AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reprocessed Masimo Pulse Oximeter sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.

Device Description

In a clinical setting, a pulse oximeter sensor measures the oxygen saturation of arterial blood (SpO2). A pulse oximeter sensor is composed of a light emitting diode (LED) and a sensor that are placed on opposite sides of a patient's finger or foot. The LED contains a red light and an infrared light that are differentially absorbed by oxygenated hemoglobin. Based on the relative absorption of the two wavelengths that is determined by the sensor, the POX determines the relative amount of oxygenated and deoxygenated hemoglobin, which is calculated as SpO2. In order to make the SpO2 calculation independent of skin color, finger size, etc., the pulse oximeter sensor uses only the time varying light absorption component generated by the patient's pulse. The sensor also uses the period of pulsation to measure patient pulse rate. The pulse oximeter can estimate the amount of oxygen in the blood without having to draw a blood sample.

The primary components of an oxygen transducer, or Pulse Oximeter (POX) Sensor, are light-emitting diodes (red and infrared LED) and a photo sensor. These components (with their wiring system) are embedded within a taping system designed for wrapping the POX Sensor around a patient's finger, foot, or hand so that the LED and photo sensor are directly opposite to each other. As the lights are emitted and received across a vascular bed, the rates of absorption at the two wavelengths vary depending upon the ratios of oxygenated hemoglobin within the blood.

The proposed devices of this submission do not differ from the predicate device. The only difference is that the proposed devices will be exposed to vaporized hydrogen peroxide instead of ethylene oxide.

AI/ML Overview

The provided text describes the 510(k) premarket notification for reprocessed Masimo Pulse Oximeters. It focuses on demonstrating substantial equivalence to a predicate device, particularly regarding the change in the reprocessing sterilization method (from ethylene oxide to vaporized hydrogen peroxide).

Here's an analysis of the acceptance criteria and the study proving the device meets them, based only on the provided text.

Acceptance Criteria and Study Details for Reprocessed Masimo Pulse Oximeters

The primary acceptance criteria for this reprocessed device appear to be maintaining the functional performance of the original device after reprocessing, particularly in light of a change in the sterilization method. The studies described are bench and clinical validation tests to demonstrate that the reprocessed device is substantially equivalent to its predicate.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance (as tested)
No open circuits (measured by continuity testing)Met (Continuity testing was conducted to verify no open circuits)
Signal transmission between photodiode and LED (measured by sensitivity testing)Met (Sensitivity testing was conducted to detect signal transmission)
Biocompatibility requirements met (per ANSI/AAMI/ISO 10993-1: 2009 for surface contacting devices with prolonged exposure to skin)Met (All reprocessed pulse oximeter sensors materials in contact with the patient are replaced with medical grade adhesive tape during reprocessing and meet biocompatibility requirements.)
SpO2 performance under no motion (ARMS for SpO2 in the range of 70% to 100%)ARMS for SpO2 was 1.48% (for woven tape) and 1.72% (for non-woven tape) over the range of 70-100%. (This is a specific performance metric, implying an acceptable range, although the acceptable range itself is not explicitly stated as a numerical threshold beyond the reported value).

2. Sample Size Used for the Test Set and Data Provenance:

  • Bench Testing (Continuity and Sensitivity): The text does not specify the exact sample size for the bench tests. It states that "Bench and laboratory testing was conducted to determine whether a pulse oximeter sensor is functional." No specific number of sensors or repetitions are mentioned.
  • Clinical Validation Testing (SpO2 performance): The clinical validation testing was performed on "healthy, adult volunteers." The exact number of volunteers is not specified in the provided text.
  • Data Provenance: The text does not explicitly state the country of origin for the clinical validation volunteers, nor does it specify if the data was retrospective or prospective. Given the nature of clinical validation testing for a medical device submission, it is typically presumed to be prospective, but this is not explicitly stated.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document does not specify the number or qualifications of experts used to establish ground truth for any of the tests. For SpO2 performance, ground truth would typically be established by a reference oximetry device or arterial blood gas measurements, rather than expert consensus on image interpretation.

4. Adjudication Method for the Test Set:

This information is not provided in the text. Adjudication methods are typically relevant for studies involving human readers or subjective interpretations. The tests described (continuity, sensitivity, SpO2 performance) are largely objective measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not reported. This type of study is more common for diagnostic imaging AI systems where human readers interpret medical images with and without AI assistance. This submission describes a reprocessed hardware device (pulse oximeter sensor) with objective performance measurements.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

The concept of "standalone" performance typically applies to algorithms. While the device contains electrical components, its performance is measured through physical and clinical tests, not as an algorithm's output. The "functional" tests (continuity, sensitivity) and clinical SpO2 performance are essentially "standalone" device performance metrics, as they evaluate the device's output and function directly, without human interpretation in the loop beyond connecting the device and reading its output.

