K Number
K220657
Device Name
Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001)
Date Cleared
2023-02-10

(340 days)

Product Code
Regulation Number
880.2910
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Medical Infrared Thermometer is a non-sterile. reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.
Device Description
The Medical Infrared Thermometer is a non-contact, hand-held, battery powered device designed to measure human body temperature, which has the feature of simple and easy use, rapid measurement and accurate temperature feedback. The Medical Infrared Thermometer is an infrared thermometer to measure human body temperature using the infrared energy emitted in the area around the subject's forehead. The reference body site of the temperature measurement is axillary. The temperature measurement distance is 1-5 cm. The product is composed of enclosure, control panel, LCD screen, infrared sensor, PCB board, battery pack and USB charging inlet. The thermometer has three functions of digital LCD display, buzzer indication for measurement results and three color back light indication for temperature (32.0℃ to 37.3℃ with green back light; 37.4℃ to 38.0℃ with yellow back light; 38.1℃ to 43.0℃ with red back light). The device is powered by Li-ion Battery (3.7VDC; 30mAh), which can be recharged through a computer with USB port or by switching power supply (100-240V, 50/60Hz) connected with a USB charging cable. The switching power supply is not provided with the subject device and the USB charging cable will be provided with the subject device as an accessory. There are two models RST-TC59-1001 and RST-TC60-1001 included in this document. Both models share the same construction except differences on product structure including appearance of enclosure, PCB design, sensor, MCU and working condition (ambient temperature). There are three buttons on control panel of both models, the first button is for power on/off and temperature measurement function, the second button is for temperature unit shift (°C and °F) function, the third button is for memory function.
More Information

Not Found

No
The description focuses on standard infrared temperature measurement technology and does not mention any AI or ML components or capabilities.

No
The device is described as a non-sterile, re-usable clinical thermometer intended for the determination of human body temperature. Its function is to measure temperature, not to treat or alleviate a condition.

No

A diagnostic device determines the nature or cause of a disease or condition. This device is a thermometer, which measures body temperature, a physiological parameter, but does not diagnose a disease.

No

The device description explicitly lists multiple hardware components including enclosure, control panel, LCD screen, infrared sensor, PCB board, battery pack, and USB charging inlet.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens (like blood, urine, or tissue) taken from the human body to provide information for diagnosis, monitoring, or screening.
  • Device Function: The Medical Infrared Thermometer measures human body temperature directly from the forehead using infrared energy. It does not analyze any biological specimens.
  • Intended Use: The intended use is for the "intermittent determination of human body temperature," which is a direct measurement on the body, not an analysis of a sample.

Therefore, this device falls under the category of a clinical thermometer, which is a medical device but not an IVD.

N/A

Intended Use / Indications for Use

The Medical Infrared Thermometer is a non-sterile. reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.

Product codes

FLL

Device Description

The Medical Infrared Thermometer is a non-contact, hand-held, battery powered device designed to measure human body temperature, which has the feature of simple and easy use, rapid measurement and accurate temperature feedback. The Medical Infrared Thermometer is an infrared thermometer to measure human body temperature using the infrared energy emitted in the area around the subject's forehead. The reference body site of the temperature measurement is axillary. The temperature measurement distance is 1-5 cm.

The product is composed of enclosure, control panel, LCD screen, infrared sensor, PCB board, battery pack and USB charging inlet.

The thermometer has three functions of digital LCD display, buzzer indication for measurement results and three color back light indication for temperature (32.0℃ to 37.3℃ with green back light; 37.4℃ to 38.0℃ with yellow back light; 38.1℃ to 43.0℃ with red back light).

The device is powered by Li-ion Battery (3.7VDC; 30mAh), which can be recharged through a computer with USB port or by switching power supply (100-240V, 50/60Hz) connected with a USB charging cable. The switching power supply is not provided with the subject device and the USB charging cable will be provided with the subject device as an accessory.

There are two models RST-TC59-1001 and RST-TC60-1001 included in this document. Both models share the same construction except differences on product structure including appearance of enclosure, PCB design, sensor, MCU and working condition (ambient temperature). There are three buttons on control panel of both models, the first button is for power on/off and temperature measurement function, the second button is for temperature unit shift (°C and °F) function, the third button is for memory function.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

center of the forehead

Indicated Patient Age Range

one month and above.

