Intended Use

This is a digital mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Device Description

This is a modified version of our previous predicate mobile PHOENIX. The predicate PHOENIX mobile is interfaced with Konica – Minolta Digital X-ray panels and CS-7 or Ultra software image acquisition. PHOENIX mobile systems will be marketed in the USA by KONICA MINOLTA. Models with the CS-7 Software will be marketed as AeroDR TX m01. Models with the Ultra software will be marketed as mKDR Xpress. The modification adds two new models of compatible Konica-Minolta digital panels, the AeroDR P-65 and AeroDR P-75, cleared in K210619. These newly compatible models are capable of a mode called DDR, Dynamic Digital Radiography wherein a series of radiographic exposures can be rapidly acquired, up to 15 frames per seconds maximum (300 frames).

AI/ML Overview

The provided text describes a 510(k) premarket notification for a mobile x-ray system. The document focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a study to prove the device meets specific performance-based acceptance criteria for an AI/algorithm.

Therefore, many of the requested details, such as specific acceptance criteria for algorithm performance, sample sizes for test sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance, are not applicable or not present in this type of submission.

The essence of this submission is that the entire mobile x-ray system, including its components (generator, panels, software), is deemed safe and effective because it is substantially equivalent to a previously cleared device, with only minor modifications (adding two new compatible digital panels and enabling a DDR function in the software, which is stated to be "unchanged firmware" and "moderate level of concern").

Here's an attempt to address your questions based on the provided text, while acknowledging that many of them pertain to AI/algorithm performance studies, which are not the focus of this 510(k):

1. A table of acceptance criteria and the reported device performance

The document does not specify performance-based acceptance criteria for an AI/algorithm. Instead, it demonstrates substantial equivalence to a predicate device by comparing technical specifications. The "acceptance criteria" in this context are implicitly met if the new device's specifications (kW rating, kV range, mA range, collimator, power source, panel interface, image area sizes, pixel sizes, resolutions, MTF, DQE) are equivalent to or improve upon the predicate, and it remains compliant with relevant international standards.

CharacteristicPredicate: K212291 PHOENIXPHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress.Acceptance Criterion (Implicit)Reported Performance
Indications for UseDigital mobile diagnostic x-ray for adults/pediatrics, skull, spine, chest, abdomen, extremities. Not for mammography.SAMEMust be identical to predicate.SAME (Identical)
ConfigurationMobile System with digital x-ray panel and image acquisition computerSAMEMust be identical to predicate.SAME (Identical)
X-ray Generator(s)kW: 20, 32, 40, 50 kW; kV: 40-150 kV (1 kV steps); mA: 10-650 mASAMEMust be identical to predicate.SAME (Identical)
CollimatorRalco R108FSAMEMust be identical to predicate.SAME (Identical)
Meets US Performance StandardYES 21 CFR 1020.30SAMEMust meet this standard.YES (Identical)
Power SourceUniversal, 100-240 V~, 1 phase, 1.2 kVASAMEMust be identical to predicate.SAME (Identical)
SoftwareKonica-Minolta CS-7 or UltraCS-7 and Ultra modified for DDR modeFunctions must be equivalent/improved; DDR enabled.CS-7 and Ultra modified for DDR mode
Panel InterfaceEthernet or Wi-Fi wirelessSAMEMust be identical to predicate.SAME (Identical)
Image Area Sizes (Panels)Listed AeroDR P-seriesListed AeroDR P-series + P-65, P-75Expanded range must be compatible and cleared.Expanded range compatible, previously cleared.
Pixel Sizes (Panels)Listed AeroDR P-seriesListed AeroDR P-series + P-65, P-75Expanded range must be compatible and cleared.Expanded range compatible, previously cleared.
Resolutions (Panels)Listed AeroDR P-seriesListed AeroDR P-series + P-65, P-75Expanded range must be compatible and cleared.Expanded range compatible, previously cleared.
MTF (Panels)Listed AeroDR P-seriesListed AeroDR P-series + P-65, P-75Performance must be equivalent or better.P-65 (Non-binning) 0.62, (2x2 binning) 0.58; P-75 (Non-binning) 0.62, (2x2 binning) 0.58
DQE (Panels)Listed AeroDR P-seriesListed AeroDR P-series + P-65, P-75Performance must be equivalent or better.P-65 0.56 @ 1 lp/mm; P-75 0.56 @ 1 lp/mm
Compliance StandardsN/AIEC 60601-1, -1-2, -1-3, -2-54, -2-28, -1-6, IEC 62304Must meet relevant international safety standards.Meets all listed IEC standards.
Diagnostic Quality ImagesN/AProduced diagnostic quality images as good as predicateMust produce images of equivalent diagnostic quality.Verified