7. The Type of Ground Truth Used:

  • Bench Testing: The ground truth for continuity and sensitivity is based on engineering specifications and direct measurement (e.g., current flow, signal detection).
  • Clinical Validation Testing: For SpO2 performance, the ground truth would typically be established by a primary reference method for oxygen saturation, such as co-oximetry (from arterial blood samples) or a highly accurate reference pulse oximeter. The text states "clinical validation testing of the SpO2 performance under no motion on healthy, adult volunteers in the range of 70% to 100%," implying comparison to a reference standard, though the specific reference standard is not detailed. It is highly likely based on arterial blood gas analysis with co-oximetry, which is the standard for SpO2 ground truth.

8. The Sample Size for the Training Set:

This information is not applicable or not provided. This submission concerns a reprocessed medical device, not an AI/ML algorithm that requires a training set. The term "training set" is generally used in the context of machine learning model development.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable as there is no "training set" for this type of medical device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

March 8, 2022

Stryker Sustainability Solutions Moira Barton-Varty Senior Principal Regulatory Affairs 1810 West Drake Drive Tempe, Arizona 85283

Re: K211140

Trade/Device Name: Reprocessed Masimo Pulse Oximeter (1859 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (1860 Pediatric O2 Transducer), Reprocessed Masimo Pulse Oximeter (1861 Infant O2 Transducer), Reprocessed Masimo Pulse Oximeter (1862 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (2317 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (2319 Infant O2 Transducer), Reprocessed Masimo Pulse Oximeter (2320 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (2328 Infant O2 Transducer), Reprocessed Masimo Pulse Oximeter (2329 Adult O2 Transducer) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: NLF Dated: February 10, 2022 Received: February 11, 2022

Dear Moira Barton-Varty:

(NOTE: Reprocessed SUD device types require a separate attachment of the list of all models cleared in the submission. A corrected SE letter will be required if the attachment is omitted.)

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and

{1}------------------------------------------------

adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K211140

Device Name Reprocessed Masimo Pulse Oximeter

Indications for Use (Describe)

Reprocessed Masimo Pulse Oximeter sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Stryker logo with the words "Sustainability Solutions" underneath. The Stryker logo is in bold, black font. Above the logo, the text "Page 1 of 3" is visible. The words "Sustainability Solutions" are also in bold, black font but are smaller than the Stryker logo.

510(k) Summary

Contact Details

Applicant Name:Stryker Sustainability Solutions
Applicant Address:1810 West Drake Drive Tempe AZ 85283 United States of America
Applicant ContactTelephone:6024100204
Applicant Contact:Mrs. Moira Barton-Varty
Applicant Contact Email:moira.barton@stryker.com

Device Name

Device Trade Name:

Reprocessed Masimo Pulse Oximeter (1859 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (1860 Pediatric O2 Transducer) Reprocessed Masimo Pulse Oximeter (1861 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (1862 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2317 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2319 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2320 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2328 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2329 Adult O2 Transducer)

Common Name: Oximeter Classification Name: Oximeter, Reprocessed Regulation Number: 870.2700 Product Code: NLF

Legally Marketed Predicate Devices

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K081238Reprocessed Masimo Pulse Oximeter SensorsNLF
K060143LNCS and SPO2.COM SensorsDQA

Device Description Summary

In a clinical setting, a pulse oximeter sensor measures the oxygen saturation of arterial blood (SpO2). A pulse oximeter sensor is composed of a light emitting diode (LED) and a sensor that are placed on opposite sides of a patient's finger or foot. The LED contains a red light and an infrared light that are differentially absorbed by oxygenated hemoglobin. Based on the relative absorption of the two wavelengths that is determined by the sensor, the POX determines the relative amount of oxygenated and deoxygenated hemoglobin, which is calculated as SpO2. In order to make the SpO2 calculation independent of skin color, finger size, etc., the pulse oximeter sensor uses only the time varying light absorption component generated by the patient's pulse. The sensor also uses the period of pulsation to measure patient pulse rate. The pulse oximeter can estimate the amount of oxygen in the blood without having to draw a blood sample.

The primary components of an oxygen transducer, or Pulse Oximeter (POX) Sensor, are light-emitting diodes (red and infrared LED) and a photo sensor. These components (with their wiring system) are embedded within a taping system designed for wrapping the POX Sensor around a patient's finger, foot, or hand so that the LED and photo sensor are directly opposite to each other. As the lights are emitted and received across a vascular bed, the rates of absorption at the two wavelengths vary depending upon the ratios of oxygenated hemoglobin within the blood.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the Stryker logo with the words "Sustainability Solutions" written below it. The Stryker logo is in bold, black letters. Above the logo, the text "Page 2 of 3" is visible. The text is smaller than the Stryker logo and is located in the upper right corner of the image.