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Clinical accuracy validation for the Medical Infrared Thermometer is designed and performed according to standard ISO 80601-2-56: 2017+A1: 2018 and ISO 14155: 2011 under clinical setting.

Both models RST-TC59-1001 and RST-TC60-1001 are selected to implement clinical accuracy validation respectively. A blackbody radiator and a Platinum resistance thermometer are selected to validate the laboratory accuracy of the infrared thermometer under testing before and after the clinical trial. And total of 165 subjects are selected to temperature measurement with the infrared thermometer under testing (DUT) and the reference clinical thermometer (RCT) respectively. RCT used in this clinical trial is mercury axillary thermometer (Geratherm Medical AG). The whole trial procedure and the clinical accuracy validation is in accordance with ISO 80601-2-56: 2017+A1: 2018.

The clinical tests were evaluated on 165 subjects and the thermometer was evaluated in three age groups including subgroup A1 and A2: A1 – one month up to three months; A2 - three months to one year; B older than one years and younger than five years; and C - older than five years old.

Subjects selected for each model in this clinical trial are: A1) group 29 b) group: 55 c) group: 55. Total number of subjects is 165.

The clinical performance test protocol and data analysis were conducted in accordance with the requirement of ISO 80601-2-56: 2017+A1: 2018.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Study Type: Non-clinical testing (Performance testing-bench), Clinical Accuracy Validation Test

Sample Size:

  • Non-clinical: Not explicitly stated beyond general verification that the device met design specifications.
  • Clinical Accuracy Validation Test: Total of 165 subjects (A1: 29 subjects, B: 55 subjects, C: 55 subjects).

Key Results:

  • Non-clinical Testing:
    • Risk Analysis developed in accordance with ISO 14971:2019.
    • Software Evaluation according to Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Software validation and verification tests concluded no defects found.
    • Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98, IEC 62133-2, ISO 10993-1.
    • Biocompatibility tests (In vitro cytotoxicity, Skin sensitization, Skin irritation) passed and met the specified criteria.
  • Clinical Accuracy Validation Test:
    • Performed according to ISO 80601-2-56: 2017+A1: 2018 and ISO 14155: 2011.
    • Laboratory accuracy validated using a blackbody radiator and a Platinum resistance thermometer.
    • The clinical test report showed the clinical performance of the subject devices complied with the requirement of ISO 80601-2-56: 2017+A1: 2018.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K163516

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

February 10, 2023

Zhejiang Rongsheng Industrial Co., Ltd. Wu Bentuan Chief Engineer 2 Jinheng Road Xiaoshun Town, Jinhua, Zhejiang 321035 China

Re: K220657

Trade/Device Name: Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 5, 2023 Received: January 12, 2023

Dear Wu Bentuan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K220657

Device Name

Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001)

Indications for Use (Describe)

The Medical Infrared Thermometer is a non-sterile. reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

K220657

510(k) Summary

SUBMITTER

Name: Zhejiang Rongsheng Industrial Co., Ltd.

Address: NO. 2 Jinheng Road, Xiaoshun Town, Jindong District Jinhua City Zhejiang, CN 321035

Name of contact person: Wu Bentuan

Telephone: +86-579-82953477

Preparation date: 2/1/2023

Device

Device trade name: Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) Classification name: Clinical Electronic Thermometer Regulation class: II Regulation number: 21CFR 880.2910 Panel: General Hospital Product code: FLL

Predicative device

Predicate Submission Number: K163516

Predicate Device Trade Name: Braun No Touch + Forehead NTF3000 Thermometer

Classification name: Clinical Electronic Thermometer

Regulation class: II

Regulation number: 21CFR 880.2910

Panel: General Hospital

Product code: FLL

Device description

The Medical Infrared Thermometer is a non-contact, hand-held, battery powered device designed to measure human body temperature, which has the feature of simple and easy use, rapid measurement and accurate temperature feedback. The Medical Infrared Thermometer is an infrared thermometer to measure human body temperature using the infrared energy emitted in the area around the subject's forehead. The reference body site of the temperature measurement is axillary. The temperature

4

measurement distance is 1-5 cm.

