K Number
K220738
Device Name
Motive Knee Wrap
Date Cleared
2022-05-12

(59 days)

Product Code
Regulation Number
890.5850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Motive™ Knee Wrap device is intended to strengthen the quadriceps muscle stimulation to provide symptomatic temporary pain relief associated with knee arthritis.

The Motive™ Knee Wrap device is indicated for adults of 22 years of age and older.

Device Description

The Motive Knee Wrap device (Motive device) is an over-the-counter (OTC) electrical muscle stimulator (or Neuromuscular Electrical Stimulation (NMES) therapy) for relief of knee pain associated with knee arthritis. The Motive device is used to apply an electrical current through a power regulated output closed-loop feedback NMES system to provide relief of knee pain associated with arthritis. The stimulator continuously contracts quadricep muscle groups for strengthening of the muscles and knee pain relief. The stimulator is placed over the thigh muscle and just above the knee.

AI/ML Overview

The provided text describes CyMedica Orthopedics, Inc.'s Motive Knee Wrap device and its 510(k) summary (K220738). It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study. The document primarily details non-clinical bench testing, software validation, usability studies, and a comparison of technological characteristics with predicate devices.

Therefore, many of the requested details about acceptance criteria and a study proving those criteria are not explicitly present as a separate clinical study for the Motive Knee Wrap itself. Instead, the justification for effectiveness relies on the equivalence to a predicate device (CyMedica IntelliHab System, K210604) for which clinical data was previously submitted.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of "acceptance criteria" for a clinical study with corresponding reported device performance outcomes for the Motive Knee Wrap. The performance information provided is primarily related to non-clinical (bench) testing of electrical and physical characteristics to meet safety and performance standards.

Table: Performance Summary from Non-Clinical Testing

Acceptance Criteria Category (Implied by testing)Reported Device PerformanceReference Section
Electrical SafetyComplies with IEC 60601-1, IEC 60601-2-10, and 60601-11Section 9
Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2 standardSection 9
Software ValidationMeets design requirements, in accordance with FDA Guidance (May 11, 2005) and IEC 62304:2006Section 10
CybersecuritySystem assessment and mitigations for hazards and risks included in submission; complies with FDA guidanceSection 11
Usability/Human FactorsUsability supported by self-selection and human factors study; complies with IEC 60601-1-6:2010 and IEC 62366Section 12
BiocompatibilityMaterials (Wrap, electrodes, gel) previously evaluated and cleared under K163067; comply with ISO 10993Section 13
Performance Testing (NMES Waveform)All features and output specifications verified by individual pulse output waveform tracings for various loads (500, 2k, 10k ohms)Section 15
Wireless Co-existence (BLE module)Met all specified requirements (FCC CFR47 Part 15, Subpart C, FCC 47CFR PT 15 SPT B)Section 8
Battery SafetyComplies with IEC 62133:2012Section 8

2. Sample size used for the test set and the data provenance

  • Test Set for Clinical Performance: Not applicable as a new clinical study was not conducted for the Motive Knee Wrap itself. The document states: "Clinical data submitted in the predicate device application, IntelliHab System, K210604 applies to the subject Motive device." No details of that predicate study's sample size or provenance are provided in this document.
  • Test Set for Human Factors/Usability Study: A "self-selection study and human factors study" was conducted to validate usability for OTC and home use. The specific sample size is not provided in this document.
  • Data Provenance: Not explicitly stated for any clinical data. For the human factors study, it doesn't specify country of origin or whether it was retrospective/prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not provided as a specific clinical study for establishing ground truth for the Motive Knee Wrap was not conducted. The clinical effectiveness is deferred to the predicate device's data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This information is not provided as a specific clinical study for the Motive Knee Wrap was not conducted directly.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not conducted. This device is an electrical muscle stimulator, not an AI-assisted diagnostic imaging device for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable in the context of typical algorithm performance for diagnostics. The device's "performance" is its ability to deliver the specified electrical stimulation consistently and safely, which was assessed via bench testing (Section 15) and regulated power output/closed-loop feedback (Section 11).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the non-clinical performance tests (electrical, EMC, battery), the ground truth is defined by the standards themselves (e.g., IEC 60601-1, FCC CFR47 Part 15).
  • For the usability study, the ground truth is derived from the study participants' ability to successfully use the device per instructions and labeling.
  • For the clinical effectiveness, the document states that "Clinical data submitted in the predicate device application, IntelliHab System, K210604 applies to the subject Motive device." The type of ground truth for that predicate study is not described in this document.

8. The sample size for the training set

  • Not applicable. This document describes a medical device, not a machine learning model that requires a "training set" in the conventional sense. The "training" for this device involved its design, development, and bench testing to meet engineering specifications and regulatory standards.

