K Number
K220530

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-08-17

(174 days)

Product Code
Regulation Number
868.2775
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.

Device Description

The TetraGraph Neuromuscular Transmission (NMT) Monitor (TetraGraph) is a portable, batteryoperated EMG-based neuromuscular transmission monitor for use perioperative and in recovery and critical care environments following or during the application of Neuromuscular block. TetraGraph undertakes this function by electrical stimulation of the peripheral nerve and directly measuring the evoked response of the muscles (Muscle Action Potential (MAP)), thus providing a quantitative and automatic measurement of muscle response to a stimulus using electromyography (EMG). The TetraGraph is a prescription-only medical device and is indicated for use in hospitals. TetraGraph consists of the following main components: TetraGraph Monitor, TetraSens Electrode, TetraSens Pediatric Electrode (new accessory), Philips Interface (new and optional accessory), and Pole clamp kit (optional accessory).

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Senzime AB TetraGraph Neuromuscular Transmission Monitor, specifically for the addition of new accessories: the TetraSens Pediatric electrode and the Philips Interface. The submission aims to demonstrate substantial equivalence to a legally marketed predicate device (TetraGraph Neuromuscular Transmission Monitor, K190795).

Based on the provided document, here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance:

The document does not explicitly present a table of acceptance criteria with corresponding performance results in a clear, summarized format. Instead, it refers to generalized performance testing.

However, it implicitly states that performance testing was conducted for the TetraGraph with the new TetraSens Pediatric electrodes and that "all of the acceptance criteria for this device" were met. These criteria are noted to be "the same or very similar to the predicate device's specifications."

The types of performance testing mentioned are:

Acceptance Criteria TypeReported Device Performance
BiocompatibilityMet compliance in accordance with ISO 10993-1, including cytotoxicity, sensitization, and irritation.
Electrode Performance (General)Met compliance with ANSI/AAMI EC12: 2000 for Disposable ECG Electrodes.
EMG Evoked Response DetectionCompleted testing, and acceptance criteria were met.
Electrode Tensile StrengthCompleted testing, and acceptance criteria were met.
Other Performance TestingCompleted and met all acceptance criteria (e.g., shelf life, software, electrical and EMC).

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

The document does not specify the sample sizes used for any of the performance tests mentioned (biocompatibility, electrode testing, EMG detection, tensile strength, shelf life, software, electrical, EMC).

The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It implies internal testing conducted by the manufacturer or their designated testing facilities.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not provided in the document. The tests described are largely bench-top and engineering performance tests, which typically do not involve experts to establish ground truth in the same way clinical studies or image-based diagnostics would.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable/not provided as the tests described are technical performance tests rather than clinical evaluations requiring adjudication of subjective outcomes.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This type of study is not applicable and was not conducted. The device is a neuromuscular transmission monitor, not an AI-assisted diagnostic tool that aids human readers. The document explicitly states: "There was no clinical testing required to demonstrate that TetraGraph with the new TetraSens Pediatric electrodes is substantial equivalent to the predicate device as the predicate has similar technological characteristics."

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device itself is a measurement and monitoring device, an "algorithm only" or "standalone" performance would refer to its accuracy and reliability in measuring neuromuscular transmission. The performance testing (EMG evoked response detection, compliance with standards) serves this purpose by demonstrating the technical capabilities of the device in isolation, even though it's used by humans. The document confirms that "performance testing made for the device" was completed and met acceptance criteria.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the described performance tests, the "ground truth" would be established by:

  • Scientific Standards: For biocompatibility (ISO 10993-1) and electrode testing (ANSI/AAMI EC12: 2000), the ground truth is defined by the requirements and test methods outlined in these international and national standards.
  • Engineering Specifications: For EMG evoked response detection, electrode tensile strength, and other performance tests, the ground truth is based on the device's design specifications and expected technical performance against which the measured results are compared.

8. The sample size for the training set:

This information is not applicable/not provided. The TetraGraph is a measurement device and the data provided does not suggest it uses machine learning or AI models that require a "training set" in the conventional sense. The submission focuses on hardware accessories and their technical performance.

9. How the ground truth for the training set was established:

This information is not applicable/not provided for the same reasons as in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger, bolder font than the words "U.S. Food & Drug Administration" below it.

