(190 days)
The indications for use for the Passport 2® with View 12™ ECG Analysis Module include the monitoring of the following human physiological parameters:
- ECG waveform derived from 3, 5 or 12 lead measurements
- Heart Rate derived from selected sources (SpO2, ECG, IBP, NIBP)
- Blood Oxygenation (SpO2 )* measurement/waveform
- ST Segment Analysis derived from 5 or 12 lead measurements
- Lethal Arrhythmia Detection derived from 5 or 12 lead measurements
- Non Invasive Blood Pressure (NIBP) measurement
- Invasive Blood Pressure (IBP) measurement/waveform measurable at two sites
- Respiration Rate/ waveform derived from ECG or CO2
- CO2, , Inspired and end tidal microstream/waveform
- Temperature measurement via YSI 400/700 series probes
- Interpretation of Resting 12 lead ECG
The target populations are adult, pediatric and neonate with the exception of the:
- Lethal Arrhythmia Detection and ST Segment Analysis, for which the target populations are adult and pediatric only, and
- Interpretation of Resting 12 Lead ECG, for which the target population is adult only.
The monitor is intended for use in the health care facility setting.
The device has the capability of interfacing with Datascope's Gas Module II, displaying the measurements of Anesthetic Gases, O3, N2O, and CO2.
- The device monitors the SpO2 parameter via the Masimo SET® 2000 Pulse Oximeter Technology and Accessories (K990966). The Masimo SET® 2000 Pulse Oximeter Technology and Accessories are indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO3) and pulse rate (measured by an SpO3 sensor) and are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
The View 12 ECG Analysis Module provides means for performing diagnostic ECG analysis in conjunction with the monitoring functions of Datascope's Passport 2 Vital Signs Monitor (K993531). The View 12 ECG Analysis Module for the Passport 2 Vital Signs Monitor enables 12-Lead ECG Acquisition. Continuous 12-Lead ST Segment Analysis and Arrhythmia Analysis all with print capabilities. The View 12 ECG Analysis Module consists of a PCMCIA ECG processor card that interfaces with the Passport 2, a front-end electronics cable, and a detachable patient cable with integral leadwire set. The data acquired by this module is interpreted utilizing Mortara Instrument's algorithm (cleared by FDA under notification numbers: K920627, K920626A and K933143) for interpretive electrocardiographs. With the View 12 ECG Analysis Module installed and enabled, the Passport 2 makes interpretive statements on printouts. These statements are categorized into three (3) sections: an interpretive statement, a condition statement, and a rhythm statement.
The provided 510(k) summary (K020532) for the Datascope Passport 2® Vital Signs Monitor with View 12™ ECG Analysis Module does not directly provide acceptance criteria or detailed results from a study specifically conducted for this 510(k) submission to prove the device meets acceptance criteria.
Instead, the submission relies on the concept of substantial equivalence to predicate devices. Here's what can be extracted and inferred regarding the study that proves the device meets acceptance criteria, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not contain a specific table outlining acceptance criteria and reported performance for the View 12 ECG Analysis Module itself.
However, the submission states:
- "The diagnostic ECG algorithm used in Datascope's View 12 ECG Analysis Module is identical to that found in the Spacelabs' Integrated Multiparameter Module 90496, as well as in the Mortara Instrument ELI-100 (K920627), ELI-200 (K920626A) and ELI-300 (K933143.)"
- "The functionality of the Passport 2 Vital Signs monitor with View 12 ECG Analysis Module is identical to that of the Spacelabs Ultraview 1050 monitor equipped with Integrated Multiparameter Module 90496."
Therefore, the "acceptance criteria" for the ECG analysis algorithm are implicitly the performance specifications already met and cleared for the predicate Mortara Instrument devices (K920627, K920626A, K933143) and the Spacelabs Module 90496. The reported device performance is, by definition of substantial equivalence in this context, assumed to be identical to these predicate devices. The submission asserts that the Datascope device's algorithm performs as well as the previously cleared Mortara algorithm because it is the Mortara algorithm.
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample size (number of patients/ECGs) for a test set for this specific 510(k) submission to validate the ECG algorithm.