2. Sample size used for the test set and the data provenance

No specific test set or data provenance (country, retrospective/prospective) is mentioned for AI/algorithm performance. The "testing" involved "bench and non-clinical tests" to verify proper system operation and safety, and that the modified combination of components produced diagnostic quality images "as good as our predicate generator/panel combination." This implies physical testing of the device rather than a dataset for algorithm evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. There was no specific test set requiring expert-established ground truth for an AI/algorithm evaluation. The determination of "diagnostic quality images" likely involved internal assessment by qualified personnel within the manufacturer's testing process.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No adjudication method is described as there was no formal expert-read test set for algorithm performance.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. An MRMC study was not conducted as this submission is not about an AI-assisted diagnostic tool designed to improve human reader performance. It is for a mobile x-ray system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This submission is for a medical device (mobile x-ray system), not a standalone AI algorithm. The software components (CS-7 and Ultra) are part of the image acquisition process, and the only software "modification" mentioned is enabling the DDR function, which is a feature of the new panels, not an AI for diagnosis.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. The substantial equivalence argument relies on comparing technical specifications and demonstrating that the physical device produces images of "diagnostic quality" equivalent to the predicate, rather than an AI producing diagnostic outputs against a specific ground truth.

8. The sample size for the training set

Not applicable. This is not an AI/ML algorithm submission requiring a training set. The software components are for image acquisition and processing, not for AI model training.

9. How the ground truth for the training set was established

Not applicable, as no training set for an AI/ML algorithm was used or mentioned.

{0}------------------------------------------------

July 18, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Sedecal SA % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

Re: K221803

Trade/Device Name: PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress. Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB Dated: June 16, 2022 Received: June 22, 2022

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221803

Device Name

PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress.

Indications for Use (Describe)

This is a digital mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: 510(k) Number K221803

Image /page/3/Picture/1 description: The image is a logo for SEDECAL. The logo is a teal rounded rectangle with a white border. Inside the rectangle is a white icon of a person sitting at a desk, followed by the word "SEDECAL" in white, bold letters.

SEDECAL SA C/ Pelaya, 9 - 13, Pol. Ind. Río de Janeiro 28110 Algete, Madrid España (Spain) Tel .- +34 91 6280544 Fax.- +34 91 6280574 Date Prepared: July 7, 2022 Contact: Mª Luisa Gómez de Agüero, Quality and Regulatory Manager

    1. Identification of the Device: Trade/Device Name: PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress. Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
    1. Equivalent legally marketed device: K212291 Trade/Device Name: PHOENIX Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
    1. Reference devices: We employ these cleared devices without modification: Trade/Device Name: (K151465, K172793, K210619) Konica-Minolta Digital X-ray Panels AeroDR SYSTEM 2 includes P-51 & P-52 and software CS-7 (K151465) SKR3000 includes panels P-71 & P-81 and software CS-7 (K172793) SKR3000 includes panels P-65 & P-75 with software CS-7 (K210619) (The Ultra software was cleared in our predicate K212291) Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
    1. Indications for Use: This is a digital mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

