K Number
K153483

Validate with FDA (Live)

Date Cleared
2015-12-15

(12 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is a balloon pump apparatus intended to inflate or deflate a balloon to assist with endoscope and over tube insertion. This product is not intended for use for any neonates, infants or children.

Device Description

The Fujifilm Balloon Controller PB-30 is a modified version of the legally marketed Fujifilm Balloon Controller PB-20 in K143556. Just like K143556, the proposed PB-30 is a balloon pump apparatus intended to inflate or deflate a balloon to assist with endoscope and over tube insertion. This product is not intended for use for any neonates, infants or children.

PB-30 is connected to the over-tube and double balloon endoscope via a tube kit. PB-30 feeds /evacuates air to the balloon of over-tube and to the balloon at the tip of the endoscope. All of the PB-30's accessories are legally marketed in the US.

AI/ML Overview

This document is a 510(k) premarket notification for the Fujifilm Balloon Controller PB-30. It states that the PB-30 is a modification of the legally marketed Fujifilm Balloon Controller PB-20. As such, the purpose of the submission is to demonstrate that the modified device performs equivalently to the predicate.

Here's an analysis of the provided information concerning acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
EMC and Electrical Safety: Based on consensus standards: ANSI/AAMI ES60601-1:2005 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)IEC 60601-1-2:2007 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests)IEC 60601-1-6:2010 (Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability)IEC 60601-2-18:2009 (Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment)"In all cases, the device met the pre-defined acceptance criteria for the test."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a separate "test set" in the context of clinical data for performance evaluation. The performance testing described relates to EMC and electrical safety, likely involving single units or a small sample of the device, rather than a large patient-based test set.

  • Sample Size: Not explicitly stated, but typically involves a limited number of devices (e.g., 1-3) for electrical and EMC testing.
  • Data Provenance: Not specified, but likely from laboratory testing conducted by Fujifilm or a certified testing facility. It is not clinical data, so country of origin or retrospective/prospective does not apply in the usual sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not applicable. The performance testing for this device (EMC and electrical safety) does not involve human expert interpretation or ground truth establishment in the way a diagnostic imaging device might. The "ground truth" for these tests would be the established international standards themselves.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are used for reconciling discordant expert interpretations in clinical studies. The testing performed here (EMC and electrical safety) does not involve such a process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence through technical performance testing (EMC and electrical safety) and functional equivalence to the predicate device (PB-20), not on human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, this question is not applicable to the device described. The Balloon Controller PB-30 is a hardware device (a balloon pump apparatus) and does not involve an AI algorithm or human-in-the-loop performance testing.

7. The Type of Ground Truth Used

The "ground truth" for the performance testing described (EMC and electrical safety) is defined by established international consensus standards (e.g., ANSI/AAMI ES60601-1, IEC 60601-1-2). Compliance with these standards serves as the ground truth for safety and performance in these specific areas.

8. The Sample Size for the Training Set

This information is not applicable. The device is a hardware product, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for this hardware device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, wave-like appearance. Encircling the graphic is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 15, 2015

Fujifilm Medical Systems USA, Inc. Aaron Ge RA Manager, Regulatory Affairs and Quality Assurance 10 High Point Drive Wayne, New Jersey 07470

Re: K153483

Trade/Device Name: Balloon Controller PB-30 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: FDF, FDA Dated: November 30, 2015 Received: December 3, 2015

Dear Aaron Ge:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153483

Device Name Balloon Controller PB-30

Indications for Use (Describe)

This product is a balloon pump apparatus intended to inflate or deflate a balloon to assist with endoscope and over tube insertion. This product is not intended for use for any neonates, infants or children.

Type of Use (Select one or both, as applicable)

Commercial and/or Industrial (C&I) Vehicle Registration
Other - This transaction requires a Utah STID

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Fujifilm's Balloon Controller PB-30

Submitter's Information

FUJIFILM Medical Systems U.S.A., Inc. Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Reqistration Number: 2431293

Contact Person:

Aaron Ge Regulatory Manager, Regulatory Affairs and Quality Assurance Telephone: (973) 686-2636 Ext. 522636 Facsimile: (973) 633-8818 E-Mail age@fujifilm.com

Date Prepared: November 30, 2015

ldentification of the Subject Device:

Proprietary/Trade Name:PB-30
Common Name:Balloon Controller
Device Class:Class II
Review Panel:Gastroenterology/Urology

Classification Information:

Colonoscope and Accessories (Flexible/Rigid), 21 C.F.R. § 876.1500 Product Code: FDF, FDA

Primary Predicate

Fujifilm Balloon Controller PB-20 (K143556)

Purpose of the Special 510(k) notice.

The PB-30 is a modification to PB-20.

Intended Use

{4}------------------------------------------------

This product is a balloon pump apparatus intended to inflate or deflate a balloon to assist with endoscope and over tube insertion. This product is not intended for use for any neonates, infants or children.

Technological Characteristics

The Fujifilm Balloon Controller PB-30 is a modified version of the legally marketed Fujifilm Balloon Controller PB-20 in K143556. Just like K143556, the proposed PB-30 is a balloon pump apparatus intended to inflate or deflate a balloon to assist with endoscope and over tube insertion. This product is not intended for use for any neonates, infants or children.

PB-30 is connected to the over-tube and double balloon endoscope via a tube kit. PB-30 feeds /evacuates air to the balloon of over-tube and to the balloon at the tip of the endoscope. All of the PB-30's accessories are legally marketed in the US.

Performance Data

Fujifilm conducted the following performance testing of the PB-30 to ensure that the modified device performs equivalently to the predicate PB-20:

EMC and electrical safety of the subject devices were evaluated using the following consensus standards: ANSI/AAMI ES60601-1:2005; IEC 60601-1-2:2007; IEC 60601-1-6:2010; and IEC 60601-2-18:2009.

In all cases, the device met the pre-defined acceptance criteria for the test.

Substantial Equivalence

PB-30 has the same intended use and similar indications, principles of operation, and technological characteristics as the predicate PB-20. The minor differences in the modified device's technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the PB-30 is as safe and effective as PB-20. Thus, the PB-30 is substantially equivalent to its predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.