AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The nitrile examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device.

The Nitrile Patient Examination Gloves Blue Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The tested chemotherapy drug are as follows:

Bleomycin Sulfate 15 mg/ml >240 min. Busulfan 6 mg/ml >240 min. Carboplatin 10 mg/ml >240 min. Carmustine (BCNU) 3.3 mg/ml 17.2 min. Cisplatin 1.0 mg/ml >240 min. Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. Cytarabine HCI 100 mg/ml >240 min. Dacarbazine (DTIC)10.0 mg/ml >240 min. Daunorubicin 5 mg/mL >240 min. Docetaxel 10.0 mg/ml >240 min Doxorubicin HCI 2.0 mg/ml >240 min. Epirubicin HCI 2.0 mg/ml >240 min. Etoposide (Toposar) 20.0 mg/ml >240 min. Fludarabine 25.0 mg/ml >240 min. Fluorouracil 50.0 mg/ml >240 min. Gemcitabine 38 mg/ml>240 min. Idarubicin 1 mg/ml >240 min. Ifosfamide 50.0 mg/ml >240 min. Irinotecan 20.0 mg/ml >240 min. Mechlorethamine HCI 1.0 mg/ml>240 min. Melphalan 5 mg/ml >240 min. Methotrexate 25 mg/ml >240 min. Mitromvcin C. 0.5 mg/ml >240 Mitoxantrone 2.0 mg/ml >240 min. Paclitaxel (Taxol) 6.0 mg/ml >240 min. Rituximab 10.0 mg/ml >240 min. Thiotepa 10.0 mg/ml 13.9min. Trisenox 1.0 mg/ml >240 min. Vincristine Sulfate 1.0 mg/ml >240 min. Fentanyl Citrate 100mcg/2ml >240 minutes

Warning: Do not use with Carmustine and Thiotepa.

The Nitrile Patient Examination Gloves Blue Violet Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The tested chemotherapy drug are as follows: Bleomycin Sulfate 15 mg/ml >240 min. Busulfan 6 mg/ml >240 min. Carboplatin 10 mg/ml >240 min. Carmustine (BCNU) 3.3 mg/ml 65.3 min. Cisplatin 1.0 mg/ml >240 min. Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. Cytarabine HCI 100 mg/ml >240 min. Dacarbazine (DTIC)10.0 mg/ml >240 min. Daunorubicin 5 mg/mL >240 min. Docetaxel 10.0 mg/ml >240 min Doxorubicin HCI 2.0 mg/ml >240 min. Epirubicin HCl 2.0 mg/ml >240 min. Etoposide (Toposar) 20.0 mg/ml >240 min. Fludarabine 25.0 mg/ml >240 min. Fluorouracil 50.0 mg/ml >240 min. Gemcitabine 38 mg/ml>240 min. Idarubicin 1 mg/ml >240 min. Ifosfamide 50.0 mg/ml >240 min. Irinotecan 20.0 mg/ml >240 min. Mechlorethamine HCI 1.0 mg/ml>240 min. Melphalan 5 mg/ml >240 min. Methotrexate 25 mg/ml >240 min. Mitromvcin C. 0.5 mg/ml >240 Mitoxantrone 2.0 mg/ml >240 min. Paclitaxel (Taxol) 6.0 mg/ml >240 min. Rituximab 10.0 mg/ml >240 min. Thiotepa 10.0 mg/ml 58.3min. Trisenox 1.0 mg/ml >240 min. Vincristine Sulfate 1.0 mg/ml >240 min. Fentanyl Citrate 100mcg/2ml >240 minutes

Warning: Do not use with Carmustine and Thiotepa.

The Nitrile Patient Examination Gloves Black Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The tested chemotherapy drug are as follows: Carmustine (BCNU) 3.3 mg/ml 49.2 min. Cisplatin 1.0 mq/ml >240 min. Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. Dacarbazine (DTIC)10.0 mg/ml >240 min. Doxorubicin HCI 2.0 mg/ml >240 min. Etoposide (Toposar) 20.0 mg/ml >240 min. Fluorouracil 50.0 mg/ml >240 min. Fentanyl Citrate 100mcg/2ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 min. Thiotepa 10.0 mg/ml 87.1min.

Warning: Do not use with Carmustine and Thiotepa.

Device Description

Powder-Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs) are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are offered in six sizes, extra-small, medium, large, extralarge, extra-extra-large. Three colors are available for all size, includes blue, blue violet and black.

The gloves are designed and manufactured in accordance with the ASTM D6319-10 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019).

AI/ML Overview

This document describes the premarket notification (510(k)) for Nitrile Patient Examination Gloves, with specific testing for chemotherapy drug permeation. Based on the provided text, the "device" is a physical product (gloves) and not an AI/software device, therefore many of the requested criteria (MRMC study, AI assistance, ground truth establishment, training set, etc.) are not applicable.