9. How the ground truth for the training set was established

  • Not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

May 12, 2022

CyMedica Orthopedics, Inc. Kereshmeh Shahriari Vice President, Regulatory, Clinical, & Quality 2120 East 6th Street Tempe, Arizona 85281

Re: K220738

Trade/Device Name: Motive Knee Wrap Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF Dated: March 7, 2022 Received: March 14, 2022

Dear Kereshmeh Shahriari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For: CDR Jitendra Virani, MS. Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220738

Device Name Motive™ Knee Wrap Device

Indications for Use (Describe)

The Motive™ Knee Wrap device is intended to strengthen the quadriceps muscle stimulation to provide symptomatic temporary pain relief associated with knee arthritis.

The Motive™ Knee Wrap device is indicated for adults of 22 years of age and older.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Motive™ Knee Wrap CyMedica Orthopedics, Inc.

1 Regulatory Information
1.1 Trade/Proprietary Name:Motive™ Knee Wrap
1.2 Common Name:Powered muscle stimulator
1.3 Regulation Names & Numbers:Powered muscle stimulator, 21 CFR 890.5850
Product Codes:Code- IPF; Powered muscle stimulator
1.4 Classification:Powered muscle stimulator: II
1.5 Manufacturer Name:CyMedica Orthopedics, Inc.2120 East 6th Street, Suite 8Tempe, AZ 85281Telephone (480) 664-1282FAX (866) 296-2772

These devices are reviewed by the Division of Neurological and Physical Medicine Devices.

2 Submission Information

Submission Number:K220738
Date:March 31, 2022
Contact:Kereshmeh Shahriari2120 East 6th Street, Suite 8Tempe, AZ 85281kereshmeh@cymedicaortho.comTelephone (480) 664-1282FAX (866) 296-2772

{4}------------------------------------------------

3 Indications for Use

The Motive Knee Wrap device is intended to strengthen the quadriceps muscle using powered muscle stimulation to provide symptomatic temporary pain relief associated with knee arthritis. The Motive Knee Wrap device is indicated for adults of 22 years of age and older.

4 Device Description

The Motive Knee Wrap device (Motive device) is an over-the-counter (OTC) electrical muscle stimulator (or Neuromuscular Electrical Stimulation (NMES) therapy) for relief of knee pain associated with knee arthritis. The Motive device is used to apply an electrical current through a power regulated output closed-loop feedback NMES system to provide relief of knee pain associated with arthritis. The stimulator continuously contracts quadricep muscle groups for strengthening of the muscles and knee pain relief. The stimulator is placed over the thigh muscle and just above the knee.

4.1 Electrical Muscle Stimulator (NMES therapy)

The device is used to apply a patented unique waveform by means of an electrical current through a power regulated output and closed-loop feedback system to provide relief of knee pain associated with arthritis. The stimulator continuously contracts quadricep muscle groups for strengthening of the muscles and for knee pain relief. Electrical stimulation waveform is generated using a pulse generator or Controller. Electrical stimulation is delivered non-invasively to the treatment site (quadricep muscles of the knee joint) using connected electrodes placed directly on the skin.

4.2 Mobile Application (App)

The electrical stimulation therapy is initiated and managed wirelessly by the User using a mobile application (App) developed for use on smart phones or tablets that interact with the system's pulse generator or Controller.

4.3 Wrap with Smart Panel

The electrical muscle stimulator along with the cutaneous electrodes and wires are incorporated in a Wrap with a Smart Panel™ placed on the thigh and above the knee. Wrap with Smart Panel is worn on the thigh covering quadriceps muscles to apply the NMES therapy. The Wrap is comprised of a modular design including a textile-based wrap and a Smart Panel. The modular Smart Panel is attached to the Wrap by means of a hook and loop interface or other mechanisms. The Smart Panel is an assembly of a Controller (NMES pulse generator), cradle, plastics for hosting the electronics, and four skin contacting snap type electrodes. The Wrap with Smart Panel design is a single size for all size Users and intended for a bilateral use (right and left thigh use independently).

4.4 Wireless Communications

The Motive device incorporates a Bluetooth Low Energy, BLE 4.2 connection module to enable wireless communication that can be paired with a Bluetooth enabled mobile device running the App, available from the App stores. The App implements a virtual control panel on the screen of the smart device where on-screen buttons are provided to the User.

The Controller is connected to the Smart Panel via a multi-pin interface. The Controller contains the primary safety controls for operation of the device and a push button is available for switching the unit on or off. The power button remains active and can be used in the event

{5}------------------------------------------------

of loss of Bluetooth connection to power on/off the device or start/stop stimulation while performing stimulation sessions. The Controller contains light emitting diodes (LED) which indicate status relating to battery charge, stimulation, and Bluetooth™ activity. Power is derived from a 3.7V Li- Po rechargeable battery pack and the unit can be recharged by using the supplied USB cable.