{4}------------------------------------------------

    1. Description of the Device: This is a modified version of our previous predicate mobile PHOENIX. The predicate PHOENIX mobile is interfaced with Konica – Minolta Digital X-ray panels and CS-7 or Ultra software image acquisition. PHOENIX mobile systems will be marketed in the USA by KONICA MINOLTA. Models with the CS-7 Software will be marketed as AeroDR TX m01. Models with the Ultra software will be marketed as mKDR Xpress. The modification adds two new models of compatible Konica-Minolta digital panels, the AeroDR P-65 and AeroDR P-75, cleared in K210619. These newly compatible models are capable of a mode called DDR, Dynamic Digital Radiography wherein a series of radiographic exposures can be rapidly acquired, up to 15 frames per seconds maximum (300 frames).
      The CS-7 software was cleared under K151465/K172793, while the Ultra software was cleared within our predicate K212291. The CS-7 is a DIRECT DIGITIZER used with an image diagnosis device, medical imaging device and image storage device connected via the network. This device digitally processes patient images collected by the medical imaging device to provide image and patient information. By contrast the Ultra software is designed as an exam-based modality image acquisition tool. Ultra software and its accompanying Universal Acquisition Interface (UAI) were developed to be acquisition device independent. Basic Features of the software include Modality Worklist Management (MWM) / Modality Worklist (MWL) support, DICOM Send, CD Burn, DICOM Print, and Exam Procedure Mapping. Ultra Software is designed to increase patient throughput while minimizing data input errors. Ultra is made up of multiple components that increase efficiency while minimizing errors. The main components of Ultra are the Worklist, Acquisition Interface and Configuration Utility. These components combine to create a Stable, Powerful, and Customizable Image capture system. The intuitive graphical user interface is designed to improve Radiology, Technologist accuracy, and image quality. Worklist and Exam screens were developed to allow site specific customizations to seamlessly integrate into existing practice workflows. PHOENIX/AeroDR TX m01 employs the "CS-7" software. PHOENIX/mKDR Xpress employs the "Ultra" software. We decided to continue to offer the CS-7 package as a lower cost option for those users who do not need the enhanced features of the Ultra software. Both software packages are of Moderate level of concern and are based on the predicate software. The ONLY DIFFERENCE between the software offered and the predicate is the fact that the DDR function has been enabled because the two newly supported panels (P65, P75) can support that function. As noted in section 3, above, reference devices: All digital flat panel detectors and software have received FDA clearance.
CharacteristicPredicate: K212291 PHOENIXPHOENIX/AeroDR TX m01 andPHOENIX/mKDR Xpress.
Indications forUse:This is a digital mobile diagnostic x-ray systemintended for use by a qualified/trained doctoror technician on both adult and pediatricsubjects for taking diagnostic radiographicexposures of the skull, spinal column, chest,abdomen, extremities, and other body parts.Applications can be performed with the patientsitting, standing, or lying in the prone or supineposition. Not for mammographyThis is a digital mobile diagnostic x-ray systemintended for use by a qualified/trained doctoror technician on both adult and pediatricsubjects for taking diagnostic radiographicexposures of the skull, spinal column, chest,abdomen, extremities, and other body parts.Applications can be performed with the patientsitting, standing, or lying in the prone or supineposition. Not for mammography. (Same)
ConfigurationMobile System with digital x-ray panel andimage acquisition computerSAME
X-rayGenerator(s)kW rating: 20 kW, 32 kW, 40 kW and 50 kW.kV range: from 40 kV to 150 kV in 1 kV steps.SAME
CharacteristicPredicate: K212291 PHOENIXPHOENIX/AeroDR TX m01 andPHOENIX/mKDR Xpress.
mA range: from 10 mA to 630 mA / 640 mA /650 mA.
CollimatorRalco R108FSAME
PhotosImage: K212291 PhoenixImage: PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress
Meets USPerformanceStandardYES 21 CFR 1020.30SAME
Power SourceUniversal power supply, from 100 V~ to 240 V~.1 phase, 1.2 kVASAME
SoftwareKonica-Minolta control software CS-7 (K151465or K172793) OR Konica-Minolta controlsoftware Ultra.CS-7 and Ultra software programs have beenmodified to accommodate DDR mode.PHOENIX/AeroDR TX m01 employs the "CS-7"software.PHOENIX/mKDR Xpress employs the "Ultra"software.
Panel InterfaceEthernet or Wi-Fi wirelessSAME
DETECTOR TECHNOLOGICAL COMPARISONS
Image areasizesAeroDR P-51 14" x 17"AeroDR P-52 14" x 17"AeroDR P-61 14" x 17"AeroDR P-71 17" x 17"AeroDR P-81 10" x 12"AeroDR P-51 14" x 17"AeroDR P-52 14" x 17"AeroDR P-61 14" x 17"AeroDR P-71 17" x 17"AeroDR P-81 10" x 12"Similar range of sizes (All previously cleared)Added:AeroDR P-65 14" x 17"
CharacteristicPredicate: K212291 PHOENIXPHOENIX/AeroDR TX m01 andPHOENIX/mKDR Xpress.
Pixel sizesAeroDR P-51 175 μmAeroDR P-52 175 μmAeroDR P-61 100/200 μmAeroDR P-71 100/200 μmAeroDR P-81 100/200 μmAeroDR P-51 175 μmAeroDR P-52 175 μmAeroDR P-61 100/200 μmAeroDR P-71 100/200 μmAeroDR P-81 100/200 μmAdded:AeroDR P-65 100 μm / 200 μm / 400 μmAeroDR P-75 100 μm / 200 μm / 400 μmSimilar pixel sizes (All previously cleared)
ResolutionsAeroDR P-51 1994 x 2430 pixelsAeroDR P-52 1994 x 2430 pixelsAeroDR P-61 3488 × 4256 pixelsAeroDR P-71 4248 × 4248 pixelsAeroDR P-81 2456 × 2968 pixelsAeroDR P-51 1994 x 2430 pixelsAeroDR P-52 1994 x 2430 pixelsAeroDR P-61 3488 × 4256 pixelsAeroDR P-71 4248 × 4248 pixelsAeroDR P-81 2456 × 2968 pixelsAdded:AeroDR P-65 3,488 × 4,256 pixelsAeroDR P-75 4,248 × 4,248 pixels
MTFAeroDR P-51 0.30 @ 2cy/mmAeroDR P-52 0.30 @ 2cy/mmAeroDR P-61 0.30 @ 2cy/mmAeroDR P-71 0.30 @ 2cy/mmAeroDR P-81 0.30 @ 2cy/mmAeroDR P-51 0.30 @ 2cy/mmAeroDR P-52 0.30 @ 2cy/mmAeroDR P-61 0.30 @ 2cy/mmAeroDR P-71 0.30 @ 2cy/mmAeroDR P-81 0.30 @ 2cy/mmAdded:AeroDR P-65 (Non-binning) 0.62(2x2 binning) 0.58AeroDR P-75 (Non-binning) 0.62(2x2 binning) 0.58
DQEAeroDR P-51 0.62 @ 0 lp/mmAeroDR P-52 0.62 @ 0 lp/mmAeroDR P-61 0.56 @ 1 lp/mmAeroDR P-71 0.56 @ 1 lp/mmAeroDR P-81 0.56 @ 1 lp/mmAeroDR P-51 0.62 @ 0 lp/mmAeroDR P-52 0.62 @ 0 lp/mmAeroDR P-61 0.56 @ 1 lp/mmAeroDR P-71 0.56 @ 1 lp/mmAeroDR P-81 0.56 @ 1 lp/mmAdded:AeroDR P-65 0.56 @ 1 lp/mmAeroDR P-75 0.56 @ 1 lp/mm
SoftwareKonica-Minolta control software CS-7 (K151465or K172793) OR Konica-Minolta controlsoftware Ultra.Konica-Minolta control software CS-7 (K151465or K172793) OR Konica-Minolta controlsoftware Ultra (K212291). Both support DDR.
Panel InterfaceEthernet or Wi-Fi wirelessSAME