Here's the breakdown of the acceptance criteria and the study proving the device meets them, based entirely on the provided text:


Acceptance Criteria and Reported Device Performance

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodologyPurposeAcceptance CriteriaResults
ASTM D6319 (Physical Dimensions)Physical Dimensions TestLength:Extra-Small: ≥ 220mmSmall: ≥ 220mmMedium: ≥ 230mmLarge: ≥ 230mmExtra-Large: ≥ 230mmExtra-Extra-Large: ≥ 230mmWidth:Extra-Small: 70±10mmSmall: 80±10mmMedium: 95±10mmLarge: 110±10mmExtra-Large: 120±10mmExtra-Extra-Large: ≥120mmThickness (mm):Finger: ≥0.05Palm: ≥0.08Pass
ASTM D6319 (Physical Properties)Physical Properties (Tensile Strength & Elongation)Before Aging:Tensile Strength: ≥14MPaUltimate Elongation: ≥500%After Aging:Tensile Strength: ≥14MPaUltimate Elongation: ≥500%Pass
ASTM D5151Freedom from pinholesMeet the requirements of ASTM D5151 Test for AQL 2.5Pass
ASTM D6124Powder ResidueMeet the requirements of ASTM D6124 < 2.0mgPass
ISO 10993-10To determine if the finished device material is an irritantNon-irritatingUnder the conditions of the study, not an irritant/ Pass
ISO 10993-10To determine if the finished device material is a sensitizerNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass
ISO 10993-11To determine if the finished device material extracts pose a systemic toxicity concernNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo / Pass
ASTM D6978-05 (Reapproved 2019)Assessment of Medical Gloves to Permeation by Chemotherapy Drugs and Fentanyl CitrateNitrile Patient Examination Gloves Blue (Tested Drugs & Permeation Time):- Bleomycin Sulfate 15 mg/ml: >240 min.- Busulfan 6 mg/ml: >240 min.- Carboplatin 10 mg/ml: >240 min.- Carmustine (BCNU) 3.3 mg/ml: 17.2 min.- Cisplatin 1.0 mg/ml: >240 min.- Cyclophosphamide (Cytoxan) 20.0 mg/ml: >240 min.- Cytarabine HCI 100 mg/ml: >240 min.- Dacarbazine (DTIC)10.0 mg/ml: >240 min.- Daunorubicin 5 mg/mL: >240 min.- Docetaxel 10.0 mg/ml: >240 min- Doxorubicin HCI 2.0 mg/ml: >240 min.- Epirubicin HCI 2.0 mg/ml: >240 min.- Etoposide (Toposar) 20.0 mg/ml: >240 min.- Fludarabine 25.0 mg/ml: >240 min.- Fluorouracil 50.0 mg/ml: >240 min.- Gemcitabine 38 mg/ml: >240 min.- Idarubicin 1 mg/ml: >240 min.- Ifosfamide 50.0 mg/ml: >240 min.- Irinotecan 20.0 mg/ml: >240 min.- Mechlorethamine HCI 1.0 mg/ml: >240 min.- Melphalan 5 mg/ml: >240 min.- Methotrexate 25 mg/ml: >240 min.- Mitomycin C. 0.5 mg/ml: >240- Mitoxantrone 2.0 mg/ml: >240 min.- Paclitaxel (Taxol) 6.0 mg/ml: >240 min.- Rituximab 10.0 mg/ml: >240 min.- Thiotepa 10.0 mg/ml: 13.9 min.- Trisenox 1.0 mg/ml: >240 min.- Vincristine Sulfate 1.0 mg/ml: >240 min.- Fentanyl Citrate 100mcg/2ml: >240 minutesWarning: Do not use with Carmustine and Thiotepa.Nitrile Patient Examination Gloves Blue Violet (Tested Drugs & Permeation Time):- Bleomycin Sulfate 15 mg/ml: >240 min.- Busulfan 6 mg/ml: >240 min.- Carboplatin 10 mg/ml: >240 min.- Carmustine (BCNU) 3.3 mg/ml: 65.3 min.- Cisplatin 1.0 mg/ml: >240 min.- Cyclophosphamide (Cytoxan) 20.0 mg/ml: >240 min.- Cytarabine HCI 100 mg/ml: >240 min.- Dacarbazine (DTIC)10.0 mg/ml: >240 min.- Daunorubicin 5 mg/mL: >240 min.- Docetaxel 10.0 mg/ml: >240 min- Doxorubicin HCI 2.0 mg/ml: >240 min.- Epirubicin HCl 2.0 mg/ml: >240 min.- Etoposide (Toposar) 20.0 mg/ml: >240 min.- Fludarabine 25.0 mg/ml: >240 min.- Fluorouracil 50.0 mg/ml: >240 min.- Gemcitabine 38 mg/ml: >240 min.- Idarubicin 1 mg/ml: >240 min.- Ifosfamide 50.0 mg/ml: >240 min.- Irinotecan 20.0 mg/ml: >240 min.- Mechlorethamine HCI 1.0 mg/ml: >240 min.