The proposed devices of this submission do not differ from the predicate device. The only difference is that the proposed devices will be exposed to vaporized hydrogen peroxide instead of ethylene oxide.

Intended Use/Indications for Use

Reprocessed Masimo Pulse Oximeter sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.

Indications for Use Comparison

The indications for use are the same as the predicate device.

Technological Comparison

The design of the reprocessed device is the same as the predicate device. There are no changes to the claims, clinical applications, patient population, or performance specifications for use does not change from the predicate device (K081238). The same standard mechanical design and equivalent materials are utilized. There are no changes to the claims, clinical applications, patient population, or performance specifications.

The subject device, LNCS Pulse Oximeter Sensors and the predicate device, have the following key similarities:

  • intended use;
  • principle of operation;
  • form factor and components;
  • measurement application sites;
  • performance specifications;
  • environmental and mechanical specifications- Both devices have the same indicated populations;

Principle of Operation

The principle of operation of pulse oximetry is based upon the fundamental principle that hemoglobin bound to oxygen (oxyhemoglobin) and hemoglobin unbound to oxygen (deoxyhemoglobin) vary in the absorption of different wavelengths of the light and the absorptions can be used to estimate SpO2 and pulse rate. The mechanism by which this process occurs is the use of red and infrared wavelengths of light delivered by an emitter and the signal from the light absorption of oxygenated blood and deoxygenated blood to determine functional oxygen saturation of hemoglobin (SpO2).

Mechanism of Action for Achieving the Intended Effect

The Reprocessed Pulse Oximeter Sensor provides the intended effect equivalent to the previously cleared pulse oximeter sensor in that it utilizes an optical sensor that is applied to the patient's finger or toe through which light is transmitted to the photodetector that detects the signal transmissionis processed by the Pulse Oximeter to provide SpO2 and pulse rate.The principles of operation are still based upon the use of red and infrared wavelengths of light provided by an emitter and the detection of the signal from the light absorption of oxygenated blood and deoxygenated blood to provide functional oxygen saturation of hemoglobin (SpO2).

A functional pulse oximeter sensor when connected to a pulse oximeter console will have the ability to monitor SpO2 and Pulse Rate. Bench and laboratory testing was conducted to determine whether a pulse oximeter sensor is functional and is assessed by performing continuity and sensitivity testing. This included the following tests:

  • Continuity testing to verify there are no open circuits and the current along the path of a circuit is continuous
  • Sensitivity testing to detect the signal transmission between the photodiode and the LED (light emitting diode)

Non-Clinical and/or Clinical Tests Summary & Conclusions

To support the substantial equivalence of product performance after by vaporized hydrogen peroxide to that of the predicate devices, non-clinical bench simulation testing was conducted using a stand-in device allows for SpO2 sensor verification by passing the light source of the pulse oximeter sensor (LED) thru one side of

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Stryker logo with the words "Sustainability Solutions" underneath. The Stryker logo is in black and is a stylized version of the company name. The words "Sustainability Solutions" are in black and are in a smaller font than the Stryker logo. The words "Page 3 of 3" are in the upper right corner of the image.

the stand-in with the signal transmission measured by the photocell (photo-detector) of the pulse oximeter sensor on the opposing side of the stand-in.

A functional pulse oximeter sensor when connected to a pulse oximeter console will have the ability to monitor SpO2 and Pulse Rate. Benchand laboratory testing was conducted to determine whether a pulse oximeter sensor is functional and is assessed by performing continuity and sensitivity testing. This included the following tests:

  • Continuity testing to verify there are no open circuits and the current along the path of a circuit is continuous
  • Sensitivity testing to detect the signal transmission between the photodiode and the LED (light emitting diode)

Performance testing , incorporated by reference K081238, also included the following which are not altered by the use of vaporized hydrogen peroxide.

All reprocessed pulse oximeter sensors materials that are normally in contact with the patient are replaced with medical grade adhesive tape during reprocessing by Stryker Sustainability Solutions. The reprocessed pulse oximeter sensors meet biocompatibility requirements of ANSI/AAMI/ISO 10993-1: 2009: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process for surface contacting devices with prolonged exposure (> 24 hours, but less than 30 days) with skin.

Clinical Testing

Stryker Sustainability Solutions performed the clinical validation testing of the SpO2 performance under no motion on healthy, adult volunteers in the range of 70% to 100%. The ARMS for SpO2 under no motion was found to be 1.48% and 1.72% for woven and non-woven tape over the range of 70-100%.

The results of the non-clinical and clinical testing demonstrate that all requirements and performance specifications were satisfied and support the subject device is substantially equivalent to its predicate.

{7}------------------------------------------------

Reprocessed Masimo Pulse Oximeter (1859 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (1860 Pediatric O2 Transducer) Reprocessed Masimo Pulse Oximeter (1861 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (1862 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2317 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2319 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2320 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2328 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2329 Adult O2 Transducer)

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).