K Number
K221416
Date Cleared
2022-08-31

(107 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Safety Syringes for Single Use are intended for use in the aspiration of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to mimize risk of accidental needle sticks. The Sterile Safety Syringes for Single Use are single use, disposable devices, provided sterile.

Device Description

The proposed device consists of a safety hypodermic needle and a luer slip or luer lock syringe. Models of Sterile Safety Syringes for Single Use shown in Table 11.1 and 11.2 are available in various models according to different syringe volume and needle specifications.

AI/ML Overview

The provided text describes the 510(k) premarket notification for "Sterile Safety Syringes for Single Use." This document primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a standalone study of the device's performance against specific acceptance criteria in the manner one would see for an AI/ML medical device or a novel product requiring a clinical trial.

Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" (especially those related to clinical studies, human expert involvement, ground truth, and training data for AI/ML) are not applicable to this type of submission.

However, I can extract information related to the technical performance tests conducted to show equivalence.

Here's a breakdown based on the provided document:

1. A table of acceptance criteria and the reported device performance:

The document doesn't explicitly state "acceptance criteria" as pass/fail thresholds for clinical outcomes, but rather lists the standards to which the device conforms and the results of the tests confirming that conformity. The "acceptance" is that the device's performance meets the requirements of these recognized standards.

Acceptance Criterion (Standard Compliance)Reported Device Performance (Test Conclusion)
Biocompatibility:
ISO 10993-1 (Evaluation & testing within risk management)Conforms (device not adverse to human tissue)
ISO 10993-2 (Animal welfare)Conforms
ISO 10993-4 (Interactions with blood)Conforms
ISO 10993-5 (In vitro cytotoxicity)Conforms (No Cytotoxicity)
ISO 10993-10 (Irritation & delayed-type hypersensitivity)Conforms (No Irritation to Skin, No significant evidence of sensitization)
ISO 10993-11 (Systemic toxicity)Conforms (No systemic toxicity)
ISO 10993-12 (Sample preparation/reference)Conforms
ASTM 756-17 (Hemolytic properties)Conforms (No Hemolysis)
Sterilization & Packaging:
ISO 10993-7 (Ethylene Oxide Sterilization Residuals)Conforms
USP 43-NF38:2020 <85> (Bacterial Endotoxins)Conforms (Non pyrogenic)
USP 43-NF38:2020 <151> (Pyrogen Test)Conforms (Non pyrogenic)
ASTM F88/F88M-15 (Seal Strength)Conforms
ASTM F1929-15 (Detecting Seal Leaks)Conforms
ASTM F1980-16 (Accelerated Aging)Conforms
Device Performance & Safety Feature:
ISO 7864:2016 (Sterile hypodermic needles)Conforms
ISO 7886-1:2017 (Sterile hypodermic syringe)Conforms
ISO 9626:2016 (Stainless steel needle tubing)Conforms
ISO 80369-7:2021 (Small-bore connectors)Conforms
ISO 23908:2011 (Sharps injury protection)Conforms
FDA Guidance "Medical Devices with Sharps Injury Prevention Features"Simulated clinical use study conducted; device complies with requirements.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the exact number of units tested for each standard. This level of detail is typically found in the full test reports, which are summarized in the 510(k) submission.
  • Data Provenance: The tests were non-clinical performance tests conducted for the purpose of demonstrating substantial equivalence. The document doesn't explicitly state the country of origin for the testing data, but the manufacturer is based in Shanghai, China. The tests are "retrospective" in the sense that they were completed prior to this submission for regulatory review.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not applicable. This submission refers to a traditional medical device (syringes), not an AI/ML device or one requiring human expert interpretation of complex data for "ground truth" establishment. The "ground truth" for these tests is defined by the objective measurements and pass/fail criteria specified within the referenced ISO and ASTM standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This type of study is relevant for diagnostic devices interpreting complex data (e.g., radiology images) and involving human readers. It is not applicable to a safety syringe.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This isn't an algorithm. The "performance" is mechanical and material integrity.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For this device, the "ground truth" is established by the objective, quantitative requirements and methods defined in the listed international and national standards (ISO, ASTM, USP) for sterility, biocompatibility, and physical performance of hypodermic needles and syringes. For instance, a test for "Force to activate safety mode" would have a defined NMT (Not More Than) value (e.g., 10N), and passing means the device operates within that mechanical specification.