- Melphalan 5 mg/ml: >240 min.- Methotrexate 25 mg/ml: >240 min.- Mitomycin C. 0.5 mg/ml: >240- Mitoxantrone 2.0 mg/ml: >240 min.- Paclitaxel (Taxol) 6.0 mg/ml: >240 min.- Rituximab 10.0 mg/ml: >240 min.- Thiotepa 10.0 mg/ml: 58.3 min.- Trisenox 1.0 mg/ml: >240 min.- Vincristine Sulfate 1.0 mg/ml: >240 min.- Fentanyl Citrate 100mcg/2ml: >240 minutesWarning: Do not use with Carmustine and Thiotepa.Nitrile Patient Examination Gloves Black (Tested Drugs & Permeation Time):- Carmustine (BCNU) 3.3 mg/ml: 49.2 min.- Cisplatin 1.0 mg/ml: >240 min.- Cyclophosphamide (Cytoxan) 20.0 mg/ml: >240 min.- Dacarbazine (DTIC)10.0 mg/ml: >240 min.- Doxorubicin HCI 2.0 mg/ml: >240 min.- Etoposide (Toposar) 20.0 mg/ml: >240 min.- Fluorouracil 50.0 mg/ml: >240 min.- Fentanyl Citrate 100mcg/2ml: >240 minutes- Paclitaxel (Taxol) 6.0 mg/ml: >240 min.- Thiotepa 10.0 mg/ml: 87.1 min.Warning: Do not use with Carmustine and Thiotepa.The permeation resistance was tested and the break-through times for each listed chemotherapy drug are reported as the "Results" for each glove colored variation. The tables within the document (pages 8-10, and 12-14) explicitly list these times. The general acceptance criterion for these types of tests is generally that the glove provides a minimum specified protection time (e.g., >240 minutes for many drugs, unless a shorter warning is given), and these results are directly reported. The "Pass" in the summary table (page 18) indicates overall compliance with the standard for specific tests.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the exact sample size (number of gloves) used for each individual test (e.g., how many gloves were tested for permeation by each drug). It generally refers to "tests" being performed.
  • Data Provenance: The tests were conducted according to established ASTM and ISO standards for medical devices. The manufacturer is "Guang Dong Kingfa Sci. & Tech.Co., Ltd." in China. The data itself is from laboratory testing, not from patient data. The studies are non-clinical (in-vitro, physical/chemical tests).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This is a physical device (medical gloves) and not an AI/software device that requires expert human interpretation or consensus for ground truth establishment. The "ground truth" here is the measured performance of the glove against established physical and chemical standards (e.g., tensile strength, pinhole presence, chemical permeation time).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. As this is not an AI/software device involving human-in-the-loop decisions or human interpretive tasks, no adjudication method for ground truth was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is a physical device, not an AI/software system for medical image analysis or diagnosis.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical device, not an algorithm. The device's performance is standalone in that its properties (e.g., permeation resistance) are directly measured.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for these tests is defined by adherence to published international standards (ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993-10, ISO 10993-11, and particularly ASTM D6978-05 for chemotherapy drug permeation). For the chemotherapy drug permeation specifically, the ground truth is the experimentally determined breakthrough time of the specific chemicals through the glove material under controlled laboratory conditions, as specified by the standard.

