K Number
K152642
Date Cleared
2016-08-26

(346 days)

Product Code
Regulation Number
882.4725
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LCCS Radiofrequency (RF) Electrode is indicated for use in temperature-controlled RF heat lesson procedures of peripheral nerve tissue for the relief of pain.

Device Description

LCCS Radiofrequency (RF) Electrode is used in conjunction with RF Generator to create RF lesions of nerve tissue or for use in percutaneous nerve blocks. The RF Electrode connects with RF Generator and thermocouple side which fits in with RF Cannula. The RF generator applies temperature-controlled RF energy into targeted nerve tissue via an RF Electrode to create a heat lesioning of peripheral nerve tissue.

LCCS RF Electrode includes LCCS Disposable RF Electrode and LCCS Reusable RF Electrode. LCCS Disposable RF Electrode Part No. includes RE2750-D-LC, RE27100-D-LC, RE27150-D-LC, and RE27200-D-LC.

LCCS Reusable RF Electrode Part No. includes RE2750-R-LC, RE27100-R-LC, RE27150-R-LC, RE27200-R-LC, RE2550-R-LC, RE25100-R-LC, RE25150-R-LC, and RE25200-R-LC.

AI/ML Overview

This document describes the LCCS Radiofrequency (RF) Electrode, a device for temperature-controlled RF heat lesion procedures of peripheral nerve tissue for pain relief. The submission focuses on demonstrating substantial equivalence to predicate devices (OWL Sterile Single Use RF Probes and OWL Reusable RF Probes) through non-clinical testing.

Here's an analysis of the provided information concerning acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the “Passed” status in the non-clinical testing table, indicating the device met the requirements of the specified test methods and standards. The reported device performance is that all tests passed.

Test CategorySpecific TestTest Method SummaryAcceptance Criteria (Implicit)Reported Device Performance
Physical/MechanicalDimensionMeasured the electrode nominal OD and effective length.Met specified dimensionsPassed
CleannessObserved the surface cleanness under normal or corrected vision with 300LX to 700LC illumination.Met visual cleanness standardsPassed
AppearanceInspect the appearance under normal or corrected vision with 300LX to 700LC illumination.Met visual appearance standardsPassed
Corrosion resistanceISO 13402 Surgical and dental hand instruments - Determination of resistance against autoclaving, corrosion and thermal exposureWithstood corrosion/autoclavingPassed
Degree of connection rigidityVerified the connection rigidity by static tensile test.Met rigidity requirementsPassed
FunctionalTemperature accuracyVerified the accuracy by performance testing.Range: 40~100 °C, Tolerance: ± 5%Passed
BiocompatibilityCytotoxicityISO 10993-5 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicityNon-cytotoxicPassed
SensitizationISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and skin sensitizationNon-sensitizingPassed
IrritationISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and skin sensitizationNon-irritatingPassed
EtO and ECH residuals (Disposable only)ISO 10993-7 Biological evaluation of medical devices -Part 7:Ethylene oxide sterilizationMet residual limitsPassed
SterilizationSterilization (Disposable)ISO 11135 Sterilization of health-care products – Ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devicesMet sterilization efficacyPassed
Sterilization (Reusable)ISO 17665-1 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devicesMet sterilization efficacyPassed
Electrical SafetyEMCIEC 60601-2-2 Medical electrical equipment Part 2-2: Particular requirements for basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessoriesMet EMC standardsPassed

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for each of the non-clinical tests. It only states that "All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices." The data provenance is implied to be from laboratory testing conducted by the manufacturer or a third-party testing facility, likely in China (given the applicant and consultant information). The studies are evidently prospective, as they were conducted to support the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

There is no mention of experts being used to establish a "ground truth" for the non-clinical test set in the traditional sense of medical image analysis or diagnostic studies. The "ground truth" for these engineering and biocompatibility tests is established by the specified international standards (e.g., ISO, IEC) and internal performance specifications. The tests themselves are designed to objectively measure physical and chemical properties.

