EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5 have the same intended use and indications for use as the previously cleared predicates EN-580T(K143556), EC-450Bl5 (K090116), and EN-450P5/20 (K040048). In addition, the proposed devices have similar technological characteristics and principles of operation as the predicates. The minor differences between the proposed devices and their predicate devices do not raise new or additional questions of safety or effectiveness of the proposed devices. Thus, the proposed devices EN-580T, EC-450BI5, EN-450P5/20, and EN-450T5 are substantially equivalent to their predicate devices.

Conclusions:

The modified EN-580T, EC-450B15, EN-450P5/20, and EN-450T5 are substantially equivalent to the predicate EN-580T, EC-450B15, and EN-450P5/20 and conform to applicable medical device safety and performance standards.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.