(149 days)
The Reprocessed RD SET pulse oximeter sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, and infant patients during non-motions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.
In a clinical setting, a pulse oximeter sensor measures the oxygen saturation of arterial blood (SpO2). A pulse oximeter sensor is composed of a light emitting diode (LED) and a sensor that are placed on opposite sides of a patient's finger or foot. The LED contains a red light and an infrared light that are differentially absorbed by oxygenated and deoxygenated hemoglobin. Based on the relative absorption of the two wavelengths that is determined by the sensor, the POX determines the relative amount of oxygenated and deoxygenated hemoglobin, which is calculated as SpO2. In order to make the SpO2 calculation independent of skin color, finger size, etc., the pulse oximeter sensor uses only the time varying light absorption component generated by the patient's pulse. The sensor also uses the period of pulsation to measure patient pulse rate. The pulse oximeter can estimate the amount of oxygen in the blood without having to draw a blood sample.
The primary components of an oxygen transducer, or Pulse Oximeter (POX) Sensor, are lightemitting diodes (red and infrared LED) and a photo sensor. These components (with their wiring system) are embedded within a taping system designed for wrapping the POX Sensor around a patient's finger, foot, or hand so that the LED and photo sensor are directly opposite to each other. As the lights are emitted and received across a vascular bed, the rates of absorption at the two wavelengths vary depending upon the ratios of oxygenated and deoxygenated hemoglobin within the blood.
As part of the reprocessing manufacturing process, we collect used devices from hospitals, replace required components, clean the devices and test the functional performance of the devices to ensure they meet or exceed requlatory requirements and the expectations of our customers.
The provided text describes the 510(k) summary for reprocessed pulse oximeter sensors. It outlines the device, its intended use, comparison to predicate devices, and testing performed to demonstrate substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Criteria | Acceptance Standard (Stated or Implied) | Reported Device Performance | Meets? |
|---|---|---|---|
| Functional Oxygen Saturation (SpO2) Accuracy | A_rms (average root mean squared) for SpO2 under no motion conditions less than 3.0%, conforming to Clause 201.12.1.101.1 of ISO 80601-2-61:2011 and Table 3 of "Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff." (Implied acceptance standard as part of regulatory guidance and ISO.) | A_rms for SpO2 under no motion conditions: 2.04% (non-woven tape), 2.11% (woven tape) over the range of 70-100%. | Yes |
| Pulse Rate Accuracy (No Motion) | Implied to be acceptable if the device performs comparably to predicate devices. Predicate specifications: ±3 bpm (30-200 bpm). | Subject Device: ±3 bpm (30-200 bpm) | Yes (Matches predicate) |
| Pulse Rate Accuracy (Low Perfusion) | Implied to be acceptable if the device performs comparably to predicate devices. Predicate specifications: ±3 bpm (25-240 bpm). | Subject Device: ±3 bpm (30-200 bpm) | Yes (Comparable to predicate range within overlap) |
| Continuity Testing | Passed (Implied) | Conducted and satisfied requirements (Stated) | Yes |
| Sensitivity Testing | Passed (Implied) | Conducted and satisfied requirements (Stated) | Yes |
| Monitor Compatibility | Compatible with pulse oximeter consoles (Implied) | Conducted and satisfied requirements (Stated) | Yes |
| Non-clinical Bench Testing (SpO2, Pulse Rate) | Satisfied performance specifications (Implied) | Conducted and satisfied requirements (Stated) | Yes |
| Clinical Validation (Healthy Adult Volunteers) | A_rms < 3.0% (as above) | A_rms 2.04% and 2.11% | Yes |
Study Details Proving Device Meets Acceptance Criteria
-
Sample Size Used for the Test Set and Data Provenance:
- Test Set (Clinical Validation - SpO2 Accuracy): Healthy, adult volunteers. The exact number of volunteers is not specified in the provided text.
- Data Provenance: The text states, "Stryker Sustainability Solutions performed the clinical validation testing of the SpO2 performance under no motion conditions on healthy, adult volunteers." This indicates it was a prospective study conducted by the applicant. The country of origin is not explicitly stated but is implicitly within the regulatory scope of the US FDA.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
- The document describes the device performance (SpO2 accuracy and pulse rate), which is measured against an established standard (ISO 80601-2-61:2011 and FDA guidance), rather than establishing a "ground truth" through expert consensus for diagnostic purposes.