4.5 Software

The Motive device utilizes an embedded Controller firmware, App software, and Web Services software.

The treatment program embedded software is a fundamental component of the system that supports the selection, adjustment, and delivery of the treatment program. The embedded software runs on the Controller and is interfaced with the stimulation pulse generator and App.

The App runs on a mobile platform operating system and communicates with the Controller over a BLE 4.2 connection. The free Motive App is downloaded by the Users from their respective App Stores. The Motive device allows the User to control the NMES therapy program in an easy and intuitive way. It allows the User to select, administer, and adjust the NMES therapy intensities, and manage their health and wellness conditions related to their therapy.

4.6 Controller and the App

The Controller's main function is the NMES therapy stimulation generator and communicating therapy metrics to the App. The Controller contains the primary safety controls for operation of the device. A push button is available for powering the unit on/off or pausing/resuming stimulation. The push button "Stop" control remains active at all times during stimulation and can be used in the event of loss of Bluetooth connection.

Once the App is installed and a wireless connection is established between the Controller and the mobile device (User's phone or tablet), the User can interact with all of the device features. The App downloaded on the User's mobile device provides the graphical user interface and governs the interactions between the controller and the User. The screens and user interface are designed to be intuitive and interactive. Improved User's engagement and compliance through the App are significant design features of the device.

Help and informational screens are provided throughout the App and on specific screens where a User needs to be aware of a feature or condition. Upon downloading the App, the User is guided through chronological screens on the App to perform the following tasks:

  • Reading and accepting the Motive App Terms of Use and Conditions ●
  • Creating a User profile
  • Pairing the App with the Controller ●
  • Starting the treatment
  • Managing the treatment
  • Reviewing User's related data and information on the App

4.7 Web Services (API, Database)

A Medical Device Data System (MDDS), the CyMedica Web Services allows for the electronic transfer, storage, retrieval, and display of data sent or received by the Users' App

{6}------------------------------------------------

and Controller. Web Services includes an encrypted database hosted on secure cloud-based servers and an Application Programming Interface (API) allowing the App to send and receive data to read or write to the database and execute logic tasks. Web Services provides persistent storage and retrieval of data.

5 General Electrical Muscle Stimulation Device Features and Output Characteristics

The following two Tables represent the Motive device electrical stimulation features and the neuromuscular electrical stimulation (NMES) therapy output specifications.

Table 1- Motive device Therapy Program General Electrical Stimulation Features

No. of Output Modes1
No. of Output Channels2
Regulated Current, Voltage, or PowerRegulated Power
Software/Firmware/Microprocessor Control?Yes
Automatic shut off?Yes
Patient device control?Yes
Indicator display- Low Battery?Yes
Indicator display- Voltage/Current Level?Yes
Timer range (minutes)20 minutes

Table 2- Motive device NMES Therapy Program Output Specifications

WaveformMonophasic
ShapeAsymmetrical, Complex
Pulse width5 ms
Frequency50 pps
Maximum current density @ 500Ω0.7 mA/cm² (rms)
Maximum power density (using smallest electrode conductivesurface area) @ 500Ω0.006 W/cm²
Maximum phase charge @ 500Ω359.8 μC
Maximum output voltage (VRMS) (±10%) @ 500Ω9.0 V
Maximum output current (IRMS) (±10%) @ 500Ω18 mA
Power sourceLi-Polymer Battery950mAh 3.7 VDC
Contraction time1.0 s
Relaxation time1.4 s
Treatment session20 minutes

6 NMES Waveform

The NMES treatment program pulse parameters are defined below, Table 3:

{7}------------------------------------------------

Table 3- Motive NMES Waveform Pulse parameters
Pulse shapeAsymmetrical, monophasic, and complex
Treatment duration20 minutes
Frequency50 pps
Pulse width5 ms
Duty cycle25%
Work cycle12 s
Relaxation time10 s
Contraction time5 cycles1 s
Rest time1.4 s

In the treatment program, the work cycle consists of the combination of five cycles of contraction and rest. Contraction time is the actual stimulation contraction period. Rest time is the period between contractions to wait to oscillate the stimulation between the two channels. Relaxation time is a period of no stimulation between the work cycles.

7 Recommended Usage

Motive device NMES therapy is recommended to be used twice a day for a minimum of 5 days per week. Each session of NMES therapy is 20 minutes. The therapy is recommended to be applied as long as the knee pain relief is sustained.

8 Summary of Non-Clinical/ Bench Studies

To demonstrate the safety, the Motive device was tested for electrical safety, electromagnetic compatibility, usability, biocompatibility, and risk management requirements.

