K Number
K221741
Date Cleared
2023-02-22

(252 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This air-powered dental handpiece is intended for removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures. It is designed for use by a trained professional in the field of general dentistry.

Device Description

The TRAUS Air Dental handpieces are air driven dental handpieces for the use by a trained dental professional. The devices are air-powered handpieces that are reusable and designed accordance with international standard. The devices can be sterilized by the prevacuum steam autoclave methods that have been validated. Through the coupling connected to a dental unit, the proposed dental handpiece receive air for functionality of the high-speed turbine. They also receive cooling water for cutting through one port and light for illumination through another port.

Using in combination with dental unit chair, it is possible to use by installing the instrument at air handpiece according to purpose. It is possible to change the rotation speed by adjusting the flowing air.

AI/ML Overview

The provided document is a 510(k) premarket notification for a dental handpiece (TRAUS Air Dental Handpiece) and does not contain information about acceptance criteria and a study proving a device meets those criteria in the context of an AI/ML medical device.

The document discusses the substantial equivalence of the TRAUS Air Dental Handpiece to a predicate device (Maxima PRO 45L) based on technical characteristics and non-clinical performance data. It explicitly states: "No clinical performance testing was performed."

Therefore, I cannot provide the requested information, such as:

  1. Table of acceptance criteria and reported device performance: This document does not establish performance criteria in the way an AI/ML device study would (e.g., sensitivity, specificity, AUC). Instead, it compares the technical specifications and operational principles of the new device to a predicate device.
  2. Sample size, data provenance, number of experts, adjudication method, MRMC study, standalone performance: These are all concepts relevant to clinical or AI/ML performance studies, which were explicitly stated as "not performed."
  3. Type of ground truth: Not applicable as it's a mechanical device with no AI component requiring ground truth.
  4. Sample size for training set, how the ground truth for the training set was established: Not applicable as there is no training set for an AI/ML model.

The "Non-clinical Performance Data" section (page 6) details bench tests conducted to ensure conformity to ISO standards (ISO 14457, ISO 17665-1, ISO 9168) and demonstrate substantial equivalence. These are engineering and sterilization validation tests, not performance studies as typically understood for AI/ML devices.

In summary, the provided text describes a 510(k) submission for a conventional dental handpiece, and thus does not contain the information requested about acceptance criteria and a study for an AI/ML driven device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

February 22, 2023

Saeshin Precision Co., Ltd. % Sanghwa Myung Regulatory Affair Consultant E&M D-1474. 230. Simin-daero, Dongan-gu Anyang, Gyeonggi-do 14067 South Korea

Re: K221741

Trade/Device Name: TRAUS Air Dental Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: November 18, 2022 Received: November 25, 2022

Dear Sanghwa Myung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adjodha -S

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221741

Device Name TRAUS Air Dental Handpiece

Indications for Use (Describe)

This air-powered dental handpiece is intended for removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures. It is designed for use by a trained professional in the field of general dentistry.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)X
Over The Counter Use (21 CFR 801 Subpart C)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for SAESHIN. The logo consists of a blue geometric shape resembling a stylized cube or a series of interconnected squares. Below the shape, the word "SAESHIN" is written in a sans-serif font, also in blue. The overall design is clean and modern.

510(k) Summary

K221741

Submitter: Saeshin Precision Co., Ltd. # 52, Secheon-ro 1-gil, Dasa-eup, Dalseong-gun, Daegu, 42921, Republic of Korea Telephone: +82 53 587 2345 Fax: +82 53 580 0939 Email: ksqc@saeshin.com

Contact: E&M Requlatory Affair/Sang Hwa Myunq Telephone: +80-70-7807-0550 FAX: +82-31-388-9263 Cellphone: +82-10-4952-6638 E-mail: mshenmc@gmail.com Primary Contact: Sanghwa Myung

Date 510(k) summary prepared: February 21, 2023

Common Name: Handpiece, Air-Powered, Dental Trade Name: TRAUS Air Dental Handpiece Classification: I Product Code: EFB Classification Panel: Dental Requlation Numbers: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories

Description of Device:

The TRAUS Air Dental handpieces are air driven dental handpieces for the use by a trained dental professional. The devices are air-powered handpieces that are reusable and designed accordance with international standard. The devices can be sterilized by the prevacuum steam autoclave methods that have been validated. Through the coupling connected to a dental unit, the proposed dental handpiece receive air for functionality of the high-speed turbine. They also receive cooling water for cutting through one port and light for illumination through another port.

Using in combination with dental unit chair, it is possible to use by installing the instrument at air handpiece according to purpose. It is possible to change the rotation speed by adjusting the flowing air.

Indication for use:

This air-powered dental handpiece is intended for removal of carious material, cavities and crown preparations, removal of filings, processing of tooth, restoration of surfaces and as a surgical tool for third molar removal procedures.

It is designed for use by a trained professional in the field of general dentistry.

{4}------------------------------------------------

Predicate Device:

Manufacturer: KaVo do Brasil Industria e Comercio Ltda

510(k) Number: K141576 Trade Name: Maxima PRO 45L Common Name: Handpiece, Air-Powered, Dental Regulation Name: Dental handpiece and accessories Regulation Numbers: 21 CFR 872.4200 Product Code: EFB Classification: Class I

Substantial Equivalence:

Comparison table is as follows.