8. The sample size for the training set:

  • Not Applicable. This is a physical device; there is no 'training set' in the machine learning sense. The manufacturing process and quality control would involve internal testing, but this is not referred to as a "training set" in the context of typical AI/ML device submissions.

9. How the ground truth for the training set was established:

  • Not Applicable. As there is no training set for an AI/ML model, this point does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

October 7, 2022

Guang Dong Kingfa Sci. & Tech.Co., Ltd. Xiaoge Yu Director No. 28 Delong Ave., Shijiao Town, Qingcheng District Qingyuan, Guangdong 511545 China

Re: K222714

Trade/Device Name: Nitrile Patient Examination Gloves Blue Tested For Use With Chemotherapy Drugs, Nitrile Patient Examination Gloves Blue Violet Tested For Use With Chemotherapy Drugs, Nitrile Patient Examination Gloves Black Tested For Use

With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, ODO Dated: August 30, 2022 Received: September 8, 2022

Dear Xiaoge Yu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K222714

Device Name

Nitrile Patient Examination Gloves Blue Tested For Use With Chemotherapy Drugs.

Indications for Use (Describe)

The nitrile examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device.

The Nitrile Patient Examination Gloves Blue Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The tested chemotherapy drug are as follows:

Bleomycin Sulfate 15 mg/ml >240 min. Busulfan 6 mg/ml >240 min. Carboplatin 10 mg/ml >240 min. Carmustine (BCNU) 3.3 mg/ml 17.2 min. Cisplatin 1.0 mg/ml >240 min. Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. Cytarabine HCI 100 mg/ml >240 min. Dacarbazine (DTIC)10.0 mg/ml >240 min. Daunorubicin 5 mg/mL >240 min. Docetaxel 10.0 mg/ml >240 min Doxorubicin HCI 2.0 mg/ml >240 min. Epirubicin HCI 2.0 mg/ml >240 min. Etoposide (Toposar) 20.0 mg/ml >240 min. Fludarabine 25.0 mg/ml >240 min. Fluorouracil 50.0 mg/ml >240 min. Gemcitabine 38 mg/ml>240 min. Idarubicin 1 mg/ml >240 min. Ifosfamide 50.0 mg/ml >240 min. Irinotecan 20.0 mg/ml >240 min. Mechlorethamine HCI 1.0 mg/ml>240 min. Melphalan 5 mg/ml >240 min. Methotrexate 25 mg/ml >240 min. Mitromvcin C. 0.5 mg/ml >240 Mitoxantrone 2.0 mg/ml >240 min. Paclitaxel (Taxol) 6.0 mg/ml >240 min. Rituximab 10.0 mg/ml >240 min. Thiotepa 10.0 mg/ml 13.9min. Trisenox 1.0 mg/ml >240 min. Vincristine Sulfate 1.0 mg/ml >240 min. Fentanyl Citrate 100mcg/2ml >240 minutes

Warning: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K222714

Device Name

Nitrile Patient Examination Gloves Blue Violet Tested For Use With Chemotherapy Drugs

Indications for Use (Describe)

The nitrile examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device.

The Nitrile Patient Examination Gloves Blue Violet Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The tested chemotherapy drug are as follows: Bleomycin Sulfate 15 mg/ml >240 min. Busulfan 6 mg/ml >240 min. Carboplatin 10 mg/ml >240 min. Carmustine (BCNU) 3.3 mg/ml 65.3 min. Cisplatin 1.0 mg/ml >240 min. Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. Cytarabine HCI 100 mg/ml >240 min. Dacarbazine (DTIC)10.0 mg/ml >240 min. Daunorubicin 5 mg/mL >240 min. Docetaxel 10.0 mg/ml >240 min Doxorubicin HCI 2.0 mg/ml >240 min. Epirubicin HCl 2.0 mg/ml >240 min. Etoposide (Toposar) 20.0 mg/ml >240 min. Fludarabine 25.0 mg/ml >240 min. Fluorouracil 50.0 mg/ml >240 min. Gemcitabine 38 mg/ml>240 min. Idarubicin 1 mg/ml >240 min. Ifosfamide 50.0 mg/ml >240 min. Irinotecan 20.0 mg/ml >240 min. Mechlorethamine HCI 1.0 mg/ml>240 min. Melphalan 5 mg/ml >240 min. Methotrexate 25 mg/ml >240 min. Mitromvcin C. 0.5 mg/ml >240 Mitoxantrone 2.0 mg/ml >240 min. Paclitaxel (Taxol) 6.0 mg/ml >240 min. Rituximab 10.0 mg/ml >240 min. Thiotepa 10.0 mg/ml 58.3min. Trisenox 1.0 mg/ml >240 min. Vincristine Sulfate 1.0 mg/ml >240 min. Fentanyl Citrate 100mcg/2ml >240 minutes

Warning: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K222714

Device Name

Nitrile Patient Examination Gloves Black Tested For Use With Chemotherapy Drugs

Indications for Use (Describe)

The nitrile examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device.

The Nitrile Patient Examination Gloves Black Colored were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The tested chemotherapy drug are as follows: Carmustine (BCNU) 3.3 mg/ml 49.2 min. Cisplatin 1.0 mq/ml >240 min. Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. Dacarbazine (DTIC)10.0 mg/ml >240 min. Doxorubicin HCI 2.0 mg/ml >240 min. Etoposide (Toposar) 20.0 mg/ml >240 min. Fluorouracil 50.0 mg/ml >240 min. Fentanyl Citrate 100mcg/2ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 min. Thiotepa 10.0 mg/ml 87.1min.

Warning: Do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{7}------------------------------------------------

510(k) Summary – K222714

l. Submitter

GUANG DONG KINGFA SCI. & TECH.CO., LTD. No. 28 Delong Avenue, Shijiao Town, Qingcheng District, Qingyuan City, Guangdong Province 511545, China

Contact person: Xiaoge Yu Position: Manager Tel.: +86-13570952157 E-mail: yuxiaoge@kingfa.com.cn

Preparation date: August. 30, 2022

II. Proposed Device

Device Trade NameNitrile Patient Examination Gloves Blue Tested For Use WithChemotherapy Drugs
Nitrile Patient Examination Gloves Blue Violet Tested For UseWith Chemotherapy Drugs
Nitrile Patient Examination Gloves Black Tested For Use WithChemotherapy Drugs
Common name:Patient Examination Glove (Tested For Use With ChemotherapyDrugs)
Regulation Number:21 CFR 880.6250
Regulatory Class:Class I
Product code:LZA, LZC, OPJ, QDO
Review PanelGeneral Hospital

III. Predicate Devices

510(k) Number:K213040
Trade name:Nitrile Patient Examination Gloves Blue Tested For Use WithChemotherapy DrugsNitrile Patient Examination Gloves Blue Violet Tested For UseWith Chemotherapy DrugsNitrile Patient Examination Gloves Black Tested For Use WithChemotherapy Drugs
Common name:Patient Examination Glove
Classification:Class I

{8}------------------------------------------------

Product Code: LZA, LZC,OPJ, QDO

IV. Device description

Powder-Free Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs) are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are offered in six sizes, extra-small, medium, large, extralarge, extra-extra-large. Three colors are available for all size, includes blue, blue violet and black.