4. Adjudication Method for the Test Set

Not applicable. As described above, the non-clinical tests are objective measurements against established standards, not subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This is a medical device for therapeutic intervention (radiofrequency ablation), not a diagnostic device involving human readers or AI assistance in interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical electrode used in conjunction with an RF generator and is operated by a medical professional; it is not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical studies is based on established international standards (ISO, IEC) for medical devices, materials, and electrical safety, along with the device's own design specifications (e.g., dimensions, temperature range and tolerance). For instance, for biocompatibility, the ground truth is defined by the Pass/Fail criteria outlined in the ISO 10993 series. For temperature accuracy, the ground truth is the ±5% tolerance within the 40-100 °C range.

8. The Sample Size for the Training Set

Not applicable. There is no training set mentioned, as this is not an artificial intelligence/machine learning device.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings forming a flowing, abstract design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 26, 2016

LCCS Products Limited % Mr. Field Fu Shenzhen Joyantech Consulting Co., Ltd. No. 55 Shizhou Middle Road, Nanshan District Shenzhen, Guangdong GD755 China

Re: K152642

Trade/Device Name: LCCS Radiofrequency (RF) Electrode Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: Class II Product Code: GXI Dated: August 9, 2016 Received: August 22, 2016

Dear Mr. Fu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S FD/Δ

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152642

Device Name LCCS Radiofrequency (RF) Electrode

Indications for Use (Describe)

LCCS Radiofrequency (RF) Electrode is indicated for use in temperature-controlled RF heat lesson procedures of peripheral nerve tissue for the relief of pain.

Type of Use (Select one or both, as applicable)

Research Use (Part 21 CFR 361 Subject B)
Same As The Control Use (21 CFR 361 Subject B)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Contact Details 1.

1.1 Applicant information

Applicant NameLCCS Products Limited
AddressFlat A, 9/F., Hennessy Plaza, 164-166 Hennessy Road, Wanchai, HongKong
Phone No.+852-82321050
Fax No.+852-35430978
Contact personMing Lin
Contact person's e-mailminglin2006@yahoo.com.cn
Company e-maillccsmed@126.com
Date PreparedJuly 8, 2016
Websitewww.lccsmed.com

1.2 Consultant information

NameShenzhen Joyantech Consulting Co., Ltd
Image: logo
卓远天成
AddressRoom 2032, International Mayors Communication Centre, NO. 55Shizhou middle road, Nanshan District, Shenzhen
Phone No.+86-755-86069197
Contact personField Fu
Contact person's e-mailcefda13485@163.com
Websitehttp://www.cefda.com

2. Device information

.

Trade nameLCCS Radiofrequency (RF) Electrode
Common nameRadiofrequency Electrode
ModelRE-D (D for Disposable), RE-R (R for Reusable)
ClassificationII
Classification nameProbe, Radiofrequency Lesion
Product codeGXI
Regulation No.882.4725

Legally Marketed Predicate Device 3.

Trade NameOWL Sterile Single Use RF Probes
510(k) NumberK110593
Product CodeGXI
ManufacturerDiros Technology Inc.

{4}------------------------------------------------

Trade NameOWL Reusable RF Probes
510(k) NumberK010202
Product CodeGXI
ManufacturerDiros Technology Inc.

Device Description 4.

LCCS Radiofrequency (RF) Electrode is used in conjunction with RF Generator to create RF lesions of nerve tissue or for use in percutaneous nerve blocks. The RF Electrode connects with RF Generator and thermocouple side which fits in with RF Cannula. The RF generator applies temperature-controlled RF energy into targeted nerve tissue via an RF Electrode to create a heat lesioning of peripheral nerve tissue.

LCCS RF Electrode includes LCCS Disposable RF Electrode and LCCS Reusable RF Electrode. LCCS Disposable RF Electrode Part No. includes RE2750-D-LC, RE27100-D-LC, RE27150-D-LC, and RE27200-D-LC.