- For a pulse oximeter, the ground truth for SpO2 accuracy is typically established by co-oximetry blood gas analysis, performed by trained laboratory personnel, not by subjective expert interpretation. The text does not elaborate on how exactly the "ground truth" reference values for SpO2 were obtained during the clinical validation, only that the A_rms values were calculated for the device's performance.
-
Adjudication Method for the Test Set:
- Not applicable in the context of a pulse oximeter accuracy study. The assessment is quantitative (numerical measurement vs. reference).
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a study for a medical device (pulse oximeter sensor), not an AI-based diagnostic imaging tool. Therefore, MRMC studies and "human readers" are not relevant to this submission.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, in a sense. The SpO2 accuracy and pulse rate accuracy measurements are characteristics of the device itself (sensor and its interaction with a console). The stated A_rms values reflect the device's intrinsic performance against a reference standard. While the device requires a human (patient) to apply it and a human (clinician) to interpret the readings, the performance metrics are fundamentally about the sensor's ability to accurately measure physiological parameters.
-
The type of ground truth used:
- For SpO2 accuracy, the ground truth is typically measured physiological parameters obtained from a reliable reference method, usually a co-oximeter (invasive arterial blood gas analysis), which is considered the gold standard for SpO2. The text states the study was conducted on healthy volunteers in a hospital setting, which is consistent with such an approach, though the explicit method of ground truth establishment is not detailed beyond the discussion of A_rms as per ISO standards.
-
The sample size for the training set:
- Not applicable. This is not an AI/machine learning device that requires a "training set" in the conventional sense. The "training" for this reprocessed device is ensuring that the reprocessing manufacturing process allows it to function equivalently to the original device.
-
How the ground truth for the training set was established:
- Not applicable, as there is no "training set" in the context of this device. The validation process involves demonstrating that the reprocessed device performs according to established safety and effectiveness standards, similar to a new device or its predicate.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, and the full name of the agency, "U.S. Food & Drug Administration," is in a smaller font size below the acronym.
December 4, 2022
Stryker Sustainability Solutions Michael Wong Regulatory Affairs Specialist 1810 W Drake Drive Tempe, Arizona 85283
Re: K222019
Trade/Device Name: Reprocessed RD SET Adt Pulse Oximeter Sensor (4000), Reprocessed RD SET Pdt Pulse Oximeter Sensor (4001), Reprocessed RD SET Inf Pulse Oximeter Sensor (4002), Reprocessed RD SET Adt Pulse Oximeter Sensor (4003) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: NLF Dated: November 4, 2022 Received: November 4, 2022
Dear Michael Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
James J. Lee -S
James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K222019
Device Name
Reprocessed RD SET Adt Pulse Oximeter Sensor (4000): Reprocessed RD SET Pat Pulse Oximeter Sensor (4001): Reprocessed RD SET Inf Pulse Oximeter Sensor (4002); Reprocessed RD SET Adt Pulse Oximeter Sensor (4003)
Indications for Use (Describe)
The Reprocessed RD SET pulse oximeter sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, and infant patients during non-motions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
Submitter:
Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, AZ 85283
Contact:
Michael Wong Regulatory Affairs Specialist (480) 213-1342 Michael.wong@stryker.com
Device Name:
Device Trade Name:
Reprocessed RD SET® Adt Pulse Oximeter Sensor (4000); Reprocessed RD SET® Pdt Pulse Oximeter Sensor (4001); Reprocessed RD SET® Inf Pulse Oximeter Sensor (4002); Reprocessed RD SET® Adt Pulse Oximeter Sensor (4003)
Common Name: Oximeter
Classification Name: Oximeter, Reprocessed
Regulation Number: 870.2700
Product Code: NLF
Legally Marketed Predicate Devices:
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|---|---|---|
| K211140 | Reprocessed Masimo Pulse Oximeter (1859 Adult O2Transducer), ReprocessedMasimo Pulse Oximeter (1860 Pediatric O2 Transducer),Reprocessed MasimoPulse Oximeter (1861 Infant O2 Transducer), ReprocessedMasimo PulseOximeter (1862 Adult O2 Transducer), Reprocessed MasimoPulse Oximeter(2317 Adult O2 Transducer), Reprocessed Masimo PulseOximeter (2319 InfantO2 Transducer), Reprocessed Masimo Pulse Oximeter (2320Adult O2 | NLF |
{4}------------------------------------------------
| Transducer), Reprocessed Masimo Pulse Oximeter (2328Infant O2 Transducer),Reprocessed Masimo Pulse Oximeter (2329 Adult O2Transducer) | ||
|---|---|---|
| K180046 | Masimo Rad-97 and Accessories | DQA |
Device Description Summary:
In a clinical setting, a pulse oximeter sensor measures the oxygen saturation of arterial blood (SpO2). A pulse oximeter sensor is composed of a light emitting diode (LED) and a sensor that are placed on opposite sides of a patient's finger or foot. The LED contains a red light and an infrared light that are differentially absorbed by oxygenated and deoxygenated hemoglobin. Based on the relative absorption of the two wavelengths that is determined by the sensor, the POX determines the relative amount of oxygenated and deoxygenated hemoglobin, which is calculated as SpO2. In order to make the SpO2 calculation independent of skin color, finger size, etc., the pulse oximeter sensor uses only the time varying light absorption component generated by the patient's pulse. The sensor also uses the period of pulsation to measure patient pulse rate. The pulse oximeter can estimate the amount of oxygen in the blood without having to draw a blood sample.