August 17, 2022

Senzime AB % Elisa Maldonado-Holmertz RA/OA Consultant Obelix Consulting 12416 Fairfax Ridge Place Austin, Texas 78738

Re: K220530

Trade/Device Name: Tetragraph Neuromuscular Transmission Monitor Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI Dated: April 18, 2022 Received: April 19, 2022

Dear Elisa Maldonado-Holmertz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Todd Courtney Assistant Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthsia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220530

Device Name

TetraGraph Neuromuscular Transmission Monitor

Indications for Use (Describe)

The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submission Sponsor and contact person 1

Submisson Sponsor: Senzime AB Ulls väg 41, 756 51 Uppsala Sweden Johanna Faris, QA/RA Director Email: johanna.faris@senzime.com Phone number: +46 (0)705286330

Submission Correspondent:Elisa Maldonado-Holmertz
RA/QA Consultant
Obelix Consulting, LLC
12416 Fairfax Ridge Place
Austin, TX 78738
Email:elisamh@obelixconsult.com
Tel number:+1 512 431 6069

Date Prepared 2

25 March 2022

Device Identification 3

Table 1. Device identification

The 510(k) number under which the legallymarketed (existing) device was clearedTraditional 510(k): K190795, cleared October 18,2019
Type of this 510(k) Submission:Traditional
Trade or Proprietary Name:Tetragraph Neuromuscular Transmission Monitor
Common or Usual Name:Tetragraph
Regulation Description:Stimulator, Nerve, Peripheral, Electric
Regulation Classification:868.2775
Product CodeKOI
Class of Device:Class II
Review Panel:Division of Anesthesia, Respiratory, and SleepDevices (DHT1C)
Reason for Submission:New accessories: electrode size and interface
Prior Related Submissions:K190795K220530 (Special 510(k) converted to Traditional510(k))

{4}------------------------------------------------

Multiple Devices: | None; this is the only device in the submission

বা Legally Marketed Predicate Device

Predicate Device - K190795 Tetragraph Neuromuscular Transmission Monitor.

ഗ Device Description

Title:

The TetraGraph Neuromuscular Transmission (NMT) Monitor (TetraGraph) is a portable, batteryoperated EMG-based neuromuscular transmission monitor for use perioperative and in recovery and critical care environments following or during the application of Neuromuscular block.

Neuromuscular Transmission (NMT) is the transfer of an electrical impulse between a motor nerve and its associated muscle. The NMT is blocked by neuromuscular blocking agents ("NMBAs") which cause transient muscle paralysis preventing the patient from moving and breathing spontaneously.

Muscle relaxation is used during general anesthesia to enable endotracheal intubation and mechanical ventilation and to provide optimal surgical conditions. Muscle relaxation may also be used in critical care during mechanical ventilation. In these circumstances, TetraGraph can be used as an objective monitor of neuromuscular transmission. TetraGraph undertakes this function by electrical stimulation of the peripheral nerve and directly measuring the evoked response of the muscles (Muscle Action Potential (MAP)), thus providing a quantitative and automatic measurement of muscle response to a stimulus using electromyography (EMG). The TetraGraph is a prescription-only medical device and is indicated for use in hospitals.

The level of neuromuscular block is routinely measured by stimulating a peripheral nerve, by evaluating the muscle response. The TetraGraph controls the level of electrical stimulation applied to the nerve and monitors the muscle response by the use of Electromyography (EMG) detected by electrodes on the muscle.

TetraGraph consists of the following main components:

5.1 TetraGraph Monitor

The TetraGraph Monitor is used to control the electrical stimulation and to measure the EMGresponse. The Monitor is controlled via a touch screen and a power button. The TetraGraph Monitor is connected to the electrode via a cable (the TetraCord Cable).

5.2 TetraSens Electrode

The TetraSens electrode is a single-use electrode array. Each array includes two stimulating electrodes and two recording electrodes. The TetraGraph Monitor can transmit stimulation pulses to the patient and can receive EMG signals via the electrode array. The electrodes are neither supplied sterile nor intended to be sterilized by the user.

The TetraSens electrodes are sold separately from the TetraGraph monitor and are available in boxes of 20 pcs.