The product is composed of enclosure, control panel, LCD screen, infrared sensor, PCB board, battery pack and USB charging inlet.

The thermometer has three functions of digital LCD display, buzzer indication for measurement results and three color back light indication for temperature (32.0℃ to 37.3℃ with green back light; 37.4℃ to 38.0℃ with yellow back light; 38.1℃ to 43.0℃ with red back light).

The device is powered by Li-ion Battery (3.7VDC; 30mAh), which can be recharged through a computer with USB port or by switching power supply (100-240V, 50/60Hz) connected with a USB charging cable. The switching power supply is not provided with the subject device and the USB charging cable will be provided with the subject device as an accessory.

There are two models RST-TC59-1001 and RST-TC60-1001 included in this document. Both models share the same construction except differences on product structure including appearance of enclosure, PCB design, sensor, MCU and working condition (ambient temperature). There are three buttons on control panel of both models, the first button is for power on/off and temperature measurement function, the second button is for temperature unit shift (°C and °F) function, the third button is for memory function.

Indication for use

The Medical Infrared Thermometer is a non-sterile, reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.

AttributeSubject devicePredicative deviceVerdict
Device trade nameMedical Infrared
Thermometer (Model: RST-TC59-1001, RST-TC60-1001)Braun No Touch + Forehead
NTF3000 Thermometer/
Product modelRST-TC59-1001, RST-TC60-1001NTF3000/
510(k) numberK220657K163516/
Device classification nameClass IIClass IISame
Classification880.2910880.2910Same
AttributeSubject devicePredicative deviceVerdict
regulations
Product codeFLLFLLSame
thermometer typeskin IR thermometersskin IR thermometersSame
Intended
use(s)/
Indication for useThe Medical Infrared
Thermometer is a non-sterile,
reusable clinical thermometer
intended for the intermittent
determination of human body
temperature in no touch mode
on the center of the forehead
as the measurement site on
people of one month and
above.The Braun No Touch +
Forehead NTF3000
Thermometer is a non-sterile,
reusable clinical thermometer
intended for the intermittent
determination of human body
temperature in a touch and no
touch mode on the center of the
forehead as the measurement
site on people of all ages.Similar
Operational principleThe optical system gathers
the infrared radiation energy
of the target in its field of
view, size of which depends
on the optical parts and their
positions. The gathered
infrared energy is focused on
the photoelectricity detector
and converted to be the
electrical signal accordingly.
After passing through the
amplifier and signal
processing circuit, the
electrical signal will be
converted to be the
temperature of the measured
target according to the
algorithm internally
contained in the device and
target emissivity correction.The thermometer uses a
thermopile sensor with
integrated thermistor for the
target reading, a thermistor
mounted in the head of the
thermometer for ambient
temperature readings, a
parabolic mirror to help focus
the infrared energy emitted
from the forehead, and an
infrared proximity sensor for
detection of contact or non-
contact use and compensation
of the temperature reading.Same
Sensor typethermopilethermopileSame
signal processing and
displayMCU and LCDMCU and LCDSame
Low
power
indicationyesyesSame
Measurement siteforeheadforeheadSame
Operation modeAdjusted modeAdjusted modeSame
°C/°F switchableyesyesSame
Temperature
measurement range32 °C43 °C (89.6 °F109.434.4°C (93.9°F) - 42.2°C
(108°F)Similar
AttributeSubject devicePredicative deviceVerdict
°F)
Measuring accuracy35.0 °C~42.0 °C (95.0 °F
~107.6°F): ±0.2 °C(±0.4 °F);35-42 °C (95-107.6 °F):± 0.2 °C (± 0.4 °F);
Outside the range 35-42 °C
(95-107.6 °F):Same
32.0 °C ~ 34.9 °C (89.6 °F
~94.8 °F): ±0.3 °C (±0.5°F);± 0.3 °C (± 0.5 °F)
42.1 °C~43.0 °C (107.8 °F
~109.4 °F): ±0.3 °C (±0.5 °F)
Measurement
precision0.1°C (0.1°F)0.1°C (0.1°F)Same
Electric Safety and
EMCIEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 62133-2IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11Different
BiocompatibilityMeets ISO 10993 and FDA
guidance "Use of
International Standard ISO
10993-1"
ISO 10993-5
ISO 10993-10Meets ISO 10993 and FDA
Bluebook memo G95-1
ISO 10993-5
ISO 10993-10Same
PerformanceASTM E1965-98
ISO 80601-2-56ASTM E1965-98
ISO 80601-2-56Same
Power requirementsrechargeable li-ion battery
pack; 3.7 V; 300 mAh;
recharged through USB port
by external power source, e.g.
compatible charger or
computer with USB port etc.2 AA batteriesDifferent
Patient
contact
materialsRST-TC59-1001: ABS with
colorants (white and green)
for device housing including
handle, PMMA for buttonsPatient contacting materials
include ABS (device housing /
handle and power button) and
TPR (temperature button and
nose / forehead touch bumper).Similar
RST-TC60-1001: ABS with
colorants (white and blue) for
device housing including
handle, PMMA for
Transparent LCD cover,
silicone rubber for buttons
Memory20 setsNot availableDifferent
AttributeSubject devicePredicative deviceVerdict
Ambient temperature
environment
Operating conditionRST-TC59-1001: 15°C ~ 40°C (59 °F~104 °F); Relative humidity: ≤85%; non-condensing15 °C – 40 °C (59 °F –104 °F), 15–95% non-condensingSimilar
RST-TC60-1001: 0°C ~ +40°C (32°F~104°F); Relative humidity: ≤85%; non-condensing
Ambient temperature
environment
Storage conditionRST-TC59-1001: temperature: -25°C ~ +50°C (-13 °F~+122 °F); Relative humidity: ≤90% non-condensing; Atmospheric pressure: 70kPa~106 kPatemperature: -25°C to 60°C (-13°F to 140°F)
Humidity: 15–95% non-condensing
Pressure: 700-1060hPASimilar
RST-TC60-1001: -25°C ~ +70°C (-13°F~+158°F); Relative humidity: ≤90% non-condensing; Atmospheric pressure: 70kPa~106 kPa
Measurement
distance1cm ~ 5cm0-5cmSimilar
Response time1 second2 secondsDifferent