The gloves are designed and manufactured in accordance with the ASTM D6319-10 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019).

V. Indication for use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

The Nitrile Patient Examination Gloves Blue Tested Chemotherapy Drugs are as follows:

  • Bleomycin Sulfate 15 mg/ml >240 min. .
  • Busulfan 6 mg/ml >240 min. .
  • Carboplatin 10 mg/ml >240 min. .
  • Carmustine (BCNU) 3.3 mg/ml 17.2 min. .
  • Cisplatin 1.0 mg/ml >240 min. .
  • Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. ●
  • Cytarabine HCI 100 mg/ml >240 min. .
  • Dacarbazine (DTIC)10.0 mg/ml >240 min. .
  • Daunorubicin 5 mg/mL>240 min. .
  • Docetaxel 10.0 mg/ml >240 min .
  • Doxorubicin HCI 2.0 mg/ml >240 min. .
  • Epirubicin HCI 2.0 mg/ml >240 min. .
  • Etoposide (Toposar) 20.0 mg/ml >240 min. .

{9}------------------------------------------------

K222714

  • Fludarabine 25.0 mg/ml >240 min. .
  • Fluorouracil 50.0 mg/ml >240 min. .
  • Gemcitabine 38 mg/ml>240 min. .
  • ldarubicin 1 mg/ml >240 min. .
  • lfosfamide 50.0 mg/ml >240 min. .
  • Irinotecan 20.0 mg/ml >240 min. .
  • Mechlorethamine HCl 1.0 mq/ml>240 min. .
  • Melphalan 5 mg/ml >240 min. .
  • Methotrexate 25 mq/ml >240 min. .
  • Mitromycin C. 0.5 mg/ml >240 .
  • Mitoxantrone 2.0 mg/ml >240 min. .
  • Paclitaxel (Taxol) 6.0 mg/ml >240 min. .
  • Rituximab 10.0 mg/ml >240 min. .
  • Thiotepa 10.0 mq/ml 13.9min. .
  • Trisenox 1.0 mg/ml >240 min. .
  • Vincristine Sulfate 1.0 mg/ml >240 min. .
  • Fentanyl Citrate 100mcg/2ml >240 minutes .

Warning: Do not use with Carmustine and Thiotepa.

The Nitrile Patient Examination Gloves Blue Violet Tested Chemotherapy Drugs are as follows:

  • Bleomycin Sulfate 15 mg/ml >240 min. .
  • Busulfan 6 mg/ml >240 min. .
  • Carboplatin 10 mg/ml >240 min. .
  • Carmustine (BCNU) 3.3 mg/ml 65.3 min. .
  • Cisplatin 1.0 mg/ml >240 min. .
  • Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. .
  • Cytarabine HCI 100 mg/ml >240 min. .
  • Dacarbazine (DTIC)10.0 mg/ml >240 min. .
  • Daunorubicin 5 mq/mL>240 min. .
  • Docetaxel 10.0 mg/ml >240 min .
  • Doxorubicin HCI 2.0 mg/ml >240 min. .
  • Epirubicin HCI 2.0 mg/ml >240 min. .
  • Etoposide (Toposar) 20.0 mg/ml >240 min. .

{10}------------------------------------------------

  • Fludarabine 25.0 mg/ml >240 min. .
  • Fluorouracil 50.0 mg/ml >240 min. .
  • Gemcitabine 38 mg/ml>240 min. .
  • ldarubicin 1 mg/ml >240 min. .
  • lfosfamide 50.0 mg/ml >240 min. .
  • Irinotecan 20.0 mg/ml >240 min. .
  • Mechlorethamine HCI 1.0 mg/ml>240 min. .
  • Melphalan 5 mg/ml >240 min. .
  • Methotrexate 25 mq/ml >240 min. .
  • Mitromycin C. 0.5 mg/ml >240 .
  • Mitoxantrone 2.0 mg/ml >240 min. .
  • Paclitaxel (Taxol) 6.0 mg/ml >240 min. .
  • Rituximab 10.0 mg/ml >240 min. .
  • Thiotepa 10.0 mg/ml 58.3min. .
  • Trisenox 1.0 mg/ml >240 min. .
  • Vincristine Sulfate 1.0 mg/ml >240 min. .
  • Fentanyl Citrate 100mcg/2ml >240 minutes .