To demonstrate the safety, the Motive device was tested per the following standards:

  • IEC 60601-1: 2005, 3rd Edition, Medical electrical equipment- General requirements for ● basic safety and essential performance + CORR. 1:2006 + CORR. 2:2007 + A1:2012
  • IEC 60601-1-2: 2014-02, 4th Edition, Medical electrical equipment-Part 1-2: General . requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility requirements
  • IEC 60601-2-10: 2012, 2nd Edition, Medical electrical equipment- Part 2-10: Particular ● requirements for the basic safety and essential performance of nerve and muscle stimulator
  • IEC 60601-1-11: 2015, 2nd Edition, Medical electrical equipment- Part 1-11: General requirements for basic safety and essential performance- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

{8}------------------------------------------------

  • IEC 60601-1-6: 2010+ A1:2013, 3rd Edition, Medical electrical equipment- Part 1-6: . General requirements for basic safety and essential performance- Collateral standard: Usability including IEC 62366: Application of usability engineering to medical devices
  • IEC 62366: 2007+ A1: 2014, 1st Edition, Medical devices -- Application ofusability . engineering to medical devices
  • IEC 62304: 2006, 15T Edition, Medical device software Software life cycleprocesses ●
  • ISO 10993-1: 2009, Biological evaluation of medical devices- Part 1: Evaluation and ● testing within a risk management process
  • . ISO 10993-5: 2009, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10: 2010, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization

In addition, the performance of BLE module for wireless co-existence was tested in an environment with equipment operating in the ISM band i.e., Bluetooth and Wi-Fi devices, cellphones, cordless phones, etc. The device met all specified requirements.

BLE module testing was conducted in accordance with the following standards:

  • FCC CFR47 Part 15, Subpart C, July 2019, Intentional Radiator, §15.247, Operation within the bands 902-928 MHz, 2400-2483.5 MHz, and 5725-5850MHz
  • FCC 47CFR PT 15 SPT B, August 2018 Title 47 CFR Part 15 Subpart B: Unintentional ● Radiators, FCC Part 15, Subpart B [FCC $15.107 & FCC §15.109]

Battery testing was conducted in accordance with IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications.

9 Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the Motive device and complies with the IEC 60601-1, IEC 60601-2-10, and 60601-11 standards for safety and the IEC 60601-1-2 standard for EMC.

10 Software Verification & Validation Testing

The device's software has been validated in accordance with the requirements set forth in the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) and IEC 62304: 2006, 15" Edition, Medical device software - Software life cycle processes. The software validation tests demonstrated that the software version meets its design requirements.

Cybersecurity Controls 11

Cybersecurity related activities included system assessment and mitigations for potential

{9}------------------------------------------------

cybersecurity hazards and risks on the Motive device performance and data security. Information related to system design features, processes, testing, and controls to manage and mitigate the risks are included in this submission.

Cybersecurity information and supporting documents created and submitted according to the requirements of FDA guidance documents, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2, 2014 and Postmarket Management of Cybersecurity in Medical Devices, December 28, 2016.

12 Human Factors and Usability

A self-selection study and human factors study including usability assessment was conducted to validate the usability of the Motive device as an over-the-counter device and for home use. The results of the study support the labeling content and User's Manual instructions for successfully using the device as intended. The results of the human factors and usability study substantiates the acceptability of the risks identified during the risk assessment activities.

Human Factors information and supporting documents created and submitted according to the requirements of FDA guidance documents, FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices, February 3, 2016.

Additionally, the Motive device complies with the IEC 60601-1-6: 2010 for usability and IEC 62366: Application of usability engineering to medical devices.

13 Biocompatibility/ Material

The patient contacting materials used in the Motive device components include the Wrap with Smart Panel, electrodes, and electrode gel. The biocompatibility of the Wrap, electrodes, and electrode gel were previously evaluated and cleared under 510(K) submission K163067. The biocompatibility testing had been conducted according to the requirements of ISO 10993: 2009 and tested as a surface contact/ >24 hour to 30 days prolonged exposure. From the evaluation and submitted information, the components of the Motive device were found to be biocompatible for its use.

14 Shelf life/ Sterility

The non-invasive nature of the device obviates the need for sterile components; however, patientcontacting surfaces should be capable of being cleaned as needed. The Motive device is provided for single person use and does not require any of the components to be sterilized by the end user. It is intended for external use only. The electrodes are disposable and can be replaced as needed.

15 Performance Testing- Bench Testing of Electrical Muscle Stimulation (NMES) Waveform

All features and output specifications of the device, including those identified in Tables 1 and 2, were verified by individual pulse output waveform tracings for loads of 500, 2k, and 10k ohms, to simulate conditions that the device could counter during use.