Table 1: Substantial equivalence comparison 1) Predicate Device

1) Predicate DeviceContentsSubject DevicePredicate Device
510(k)NumberNot availableK141576
Trade NameTRAUS Air Dental HandpieceMaxima PRO 45L
Indication for UseThis air-powered dentalhandpiece is intended forremoval of carious material,cavities and crownpreparations, removal of filings,processing of tooth, restorationof surfaces and as a surgicaltool for third molar removalprocedures.It is designed for use by atrained professional in the fieldof general dentistry.This air-powered dentalhandpiece is intended forremoval of cariousmaterial, cavities and crowpreparations, removal offilings, processing of tooth,restoration of surfaces andas a surgical tool for thirdmolar removal procedures.It is designed for use by atrained professional in thefield of general dentistry.
Operation principalThrough the tube and thecoupling connected to adental unit, the airpoweredhandpiecereceives the air foroperating the high speedturbine, the cooling waterfor cutting treatmentthrough one port and lightfor illumination theoperation area.Through the tube and thecoupling connected to adental unit, the airpoweredhandpiecereceives the air foroperating the high speedturbine, the cooling waterfor cutting treatmentthrough one port and lightfor illumination theoperation area.
Head SizeATN400: Ø11.2 x13.55mmCAB10LK: Ø11.2 x13.7mCAB10LN: Ø11.2 x14.2mCAB20ND M4: Ø11.2 x13.55mCAB20ND B2: Ø11.2 x13.55mISO1797-20174.ClassificationHead size-Height: 14,6mmHead size-Diameter: 12,5mm
Type of chuck-Type 3: FG(Friction Grip)CAB10LN/CAB10LK/CAN10NN/CAB20ND M4 : 4holePush Button
Water portsCAB20ND B2: 1hole1 hole
Coupling DimensionsLength with coupling:Length with coupling:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for SAESHIN. The logo consists of a blue cube-like shape on top of the word "SAESHIN" in blue block letters. The cube is made up of several smaller cubes, and the word "SAESHIN" is in a sans-serif font.

TRAUS CAN10NN + TRAUSCP10NQApprox. 129.59 mmTRAUS CAB10LN + TRAUSCP10LNApprox. 143.53 mmTRAUS CAB10LK + TRAUSCP10LKApprox. 138.97 mmApprox. 121 mm
Chemical composition ofthe patient- contactingportions of the device303F303F
Light IntensityApproximately 7,000 LUXApproximately 25,000 LUX
Bur retention forceup to 24 Ncmup to 24 Ncm
Operating Pressure43.5 psi (0.3Mpa)recommended40 ± 1.45 psirecommended
Idling Speed360,000 - 440,000 rpm380,000 - 420,000 rpm
Head angle90-degree45-degree
Compliance to StandardsComplies with ISO 14457and ISO 9168Complies with ISO 14457and ISO 9168
LubricantPana Plus SprayKaVo SprayHenry Schein Spray &Clean

Biocompatibility: Materials tested in accordance with 10993-1 are used and materials that have been confirmed to be biologically safe are used.

Patient contact parts of the TRAUS Air Dental Handpiece are as follows.

PartsRaw MaterialSurface TreatmentPatient contactDuration of contact by ISO 10993-1
HeadSUS303FHard Chrome coatedYesLimited(<24 hours)
    1. ISO 10993-1: 2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
    1. ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
    1. ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
    1. ISO 10993-12:2012 Biological evaluation of medical devices Part 12: Sample preparation and reference materials

{6}------------------------------------------------

Non-clinical Performance Data:

  • . The bench tests were performed in order to ensure the performance of the TRAUS Dental Air Handpieces (ATN400,CAB10LK,CAB10LN,CAB20ND M4,CAB20ND B2) verify conformity to ISO 14457 and demonstrate substantial equivalence to the predicates. ATN400,CAB10LK,CAB10LN,CAB20ND M4,CAB20ND B2 samples were compliant with ISO 14457: 2017 Dentistry -Handpieces and Motors and demonstrated substantial equivalence to the predicates.
  • . Sterilization has been validated in conformance to the FDA recognized consensus standard ISO 17665-1:2006 Sterilization of health care products moist heat - Part 1: requirements for the development, validation and routine control of a sterilization process for medical devices. Cleaning validation testing is performed in accordance with recommended evaluations as listed in AAMI TIR30, AAMI TIR12, and Guidance for Industry and FDA Staff -Processing/Reprocessing Medical Devices in Health Care Settings.
  • . ISO 9168:2009 Dentistry— Hose connectors for air driven dental handpieces, device's hose connection has been verified by standard of ISO 9168.

Clinical Data: No clinical performance testing was performed.

Conclusion

The performance of TRAUS Air Dental Handpiece meets the requirements of the non-clinical bench testing conducted to support substantial equivalence. Based on the available information, the subject device and the predicates are similar indication for use, operational principal, performance data.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that we conclude that substantially equivalent with predicate device.

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.