Warning: Do not use with Carmustine and Thiotepa.

The Nitrile Patient Examination Gloves Black Tested Chemotherapy Drugs are as follows:

  • Carmustine (BCNU) 3.3 mg/ml 49.2 min. .
  • Cisplatin 1.0 mg/ml >240 min. .
  • Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 min. .
  • Dacarbazine (DTIC)10.0 mg/ml >240 min. .
  • Doxorubicin HCI 2.0 mg/ml >240 min. .
  • Etoposide (Toposar) 20.0 mg/ml >240 min. .
  • Fluorouracil 50.0 mg/ml >240 min. .
  • Fentanyl Citrate 100mcg/2ml >240 minutes .
  • Paclitaxel (Taxol) 6.0 mg/ml >240 min. .
  • Thiotepa 10.0 mg/ml 87.1min. .

Warning: Do not use with Carmustine and Thiotepa.

{11}------------------------------------------------

VI. Comparison of technological characteristics with the predicate devices

ItemProposed device (K222714)Predicate device (K213040)Discussion
Product nameNitrile Patient Examination GlovesBlue Tested For Use With Chemotherapy Drugs.Nitrile Patient Examination GlovesBlue Violet Tested For Use With Chemotherapy Drugs.Nitrile Patient Examination GlovesBlack Tested For Use With Chemotherapy Drugs.Nitrile Patient Examination Gloves Blue Tested For Use With Chemotherapy Drugs.Nitrile Patient Examination Gloves Blue Violet Tested For Use With Chemotherapy Drugs.Nitrile Patient Examination Gloves Black Tested For Use With Chemotherapy Drugs.-
Product CodeLZA, LZC,OPJ, QDOLZA, LZC,OPJ, QDOSame
Regulation No.21 CFR 880.625021 CFR 880.6250Same
ClassificationClass IClass ISame
Powder freeYesYesSame
Indication for useA patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05Same
(Reapproved 2013)Standard Practice forAssessment of MedicalGloves to Permeation byChemotherapy Drugs(Reapproved 2013)Standard Practice forAssessment of MedicalGloves to Permeation byChemotherapy Drugs
Main MaterialPowder-Free NitrilePowder-Free NitrileSame
ColorBlue, Blue violet, BlackBlue, Blue violet, BlackSame
Sizex-small, small, medium,large, x-large, XX-largex-small, small, medium,large, x-large, XX-largeSame
Dimensions –LengthComplies with ASTMD6319-10XS (220mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)XXL (230mm min)Complies with ASTMD6319-10XS (220mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)XXL (230mm min)Same
Dimensions –WidthComplies with ASTMD6319-10XS (70±10mm)S (80±10mm)M (95±10mm)L (110±10mm)XL (120±10mm)XXL (≥120mm)Complies with ASTMD6319-10XS (70±10mm)S (80±10mm)M (95±10mm)L (110±10mm)XL (120±10mm)XXL (≥120mm)Same
Dimensions –ThicknessComplies with:ASTM D6319-10Palm: 0.05mm minFinger: 0.11mm minComplies with:ASTM D6319-10Palm: 0.05mm minFinger: 0.11mm minSame
PhysicalPropertiesComplies with:ASTM D6319-10minimum:Tensile Strength:Before Aging ≥14 MPa,min.After Aging ≥14 MPa,min.Elongation:Complies with:ASTM D6319-10minimum:Tensile Strength:Before Aging ≥14 MPa,min.After Aging ≥14 MPa,min.Elongation:Same
After Aging 400%, min.After Aging 400%, min.
Freedom fromHolesComplies with ASTMD6319-10 and ASTMD5151-06 G-1, AQL 1.5Complies with ASTMD6319-10 and ASTMD5151-06 G-1, AQL 1.5Same
ResidualPowderComplies with:ASTM D6319-10<2mg per gloveComplies with:ASTM D6319-10<2mg per gloveSame
ContactDurationsLimited <24 hoursLimited <24 hoursSame
BiocompatibilityISO 10993-10:Not a skin irritantNot a skin sensitizerAt the neat extraction,the test article isconsidered cytotoxic, butthe acute systemictoxicity resultsdemonstrate the devicewill not cause a systemiceffect.ISO 10993-10:Not a skin irritantNot a skin sensitizerAt the neat extraction, thetest article is consideredcytotoxic, but the acutesystemic toxicity resultsdemonstrate the devicewill not cause a systemiceffect.Same
SterilityNon-sterileNon-sterileSame
Rx Only orOTCOver the CounterOver the CounterSame
TestedChemotherapyDrugsNitrile PatientExamination GlovesBlue ColoredTested For Use WithChemotherapy DrugsBleomycin Sulfate 15mg/ml >240 min.Busulfan 6 mg/ml >240min.Carboplatin 10mg/ml >240 min.Carmustine (BCNU) 3.3mg/ml 17.2 min.Cisplatin 1.0mg/ml >240 minNitrile PatientExamination GlovesBlue ColoredTested For Use WithChemotherapy DrugsBleomycin Sulfate 15mg/ml >240 min.NACarboplatin 10mg/ml >240 min.Carmustine (BCNU) 3.3mg/ml 17.2 min.Cisplatin 1.0 mg/ml >240min/Similar¹

