K Number
K221886
Manufacturer
Date Cleared
2022-10-14

(107 days)

Product Code
Regulation Number
870.1250
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sublime™ Microcatheter is intended to access the peripheral vasculature in order to facilitate the placement and/or the exchange of guidewires. The Sublime Microcatheter is also intended to provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

Device Description

The Sublime Microcatheter is a sterile, single-use, disposable intravascular catheter with an atraumatic distal tip and a proximal hub with a luer fitting. The Sublime Microcatheter is a single lumen support catheter, compatible with 014", 018", and 035" guide wire platforms and a 4F guide catheter. It has a working length of up to 200 cm, and both straight and angled tip configurations. The catheter has a stainless-steel braid supported construction with hydrophilic coating. The distal end of the catheter includes a radiopaque markerband to facilitate visualization under fluoroscopy.

AI/ML Overview

This document is a 510(k) Premarket Notification from the U.S. Food and Drug Administration (FDA) for the Sublime™ Microcatheter. It describes the device, its intended use, and the studies conducted to demonstrate its substantial equivalence to a legally marketed predicate device.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Important Note: This document pertains to a medical device (Microcatheter), not an AI/software device. Therefore, many of the typical questions for AI/software such as "Number of experts used to establish ground truth," "Adjudication method," "MRMC study," "Standalone performance," "Training set sample size," and "How ground truth for training set was established" are not applicable in this context. The acceptance criteria and studies are for a physical medical device's performance characteristics.


Acceptance Criteria and Reported Device Performance

The document states that "All test results met documented acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each test. It only lists the tests performed, implying that specific criteria were defined and met internally by the manufacturer.

Table 1: Acceptance Criteria and Reported Device Performance (as inferred from the text)

Performance Characteristic/TestAcceptance Criteria (Not explicitly stated in document, but implied meeting criteria)Reported Device Performance (Not explicitly quantified, but stated as "met documented acceptance criteria")
Flow RateMust meet intended flow rates for saline/contrast deliveryMet documented acceptance criteria
Dimensional EvaluationsMust conform to specified dimensions (e.g., ID, OD, length)Met documented acceptance criteria
RadiopacityMust be sufficiently visible under fluoroscopyMet documented acceptance criteria
Tensile StrengthMust withstand specified tensile forces without failureMet documented acceptance criteria
Freedom from LeakageMust not leak under specified pressure/conditionsMet documented acceptance criteria
Hub/Luer connector compatibilityMust be compatible with standard luer fittingsMet documented acceptance criteria
Guidewire MovementMust allow smooth guidewire placement and exchangeMet documented acceptance criteria
Burst TestingMust withstand specified internal pressures without burstingMet documented acceptance criteria
Flexibility and Kink ResistanceMust exhibit appropriate flexibility and resist kinking during insertion/maneuveringMet documented acceptance criteria
Track ForceMust achieve specified trackability over a guidewireMet documented acceptance criteria
Torque StrengthMust withstand specified torque forces without failureMet documented acceptance criteria
Hydrophilic CoatingMust maintain lubricity and integrity as specifiedMet documented acceptance criteria
Particulate TestingMust demonstrate minimal particulate releaseMet documented acceptance criteria
Atraumatic SurfacesMust have smooth surfaces to prevent vessel damageMet documented acceptance criteria
Simulated Use TestingMust perform acceptably in simulated clinical scenariosMet documented acceptance criteria
Biocompatibility
- CytotoxicityNon-cytotoxicMet documented acceptance criteria
- HemocompatibilityHemocompatibleMet documented acceptance criteria
- SensitizationNon-sensitizingMet documented acceptance criteria
- Acute System ToxicityNon-toxic acute system responseMet documented acceptance criteria
- Irritation/ReactivityNon-irritating/non-reactiveMet documented acceptance criteria
- PyrogenicityNon-pyrogenicMet documented acceptance criteria

Study Details

As this is a physical medical device, the "study" primarily consists of design verification testing rather than clinical performance studies typically seen with diagnostic or AI devices. The goal is to demonstrate "substantial equivalence" to a predicate device based on performance characteristics.

  1. Sample sizes used for the test set and the data provenance:

    • Sample Size: The document does not specify the sample sizes used for each of the listed tests (e.g., number of catheters tested for flow rate, tensile strength, etc.).
    • Data Provenance: This would typically be from laboratory bench testing and simulated use, conducted by the manufacturer (Surmodics Inc.). The country of origin is implied to be the USA, as the company is based in Eden Prairie, Minnesota, and submitting to the US FDA. The testing is prospective in nature, as it's part of the premarket notification process for a new device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable for this type of medical device clearance. "Ground truth" in this context refers to engineering specifications and performance standards established by the manufacturer and recognized industry standards (e.g., ISO standards for medical devices). The expertise would be in materials science, mechanical engineering, and quality control, rather than medical interpretation by radiologists.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable for engineering bench testing. Adjudication methods are typically for human interpretation of medical images or clinical outcomes. Deviations from expected test results would be handled through a non-conformance process and investigation by engineers and quality personnel.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. This is a physical microcatheter, not an AI software. No human reader studies (MRMC) were conducted or required for this type of device clearance.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This is a physical medical device, not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" (or more appropriately, the reference standard) for these tests is based on defined engineering specifications, performance standards (e.g., ISO standards for catheters), and validation methods for each physical characteristic (e.g., a calibrated flow meter for flow rate, a universal testing machine for tensile strength, a pressure gauge for burst testing). Biocompatibility is assessed against established biological safety standards.
  7. The sample size for the training set:

    • This is not applicable. This is a physical medical device, not a machine learning model that requires a training set.
  8. How the ground truth for the training set was established:

    • This is not applicable for the same reason as above.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).