(421 days)
Intended for use for lesioning of neural tissue and for pain management. It is indicated for use in the peripheral nervous system. The APEX 6 is to be used with LCCS electrodes and cannulae and Conmed Thermogard Dispersive Electrodes.
The RF Innovations APEX 6 is a desktop RF lesioning generator which is used for the lesioning of neural tissue. The APEX 6 is a multi-lesioning, 6 channel portable generator that can provide continuous or pulsed RF output at 460 kHz, and monopolar or dual electrode modes. The device includes sensory and motor stimulation functions to fine tune electrode placement for procedures. Based on performance testing the device is designed to connect to LCCS FDA cleared lesioning probes which are inserted into patients for lesioning of neural tissue during medical procedures. Device features include a touch screen monitor incorporating microprocessor and graphics display for user interface as well as self-diagnostics, calibration checks, and recordkeeping functions.
This document is a 510(k) summary for the APEX 6 Lesioning Generator. It outlines the device's characteristics, its intended use, and how it compares to a predicate device (NeuroTherm NT 2000 RF Lesioning Generator) to establish substantial equivalence.
Here's an analysis of the acceptance criteria and the supporting study, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Lesion Size (comparison to predicate device) | Affected area differences between APEX 6 and predicate device should be minimal (implied: insignificant for clinical equivalence). | "All affected area differences were less than 1 mm." The study confirms "the Apex 6 and the predicate device are substantially equivalent in regards to affected tissue size." |
| Comparison of Treatment Times (comparison to predicate device) | Time to ramp up to set temperature and treatment times for APEX 6 should be substantially equivalent to the predicate device under worst-case conditions. | "Results demonstrated that the APEX 6 was substantially equivalent in terms of treatment times compared to the predicate device." |
| Design Validation Review | All documented device requirements (Requirement Specification, Traceability Matrix, Software Specifications, Validation Plans, Instructions for Use) must be met. | "The finished unit design review verified that all documented device requirements were met." |
| Every Unit Functional Test | Each manufactured unit must pass a full functional test regimen (Voltage checks, Program and Impedance Testing, Software testing, Main GUI, Final testing: Impedance/Temperature measurement, Electrical safety testing) prior to shipment. | "Each unit must pass all tests prior to shipment." |
| Safety (Bench Testing) | Compliance with ANSI/AAMI 60601-1, CAN/CSA-C22.2 No. 60601-1, ANSI/AAMI 60601-2-2, and CAN/CSA-C22.2 No. 60601-2-2. | "Tested to ANSI/AAMI 60601-1:2005 + C1:2009 + A2:2010 + A1:2012, CAN/CSA-C22.2 & 60601-1:2014 AND ANSI/AAMI 60601-2-2:2017, CAN/CSA-C22.2 No. 60601-2-2:2019." The document states compliance implicitly by reporting "Tested to...". |
| Electromagnetic Compatibility (EMC) (Bench Testing) | Compliance with IEC 60601-1-2 / EN 60601-1-2. | "EMC was tested in accordance with IEC 60601-1-2:2014/ EN 60601-1-2:2015/IEC 60601-2-2." The document states compliance implicitly by reporting "Tested in accordance with...". |
| Software Performance (Validation to FDA Moderate Level of Concern) | Software performs as expected per validation to FDA guidance for Moderate Level of Concern. | "Software testing supports that the APEX 6 performs as expected. Validation was performed to the FDA Moderate Level of Concern per the FDA software guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." |
Study Details:
The provided document describes bench testing to demonstrate the substantial equivalence of the APEX 6 to its predicate device, primarily focusing on physical performance characteristics and safety standards. There is no clinical study (human-in-the-loop or standalone AI performance) described for this device, as it is a medical device generator and not an AI/Software as a Medical Device (SaMD).
