K Number
K220250
Date Cleared
2022-05-11

(100 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards.

Device Description

The subject device is Nitrile Examination Gloves Sterile with Chemotherapy drugs claims. The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per 10993-10, 10993-11 and permeability to chemotherapy drugs per ASTM D6978-05(2019).

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the FDA regarding Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs.

This document does not describe the acceptance criteria and study proving a software algorithm device meets these criteria, but rather the acceptance criteria and test results for a physical medical device (examination gloves).

Therefore, many of the requested categories in your prompt (such as sample size for test/training sets, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, etc.) are not applicable to this type of device and the information provided in the 510(k) summary.

However, I can extract the relevant information about the acceptance criteria and the device's performance based on the provided document.


Device: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

Type of Device: Physical Medical Device (Examination Gloves), Class I, reserved

Study Purpose: To demonstrate substantial equivalence to a predicate device (Riverstone Resources SDN BHD. Nitrile Examination Glove, Sterile Tested for use with Chemotherapy Drugs Claim, White (K190725)) by verifying that the subject device meets specific performance standards and has similar technological characteristics.

1. Table of Acceptance Criteria and Reported Device Performance

The document presents several tables detailing the acceptance criteria (standards/specifications) and the results indicating that the device meets these criteria.

Item/TestAcceptance Criteria (Standard/Specification)Reported Device Performance (Results)
Dimensions
LengthMinimum 230mmAll size ≥290mm
Palm Width (XS)70 ± 10mm76-78mm
Palm Width (S)80 ± 10mm86-88mm
Palm Width (M)95 ± 10mm96-98mm
Palm Width (L)110 ± 10mm106-108mm
Palm Width (XL)120 ± 10mm116-118mm
Palm Width (XXL)130 ± 10mm126-128mm
Physical Properties (Thickness)
Thickness (Palm)0.05mm (min)0.09-0.11mm
Thickness (Finger)0.05mm (min)0.13-0.14mm
Physical Properties (Tensile Strength)
Tensile Strength, Before Aging14MPa, min15.8-20.8 MPa
Tensile Strength, After Accelerated Aging14MPa, min15.6-19.8 MPa
Physical Properties (Ultimate Elongation)
Ultimate Elongation, Before Aging500%, min500-560%
Ultimate Elongation, After Accelerated Aging400%, min400-500%
Integrity
Freedom from holesG-I, AQL 2.5 (per ASTM D5151-19, ASTM D6319-19)Meet and above AQL 2.5 requirements
Powder Content≤ 2 mg per glove (per ASTM D6124-06(2017), ASTM D6319-19)≤ 2 mg, meet requirements
Biocompatibility
Primary Skin Irritation TestNot an irritant (per ISO 10993-10:2010)Not an irritant
Dermal Sensitization AssayNot a sensitizer (per ISO 10993-10:2010)Not a sensitizer
Acute Systemic Toxicity TestNo evidence of systemic toxicity (per ISO 10993-11:2017)No evidence of systemic toxicity
Chemotherapy Drug Permeation (Minimum Breakthrough Time - BDT)
Carmustine (BCNU) (3.3mg/ml)- (Predicate: 39 min)54.3 minutes
Cyclophosphamide (20mg/ml)- (Predicate: >240 min)>240 minutes
Doxorubicin Hydrochloride (2 mg/ml)- (Predicate: >240 min)>240 minutes
Etoposide (20mg/ml)- (Predicate: >240 min)>240 minutes
Fluorouracil (50mg/ml)- (Predicate: >240 min)>240 minutes
Paclitaxel (6mg/ml)- (Predicate: >240 min)>240 minutes
Thiotepa (10mg/ml)- (Predicate: 97.6 min)196.7 minutes
Methotrexate (25mg/ml)- (Predicate: >240 min)>240 minutes
Cisplatin (1mg/ml)- (Predicate: N/A)>240 minutes
Vincristine Sulfate (1.0 mg/ml)- (Predicate: N/A)>240 minutes
Cytarabine HCL (100mg/ml)- (Predicate: N/A)>240 minutes
Mitoxantrone (2.0 mg/ml)- (Predicate: N/A)>240 minutes
Mitomycin C (0.5 mg/ml)- (Predicate: N/A)>240 minutes
Ifosfamide (50.0 mg/ml)- (Predicate: N/A)>240 minutes
Dacarbazine (DTIC) (10.0 mg/ml)- (Predicate: N/A)>240 minutes

(Note: For chemotherapy drug permeation, the "Acceptance Criteria" are implied by comparison to the predicate device and showing equivalent or better performance. The exact criteria are not explicitly stated for each drug beyond the measured BDT.)