Given the reliance on the predicate Mortara algorithm, the testing data would have been part of the original 510(k) submissions for Mortara Instrument ELI-100 (K920627), ELI-200 (K920626A) and ELI-300 (K933143). Without access to those original submissions, the sample size and data provenance remain unknown from this document. It is highly likely the data would have been retrospective, as is common for established ECG algorithm validation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
This information is not provided in the K020532 document, as the ECG algorithm itself was cleared in prior 510(k)s (Mortara Instrument devices). The original Mortara submissions would contain this detail. Typically, expert cardiologists or electrophysiologists with significant experience would be used for manual adjudication of ECGs to establish ground truth.
4. Adjudication Method for the Test Set:
This information is not provided in the K020532 document. It would have been part of the original Mortara submissions. Standard adjudication methods for ECGs often involve a consensus of multiple cardiologists (e.g., 2-out-of-3 agreement, or a tie-breaker by a senior expert).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
No MRMC comparative effectiveness study is mentioned for the View 12 ECG Analysis Module in this 510(k) summary. The submission's argument is that the algorithm is the predicate algorithm, thus no new large-scale comparative study is required for algorithm performance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
Yes, implicitly. The performance of the diagnostic ECG algorithm (Mortara's algorithm) was assessed in a standalone manner when it was originally cleared under K920627, K920626A, and K933143. This 510(k) leverages that prior clearance.
7. The Type of Ground Truth Used:
For ECG interpretation algorithms, the ground truth is almost exclusively expert consensus by qualified cardiologists or electrophysiologists. This would have been the case for the underlying Mortara algorithm.
8. The Sample Size for the Training Set:
This information is not provided in the K020532 document. The training data for the Mortara algorithm would have been developed prior to its initial 510(k) clearances.
9. How the Ground Truth for the Training Set Was Established:
This information is not provided in the K020532 document. For ECG algorithms, ground truth for training data would also typically be established through expert consensus, often with a similar process to, but distinct from, the ground truth establishment for the independent test set.
Summary of Device Acceptance Rationale (based on the provided text):
The Datascope Passport 2® Vital Signs Monitor with View 12™ ECG Analysis Module's acceptance is based on its substantial equivalence to legally marketed predicate devices. Specifically:
- Functionality: The overall device functionality (Passport 2 with View 12) is stated to be identical to the Spacelabs Ultraview 1050 monitor with Integrated Multiparameter Module 90496.
- ECG Algorithm: The diagnostic ECG algorithm itself is identical to one already cleared and used in several Mortara Instrument devices (K920627, K920626A, K933143) and the Spacelabs module. These predicate devices have already demonstrated compliance with performance standards and safety through their own clearance processes.
- Safety and Performance Testing: The Datascope device underwent "extensive safety and performance testing" and complies with various industry and safety standards (e.g., ANSI/AAMI EC11, EC13, IEC 60601 series, ISO 10993-1, EN 1441). This testing would focus on the integration of the module, electrical safety, EMC, and ensuring that the overall device still meets general performance requirements, not necessarily re-validating the intrinsic accuracy of the ECG interpretation algorithm, which is assumed from its predicate status.
In conclusion, the "study that proves the device meets the acceptance criteria" for the ECG analysis algorithm refers to the studies conducted for the original Mortara Instrument devices (K920627, K920626A, K933143) because the Datascope device uses the identical, previously cleared algorithm. This 510(k) primarily establishes that the integration of this already cleared algorithm into a new monitor platform does not raise new questions of safety or effectiveness.
{0}------------------------------------------------
KC20532
AUG 2 8 2002
510(k) Summary Datascope Passport 2° Vital Signs Monitor with View 12™ ECG Analysis Module
| Submitter: | Datascope Corp. Patient Monitoring Division 800 MacArthur Boulevard Mahwah, New Jersey 07430-0619 Tel: (201) 995-8391 Fax: (201) 995-8605 |
|---|---|
| Contact Person: | Russell Olsen Director, Quality and Regulatory Affairs Datascope Corp. Patient Monitoring Division Tel: (201) 995-8391 Fax: (201) 995-8605 Email: russell_olsen@datascope.com |
| Date Prepared: | February 5, 2002 |
Name of the device:
-
Trade/Proprietary Name: Passport 2" Vital Signs Monitor with View 12™ ECG Analysis ● Module
Please note that during the product development process, the device was also referred to as "Passport 2 12-lead", and the Passport 2 Vital Signs Monitor was also referred as "Enterprise" and "Enterprise Configured Monitor" (or "ECM"). These names will be found in some of the supporting documentation included in this submission. -
Common Name: Multi-parameter patient monitoring system
{1}------------------------------------------------
| • Classification: | |||
|---|---|---|---|
| 21 CFR 870.1025 | Arrhythmia detector and alarm | 74 DSI | Class III |
Legally Marketed Predicate Devices:
The View 12 ECG Analysis Module is an addition to the existing Passport 2 Vital Signs Monitor (K993531). This submission compares the performance specifications and functionality of the Passport 2 Vital Signs Monitor with View 12 ECG Analysis Module to those of the Spacelabs Ultraview 1050 Monitor (K972282) equipped with Integrated Multiparameter Module 90496 (K972502). The functionality of the Passport 2 Vital Signs monitor with View 12 ECG Analysis Module is identical to that of the Spacelabs Ultraview 1050 monitor equipped with Integrated Multiparameter Module 90496. In addition, the diagnostic ECG algorithm used in Datascope's View 12 ECG Analysis Module is identical to that found in the Spacelabs Integrated Multiparameter Module 90496, as well as in the Mortara Instrument ELI-100 (K920627), ELI-200 (K920626A) and ELI-300 (K933143).