Here's an breakdown of the specific points requested, based on the provided text:
-
A table of acceptance criteria and the reported device performance: (See table above)
-
Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify exact sample sizes for the "Lesion size" or "Comparison of treatment times" tests. It refers to "minimal, typical, and maximum energy delivery" for lesion size and "worse case conditions" for treatment times, implying a set of controlled experimental conditions rather than a large clinical "test set" in the common AI/SaMD sense. For "Every unit functional test," the sample size is "Every unit manufactured."
- Data Provenance: The tests are described as "Bench testing," indicating they were performed in a laboratory or controlled environment. There is no information regarding country of origin or whether it's retrospective or prospective, as it's not a clinical data study.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable to the type of testing described. The "ground truth" for a device like a radiofrequency lesion generator is based on physical measurements (e.g., lesion dimensions, electrical parameters, time) and established engineering/safety standards. It's not about expert interpretation of medical images or clinical outcomes that would require human expert consensus for ground truth.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This is not applicable. Adjudication methods are typically used in clinical studies where multiple human readers interpret data, and significant disagreements require a tie-breaking mechanism or consensus. The tests described are bench tests with objective physical measurements.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This type of study is relevant for AI/SaMD devices where the AI assists human readers in diagnostic or screening tasks. The
APEX 6is a physical medical device generator, not an AI software.
- No, an MRMC study was not done. This type of study is relevant for AI/SaMD devices where the AI assists human readers in diagnostic or screening tasks. The
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- While the
APEX 6itself performs functions autonomously (e.g., generating RF energy), the "standalone" concept usually applies to AI algorithms whose performance is evaluated independently of human input. No such "algorithm-only" performance study is described here, as the device's function is the generation and control of RF energy, which is evaluated through bench tests against physical parameters and predicate device performance. Software functions were validated, but this is different from a standalone AI performance study.
- While the
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for the bench tests was based on physical measurements (e.g., linear dimensions in millimeters, time in seconds, electrical parameters like impedance, voltage, current, power) and compliance with established international and national standards (e.g., IEC 60601 series, ANSI/AAMI, CAN/CSA).
- For "Lesion size" and "Comparison of treatment times," the "ground truth" was derived from the performance of the legally marketed predicate device, used as a benchmark for substantial equivalence.
-
The sample size for the training set:
- This information is not applicable. The device is not an AI/machine learning model that requires a "training set" in the sense of data used to train a statistical or learning algorithm. Its design and validation rely on engineering principles, product specifications, and comparisons to predicate devices.
-
How the ground truth for the training set was established:
- This information is not applicable, as there is no training set for an AI/ML model for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
March 15, 2023
RF Innovations, Inc % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114
Re: K220122
Trade/Device Name: Apex 6 Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency Lesion Generator Regulatory Class: Class II Product Code: GXD Dated: April 27, 2022 Received: April 28, 2022
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed am D. bv Adam D. Pierce -S Pierce -Spate: 2023.03.15 13:24:54 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K220122
Device Name APEX 6
Indications for Use (Describe)
Intended for use for lesioning of neural tissue and for pain management. It is indicated for use in the peripheral nervous system. The APEX 6 is to be used with LCCS electrodes and cannulae and Conmed Thermogard Dispersive Electrodes.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K220122
Date Prepared: March 10, 2023
| SUBMITTED BY: | RF Innovations, IncMiddleton, MA 01940 |
|---|---|
| CONTACT: | William Rittman, Director.RF Innovations, Inc |
| Middleton, MA 01949 | |
| 781-640-1212 | |
| DEVICE NAME: | APEX 6 Lesioning Generator |