Table 1 Comparison of Nitrile Examination Gloves

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

Cyclophosphamide(Cytoxan) 20.0mg/ml >240 min.Cyclophosphamide(Cytoxan) 20.0mg/ml >240 min.
Cytarabine HCI 100mg/ml >240 min.Cytarabine HCI 100mg/ml >240 min.
Dacarbazine (DTIC)10.0mg/ml >240 min.Dacarbazine (DTIC)10.0mg/ml >240 min.
Daunorubicin 5mg/mL>240 min.Daunorubicin 5mg/mL>240 min.
Docetaxel 10.0 mg/ml>240 minDocetaxel 10.0 mg/ml>240 min
Doxorubicin HCI 2.0mg/ml >240 min.Doxorubicin HCI 2.0mg/ml >240 min.
Epirubicin HCI 2.0mg/ml >240 min.NA
Etoposide (Toposar)20.0 mg/ml >240 min.Etoposide (Toposar) 20.0mg/ml >240 min.
Fludarabine 25.0mg/ml >240 min.NA
Fluorouracil 50.0mg/ml >240 min.Fluorouracil 50.0mg/ml >240 min.
Gemcitabine 38mg/ml>240 min.Gemcitabine 38mg/ml>240 min.
Idarubicin 1 mg/ml >240min.Idarubicin 1 mg/ml >240min.
Ifosfamide 50.0mg/ml >240 min.Ifosfamide 50.0mg/ml >240 min.
Irinotecan 20.0 mg/ml>240 min.Irinotecan 20.0 mg/ml>240 min.
Mechlorethamine HCI1.0 mg/ml>240 min.Mechlorethamine HCI 1.0mg/ml>240 min.
Melphalan 5 mg/ml >240min.Melphalan 5 mg/ml >240min.
Methotrexate 25mg/ml >240 min.Methotrexate 25mg/ml >240 min.
Mitromycin C. 0.5mg/ml >240Mitromycin C. 0.5mg/ml >240
Mitoxantrone 2.0 mg/mlMitoxantrone 2.0 mg/ml
>240 min.>240 min.
Paclitaxel (Taxol) 6.0mg/ml >240 min.Paclitaxel (Taxol) 6.0mg/ml >240 min.
Rituximab 10.0mg/ml >240 min.NA
Thiotepa 10.0 mg/ml58.3min.Thiotepa 10.0 mg/ml58.3min.
Trisenox 1.0 mg/ml >240min.NA
Vincristine Sulfate 1.0mg/ml >240 min.Vincristine Sulfate 1.0mg/ml >240 min.
Fentanyl Citrate100mcg/2ml >240minutesNA
Nitrile PatientExamination GlovesBlue Violet ColoredTested For Use WithChemotherapy DrugsNitrile PatientExamination GlovesBlue Violet ColoredTested For Use WithChemotherapy Drugs/
Bleomycin Sulfate 15mg/ml >240 min.NASimilar¹
Busulfan 6 mg/ml >240min.NA
Carboplatin 10mg/ml >240 min.NA
Carmustine (BCNU) 3.3mg/ml 65.3 min.Carmustine (BCNU) 3.3mg/ml 65.3 min.
Cisplatin 1.0mg/ml >240 min.Cisplatin 1.0 mg/ml >240min.
Cyclophosphamide(Cytoxan) 20.0mg/ml >240 min.Cyclophosphamide(Cytoxan) 20.0mg/ml >240 min.
Cytarabine HCI 100mg/ml >240 min.NA
Dacarbazine (DTIC)10.0mg/ml >240 min.Dacarbazine (DTIC)10.0mg/ml >240 min.
Daunorubicin 5mg/ml >240 min.NA
Docetaxel 10.0 mg/ml>240 minNA
Doxorubicin HCI 2.0mg/ml >240 min.Doxorubicin HCI 2.0mg/ml >240 min.
Epirubicin HCI 2.0mg/ml >240 min.NA
Etoposide (Toposar)20.0 mg/ml >240 min.Etoposide (Toposar) 20.0mg/ml >240 min.
Fludarabine 25.0mg/ml >240 min.NA
Fluorouracil 50.0mg/ml >240 min.Fluorouracil 50.0mg/ml >240 min.
Gemcitabine 38mg/ml>240 min.NA
Idarubicin 1 mg/ml >240min.NA
Ifosfamide 50.0mg/ml >240 min.NA
Irinotecan 20.0 mg/ml>240 min.NA
Mechlorethamine HCI1.0 mg/ml>240 min.NA
Melphalan 5 mg/ml >240min.NA
Methotrexate 25mg/ml >240 min.Methotrexate 25mg/ml >240 min.
Mitromycin C. 0.5mg/ml >240.NA
Mitoxantrone 2.0 mg/ml>240 min.NA
Paclitaxel (Taxol) 6.0mg/ml >240 min.Paclitaxel (Taxol) 6.0mg/ml >240 min.
Rituximab 10.0mg/ml >240 min.NA
Thiotepa 10.0 mg/ml58.3min.Thiotepa 10.0 mg/ml58.3min.
Trisenox 1.0 mg/ml >240min.NA
Vincristine Sulfate 1.0mg/ml >240 min.NA
Fentanyl Citrate100mcg/2ml >240 minutesNA
Nitrile PatientExamination GlovesBlack ColoredTested For Use WithChemotherapy DrugsNitrile PatientExamination GlovesBlue Violet ColoredTested For Use WithChemotherapy Drugs/
Cisplatin 1.0mg/mL>240 min.NASimilar¹
Carmustine (BCNU) 3.3mg/ml 49.2 min.NA
Cyclophosphamide(Cytoxan) 20.0mg/ml >240 min.NA
Dacarbazine (DTIC)10.0mg/ml >240 min.NA
Doxorubicin HCI 2.0mg/ml >240 min.NA
Etoposide (Toposar)20.0 mg/ml >240 min.NA
Fluorouracil 50.0mg/ml >240 min.NA
Paclitaxel (Taxol) 6.0mg/ml >240 min.NA
Thiotepa 10.0 mg/ml87.1min.NA
Fentanyl Citrate100mcg/2ml >240minutesFentanyl Citrate100mcg/2ml >240minutes