LCCS Reusable RF Electrode Part No. includes RE2750-R-LC, RE27100-R-LC, RE27150-R-LC, RE27200-R-LC, RE2550-R-LC, RE25100-R-LC, RE25150-R-LC, and RE25200-R-LC.

5. Intended Use

LCCS Radiofrequency (RF) Electrode is indicated for use in temperature-controlled RF heat lesion procedures of peripheral nerve tissue for the relief of pain.

6. Indication for Use

LCCS Radiofrequency (RF) Electrode is indicated for use in temperature-controlled RF heat lesion procedures of peripheral nerve tissue for the relief of pain.

7. Substantial Equivalence Comparison

  • 7.1 Comparison between LCCS Disposable RF Electrode and OWL Sterile Single Use RF Probes
CharacteristicLCCS DisposableRF ElectrodeOWL Sterile SingleUse RF Probes(K110593)Comments
Intended UseLCCS Radiofrequency(RF) Electrode isindicated for use intemperature-controlledRF lesion procedures ofperipheral nerve tissuefor the relief of pain.The OWL SterileSingle use R.F.Probe/TemperatureSensor is intended foruse inRadio-FrequencyHeat Lesionprocedures forrelief of pain.Equivalent

{5}------------------------------------------------

LCCS Products Limited

Product: LCCS RF Electrode

CharacteristicLCCS DisposableRF ElectrodeOWL SterileSingle Use RFProbes(K110593)Comments
Diameter0.4mm (27gauge)25 AWG(25gauge)27 AWG(27gauge)Equivalent
Lengths Available(Cannulae length to beused with)69mm (5cm)117mm (10cm)160mm (15cm)220mm (20cm)65.8mm (5cm)116.5mm(10cm)161mm (15cm)194mm (20cm)Equivalent
TemperatureMeasurementCapabilitiesTemperaturemeasurementdeviceThermocoupleThermistorThermocoupleTemperature accuracy ofthermocouple has beendemonstrated throughcomparison testing withpredicate. The LCCS DisposableRF Electrode device is asaccurate as the predicate device.
Range40~100 °CUnknown
Tolerance$\pm$ 5%Unknown
Single UseYesYesSame

7.2 Comparison between LCCS Reusable RF Electrode and OWL Reusable RF Probes and Cannulae

CharacteristicLCCS Reusable RFElectrodeOWL ReusableRF Probes(K010202)Comments
Intended UseLCCS Radiofrequency(RF) Electrode isindicated for use intemperature-controlledRF lesion procedures ofperipheral nerve tissuefor the relief of pain.The OWL R.F.Probe/TemperatureSensor is intended foruse inRadio-Frequency HeatLesion procedures forrelief of pain.Equivalent
MaterialElectrode tube: 304Stainless SteelTubeElectrode tube:304 StainlessSteel TubeSame
CharacteristicLCCS Reusable RF ElectrodeOWL Reusable RF Probes (K010202)Comments
Diameter0.4mm (27gauge)0.5mm (25gauge)25 gauge(0.5mm OD)Although the diameter of the predicate OWL Reusable RF Probes do not contain 27gauge, but the OWL Sterile Single Use RF Probes (K110593) contain the diameter of 27gauge, the difference between the predicate OWL Reusable RF Probes and the OWL Sterile Single Use RF Probes (K110593) is just for single use or not. Also, diameter of probe is determined by diameter of cannula. So the difference of electrode's diameter does not change the intended use of the device and does not introduce any new issues of safety and effectiveness.
Lengths Available(Cannulae length to be used with)69mm (5cm)117mm (10cm)160mm (15cm)220mm (20cm)65.8mm (5cm)116.5mm (10cm)161mm (15cm)Although the cannulae length to be used with of the predicate OWL Reusable RF Probes do not contain 200mm, but the OWL Sterile Single Use RF Probes (K110593) contain the length of 200mm, the difference between the predicate OWL Reusable RF Probes and the OWL Sterile Single Use RF Probes (K110593) is just for single use or not. Also, length of probe is determined by length of cannula. So the difference of length does not change the intended use of the device and does not introduce any new issues of safety and effectiveness.
TemperatureMeasurementTemperature measurement deviceThermistorThermocoupleTemperature accuracy of thermocouple has been demonstrated through comparison testing with
MeasurementRange40~100 °CUnknown
CharacteristicLCCS Reusable RF ElectrodeOWL Reusable RF Probes (K010202)Comments
CapabilitiesTolerance$\pm$ 5%Unknownpredicate. The LCCS Reusable RF Electrode device is as accurate as the predicate device.
Single UseNoNoSame