Comparison of technological characteristics with the predicate device

5

6

7

Discussion for similarities and differences:

    1. Intended use(s)/Indication for use: indication for use of the subject device has the similar meaning with different words as the predicate device. In addition, the subject device has more restricted people age range than the predicate device. Such difference will not affect the safety and effectiveness of the subject devices.
    1. Temperature measurement range: the subject devices measurement range from 32 ℃43 ℃ (89.6 °F109.4 °F), which is different from the predicate device's measurement range, 34.4℃ ( 93.9F ) – 42.2℃ (108°F ). While the subject devices have been tested and validated according to standard ISO 80601-2-56 including clinical accuracy and the measuring range of the subject devices meet the minimum rated output range of clinical thermometer requirement, from 35°C to 42°C, therefore, such difference will not affect the safety and effectiveness of the subject devices.

8

  • Electric safety and EMC: the batteries used in the subject device are rechargeable batteries, different 3. from the predicate device, and tested according to IEC 62133-2. Such difference between the predicate device and the subject device will not affect the safety and effectiveness of the subject device.
  • Power requirements: the subject devices are powered with rechargeable li-ion battery pack and 4. different from the predicate device, which is powered with 2 AA batteries. The rechargeable battery pack has been tested according to IEC 62133-2: 2017, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
  • Patient contact materials: there are some material differences between the subject devices and the 5. predicate device, but biocompatibility of both the subject devices and the predicate device has been evaluated and tested according to ISO 10993 series and FDA guidance. The patient contact materials have passed the biocompatibility testing in accordance with ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010. Such difference between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
  • Memory function: The subject devices have the memory function to store 20 sets history 6. measurement records while the predicate device has not such memory function. The validation test of the software used in the subject device was conducted in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The difference will not raise safety or effectiveness issues on the subject device.
  • Ambient temperature environment operating condition: there are differences between ambient 7. temperature of the subject devices and the predicate device, but performance and safety tests of the subjective device have been conducted and passed under the specific operating condition according to standards IEC 60601-1, ISO 80601-2-56 and ASTM E1965-98, and the user manuals of the subjective devices provide the operating and storage condition as well. Therefore, such difference will not affect the safety and effectiveness of the subject devices.
  • Ambient temperature environment Storage condition: Performance and safety tests of the 8. subjective device have been passed under the specific storage condition, and the user manual of the subjective device provides the storage condition as well, therefore, the difference of the storage conditions between subject device and predicate device will not affect the safety and effectiveness of subject device.