{10}------------------------------------------------

16 Comparison of Technological Characteristics with the Predicate Device

CyMedica Orthopedics, Inc. demonstrated that, for the purposes of FDA's regulation of medical devices, the Motive device is substantially equivalent in design principles to the predicate device, CyMedica Orthopedics, Inc. IntelliHab System, K210604, the primary predicate device. The data included in this submission demonstrates design substantial equivalence to the predicate device, IntelliHab System. Both devices utilize the same NMES therapy electrical stimulation waveform and electrodes to deliver the therapy to the quadriceps muscles of the knee. Both devices are indicated for the treatment of knee pain and strengthening of the quadriceps muscle.

The primary predicate IntelliHab system is intended for prescription use and the subject Motive device is intended for over-the-counter (OTC) use. However, both devices are intended for home-use. Data from a self-selection study and human factors study were provided to support the Motive device safety for the OTC use. Additionally, a second predicate device is included, RelieforMe TENS/EMS Device Model UPK-GE01 (K192733) as an OTC device.

The intended use, design, materials, and functional characteristics of the Motive device and the primary predicate device are substantially the same. The subject device and primary predicate device are both intended as home-based use devices. Both devices have the same general purpose and function. Both devices are intended for the same general health population. Both devices utilize the same electrical stimulation therapy to strengthen the quadricep muscle to provide symptomatic temporary knee pain relief. Similar to the primary predicate device, Motive device provides the same NMES waveform, protocol, and pulse amplitude for the purpose of muscle strengthening and knee pain relief. Similar to the primary predicate device, Motive device includes a Wrap. Controller (pulse generator), RF communications, electrodes, and an App for delivery of the NMES therapy.

The Motive device is substantially equivalent in design and labeling to the primary predicate device and secondary predicate device. Nonclinical testing including self-selection study and human factors study performed on the Motive device are sufficient to demonstrate that the subject device is as safe and effective as the legally marketed predicate devices. The technological and labeling differences do not raise new or different questions about safety or effectiveness. Motive device is substantially equivalent to the predicate devices.

The following table demonstrates the similarities and differences between the subject Motive device and the predicate devices, IntelliHab System and RelieforMe TENS/EMS Device.