8. The sample size for the training set:

  • Not applicable. This refers to machine learning. No training set was used for a physical syringe.

9. How the ground truth for the training set was established:

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

Shanghai Kindly Enterprise Development Group Co., Ltd % Evan Hu Official Correspondent Shanghai Mind-Link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District Shanghai, 200040 Cn

Re: K221416

Trade/Device Name: Sterile Safety Syringes for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG, FMF, FMI Dated: August 05, 2022 Received: August 08, 2022

Dear Evan Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Section 4

Indications for Use Statement

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K221416

Device Name Sterile Safety Syringes for Single Use

Indications for Use (Describe)

The Sterile Safety Syringes for Single Use are intended for use in the aspiration of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to mimize risk of accidental needle sticks. The Sterile Safety Syringes for Single Use are single use, disposable devices, provided sterile.

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

{4}------------------------------------------------

Section 5

510(K) Summary

{5}------------------------------------------------

I. SUBMITTER:

Shanghai Kindly Enterprise Development Group Co., Ltd. No.658 Gaochao Road, Shanghai, 201803,China Contact Person: Liu Hualong Title: Technical Director Tel: 86-69118232-8125 Email: henry liu@kdlchina.net

Submission Correspondent: Evan Hu Email: evan.hu@mind-link.net Tel:+86 18616124827

II. Proposed Device

Name of Device: Sterile Safety Syringes for Single Use Regulation Number: 21 CFR PART 880.5860 Common Name: Syringe, Antistick Syringe, Piston Needle, Hypodermic, Single Lumen Classification Name: Piston syringe Regulatory Class: II Product Code: MEG, FMF, FMI

{6}------------------------------------------------

III. PREDICATE DEVICE

Name of Device: Safety Needles

Sterile Syringe Sterile Syringe with Safety Needle Sterile Syringe with Needle Hypodermic Needle for Single Use (K183665) Regulation Number: 21 CFR 880.5860 Common Name: Syringe, Antistick Syringe, Piston Needle, Hypodermic, Single Lumen Classification Name: Piston syringe Regulatory Class: II Product Code: MEG, FMF, FMI

IV. DEVICE DESCRIPTION

The proposed device consists of a safety hypodermic needle and a luer slip or luer lock syringe. Models of Sterile Safety Syringes for Single Use shown in Table 11.1 and 11.2 are available in various models according to different syringe volume and needle specifications.

Table 11-1 Models of Sterile Safety Syringes for Single Use (Luer Lock Type)

SyringeMatched needle
volumeREFNeedle gaugeNeedle lengthNeedle wallBevel
1mlSRS2101232523G25mmThin wall$11\pm2^\circ$
SRS2101252525G25mmThin wall11±2°
SRS2101251625G16mmThin wall11±2°
SRS2101271327G13mmRegular wall11±2°
SRS2102232523G25mmThin wall11±2°
SRS2102252525G25mmThin wall11±2°
2mlSRS2102251625G16mmThin wall11±2°
SRS2102271327G13mmRegular wall11±2°
SRS2103183818G38mmThin wall11±2°
SRS2103202520G25mmThin wall11±2°
SRS2103203820G38mmThin wall11±2°
SRS2103212521G25mmThin wall11±2°
SRS2103213821G38mmThin wall11±2°
3mlSRS2103222522G25mmThin wall11±2°
SRS2103223822G38mmThin wall11±2°
SRS2103232523G25mmThin wall11±2°
SRS2103233823G38mmThin wall11±2°
SRS2103252525G25mmThin wall11±2°
SRS2103253825G38mmThin wall11±2°
SRS2105183818G38mmThin wall11±2°
SRS2105202520G25mmThin wall11±2°
SRS2105203820G38mmThin wall11±2°
SRS2105212521G25mmThin wall11±2°
5mlSRS2105213821G38mmThin wall11±2°
SRS2105222522G25mmThin wall11±2°
SRS2105223822G38mmThin wall11±2°
SRS2105232523G25mmThin wall11±2°
SRS2105233823G38mmThin wall$11\pm2°$
SRS2105252525G25mmThin wall$11\pm2°$
10mlSRS2105253825G38mmThin wall$11\pm2°$
SRS2110213821G38mmThin wall$11\pm2°$
SRS2110223822G38mmThin wall$11\pm2°$