6) Substantial Equivalence Chart

{5}------------------------------------------------

{6}------------------------------------------------

  1. The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness.

Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and

{7}------------------------------------------------

produced diagnostic quality images as good as our predicate generator/panel combination. Here is a summary of the changes we made to the device: We added two more compatible flat panel detectors: AeroDR P-65 and the AeroDR P-75. These two panels are capable of acquiring up to 20 seconds of serial radiography.

    1. Summary of non-clinical testing: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. Firmware was unchanged. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff. The CS-7 digital panel software employed was already reviewed by FDA in K172793 and subsequently in our predicate K212291. The Ultra digital panel software was cleared in our predicate and has not been modified.
      The PHOENIX Battery Mobile X-Ray Units have been tested to be in compliance with the following International Standards:
  • a) IEC 60601-1:2005+A1:2012 (Edition 3.1)

  • b) IEC 60601-1-2:2014 (Edition 4.0)

  • c) IEC 60601-1-3:2008+A1:2013 (Edition 2.1)

  • d) IEC 60601-2-54:2009+A1:2015 (Edition 1.1)

  • e) IEC 60601-2-28:2010 (Edition 3.0)

  • IEC 60601-1-6:2010 + A1:2013 (Edition 3.1) f)

  • g) IEC 62304:2006 + A1:2016 (Edition 1.1)

    1. Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance.
    1. Conclusion: After analyzing bench and non-clinical tests, it is the conclusion of Sedecal SA. that the modified PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress Digital Diagnostic Mobile X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.