{11}------------------------------------------------

TABLE 4-SubstantialEquivalenceComparisonProposed Subject Device, Motive™ KneeWrap (Motive device), K220738CyMedica Orthopedics, Inc.Primary Predicate Device, IntelliHab™System, K210604CyMedica Orthopedics, Inc.2nd Predicate Device, RelieforMe TENS/EMSDevice Model UPK-GE01, K192733UMEHEAL Ltd.
Indications foruseThe Motive Knee Wrap device is intended tostrengthen the quadriceps muscle using poweredmuscle stimulation to provide symptomatictemporary pain relief associated with kneearthritis.The Motive Knee Wrap device is indicated foradults of 22 years of age and older.The IntelliHab system is intended to strengthenthe quadricep muscle using powered musclestimulation to provide symptomatic temporarypain relief associated with knee osteoarthritis andimprovement of the knee joint mobility when therecommended treatment regimens are followed.In addition, the Intellihab System is indicated forthe following:• Retardation or prevention of disuse atrophy• Evaluation of joint mobility by measuring andrecording range of motion.The IntelliHab System is indicated for adults of22 years of age and older.TENS:1. It is intended for temporary relief ofpain associated with sore and achingmuscles in the shoulder, neck, back,waist, abdomen, lower extremities (legs),upper extremities (arms) due to strainfrom exercise or normal household andwork activities.2. It is also intended for symptomatic reliefand management of chronic, intractable painand relief of pain associated with arthritis.EMS:3. To stimulate healthy muscles toimprove and facilitate muscle performance.To be used for the improvement of muscletone and firmness, and for strengtheningmuscles in the intended areas onthe body.4. To be used for relaxation of musclespasm, increase of blood flow circulation,prevention or retardation of disuseatrophy, muscle re-education, maintainingor increasing range of motion, andimmediate post-surgical stimulation ofcalf muscles to prevent venous thrombosis
Intended UseOver-the-Counter (OTC)Home-usePrescriptionHome-useOTCHome-use
Regulationnames andnumbersPowered muscle stimulator, 21 CFR 890.5850Powered muscle stimulator, 21 CFR 890.5850Goniometer, AC-powered, 21 CFR 888.1500Powered muscle stimulator, 21 CFR 890.5850TENS, 21 CFR 882.5890
Product codesandclassificationsPowered muscle stimulator, IPF, Class IIPowered muscle stimulator, IPF, Class IIGoniometer, AC-powered, KQX, Class IPowered muscle stimulator, IPF, Class IITENS, GZJ and NUH, Class II
TABLE 4-SubstantialEquivalenceComparisonProposed Subject Device, Motive™ KneeWrap (Motive device), K220738CyMedica Orthopedics, Inc.Primary Predicate Device, IntelliHab™System, K210604CyMedica Orthopedics, Inc.2nd Predicate Device, RelieforMe TENS/EMSDevice Model UPK-GE01, K192733UMEHEAL Ltd.
ProductidentificationsPowered muscle stimulator, A powered musclestimulator is an electrically powered deviceintended for medical purposes that repeatedlycontracts muscles by passing electrical currentsthrough electrodes contacting the affected bodyarea.Powered muscle stimulator, A powered musclestimulator is an electrically powered deviceintended for medical purposes that repeatedlycontracts muscles by passing electrical currentsthrough electrodes contacting the affected bodyarea.Goniometer, AC-powered, A goniometer is anAC-powered or battery powered device intendedto evaluate joint function by measuring andrecording ranges of motion, acceleration, orforces exerted by a joint.Powered muscle stimulator, A poweredmuscle stimulator is an electrically powereddevice intended for medical purposes thatrepeatedly contracts muscles by passingelectrical currents through electrodes contactingthe affected body area.TENS, A transcutaneous electrical nervestimulator for pain relief is a device used toapply an electrical current to electrodes on apatient's skin to treat pain.
ElectricalstimulationmodalitiesNMES "Strength" program therapyNMES "Strength" program therapyNMES and TENS therapies
SystemcomponentsController, Controller changing cable, App,Wrap with Smart Panel, electrodes, andelectrode gelController, Controller changing cable, App,conductive wrap, electrodes, and electrode gelController and electrodes
Controller andPCBAController (pulse generator) consists of aprinted-circuit board within a plastic enclosurethat docks in a wrap docking receptacle via a 16-pin interface. The Controller is the stimulationgenerator and the tactile control device for thesystem. The User can establish a wirelessconnection between the Controller and theirmobile device.SameUnknown
WrapThe Wrap design includes a modular design of abasic textile-based wrap and a removable SmartPanel where all the electronics, wires, andelectrodes are included in the Smart Panel.Wrap with Smart Panel is single size.A single Wrap with Smart Panel canprovide treatment for both right and leftknees (not simultaneously).The Wrap is one piece only and includesintegrated electronics, wires, and accelerometersWrap is not washable.Wrap comes in different sizes.Two designated right and left wraps arerequired to provide the treatment for theright knee or left knee (notsimultaneously).No wrap
TABLE 4-SubstantialEquivalenceComparisonProposed Subject Device, Motive™ KneeWrap (Motive device), K220738CyMedica Orthopedics, Inc.Primary Predicate Device, IntelliHabTMSystem, K210604CyMedica Orthopedics, Inc.2nd Predicate Device, RelieforMe TENS/EMSDevice Model UPK-GE01, K192733UMEHEAL Ltd.
ElectrodesThree snap type connected electrodes integratedwith the Smart Panel of the Wrap. One 2"diameter electrode is required to activate theVastous Medialis Oblique (VMO) muscle andtwo 2" x 3.5" electrodes for Rectus Femoris (RF)muscle.Although only three electrodes are required toprovide the therapy, in the Motive device setupsteps, Users are instructed to install 4 electrodes(2 VMO electrodes) so they can apply the rightand left knee therapies when desired.Three wire connected or pigtail electrodesintegrated with the conductive wrap. One 2x2"electrode for VMO muscle and two 2" x 3.5"electrodes for RF muscle.Unknown
User InterfaceThe User's App serves as the primary userinterface and communication link between theController and UI. The App includes screens forapplication of NMES therapy, changingintensities, starting, pausing, and stopping atreatment. The App includes screens for creatinga profile, initial device setup guide, Help, andreminders related to critical tasks such aschanging electrodes.Additionally, the App improves therapycompliance through a multi-faceted approach ofincentivizing User engagement through amotivational User Dashboard, therapy remindersand a rewards system.Same UI designUnknown