The primary components of an oxygen transducer, or Pulse Oximeter (POX) Sensor, are lightemitting diodes (red and infrared LED) and a photo sensor. These components (with their wiring system) are embedded within a taping system designed for wrapping the POX Sensor around a patient's finger, foot, or hand so that the LED and photo sensor are directly opposite to each other. As the lights are emitted and received across a vascular bed, the rates of absorption at the two wavelengths vary depending upon the ratios of oxygenated and deoxygenated hemoglobin within the blood.
As part of the reprocessing manufacturing process, we collect used devices from hospitals, replace required components, clean the devices and test the functional performance of the devices to ensure they meet or exceed requlatory requirements and the expectations of our customers.
Indications for Use:
The Reprocessed RD SET® pulse oximeter sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, and infant patients during nonmotion conditions, and for patients who are well or poorly perfused in hospitals and hospitaltype facilities.
Indications for Use Comparison:
The indications for use for the subject device are the same as those of the predicate devices, with the exceptions that the neonatal indication has been removed from the 4003 model and
Stryker Sustainability Solutions Reprocessed RD SET® Pulse Oximeter Sensor Traditional 510(k)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and black. The background is white.
motion and mobile/home environment indications have been removed from the 4000, 4001, 4002, and 4003 models.
Technological Comparison:
The design of the reprocessed device is the same as the predicate device. The indications for use does not change from the predicate devices (K211140 and K180046), with the exception that the neonatal indication has been removed from the 4003 model and the motion and mobile/home environment indications have been removed from the 4000, 4001, 4002, and 4003 models. The same standard mechanical design and sizes and equivalent materials are utilized. There are no changes to the claims, clinical applications, or performance specifications.
The subject device and the predicate devices have the same technological characteristics, i.e., they have the same:
- intended use;
- principle of operation;
- form factor; -
- measurement application sites;
- performance specifications; -
- environmental and mechanical specifications -
Principle of Operation
The principle of operation of pulse oximetry is based upon the fundamental principle that hemoglobin bound to oxygen (oxyhemoglobin) and hemoglobin unbound to oxygen (deoxyhemoglobin) vary in the absorption of different wavelengths of the light and the absorptions can be used to estimate SpO2 and pulse rate. The mechanism by which this process occurs is the use of red and infrared wavelengths of light delivered by an emitter and the detection of the signal from the light absorption of oxygenated blood and deoxygenated blood to determine the functional oxygen saturation of hemoglobin (SpO2).
Mechanism of Action for Achieving the Intended Effect
The Reprocessed RD SET® Pulse Oximeter Sensor provides the intended effect equivalent to the previously cleared pulse oximeter sensor in that it utilizes an optical sensor that is applied to the patient's finger or toe through which light is transmitted to the photodetector that detects the signal transmission. The signal transmission is processed by the Pulse Oximeter to provide SpO2 and pulse rate.
Non-Clinical and/or Clinical Tests Summary:
To support the substantial equivalence of product performance after being reprocessed and decontaminated by vaporized hydrogen peroxide to that of the predicate devices, non-clinical bench testing was conducted using a stand-in device. The stand-in device allows for SpO2 sensor verification by passing the light source of the pulse oximeter sensor (LED) through one
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and black. The word is centered and takes up most of the frame. The background is white.
side of the stand-in with the signal transmission measured by the photocell (photo-detector) of the pulse oximeter sensor on the opposing side of the stand-in.