5.3 TetraSens Pediatric Electrode (new accessory to TetraGraph)

{5}------------------------------------------------

The TetraSens Pediatric electrode is a single-use electrode array includes two stimulating electrodes and two recording electrodes. The TetraGraph Monitor can transmit stimulation pulses to the patient and can receive EMG signals via the electrode array. The electrodes are neither supplied sterile nor intended to be sterilized by the user.

The TetraSens Pediatric electrodes are sold separately from the TetraGraph monitor and are available in boxes of 15 pcs.

5.4 Philips Interface (new and optional accessory to TetraGraph)

The Philips interface is a reusable dongle/cable to connect between TetraGraph and Philips patient monitor. Live measurement data is transmitted through the interface for display on the external patient monitor. The Philips Interface is neither supplied sterile nor intended to be sterilized by the user. This cable is not intended to have any patient contact during clinical application.

The Philips Interface Cable is an optional accessory to Tetragraph.

5.5 Pole clamp kit (optional accessory to TetraGraph)

The pole clamp kit is a reusable accessory to mount the TetraGraph monitor to a pole stand. The kit includes a mounting device and an attachment to the TetraGraph. This pole clamp kit is neither supplied sterile nor intended to be sterilized by the user. This accessory is not intended to have any patient contact during clinical application.

The Pole clamp kit is an optional accessory to TetraGraph.

6 Indication for Use Statement

The Indication for Use Statement is the same as K190795.

The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.

7 Substantial Equivalence Discussion

A device comparison is made between the TetraGraph and new accessory TetraSens Pediatric electrode and the predicate device TetraSens electrode (K190795). No characteristics of TetraGraph monitoring system are changed due to the TetraGraph Philips Interface cable that allows for connection to an external Philips monitor.

TetraSens Pediatric electrode and TetraSens electrode (K190795) are substantially equivalent with minor dimension differences, minor material differences and labeling additions. TetraSens Pediatric electrode has the same target patient group and maximum stimulating energy as TOFscan (K172690). The other reference devices are the pediatric sensor for Phillips IntelliVue Patient Monitor (K122439 and K161531).

Philips Interface are substantially equivalent with Philips IntelliVue 4-Slot Module Rack FMX-4 (K210906) as both are intended to be connected to a Philips IntelliVue monitor to allow for display of data

{6}------------------------------------------------

Device Modifications and Comparison of Technological Characteristics 7.1 with the predicate device

In this submission, a new electrode accessory called TetraSens Pediatric is introduced to the TetraGraph monitoring system. This electrode has the same indication for use, the same technological characteristics, and the same operation principle, compared to the electrode submitted in the predicate device, i.e. TetraSens electrode that is part of the cleared K190795 is still part of the system and is not changed. Both TetraSens Pediatric electrodes and the predicate TetraSens electrode are intended to be connected to TetraGraph with the same reusable cables. No software and hardware modification has been made to TetraGraph monitoring system due to the introduction of TetraSens Pediatric. There are minor technological differences in size of electrode and materials between TetraSens Pediatric electrode and predicate TetraSens electrode.

Substantial Equivalence Table:

The following Comparison of Characteristics Table compares the TetraGraph in this submission to the predicate TetraGraph (K190795) with respect to indication for use, technological characteristics, and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

No characteristics of TetraGraph monitoring system are changed due to the Philips Interface cable that allows for connection to an external Philips monitor.

Table 2- Comparison of Characteristics between TetraGraph (subject device) and K190795 (predicate device)

Trade NameTetraGraph Neuromuscular Transmission MonitorTetraGraph Neuromuscular Transmission MonitorComment
DeviceSUBJECT (K220530)PREDICATE (K190795)
Product CodeKOIKOISame
Device ClassClass IIClass IISame
Classification nameElectrical peripheral nerve stimulatorElectrical peripheral nerve stimulatorSame
Regulation Number868.2775868.2775Same
Classification PanelAnesthesiologyAnesthesiologySame
Indications for UseThe TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.Same
Muscle movement detection technologyElectromyography (EMG)Electromyography (EMG)Same