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

1 Add new kinds of the chemotherapy label claim to the previous cleared under K213040. The permeation testing was conducted per ASTM D6978-05 (Reapproved 2019) to support the addition of the labeling claim.

VII. Non-Clinical Testing

Non-clinical tests were leveraged from the predicate submission. Additional testing per

{18}------------------------------------------------

ASTM D6978-05 (2019) was performed to support inclusion of additional chemotherapy drugs to the indications for use.

Test MethodologyPurposeAcceptance CriteriaResults
ASTM D6319PhysicalDimensions TestExtra-Small:Length: ≥ 220mm Width: 70±10 mm;Small:Length: ≥ 220mm Width: 80±10mm;Medium:Length: ≥ 230mm Width: 95±10mmLarge:Length: ≥ 230mm Width: 110±10mmExtra- Large:Length: ≥ 230mm Width: 120±10mmExtra- Extra- Large:Length: ≥ 230mm Width: ≥120mmPass
Thickness (mm):Finger:≥0.05Palm: ≥0.08Pass
PhysicalpropertiesBeforeAging
TensileStrength≥14MPaPass
UltimateElongation≥500%
AfterAging
TensileStrength≥14MPaPass
UltimateElongation≥500%
ASTM D5151Freedom frompinholesMeet the requirements of ASTMD5151Test for AQL 2.5Pass
ASTM D6124Powder ResidueMeet the requirements of ASTMD6124< 2.0mgPass
ISO 10993-10To determine ifthe finished devicematerial is anirritantNon-irritatingUnder theconditions of thestudy not anirritant/ Pass
ISO 10993-10To determine ifthe finished devicematerial is asensitizerNon- sensitizingUnder conditionsof the study, not asensitizer. / Pass
ISO 10993-11To determine ifthe finished devicematerial extractspose a systemictoxicity concernNon-acute systemic toxicityUnder conditionsof the study, didnot show acutesystemic toxicityin vivo / Pass

Table 2 Summary of Non-Clinical Performance Testing

{19}------------------------------------------------

  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for ● Medical Application
  • ASTM D3767-03(2020), Practice for rubber-Measurement of Dimensions ●
  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves ●
  • . ASTM D6124-06(2017), Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D573-04(2019), Standard Test Method for Rubber—Deterioration in an Air ● Oven
  • ASTM D412-16, Standard Test Methods for Vulcanized Rubber and ● Thermoplastic Elastomers-Tension
  • ASTM D6978-05(2019), Standard Practice for Assessment of Resistance of ● Medical Gloves to Permeation by Chemotherapy Drugs
  • ISO 10993-10: 2010 Biological Evaluation Of Medical Devices Part 10: Tests ● For Irritation And Skin Sensitization.
  • ISO 10993-11:2017, Biological evaluation of medical devices Part11:Tests for ● Systemic Toxicity

VIII. Clinical Testing

No clinical study is included in this submission.

IX. Conclusion

The conclusions drawn from the nonclinical and clinical tests that demonstrate that the subject device, Nitrile Patient Examination Gloves Blue Tested For Use With Chemotherapy Drugs; Nitrile Patient Examination Gloves Blue Violet Tested For Use With Chemotherapy Drugs; Nitrile Patient Examination Gloves Black Tested For Use With Chemotherapy Drugs, are as safe, as effective, and performs as well as or better than the legally marketed predicate device (K213040).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.