A functional pulse oximeter sensor when connected to a pulse oximeter console will have the ability to monitor SpQ2 and Pulse Rate. Bench and laboratory testing was conducted to determine whether a pulse oximeter sensor is functional and is assessed by performing continuity and sensitivity testing. This included the following tests:
- Continuity testing -
- -Sensitivity testing
- Clinical laboratory testing
- -Non-clinical bench testing (SpO2, Pulse Rate, Monitor Compatibility)
Stryker Sustainability Solutions performed the clinical validation testing of the SpO2 performance under no motion conditions on healthy, adult volunteers in the range of 70% to 100%. The Ams (average root means squared) for SpO2 under no motion conditions was found to be 2.04% and 2.11% for non-woven and woven tape, respectively, over the range of 70-100%. These Arms results are less than 3.0% and are in conformance with Clause 201.12.1.101.1 of ISO 80601-2-61:2011 and Table 3 of "Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff."
The results of the non-clinical and clinical testing demonstrate that all requirements and performance specifications were satisfied and support the subject device is substantially equivalent to its predicate devices.
Comparison Table:
Please see Pages 5-7 for a table comparing the subject device with the primary (K211140) and secondary (K180046) predicate devices.
Conclusion:
The subject device has the same intended use as the proposed predicates and the differences in technological features do not raise different questions of safety and effectiveness. Based on a comparison of the intended use/indications for use, technological characteristics, and performance data to the predicate devices, the subject device is equivalent to the predicate devices. The results from bench and laboratory testing demonstrate that the Reprocessed RD SET® Pulse Oximeter Sensor is as safe and effective as the identified legally marketed predicate devices as described herein.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and black. The word is centered and takes up most of the frame.
| Description | Subject Device | K180046Secondary Predicate | K211140Primary Predicate | Comparison | |||
|---|---|---|---|---|---|---|---|
| Intended Patient Population | 4000 | Adult | 4000 Adult | 1859, 2317,1862, 2320,2329 Adult | Stryker Sustainability Solutions will not makeclaims for neonatal use. | ||
| 4001 | Pediatric | 4001 Pediatric | |||||
| 4002 | Infant | 4002 Infant | 1860 Pediatric | ||||
| 4003 | Adult | 4003 Adult/Neonatal | 1861, 2319,2328 Infant | ||||
| Patient Weight Range | • Adult: >30 kg (4000)• Adult: >40 kg (4003)• Pediatric: 10 - 50 kg (4001)• Infant: 3 - 20 kg (4002) | • Adult: >30 kg (4000)• Adult: >40 kg (4003)• Pediatric: 10-50 kg (4001)• Infant: 3 - 20 kg (4002)• Neonate: <3 kg (4003) | • Adult: >30 kg (Adtx, Adtx-3)• Pediatric: 10 - 50 kg (Pdtx, Pdtx-3)• Infant: 3-20 kg (Inf, Inf-L, Inf-3) | Stryker Sustainability Solutions will not makeclaims for neonatal use. | |||
| Intended Application Site | 4000 | Middle or ring finger | 4000 Middle or ring finger of non-dominant hand | 1859, 2317 Finger | As stated | ||
| 4001 | Middle or ring finger | 4001 Middle or ring finger of non-dominant hand | 1860 Finger or toe | ||||
| 4002 | (3 - 10 kg):Great toe, toe next to greattoe, or thumb(10 - 20 kg):Middle or ring finger | 4002 (3 - 10 kg):Great toe, toe next to greattoe, or thumb(10 - 20 kg):Middle or ring finger of non-dominant hand | 1861, 2319,2328 Thumb or great toe | ||||
| 4003 | (>40 kg):Middle or ring finger | 4003 (<3 kg):Foot or across the palm andback of the hand(>40 kg):Middle or ring finger of non-dominant hand | 1862, 2320,2329 Adult finger or toe | ||||