{7}------------------------------------------------

{8}------------------------------------------------

Table 11-2 Models of Sterile Safety Syringes for Single Use (Luer Slip Type)

SyringeMatched needle
volumeREFNeedle gaugeNeedle lengthNeedle wallBevel
1mlSRS2201232523G25mmThin wall11±2°
SRS2201252525G25mmThin wall11±2°
SRS2201251625G16mmThin wall11±2°
SRS2201271327G13mmRegular wall11±2°
2mlSRS2202232523G25mmThin wall11±2°
SRS2202252525G25mmThin wall11±2°
SRS2202251625G16mmThin wall11±2°
SRS2202271327G13mmRegular wall11±2°
3mlSRS2203183818G38mmThin wall11±2°
SRS2203202520G25mmThin wall11±2°
SRS2203203820G38mmThin wall11±2°
SRS2203212521G25mmThin wall11±2°
SRS2203213821G38mmThin wall11±2°
SRS2203222522G25mmThin wall11±2°
SRS2203223822G38mmThin wall11±2°
SRS2203232523G25mmThin wall11±2°

{9}------------------------------------------------

SRS2203233823G38mmThin wall11±2°
SRS2203252525G25mmThin wall11±2°
SRS2203253825G38mmThin wall11±2°
5mlSRS2205183818G38mmThin wall11±2°
SRS2205202520G25mmThin wall11±2°
SRS2205203820G38mmThin wall11±2°
SRS2205212521G25mmThin wall11±2°
SRS2205213821G38mmThin wall11±2°
SRS2205222522G25mmThin wall11±2°
SRS2205223822G38mmThin wall11±2°
SRS2205232523G25mmThin wall11±2°
SRS2205233823G38mmThin wall11±2°
SRS2205252525G25mmThin wall11±2°
SRS2205253825G38mmThin wall11±2°
10mlSRS2210213821G38mmThin wall11±2°
SRS2210223822G38mmThin wall11±2°

V. INDICATIONS FOR USE

The Sterile Safety Syringes for Single Use are intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks. The Sterile Safety Syringes for Single Use are single use, disposable devices, provided sterile.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

{10}------------------------------------------------

The Sterile Safety Syringes for Single Use are compared with the predicate device, Safety Needles, Sterile Syringe, Sterile Syringe with Safety Needle, Sterile Syringe with Needle, Hypodermic Needle for Single Use (K183665). The results are shown below in the Technological Characteristics Comparison Table:

DeviceSterile Safety Syringes forSingle UseSafetyNeedles,SterileSyringe, Sterile Syringe withSafety Needle, Sterile SyringewithNeedle,HypodermicNeedleforSingleUse(K183665)Remark
Indication for UseThe Sterile Safety Syringes forSingle Use are intended foruse in the aspiration andinjection of fluids for medicalpurpose. After withdrawal ofthe needle from the body, theattached needle safety shieldcan be manually activated tocover the needle immediatelyafter use to minimize risk ofaccidental needle sticks. TheSterile Safety Syringes forSingle Use are single use,disposable devices, providedsterile.The sterile syringe with safetyneedle is intended for use in theaspiration and injection of fluidsfor medical purpose.After withdrawal of the needlefrom the body, the attachedneedle safety shield can bemanually activated to cover theneedle immediately after use tominimize risk of accidentalneedle sticks.Same
ClassificationProduct CodeMEG, FMF,FMIMEG, FMF,FMISame
Prescription OnlyOverorCounterPrescriptionPrescriptionSame
Regulation No.21 CFR 880.586021 CFR 880.5860Same
ClassIIIISame
ConfigurationBarrelBarrel (Polypropylene)Same
PlungerPlunger (Polypropylene)
Plunger stopperStopper (Polyisoprene)
Needle tubeNeedle cap (Polypropylene)
Protective needle hubNeedle tube (SUS304)
Protective capHub with protector(Polypropylene)
SterilityEO SterilizedEO SterilizedSame
SizeVolume1ml, 2ml, 3ml, 5ml, 10ml1ml, 3ml, 5ml, 10ml, 20ml,30ml, 50ml, and 60mlDifferentNote 1
Syringe TypeLuer Lock/Luer SlipLuer Lock/Luer Slip/Luer slipEccentric
Needle Gauge18G, 20G, 21G, 22G, 23G,25G, 27G18G, 20G, 21G, 22G, 23G, 25G,and 26G
Needle Length13mm, 16mm, 25mm, 38mm1/2"-2"Same
Single UseYesYesSame
No CytotoxicityNo CytotoxicitySame
BiocompatibilityNo Irritation to SkinNo Irritation to SkinSame
No significant evidence ofsensitizationNo significant evidence ofsensitizationSame
No systemic toxicityNo systemic toxicitySame
No HemolysisNon pyrogenicNo HemolysisNon pyrogenicSame
Performance TestConforms with therequirements of ISO 7864,ISO 9626, ISO 7886, and ISO80369-7.Needle Safety Feature:TestingConforms with the requirementsof ISO 7864, ISO 9626, ISO7886, and ISO 80369-7.Needle Safety Feature: Testingconducted per ISO 23908:2011.Same