StimulationpulsecharacteristicsThe patented waveform is a unique,asymmetrical, complex, and monophasic shapedpulse which is designed to provide optimizedtherapeutic benefits while maximizing comfortand compliance.The unique design features including a longerpulse width (5 ms), monophasic polarity, workcycles, and regulated power output provide alonger duration of muscle contraction within a20 minute treatment session. These uniquewaveform characteristics lead to the therapeuticbenefits of the device reducing the knee pain andSameUnknown
TABLE 4-SubstantialEquivalenceComparisonProposed Subject Device, Motive™ KneeWrap (Motive device), K220738CyMedica Orthopedics, Inc.Primary Predicate Device, IntelliHab™System, K210604CyMedica Orthopedics, Inc.2nd Predicate Device, RelieforMe TENS/EMSDevice Model UPK-GE01, K192733UMEHEAL Ltd.
improving knee joint functionality.The Moive NMES stimulation pulses have threedistinct phases:Phase 1: Pulse Spike (Rise Time, Peak,and Decay)Phase 2: Pulse MesaPhase 3: Recharge (off period)
Regulatedpower outputUsing a proprietary technology, the output ofstimulation circuit delivers energy to the User ata constant power, independent of the loadimpedance, hence power regulation during themesa portion of the pulse.SameUnknown
Closed loopfeedbacksystemThe stimulation circuit employs a proprietaryclosed loop feedback technology to regulateenergy transferred to the User for stimulationsensation comfort.SameNone
WaveformAsymmetrical, monophasic, complexSameUnknown
Pulse width(µsec)5000 µsecSame250-400 µs, NMES4-300 µs, TENS
Pulse Train50 HzSame1-100 Hz, NMES1-1200, TENS
MaximumOutput Voltage(volts, rms) (+/-%) @ 2k Ω16.7 VSame65 V
MaximumOutput Current(mA, rms) (+/-%) @ 2k Ω8.3 mASame32.5 mA
Net Charge(microcoulombs(µC) per pulse)359.8 µC @500ΩSameUnknown
TABLE 4-SubstantialEquivalenceComparisonProposed Subject Device, Motive™ KneeWrap (Motive device), K220738CyMedica Orthopedics, Inc.Primary Predicate Device, IntelliHab™System, K210604CyMedica Orthopedics, Inc.2nd Predicate Device, RelieforMe TENS/EMSDevice Model UPK-GE01, K192733UMEHEAL Ltd.
MaximumPhase Charge,(μC)359.8 μC @500ΩSame48.8 μC @500Ω, NMES36.6 μC @500Ω, TENS
MaximumCurrentDensity(mA/cm²,r.m.s.)0.9 mA @500ΩSameUnknown
MaximumAverage PowerDensity,(W/cm²), (usingsmallestelectrodeconductivesurface area)0.009 W/cm² @500ΩSameUnknown
IntegrateddigitalgoniometerNo goniometerIn addition to an electrical muscle stimulator,IntelliHab System includes an integrated digitalgoniometer for measurement of knee joint rangeof motion (ROM) to evaluate joint mobility. Thedevice includes two accelerometers located in aconductive wrap and IntelliHab Controllerallowing for the measurement of knee jointextension and flexion degrees or range of motionmeasurement. The measured flexion, extension,and ROM angles are displayed and stored in theIntelliHab patient's App.No goniometer
EmbeddedsoftwareThe treatment program embedded software is afundamental component of the system thatsupports the selection, adjustment, and delivery ofthe treatment program. The embedded softwareruns on the Controller and is interfaced with thestimulation pulse generator and App.The Controller software drives the stimulationcircuits to generate appropriate waveforms for theprescribed protocols and deliver them to theSameUnknown
TABLE 4-SubstantialEquivalenceComparisonProposed Subject Device, Motive™ KneeWrap (Motive device), K220738CyMedica Orthopedics, Inc.Primary Predicate Device, IntelliHab™System, K210604CyMedica Orthopedics, Inc.2nd Predicate Device, RelieforMe TENS/EMSDevice Model UPK-GE01, K192733UMEHEAL Ltd.
electrode interface. The power button, withintegrated LED indicator, on the Controller isused for powering the Controller On/Off andPause/Resume an active treatment; the button isalways active and can be used in the event of lossof Bluetooth connection to the mobile device topause or abort therapy. The Controllercommunicates collected data and commands inreal-time to the App wirelessly over Bluetoothcommunications.
WirelesscommunicationsMotive device incorporates a Bluetooth LowEnergy, BLE 4.2 connection module to enablewireless communication that can be paired witha Bluetooth enabled mobile device running theApp, available from the App stores. The Appimplements a virtual control panel on the screenof the smart device where on-screen buttons areprovided to the user.The Controller is connected to the Smart Panelvia a multi-pin interface. The Controller containsthe primary safety controls for operation of thedevice and a push button is available forswitching the unit on or off. The power buttonremains active and can be used in the event ofloss of Bluetooth connection to power on/off thedevice or start/stop stimulation while performingstimulation sessions. The Controller containslight emitting diodes (LED) which indicatestatus relating to battery charge, stimulation, andBluetooth™ activity. Power is derived from a3.7V Li-Po rechargeable battery pack.SameNone
ParameterMotive Knee WrapIntelliHab SystemRelieforMe
510(k) NumberK220738K210604K192733
Device Name and ModeMotive NMESIntellihab NMESRelieforMe TENS/EMS
ManufacturerCyMedica OrthopedicsSameUMEHEAL Ltd.
Power Source(s)†SingleVD434053:3.7V;1000mAh;Lithium ionpolymer batterySameBattery powered
- Method of Line Current IsolationNo lineconnectionSameUnknown
- Patient Leakage Current---------
- Normal Condition(µA)4.88SameUnknown
- Single Fault Condition (μΑ)8.00SameUnknown
Number of Output ModesOne (NMES)Same2
Number of Output Channels ††††:1, AlternatingSame1
1, TransistorSame1
Regulated Current or Regulated Voltage?Regulated PowerSameVoltage
Software/Firmware/Microprocessor Control?YesYesYes
Automatic Overload Trip?NoNoNo
Automatic No-Load Trip?NoNoNo
Automatic Shut Off?YesYesYes
User Override Control?Yes, Stop ButtonsYesYes, Off button stopstreatmentimmediately
Indicator Display:On/Off Status?YesLow Battery?YesVoltage/CurrentLevel?NoYesYesNoYesYesNo
Timer Range (minutes)20-openSameUnknown
Compliance with Voluntary Standards?Yes.IEC 60601-2-10:2012Part 2-10;IEC 60601-1-11:2015Part 1-11;IEC 60601-1-6:2010;ISO 10993-1:2009Part 1Yes, SamestandardsUnknown
Weight (lbs., oz.)1.76 oz.(50g)SameUnknown
Dimensions (in.) [W x H x D]1.93" x 0.64" x3.28"SameUnknown
Housing Materials and ConstructionMolded PC\ABSPlastic BayblendFR3010SameUnknown