{7}------------------------------------------------

Trade NameTetraGraph NeuromuscularTransmission MonitorTetraGraph NeuromuscularTransmission MonitorComment
DeviceSUBJECTPREDICATE (K190795)
Electrode ConnectionReusable CableReusable CableSame
Electrode for StimulationSingle use electrodeSingle use electrodeAdditional size
array (4 electrodes on an array)Smaller dimension of the electrodearray for pediatric usearray (4 electrodes on an array)Minor changein materialused inTetraSensePediatricelectrode
Stimulation PatternsSingle Twitch (ST), Train-of-Four(TOF), Post-tetanic Count (PTC)Single Twitch (ST), Train-of-Four(TOF), Post-tetanic Count (PTC)Same
Stimulation Current Range10-60 mA10-60 mASame
Stimulation Pulse WidthSquare wave, constant current: 200us or 300 usSquare wave, constant current: 200us or 300 usSame

TetraSens Pediatric electrode has the same target patient group and maximum stimulating energy as TOFscan (K172690). The other reference devices are the pediatric sensor for Phillips IntelliVue Patient Monitor (K122439 and K161531).

Trade nameTetraSens PediatricTetraSensToFscanIntelliVue Patient Monitors
DeviceSubject devicePredicate device(K190795)Reference device(K172690)Reference device(K122439)Reference Device(K161531)
Product codeKOIKOIKOIMHX (KOI)MHX (KOI)
Patient populationgroupinfant and PediatricpatientsAdult-size handPediatric Handsensor andelectrodesconnector,"The NMT Moduleisintended to beused with adult andpediatric patients.""The monitor isintended to be usedfor monitoring andrecording of, and togenerate alarms,for,multiplephysiologicalparameters ofadults, pediatrics,and neonates."

{8}------------------------------------------------

TetraGraph Philips Interface cable has same intended use as Philips IntelliVue 4-Slot Module Rack FMX-4 as both are intended to be connected to a Philips IntelliVue monitor to allow for display of data.

Trade nameTetraGraph Philips InterfacecablePhilips IntelliVue 4-SlotModule Rack FMX-4
DeviceSubject devicePredicate device (K210906)
Product codeKOIMHX (KOI)
Outlet PortPhilip IntelliVue MonitorPhilip IntelliVue Monitor

7.2 Performance Testing

Performance testing has been made for the TetraGraph with the TetraSens Pediatric electrodes which supports substantial equivalence between the TetraGraph and the predicate device.

The performance testing made for the device is summarized below:

  • . Biocompatibility testing (in accordance with ISO 10993-1, including cytotoxicity, sensitization, and irritation)
  • Testing of electrodes against ANSI/AAMI EC12: 2000 for Disposable ECG Electrodes
  • EMG evoked response detection
  • . Electrode tensile strength
  • Other performance testing was also completed and met all acceptance criteria

The performance testing on bench showed that the new TetraSens Pediatric electrode used with the TetraGraph met all of the acceptance criteria for this device, and which are the same or very similar to the predicate device's specifications. Thus, the test data demonstrate that TetraGraph is at least as safe and effective in monitoring the relaxation of the patient when neuromuscular blockade is administered as the predicate.

7.3 Clinical Performance Data

There was no clinical testing required to demonstrate that TetraGraph with the new TetraSens Pediatric electrodes is substantial equivalent to the predicate device as the predicate has similar technological characteristics.

Statement of Substantial Equivalence 7.4

In summary, TetraGraph with the TetraSens Pediatric electrode, compared to its predicate (TetraGraph with the TetraSens electrode), has the same indications for use, same technological characteristics which do not raise questions of safety or effectiveness. Test data (shelf life, biocompatibility, software,

{9}------------------------------------------------

electrical and EMC, and performance) demonstrates that the TetraGraph with the new TetraSens Pediatric electrodes is at least as safe and effective as its predicate device for the same indication. Thus, TetraGraph with the TetraSens Pediatric electrode is substantially equivalent to its predicate device, TetraGraph with the TetraSens electrode.

The TetraGraph monitoring system, the related components, the accessories, and the software are substantially equivalent to its predicate device, TetraGraph (K190795).

§ 868.2775 Electrical peripheral nerve stimulator.

(a)
Identification. An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.(b)
Classification. Class II (performance standards).