{11}------------------------------------------------

{12}------------------------------------------------

23908:2011.Force to activate safety mode:NMT 10N
Force to disengage safety mode:
NLT 20N
Force to separate safety featurefrom needle hub: NLT 50N

Note 1

Analysis: Although there are differences between proposed device and predicate device in terms of syringe type, syringe volume, needle length and needle gauge, corresponding performance tests were conducted on all the specifications including ISO 7886-1, ISO 7864-1 and ISO 9626-1 to prove it meet the specific requirements. Therefore, this difference does not raise different questions of safety and effectiveness.

VII PERFORMANCE DATA

Non-Clinical Performance Test Conclusion

Biocompatibility

The proposed device is external communicating device, blood path indirect, contact duration is limited (≤24hrs). The biocompatibility tests were conducted to verify that the proposed devices are not adverse to human tissue based on the following standards:

  • ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and . testing within a risk management process
  • ISO 10993-2:2006 Biological evaluation of medical devices-Part 2: Animal . welfare requirements
  • . ISO 10993-4:2017 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood
  • ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity

{13}------------------------------------------------

  • ISO 10993-10:20101Biological evaluation of medical devices- Part 10: Test for ● irritation and delayed-type hypersensitivity
  • ISO 10993-11: 2017 Biological evaluation of medical devices -- Part 11: Tests ● for systemic toxicity
  • ISO 10993-12:2012 Biological evaluation of medical devices -- Part 12: Sample . preparation and reference materials
  • ASTM 756-17 Standard Practice for Assessment of Hemolytic Properties of . Material

Other non-clinical tests were conducted to verify that the proposed device met all design specifications was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the Sterile Safety Syringe for Single Use (Retractable) complies with the following standards:

  • ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene ● Oxide Sterilization Residuals
  • . USP 43-NF38:2020 <85> Bacterial Endotoxins Test
  • USP 43-NF38:2020 <151> Pyrogen Test (USP Rabbit Test) .
  • . ASTM F88/F88M-15, Standard Test Method For Seal Strength Of Flexible Barrier Materials
  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous . Medical Package by Dye Penetration
  • ASTM F1980 16 Standard Guide for Accelerated Aging of Sterile Barrier . Systems for Medical Devices
  • ISO 7864:2016 Sterile hypodermic needles for single use-requirements and test . method
  • ISO 7886-1:2017 Sterile hypodermic syringe for single use- Part 1: Syringes for manual use
  • . ISO 9626:2016 Stainless steel needle tubing for the manufacture of the medical devices-requirements and test method
  • ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare Part 7: Connectors for intravascular or hypodermic applications ---applications

{14}------------------------------------------------

  • ISO 23908:2011 Sharps injury protection -- Requirements and test methods --● Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
    In addition, a simulated clinical use study was conducted on the subject device, Sterile Safety Syringes for Single Use to evaluate the effect of safety feature per FDA Guidance "Medical Devices with Sharps Injury Prevention Features" issued on August 9, 2005. The test results demonstrated that the subject device complies with the requirements.

Clinical Test Conclusion

No clinical study is included in this submission.

VIII CONCLUSION

The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).