(51 days)
The Unistik® Touch is a puncture device to obtain micro blood samples. Unistik Touch has a sharps prevention feature to protect the user from a needlestick injury.
The Unistik® Touch contact-activated safety lancets are hand-held disposable devices intended for performing controlled skin punctures of the fingertips in adults and children, by means of a needle lancing mechanism. Unistik® Touch safety lancets are indicated for use where a capillary blood specimen is required for the purposes of performing in-vitro diagnostic (IVD) assays, e.g. for blood glucose monitoring in patients with diabetes. The Unistik® Touch contact-activated safety lancets are available in five different variants of needle gauge as follows: Needle gauge 16G, Penetration Depth 2.0 mm Needle gauge 21G, Penetration Depth 2.0 mm Needle gauge 23G, Penetration Depth 2.0 mm Needle gauge 28G, Penetration Depth 1.8 mm Needle gauge 30G, Penetration Depth 1.5 mm The Unistik® Touch contact-activated safety lancets are intended for prescription and over-the-counter use and to be used by self-testing patients, care-givers and healthcare professionals. The devices are designed to perform a controlled skin puncture on the fingertip, in order for caregivers and healthcare professionals to obtain capillary blood specimens from patients for IVD assays, and also for lay (home) users to be able to perform a skin puncture on themselves where an IVD self-testing regime is required. The intended user population includes male and female. right or left-handed self-administering patients, care givers and healthcare professionals. The frequency of use and intended patient population is dependent on the given diagnostic regime. Unistik® Touch contact-activated safety lancets are sterile single-use devices with integral sharps protection whereby the lancet blade or needle is shielded before and after use to prevent needlestick injuries, so mitigating the hazard of transmission of blood-borne infectious agents. Furthermore, the devices automatically self-disable after a single use, thus preventing any hazards of re-use. The Unistik® Touch contact-activated safety lancets achieve their intended purpose of performing controlled skin punctures of the fingertips in adults and children, by means of a needle lancing mechanism with a pre-loaded steel spring for propelling the lancet holder forward when the device is activated and subsequently automatically retracting it. After retraction, the lancet holder is automatically locked into the device such that the device cannot be re-used and the needle tip is safely shielded. The lancet needles are moulded into the lancet holder component such that the needle tip is sealed by complete encapsulation in overmoulded plastic. The complete lancet holder component is then sterilised by gamma irradiation, so after irradiation the sterility of the needle tip is maintained by encapsulation within the plastic. The sterile seal is only broken when the user twists off and removes the lancet cap immediately before use. The needle tip is the exposed needle length after the cap is removed, and this is the only part of the needle that will penetrate the patient's skin during use. Therefore, the encapsulation of the needle tip by plastic overmoulding performs the function of primary packaging, whereby a sterile seal is maintained until the point of use.
The Unistik® Touch Single-Use Safety Lancets were evaluated through a series of non-clinical performance studies, primarily bench testing and a simulated clinical use study.
1. Table of acceptance criteria and the reported device performance:
Items | Acceptance criteria | Results |
---|---|---|
Appearance | The surface shall be free of burrs and no scratches shall be visible | Pass |
Dimension | Product dimensions shall be consistent to the drawings | Pass |
Cleanness | No dust, no grease, no hair, no dirt | Pass |
Firmness | Needle should connect firmly with plastic handle | Pass |
Resistance to corrosion | Needle of lancet shall show no evidence of corrosion. | Pass |
Acidity or Alkalinity | The pH value of an extract prepared refers to ISO 9626 Annex A shall be within one pH unit of that of the control fluid. | Pass |
Limits for Extractable Metals | When corrected for the metals content of the control fluid, contain not greater than a combined total of 5 mg/l of lead, tin, zinc and iron. The cadmium content of the extract shall, when corrected for the cadmium content of the control fluid, be lower than 0.1 mg/l. | Pass |
Puncture depth | Use calipers to measure and meet the requirements. | Pass |
Launch performance | Launch performance should be good, launch button press smoothly, no jam | Pass |
Puncture force | The needle tip of the needle should have good puncture ability. | Pass |
Lubricant | No lubricant shall be visible. | Pass |
Disposable | Safety lancet should be single use, and should not be operational after use | Pass |
Drop Test | The needle tip shall not be exposed after dropping the device vertically from a height of 1.2m | Pass |
Safety Feature | The force to activate the safety feature: 4 - 15N. Test access to the needle: the needle shall not touch the sphere. | Pass |
2. Sample size used for the test set and the data provenance:
- Simulated Clinical Use Study: 500 device samples were used.
- Data Provenance: The document does not specify the country of origin of the data, nor whether it was retrospective or prospective, beyond stating it was a "simulated clinical use study."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The studies described are non-clinical, primarily bench testing, and a simulated clinical use study. The performance criteria are objective measurements against established standards (e.g., ISO, FDA guidance).
4. Adjudication method for the test set:
- This information is not applicable and not provided. The studies involved objective measurements against predefined acceptance criteria rather than subjective expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is a blood lancet, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This device is a physical medical device, not an algorithm. The "standalone" performance here refers to the device's inherent functional performance without human interaction regarding its analytical functions (e.g., accurate puncture depth, safety mechanism activation). This was assessed through the various bench tests.
7. The type of ground truth used:
- The ground truth for the bench testing was based on objective, verifiable measurements and physical characteristics outlined in the acceptance criteria (e.g., freedom from burrs, adherence to dimensions, pH values, metal content, puncture depth measurements, force to activate safety features).
- For the simulated clinical use study, the ground truth was based on demonstrating that the device "met the pre-established criteria" for its safety mechanism, as evaluated against FDA Guidance and ISO 23908.
8. The sample size for the training set:
- Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set in the conventional sense.
9. How the ground truth for the training set was established:
- Not applicable. As noted above, this is a physical medical device and does not involve a training set for an AI/ML algorithm.
§ 878.4850 Blood lancets.
(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.