| Regulation Number: | 21 CFR 882.4400 |
| Regulation Name: | Radiofrequency Lesion Generator |
| Regulatory Class: | II |
| Product Code: | GXD |
| PREDICATE DEVICE: | K111576 |
| DEVICE NAME: | NeuroTherm NT 2000 RF Lesioning Generator |
| Regulation Number: | 21 CFR 882.4400 |
| Regulation Name: | Radiofrequency Lesion Generator |
| Regulatory Class: | II |
| Product Code: | GXD |
| Reference Devices: | Compatible electrodes: |
| 510(k) number: | K191293 |
| Trade/Device Name: | LCCS RF Cannula |
| Regulation Number: | 21 CFR 882.4725 |
| Regulation Name: | Radiofrequency Lesion Probe |
| Regulatory Class: | Class II |
| Product Code: | GXI |
| 510(k) number: | K152642 |
| Trade/Device Name: | LCCS Radiofrequency (RF) Electrode |
| Regulation Number: | 21 CFR 882.4725 |
| Regulation Name: | Radiofrequency Lesion Probe |
| Regulatory Class: | Class II |
| Product Code: | GXI |
| Compatible Dispersive Grounding Pad: | |
| 510(k) number: | K140658 |
| 510(k) number: | K140658 |
| Trade/Device Name: | Thermogard & Thermogard Plus ABC dispersive electrodes |
| Regulation Number: | 21CFR878.4400 |
| Regulation Name: | Electrosurgical cutting and coagulation device and accessories |
{4}------------------------------------------------
| Regulatory Class: | Class I |
|---|---|
| Product Code: | GEI |
DEVICE DESCRIPTION:
The RF Innovations APEX 6 is a desktop RF lesioning generator which is used for the lesioning of neural tissue. The APEX 6 is a multi-lesioning, 6 channel portable generator that can provide continuous or pulsed RF output at 460 kHz, and monopolar or dual electrode modes. The device includes sensory and motor stimulation functions to fine tune electrode placement for procedures. Based on performance testing the device is designed to connect to LCCS FDA cleared lesioning probes which are inserted into patients for lesioning of neural tissue during medical procedures. Device features include a touch screen monitor incorporating microprocessor and graphics display for user interface as well as self-diagnostics, calibration checks, and recordkeeping functions.
INDICATIONS FOR USE: Intended for use for lesioning of neural tissue and for pain management. It is indicated for use in the peripheral nervous system. The APEX 6 is to be used with LCCS electrodes and cannulae and Conmed Thermogard Dispersive Electrodes.
| NeuroThermNT 2000Generator K111576 | RF InnovationsAPEX 6Generator K220122 | Discussion | |
|---|---|---|---|
| Indicationfor Use | The NT 2000 is intended for use forlesioning of neural tissue. The NT2000 is indicated for use in theperipheral nervous system. The NT2000 is to be used only, with FDAcleared NeuroTherm RF probes andSmith & Nephew SPINECATH andACUTHERM catheters, catheters. | Intended for use for lesioning of neuraltissue and for pain management. It isindicated for use in the peripheralnervous system. The APEX 6 is to beused with LCCS electrodes andcannulae and Conmed ThermogardDispersive Electrodes. | SIMILAR |
| Appearance | Image: NeuroTherm NT 2000 Generator | Image: RF Innovations APEX 6 Generator | SIMILAR |
| Touchscreendimensions | 12" diagonal LCD | 15" diagonal LCD | SIMILAR |
| Number ofChannels | 4 | 6 | Same maxpowerdelivered |
| RF OutputFrequency | 460 kHz | 460 kHz | SAME |
| Maximumoutputpower (100ohm load) | 50 watts | 50 watts | SAME |
| Energydeliveryduring multichannel RFtreatment | Independent sequential energy delivery | Independent sequential energy delivery | SAME |
| Powerdeliverymodes | Continuous and pulsed | Continuous and pulsed | SAME |
| Powersupply | AC Line | AC Line | SAME |
| Autoshutdownfor powerexceedingsafe levels | Yes | Yes | SAME |
| Electricalsafety/EMC | IEC 60601 compliant | Tested to ANSI/AAMI 60601-1:2005 +C1:2009 + A2:2010 + A1:2012, CAN/CSA-C22.2 & 60601-1:2014 AND ANSI/AAMI60601-2-2:2017, CAN/CSA-C22.2 No.60601-2-2:2019. AND IEC 60601-1-2:2014/EN 60601-1-2:2015 | Equivalent |
| RF energy delivery modes: | |||
| Continuousthermal | YES | YES | SAME |
| Stimulation-sensory andmotor | YES | YES | SAME |
| Pulsed RF | YES | YES | SAME |
| RF energy delivery channel types | |||
| Monopolar | YES | YES | SAME |
| Bipolar* | Yesaka "dual" | Yesaka "dual" | SAME |
SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION
{5}------------------------------------------------
- current between two monopolar electrodes
Any differences between the APEX 6 and the predicate 4-channel devices do not raise new issues of safety or effectiveness. The APEX 6 uses the same amount of energy output across 6 channels and delivers this energy sequentially as opposed to continuously. Therefore, the APEX 6 is substantially equivalent to the predicate device based upon the Indication for Use, technology, functional modes, hardware and software components, and performance.