| Indications for Use | The reprocessed RD SET Disposable PulseOximeter Sensors are indicated for thecontinuous noninvasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rate (measuredby an SpO2 sensor) for use with adult,pediatric, and infant patients during nomotion conditions, and for patients who arewell or poorly perfused in hospitals andhospital-type facilities. | The RD SET® Series disposable sensorsare indicated for the continuousnoninvasive monitoring of functionaloxygen saturation of arterialhemoglobin (SpO2) and pulse rate(measured by an SpO2 sensor) for usewith adult, pediatric, infant, andneonatal patients during both nomotion and motion conditions, and forpatients who are well or poorlyperfused in hospitals, hospital-typefacilities, mobile, and homeenvironments. | INDICATIONS - When used with MasimoSET Radical: Reprocessed Masimo PulseOximeter sensor is indicated for use incontinuous noninvasive arterial oxygensaturation and pulse rate monitoring. | Stryker Sustainability Solutions will not makeclaims for use in home or mobileenvironments or for use during motionconditions. | |||
| Contra-Indications | The reprocessed RD SET sensors arecontraindicated for patients who exhibit | The RD SET sensors arecontraindicated for patients who | Reprocessed Masimo Pulse Oximetersensor should not be used in patients who | Same | |||
| allergic reactions to foam rubber productsand/or adhesive tape. | exhibit allergic reactions to foamrubber products and/or adhesive tape. | exhibit allergic reactions to foam rubberproducts and/or adhesive tape. | |||||
| Form Factor | • Adhesive Bandage | • Adhesive Bandage | • Adhesive Bandage | Same | |||
| Single Use | Yes | Yes | Yes | Same | |||
| Use Environment | Hospitals and hospital-type facilities | Hospitals, hospital-type facilities,mobile, and home environments | Hospitals and hospital-type facilities | Stryker Sustainability Solutions will not makeclaims for use in home and mobileenvironments. | |||
| Sterility | Non-sterile | Non-sterile | Non-sterile | Same | |||
| Expiration Date | No | No | No | Same | |||
| Uses | Single patient use | Single patient use | Single patient use | Same | |||
| Number of ReprocessingCycles | 4 | N/A | 4 | As stated | |||
| Laminate (replaced or saved) | The laminate that encloses the opticalcomponents is replaced. | N/A | The laminate that encloses the opticalcomponents is replaced. | Same | |||
| Optical Design | Transmissive Sensor | Transmissive Sensor | Transmissive Sensor | Same | |||
| Performance Specifications | |||||||
| Saturation Accuracy, NoMotion | 2.2% | 1.5% | ±2% | As stated | |||
| Saturation Accuracy, Motion | N/A | 1.5% | N/A | As stated | |||
| Saturation Accuracy, LowPerfusion | 2.5% | 2% | N/A | As stated | |||
| Pulse Rate Accuracy, NoMotion | ±3 bpm (30-200 bpm) | ±3 bpm (25-240 bpm) | ±3 bpm | As stated | |||
| Pulse Rate Accuracy, Motion | N/A | ±5 bpm (25-240 bpm)N/A | As stated | ||||
| Pulse Rate Accuracy, LowPerfusion | ±3 bpm (30-200 bpm) | ±3 bpm (25-240 bpm) | N/A | As stated | |||
| Safety Specifications (e.g., electrical, mechanical, environmental) | |||||||
| Electrical | Universal 100-240 VAC to DC multi-voltagepower supply | Universal 100-240 VAC to DC multi-voltage power supply | Universal 100-240 VAC to DC multi-voltagepower supply | Same | |||
| Environmental | |||||||
| Operating Temperature: | 5°C to 40°C, ambient humidity | 5°C to 40°C, ambient humidity | 41°F to 104°F (5°C to 40°C) | As stated | |||
| Storage Temperature: | -20°C to +50°C, ambient humidity | -20°C to +40°C, ambient humidity | -40°F to +158°F (-40°C to +70°C) | As stated | |||
| Operating Humidity: | 15% to 95%, non-condensing | 10% to 90%, non-condensing | 5% to 95%, non-condensing | As stated | |||
| Features (e.g., alarms, display and indicators, modes) | |||||||
| Alarms | N/A | Audible and visual alarms for high andlow saturation and pulse rate. | N/A | Stryker Sustainability Solutions does notprovide or reprocess the display. Therefore,the alarms remain the same. | |||
| Displays & Indicators | N/A | Data display: %SpO2, pulse rate, alarmstatus, alarm silenced status, ACpower, Signal IQ/pleth bar, perfusionindex bar, battery status | N/A | Stryker Sustainability Solutions does notprovide or reprocess the display. Therefore,the display characteristics remain the same. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and black. The word is centered and takes up most of the frame.
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).