K Number
K222066

Validate with FDA (Live)

Date Cleared
2023-04-14

(275 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

General indications: Mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. Indications: 1) Primary and secondary osteoarthritis 2) Rheumatoid arthritis 3) Correction of functional deformities 4) Avascular necrosis 5) Femoral neck fractures 6) Revision after implant loosening dependent on bone mass and quality The MobileLink Dual Mobility Insert is additionally indicated for: 7) Dislocation risks The MobileLink Acetabular Shells are intended for cementless fixation.

Device Description

This 510k adds MobileLink TrabecuLink Acetabular Shells to the MobileLink Acetabular Cup System previously cleared in K182321, K192559, K200607. These shells are additively manufactured from Ti6Al4V alloy per ISO 5832-3 and feature a porous structure (TrabecuLink) on the bone interfacing surface for biologic fixation. The shells come in cluster-hole and in multi-hole patterns. mate with the same UHMWPE liners (in conventional PE and highly-crosslinked vitamin E PE) as previously cleared in K182321, and with the same Dual Mobility Inserts and Shell/Insert Adapters previously cleared in K200607. They can be used with the same bone screws cleared in K192559.

AI/ML Overview

I am sorry, but the provided text does not contain the specific information required to complete your request. The document is a 510(k) premarket notification clearance letter from the FDA for a medical device (LINK MobileLink Acetabular Cup System). It discusses the device's substantial equivalence to predicate devices and lists non-clinical performance testing that was performed (e.g., acetabular deformation, fatigue testing, biocompatibility).

However, it does not include:

  • A table of acceptance criteria and reported device performance.
  • Details about sample size, data provenance, number or qualifications of experts, or adjudication methods for test sets.
  • Information about multi-reader multi-case (MRMC) comparative effectiveness studies or effect sizes.
  • Details about standalone algorithm performance.
  • The type of ground truth used, training set sample size, or how ground truth for the training set was established.

The document explicitly states: "Clinical performance testing was not required to demonstrate the substantial equivalence of this device." This suggests that the clearance was based on non-clinical testing and comparison to predicate devices, rather than a clinical study with the detailed criteria you're asking for.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

April 14, 2023

Waldemar Link GmbH & Co KG % Terry Powell Regulatory Affairs LinkBio Corp. 59 King Street Dover, New Jersey 07801

Re: K222066

Trade/Device Name: LINK MobileLink Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, OQG Dated: March 20, 2023 Received: March 21, 2023

Dear Terry Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222066

Device Name

LINK MobileLink Acetabular Cup System

Indications for Use (Describe)

General indications:

Mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. Indications:

  1. Primary and secondary osteoarthritis
    1. Rheumatoid arthritis
    1. Correction of functional deformities
    1. Avascular necrosis
    1. Femoral neck fractures
    1. Revision after implant loosening dependent on bone mass and quality
  • The MobileLink Dual Mobility Insert is additionally indicated
  • for: 7) Dislocation risks

The MobileLink Acetabular Shells are intended for cementless fixation.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k)Submitter:Waldemar Link GmbH & Co. KGBarkhausenweg 1022339 Hamburg, GermanyPhone: +49-40-539950Facility Registration: 3003386935
Contact Person:Waldemar Link GmbH & Co. KGStefanie Fuchs (Regulatory Affairs)Oststraße 4-1022844 Norderstedt, GermanyPhone: +49-40 53995-530Fax: +49-40 53995-174E-Mail: st.fuchs@linkhh.de
AlternateContact:Link Bio Corp.Terry Sheridan Powell (Regulatory Affairs)69 King StreetDover, NJ 07801e-mail: t.powell@linkbio.com
Date Prepared:April 14, 2023
Trade Name:LINK MobileLink Acetabular Cup System
Common Name:Total hip replacement system
ClassificationName:Hip joint metal/polymer/metal semi-constrained porous-coated uncementedprosthesis. 21 CFR §888.3358, product code LPH, OQG
Classificationand Panel:Class II, Orthopedic / 87

Predicate

Devices:

Subect DeviceComponentsPredicate Devices510(k)Number
TrabecuLinkShells1.MobileLink Acetabular Cup SystemPro Codes: LPH, LZO, MEH, OQGWaldemar Link GmbH & Co KG[Primary Predicate]K182321,K192559,K200607
2.G7 OsseoTi Acetabular ShellPro Codes: LPH, JDI, KWZ, LZO, OQG, OQH, OQI, PBIBiomet, Inc. (Zimmer Biomet)K140669

Reason for Submission

New Device System Components – TrabecuLink Shells

Device This 510k adds MobileLink TrabecuLink Acetabular Shells to the MobileLink Description: Acetabular Cup System previously cleared in K182321, K192559, K200607.

These shells are additively manufactured from Ti6Al4V alloy per ISO 5832-3 and feature a porous structure (TrabecuLink) on the bone interfacing surface for biologic fixation. The shells come in cluster-hole and in multi-hole patterns.

{4}------------------------------------------------

mate with the same UHMWPE liners (in conventional PE and highly-crosslinked vitamin E PE) as previously cleared in K182321, and with the same Dual Mobility Inserts and Shell/Insert Adapters previously cleared in K200607. They can be used with the same bone screws cleared in K192559. Intended Use: General indications: The MobileLink Acetabular Cup System is indicated for patients with mobilitylimiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. Indications: 1) Primary and secondary osteoarthritis Rheumatoid arthritis 2) 3) Correction of functional deformities 4) Avascular necrosis Femoral neck fractures 5) 6) Revision after implant loosening dependent on bone mass and quality The MobileLink Dual Mobility Insert is additionally indicated for: 7) Dislocation risks The MobileLink Acetabular Shells are intended for cementless fixation. Comparison to The subject MobileLink TrabecuLink Acetabular Shells differ from the predicate Predicate Device: MobileLink Acetabular Shells (K182321) in that the subject shells are additive manufactured and feature a 3-D porous surface (TrabecuLink) that is integrated with the shell substrate during the additive manufacturing process. This feature (additively manufactured shell with integrated porous surface) is equivalent to the predicate Zimmer/Biomet G7 Acetabular Cup System with OsseoTi Shells (K140669). The subject MobileLink TrabecuLink Acetabular Shells are similar to the predicate MobileLink Acetabular Shells (K182321) in that they are made of Ti6Al4V alloy, and come in the same sizes and design variants (cluster hole or multi-hole), have the same interior geometry and mate with all of the same previously cleared MobileLink polyethylene inserts, Dual Mobility CoCr Inserts, Adapters (face changers) and bone screws. Performance Non-clinical performance testing and analysis were provided, including: Testing: -Acetabular deformation testing -Acetabular fatigue testing -Assessment of disassembly (Push-out, lever-out, torque out for all acetabular modular connections in MobileLink system) Assessment of range of motion --Characterization of the TrabecuLink porous surface -Biocompatibility evaluation -Particle analysis The results of non-clinical performance testing demonstrate that the device is as

safe and effective as the predicate device, and therefore Substantially Equivalent.

{5}------------------------------------------------

  • Clinical Testing: Clinical performance testing was not required to demonstrate the substantial equivalence of this device.
  • Conclusion: The subject MobileLink TrabecuLink Acetabular Shells are substantially equivalent to the predicate devices identified in this premarket notification.

N/A