{6}------------------------------------------------
PERFORMANCE TESTING
Bench - Bench testing supports that the APEX 6 performs as expected. Safety was tested in accordance with Standard for Medical electrical equipment - Part 1: General requirements for basic safety and essential performance ANSI/AAMI 60601-1:2005 + Cl:2009 + A2:2010 + Al:2012, CAN/CSA-C22.2 No. 60601-1:2014 and the Standard for Medical Electrical Equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories ANSI/AAMI 60601-2-2:2017, CAN/CSA-C22.2 No. 60601-2-2:2019. EMC was tested in accordance with IEC 60601-1-2:2014/ EN 60601-1-2:2015/IEC 60601-2-2.
| Test | Test Method Summary | Results |
|---|---|---|
| Lesion size | Purpose was to document the maximumlinear dimensions in all three planes aswell as the volume for the affected tissueregion and compare them to thepredicate device for minimal, typical, andmaximum energy delivery | Analysis: All affected area differences were lessthan 1 mm. Minor errors can introduced by:• Measuring small diameters with fuzzy borderscan increase the measurement error.• For volume, a 10% error in the lineardimensions will result in a 33% error in volume.Therefore this study confirms that the Apex 6and the predicate device are substantiallyequivalent in regards to affected tissue size. |
| Comparisonof treatmenttimesbetweenNeuroThermNT 2000and Apex 6 | The purpose of this protocol is tomeasure and compare the time to rampup to set temperature of the Apex 6 andthe predicate device under worse caseconditions to evaluate the efficiency ofenergy delivery and measure treatmenttimes. | Results demonstrated that the APEX 6 wassubstantially equivalent in terms of treatmenttimes compared to the predicate device. |
| DesignValidationReview | A documented review of deviceacceptance criteria was performed toconfirm compliance with the followingplans:• Requirement Specification• Requirements, Design and VerificationTraceability Matrix• Embedded Software RequirementSpecification• GUI Software RequirementSpecification• Embedded Software Validation Plan• GUI Software Validation Plan• Instructions for Use | The finished unit design review verified that alldocumented device requirements were met. |
| Every unitfunctionaltest | Every unit manufactured is subjected to afull functional test regimen and the resultsare recorded in the Device HistoryRecord:Voltage checksProgram and Impedance TestingSoftware testingMain GUI | Each unit must pass all tests prior to shipment. |
| Test | Test Method Summary | Results |
| Final testing: Impedance/TemperaturemeasurementElectrical safety testing |
{7}------------------------------------------------
Software - Software testing supports that the APEX 6 performs as expected. Validation was performed to the FDA Moderate Level of Concern per the FDA software guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
CLINICAL TESTING Not required.
CONCLUSION
The APEX 6 is similar to or the same as the predicate device as follows:
- . Technology
- Intended use/Indication for Use
- Technical specifications, or ranges of technical specifications
- . Functional modes compared to 4-channel devices
The fundamental technological characteristics of the APEX 6 are substantially equivalent to the predicate device based upon the Indication for Use, technology, functional modes, hardware components, and performance. Therefore, the APEX 6 is substantially equivalent to the predicate device.
§ 882.4400 Radiofrequency lesion generator.
(a)
Identification. A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.(b)
Classification. Class II (performance standards).