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

The following table summarizes the subject Motive device and the predicated devices, IntelliHab System and RelieforMe technological characteristics:

Table 5- Technological characteristics of the subject device and predicate devices

{18}------------------------------------------------

{19}------------------------------------------------

The following table summarizes the subject Motive device, IntelliHab System therapy waveform characteristics:

Table 6- Motive NMES Waveform & IntelliHab NMES Waveform Parameters
NMES Waveform ParametersSubject device, Motive NMES programPredicate device, IntelliHab NMES program, K210604Discussion of similarities and differences
Waveform (e.g., pulsed monophasic, biphasic)Pulsed MonophasicPulsed MonophasicSame
Shape (e.g., rectangular, spike, rectified sinusoidal)ComplexComplexSame
Maximum Output Voltage (volts, rms) (+/- ______ %)9.3 @500Ω9.0 @500ΩSimilar
16.7 @ 2 k Ω16.2 @ 2 k ΩSimilar
21.8 @10 k Ω21.1 @10 k ΩSimilar
Maximum Output Current (mA, rms) (+/- ______%)18.6 @500Ω18.0 @500ΩSimilar
8.3 @ 2 k Ω8.1 @ 2 k ΩSimilar
2.2 @10 k Ω2.1 @10 k ΩSimilar
Duration of primary (depolarizing) phase (μsec)50005000Same
Pulse Duration (μsec)50005000Same
Frequency (Hz) [or Rate (pps)]5050Same
For interferential modes only: Beat Frequency (Hz)N/AN/AN/A
For multiphasic waveforms only:Symmetrical phases?N/AN/AN/A
Phase durationN/AN/AN/A
Net Charge (microcoulombs (µC) per pulse) (If zero, state method of achieving zero net charge.)359.8 @500Ω359.8 @500ΩSame
Maximum Phase Charge, (µC)359.8 @500Ω359.8 @500ΩSame
Maximum Current Density (mA/cm², r.m.s.)0.9 @500Ω0.7 @500ΩSimilar
Maximum Average Current (average absolute value),mA18.6 @500Ω18.0 @500ΩSimilar
Maximum Average Power Density, (W/cm²), (usingsmallest electrode conductive surface area)0.009 @500Ω0.006 @500ΩSimilar
Burst Mode(a) Pulses per burst5050Same
(b) Bursts per second0.230.23Same
(c) Burst duration (seconds)11Same
(d) Duty Cycle: Line (b) x Line (c)0.230.23Same
ON Time (seconds)276276Same
OFF Time (seconds)924924Same
Additional Features (specify, if applicable)N/AN/AN/A

{20}------------------------------------------------

{21}------------------------------------------------

Summary of Clinical Study 17

Subject Motive device utilizes the same NMES waveform and protocol as the predicate device CyMedica IntelliHab System, K210604. Clinical data submitted in the predicate device application, IntelliHab System, K210604 applies to the subject Motive device.

18 Conclusion

The basis for substantial equivalence for the Motive device and the predicate devices is nonclinical data, a self-selection study, human factors test data, and conformity with recognized standards. The hardware and software verification and validation demonstrate that the subject device performs comparably to the predicate devices that are marketed for the same intended use. Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the Motive device is as safe and effective as, and substantially equivalent to, the predicate devices.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).