9

  • Measurement distance: there is slight difference between the subject devices and the predicate device 9. on measurement distance and the subject devices have more restricted measurement range (1cm-5cm) than the predicate device (0cm-1cm). The performance tests of the subject device are conducted in accordance with standard ISO 80601-2-56 and ASTM E1965-98, and the test results showed that the measurement accuracy of the subject device is in conformity with the standard requirement. In addition, clinical accuracy testing of the subject devices was validated according to standard ISO 80601-2-56, therefore, such difference will not affect the safety and effectiveness of the subject devices.
    1. Response time: there is slight difference between the subject devices and the predicate device on response time and the subject devices are faster in response than the predicate device. The performance tests of the subject device are conducted in accordance with standard ISO 80601-2-56 and such difference will not affect the safety and effectiveness of the subject devices.

Summary of non-clinical testing (Performance testing-bench)

The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.

  • . Risk Analysis developed in accordance with ISO 14971:2019.
  • . Software Evaluation according to Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software validation and verification tests are conducted at unit level, integration level and system level. The software and hardware system is verified that no defects are found.
  • IEC 60601-1: 2005+A1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
  • . IEC 60601-1-2: 2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -Requirements and tests
  • IEC 60601-1-11: 2015 Medical electrical equipment. Part 1-11: General requirements for basic ● safety and essential performance. Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • ISO 80601-2-56: 2017+A1: 2018 medical electrical equipment part 2-56: particular ●

10

requirements for basic safety and essential performance of clinical thermometers for body temperature measurement

  • . ASTM E1965-98(2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
  • . IEC 62133-2: 2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
  • . ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, summary level biocompatibility information sees below table:

Name of testEndpointResult
In vitro
cytotoxicity testQuantitative evaluation according to ISO 10993-5: there shall be no reduction of cell viability by more than 30 %, which is considered a cytotoxic effect.Pass
Skin sensitization
testAccording to ISO 10993-10: Magnusson and Kligman grades shall not be of 1 or greater in the test group, provided grades of less than 1 are seen in control animals.Pass
Skin irritation testAccording to ISO 10993-10: the primary irritation index (PII) determined on the test group shall be 0-0.4Pass

Summary of clinical Accuracy Validation Test

Clinical accuracy validation for the Medical Infrared Thermometer is designed and performed according to standard ISO 80601-2-56: 2017+A1: 2018 and ISO 14155: 2011 under clinical setting.

Both models RST-TC59-1001 and RST-TC60-1001 are selected to implement clinical accuracy validation respectively. A blackbody radiator and a Platinum resistance thermometer are selected to validate the laboratory accuracy of the infrared thermometer under testing before and after the clinical trial. And total of 165 subjects are selected to temperature measurement with the infrared thermometer under testing (DUT) and the reference clinical thermometer (RCT) respectively. RCT used

11

in this clinical trial is mercury axillary thermometer (Geratherm Medical AG). The whole trial procedure and the clinical accuracy validation is in accordance with ISO 80601-2-56: 2017+A1: 2018.

The clinical tests were evaluated on 165 subjects and the thermometer was evaluated in three age groups including subgroup A1 and A2: A1 – one month up to three months; A2 - three months to one year; B older than one years and younger than five years; and C - older than five years old.

Subjects selected for each model in this clinical trial are: A1) group 29 b) group: 55 c) group: 55. Total number of subjects is 165.

The clinical performance test protocol and data analysis were conducted in accordance with the requirement of ISO 80601-2-56: 2017+A1: 2018. The test report showed the clinical performance of the subject devices complied with the requirement of ISO 80601-2-56: 2017+A1: 2018.

Conclusions

Based on comparison of the intended use, technological characteristics, applicable safety standards, verification and validation testing, the differences between Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) and the predicate device do not raise new issues of safety and effectiveness. it can be concluded that Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) is substantially equivalent to the legally marketed predicate device.