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated as a number of "cases" or "samples" in the way a software algorithm test set would be. The testing is based on batches of gloves or representative samples according to the referenced ASTM/ISO standards. For example, "Freedom from holes" testing follows an AQL (Acceptance Quality Limit) which implies a sampling plan.
  • Data Provenance: The tests are non-clinical and conducted based on international standards (ASTM, ISO). The manufacturer is Grand Work Plastic Products Co., Ltd, located in Donggao Industrial Zone, Zanhuang, Hebei, China. The testing would have been performed there or at certified labs used by the manufacturer. The data is retrospective in the sense that it was collected as part of the device's development and qualification prior to the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The "ground truth" for this physical device's performance (e.g., tensile strength, breakthrough time) is established by adherence to standardized physical and chemical testing methods (ASTM, ISO standards), not by human expert consensus or clinical judgment in the traditional sense of diagnostic AI. These standards define the test procedures and acceptance limits.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a study requiring human reader adjudication. Performance is measured by laboratory equipment and standard protocols.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-based diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI-based diagnostic device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" is established through standardized physical, chemical, and biological testing methods as defined by:

  • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves)
  • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
  • ASTM D6124-06(2017) (Standard Test Method for Residual Powder on Medical Gloves)
  • ASTM D6978-05(2019) (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs)
  • ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
  • ISO 10993-11:2017 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)

8. The sample size for the training set

Not applicable. This is not a machine learning device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not a machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

May 11, 2022

Grand Work Plastic Products Co., Ltd % Kathy Liu Project Manager Hongray USA Medical Products Inc. 3973 Schaefer Avenue Chino, California 91710

Re: K220250

Trade/Device Name: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: February 12, 2022 Received: February 17, 2022

Dear Kathy Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220250

Device Name

Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards.

Chemotherapy DrugMinimum BDT (Minutes)
Carmustine (BCNU) (3.3mg/ml)54.3
Cyclophosphamide (20mg/ml)>240
Doxorubicin Hydrochloride (2 mg/ml)>240
Etoposide (20mg/ml)>240
Fluorouracil (50mg/ml)>240
Paclitaxel (6mg/ml)>240
Thiotepa (10mg/ml)196.7
Methotrexate (25mg/ml)>240
Cisplatin (1mg/ml)>240
Vincristine Sulfate (1.0 mg/ml)>240
Cytarabine HCL (100mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240
Please note that the following drugs have low permeation times:
Carmustine (BCNU): 54.3 minutes, Thiotepa: 196.7 minutes

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

510(K) SUMMARY

This summary of 510(K) is being submitted in accordance with requirements of 21 CFR $807.92.

The assigned 510(K) number is: K220250

Date Prepared: January 20, 2022

  1. Owner's Identification:

Mrs. Wu Yuli Grand Work Plastic Products Co., Ltd. Donggao Industrial Zone, Zanhuang, Hebei, 050000, China Tel: 86-311-66179668

Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Ave., Chino, CA 91710 Tel: 909-590-1611

2. Name of the Device:

Trade Name: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs Common Name: Exam Gloves Classification Name: Patient Examination Glove Classification Regulation: 880.6250 Classification Panel: 880 General Hospital and Personal Use Product Code: LZA and LZC Device Class: Class I

3. Predicate Device Information:

Riverstone Resources SDN BHD. Nitrile Examination Glove, Sterile Tested for use with Chemotherapy Drugs Claim, White (K190725)

4. Device Description:

The subject device is Nitrile Examination Gloves Sterile with Chemotherapy drugs claims. The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per 10993-10, 10993-11 and permeability to chemotherapy drugs per ASTM D6978-05(2019).