{6}------------------------------------------------

LCCS Products Limited Product: LCCS RF Electrode 510(k) Summary

{7}------------------------------------------------

LCCS Products Limited

Product: LCCS RF Electrode

8. Non-clinical Testing

All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the testing are demonstrating the standards and matching the performance of new devices to the predicate devices.

The following performance data were provided in support of the substantial equivalence determination.

TestTest Method SummaryResults
DimensionMeasured the electrode nominal OD and effective length.Passed
CleannessObserved the surface cleanness under normal or corrected vision with 300LX to 700LC illumination.Passed
AppearanceInspect the appearance under normal or corrected vision with 300LX to 700LC illumination.Passed
Corrosion resistanceISO 13402 Surgical and dental hand instruments - Determination of resistance against autoclaving, corrosion and thermal exposurePassed
Temperature accuracyVerified the accuracy by performance testing.Passed
Degree of connection rigidityVerified the connection rigidity by static tensile test.Passed
BiocompatibilityCytotoxicityISO 10993-5 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicityPassed
SensitizationISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and skin sensitizationPassed
IrritationISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and skin sensitizationPassed
EtO and ECHISO 10993-7 Biological evaluation of medical devices -Part 7:Ethylene oxide sterilizationPassed

8.1 LCCS Disposable RF Electrode nonclinical testing list

{8}------------------------------------------------

residualsresiduals
SterilizationISO 11135 ISO 11135 Sterilization of health-care products – Ethylene oxide –Requirements for the development, validationand routine control of a sterilization process formedical devicesPassed
EMCIEC 60601-2-2 Medical electrical equipmentPart 2-2: Particular requirements for basicsafety and essential performance of highfrequency surgical equipment and highfrequency surgical accessoriesPassed

8.2 LCCS Resuable RF Electrode nonclinical testing list

TestTest Method SummaryResults
DimensionMeasured the electrode nominal OD and effective length.Passed
CleannessObserved the surface cleanness under normal or corrected vision with 300LX to 700LC illumination.Passed
AppearanceInspect the appearance under normal or corrected vision with 300LX to 700LC illumination.Passed
Corrosion resistanceISO 13402 Surgical and dental hand instruments - Determination of resistance against autoclaving, corrosion and thermal exposurePassed
Temperature accuracyVerified the accuracy by performance testing.Passed
Degree of connection rigidityVerified the connection rigidity by static tensile test.Passed
BiocompatibilityCytotoxicityISO 10993-5 Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicityPassed
SensitizationISO 10993-10 Biological evaluation of medical devices –Part 10:Tests for irritation and skin sensitizationPassed
IrritationISO 10993-10 Biological evaluation of medical devices –Part 10:Tests for irritation and skin sensitizationPassed
SterilizationISO 17665-1 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devicesPassed
EMCIEC 60601-2-2 Medical electrical equipmentPassed

{9}------------------------------------------------

Part 2-2: Particular requirements for basic
safety and essential performance of high
frequency surgical equipment and high
frequency surgical accessories

Clinical Testing 9.

Substantial equivalence does not depend on the clinical test data.

10. Conclusions

The non-clinical data demonstrate that the LCCS RF Disposable and Reusable Electrode devices perform comparably to the predicate devices that are currently marketed for the same intended use.

§ 882.4725 Radiofrequency lesion probe.

(a)
Identification. A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.(b)
Classification. Class II (performance standards).