5. Indications for Use:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs and provides protection against following chemotherapy drugs and the minimum breakthrough time are listed as follows.

Chemotherapy DrugMinimum BDT (Minutes)
Carmustine (BCNU) (3.3mg/ml)54.3
Cyclophosphamide (20mg/ml)>240
Doxorubicin Hydrochloride (2 mg/ml)>240
Etoposide (20mg/ml)>240
Fluorouracil (50mg/ml)>240
Paclitaxel (6mg/ml)>240

{4}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

Thiotepa (10mg/ml)196.7
Methotrexate (25mg/ml)>240
Cisplatin (1mg/ml)>240
Vincristine Sulfate (1.0 mg/ml)>240
Cytarabine HCL (100mg/ml)>240
Mitoxantrone (2.0 mg/ml)>240
Mitomycin C (0.5 mg/ml)>240
Ifosfamide (50.0 mg/ml)>240
Dacarbazine (DTIC) (10.0 mg/ml)>240

Please note that the following drugs have low permeation times: Carmustine (BCNU): 54.3 minutes, Thiotepa: 196.7 minutes

6. Specification for Nitrile Gloves:

ItemsAcceptance CriteriaResults
LengthMinimum 230mmAll size ≥290
Palm WidthXS: 70±10mm76-78mm
S: 80±10mm86-88 mm
M:95±10mm96 -98mm
L:110±10mm106-108 mm
XL: 120±10mm116-118 mm
XXL: 130±10mm126-128 mm
ThicknessPalm: 0.05mm (min)0.09-0.11mm
Finger: 0.05mm (min)0.13-0.14mm
Tensile Strength, Before Aging14MPa, min15.8-20.8 MPa
Tensile Strength, After Accelerated Aging14MPa, min15.6-19.8 MPa
Ultimate Elongation, Before Aging500%, min500-560%
Ultimate Elongation, After Accelerated Aging400%, min400-500%
Freedom from holesG-I, AQL 2.5Meet and above AQL2.5requirements
Powder-Content≤2 mg per glove≤ 2 mg, meet requirements

Gloves meet all the specification listed in ASTM D 6319-19.

7. Technological Characteristic Comparison:

CharacteristicsSubject DevicePredicate Device K190725Comparison
Trade NameSterile Powder Free NitrileExamination Gloves (Blue), Testedfor Use with Chemotherapy DrugsNitrile Examination GlovesSterile Tested for Use withChemotherapy Drugs claim, WhiteSimilar
Product CodeLZA and LZCLZA and LZCSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same

{5}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

ClassIISame
Indications for UseSubject device is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.Predicate device is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.Same
DescriptionSterile Examination glove made ofNitrile and color Blue and testedwith chemotherapy DrugsSterile Examination glove made ofNitrile and color White and tested withchemotherapy DrugsSame
Design FeatureAmbidextrousAmbidextrousSame
Material useNitrile Synthetic LatexNitrile Synthetic LatexSame
ColorBlueWhiteDifferent
Sterility ISO 11737-2EN ISO 11137-2SterilitySterilitySame
Single usedSingle usedSingle usedSame
Non Sterile or SterileSterileSterileSame
10993-10:2010 SkinIrritation StudyUnder the conditions of thestudy, not an irritantUnder the conditions of the study,not an irritantSame
10993-10:2010Maximization SensitizationStudyUnder the conditions of thestudy, not a sensitizerUnder the conditions of the study,not a sensitizerSame
ISO 10993 Part 11Acute Systemic ToxicityTestUnder the conditions of this study,there was no evidence of systemictoxicity.Under the condition of this study,Nitrile Examination glove, sterileshows no adverse biological reactionSame

Dimensions and Performance Comparison Table:

TechnologicalCharacteristicsSubject DevicePredicate DeviceK190725Comparison
LengthMinimum 230mmMinimum 230mmSame
Palm Width (size) (mm)
XS70±1070±10Same
S80±1080±10Same
M95±1095±10Same
L110±10110±10Same
XL120±10120±10Same
XXL130±10/Different
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same
PalmMinimum 0.05Minimum 0.05Same
Tensile Strength, Before Aging14MPa, min14MPa, minSame
Ultimate Elongation,Before Aging500%, min500%, minSame

{6}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

Tensile Strength, AfterAccelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation, AfterAccelerated Aging400%, min400%, minSame
Freedom from holesIn accordance with ASTM D5151-19, following ASTMD6319- 19, G-I, AQL 2.5In accordance with ASTM D5151-19, following ASTMD6319- 19, G-I, AQL 2.5Same
Residual Powder≤ 2 mg per glove≤ 2 mg per gloveSame

Chemotherapy Permeation Comparison:

Tested Chemotherapy Drug andConcentrationMinimum BDT (Minutes)Comparison
Subject DevicePredicate DeviceK190725
Carmustine (BCNU) (3.3mg/ml)54.339Similar
Cyclophosphamide (20mg/ml)>240>240Same
Cytarabine (1.0 mg/ml)/>240Different
Doxorubicin Hydrochloride (2 mg/ml)>240>240Same
Etoposide (20mg/ml)>240>240Same
Fluorouracil (50mg/ml)>240>240Same
Paclitaxel (6mg/ml)>240>240Same
Thiotepa (10mg/ml)196.797.6Similar
Methotrexate (25mg/ml)>240>240Same
Cisplatin (1mg/ml)>240/Exceed
Vincristine Sulfate (1.0 mg/ml)>240/Exceed
Cytarabine HCL (100mg/ml)>240/Exceed
Mitoxantrone (2.0 mg/ml)>240/Exceed
Mitomycin C (0.5 mg/ml)>240/Exceed
Ifosfamide (50.0 mg/ml)>240/Exceed
Dacarbazine (10.0 mg/ml)>240/Exceed

8 Summary of Non-Clinical Testing

Non-clinical tests were performed to verify that the subject device will meet the acceptance criteria of the performance test shown below:

Test MethodologyPurposeAcceptanceCriteriaResult Summary
ASTM D6319-19The purpose of the testing is tomeasure the dimensions of thesubject device.XS-70±10S-80±10M-95±10L-110±10XL-120±10XXL-130±10Meets
ASTM D6319-19The purpose of the testing is tomeasure the physical properties ofFinger - Minimum 0.05Meets

{7}------------------------------------------------

ggao Industrial Zone, Zanhuang, Hebei, 050000, China

the subject device.Palm - Minimum 0.05
Tensile Strength, before aging -14MPa, min
Ultimate Elongation, Before Aging -500%, min
Tensile Strength, After AcceleratedAging - 14MPa, min
UltimateElongation, AfterAcceleratedAging - 400%,mi
ASTM D6319- 19ASTM D5151-19The purpose of the testing is tomeasure the freedom from holes ofthe subject device.In accordance with ASTM D5151-19, following ASTMD6319- 19, G-I,AQL 2.5Meets
ASTM D 6124-06(2017)ASTM D6319- 19The purpose of the testing is tomeasure the residual powder on thesubject device.≤2 mg per gloveMeets
ISO 10993-10:2010The purpose of the testing is tomeasure the primary skin irritationtest of the subject device.Under the conditions of the study thedevice extracts were not an irritantMeets
ISO 10993-10:2010The purpose of the testing is tomeasure the Dermal sensitizationassay of the subject device.Under the conditions of the studythe device extracts were not asensitizerMeets
ISO 10993-11:2017The purpose of the testing is tomeasure the Acute SystemicToxicity of the subject device.Under the conditions of this study,there was no evidence of systemictoxicity for the device extracts.Meets

{8}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

The subject device meets the applicable requirements for patient examination gloves with regards to dimension and sizes, physical properties, freedom from holes and powder Residuals as found in the following standards: ASTM D6139, ASTM D5151, ASTM D6124.The subject device passes biological reactivity testing for dermal sensitization, irritation and acute systemic toxicity with accordance ISO10993-10 and ISO 10993-11.

9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.