Description:
The View 12 ECG Analysis Module provides means for performing diagnostic ECG analysis in conjunction with the monitoring functions of Datascope's Passport 2 Vital Signs Monitor (K993531). The View 12 ECG Analysis Module for the Passport 2 Vital Signs Monitor enables 12-Lead ECG Acquisition. Continuous 12-Lead ST Segment Analysis and Arrhythmia Analysis all with print capabilities. The View 12 ECG Analysis Module consists of a PCMCIA ECG processor card that interfaces with the Passport 2, a front-end electronics cable, and a detachable patient cable with integral leadwire set. The data acquired by this module is interpreted utilizing Mortara Instrument's algorithm (cleared by FDA under notification numbers: K920627, K920626A and K933143) for interpretive electrocardiographs. With the View 12 ECG Analysis Module installed and enabled, the Passport 2 makes interpretive statements on printouts. These statements are categorized into three (3) sections: an interpretive statement, a condition statement, and a rhythm statement.
Statement of Intended Use:
In addition to the indications for use identified for the Passport 2 Vital Signs Monitor (K993531), the indications for use for the Passport 2 equipped with the View 12 ECG Analysis Module include the monitoring of the ECG waveform derived from 12 lead measurements, 12 lead ST Segment Analysis, 12 lead Ventricular Arrhythmia Detection and Interpretation of Resting 12 lead ECG.
The target populations are adult, pediatric and neonate, with the exceptions of the:
- Ventricular Arrhythmia Detection and ST Segment Analysis functions, for which . the target populations are adult and pediatric only, and
Interpretation of Resting 12 lead ECG, for which the target population is adult only. The monitor is intended for use in the health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The monitor is not recommended for use in a patient's home or residence, during patient transport other than intra-hospital, or when it has not been ordered by a physician.
{2}------------------------------------------------
Comparison of Technological Characteristics
The Passport 2 Vital Signs monitor with View 12 ECG Analysis Module is substantially equivalent to Spacelabs Medical Ultraview 1050 Monitor with Integrated Multiparameter Module 90496. The diagnostic ECG algorithm used in Datascope's View 12 ECG Analysis Module is identical to that found in the Spacelabs' Integrated Multiparameter Module 90496, as well as in the Mortara Instrument ELI-100 (K920627), ELI-200 (K920626A) and ELI-300 (K933143.) The design, components, storage technology and energy source of the Passport 2 are similar to those of its predicate device.
Testing:
The Passport 2 Vital Signs monitor with View 12 ECG Analysis Module has been subjected to extensive safety and performance testing. Final testing for the monitor equipped with the View 12 ECG Analysis Module included various performance tests designed to ensure that the device meets all functional requirements and performance specifications. Certain safety testing has been performed by third party agencies to ensure the device complies with applicable industry and safety standards. The Passport 2 Vital Signs monitor with View 12 ECG Analysis Module has also been tested to assure compliance with the requirements of various published standards, including ANSVAAMI EC11 and EC 13, IEC 60601-1 (1988-12) with Amendment 1 (1991-11) & Amendment 2 (1995-03), IEC 60601-1-1 (1992-06) with Amendment 1 (1995-11), IEC 60601-1-2 (1993-04), IEC 60601-1-4 (1996-05), IEC 60601-2-25:1993, ISO 10993-1 (1997), and EN 1441 (1997).
In conclusion, the Passport 2 Vital Signs monitor with View 12 ECG Analysis Module is substantially equivalent to the predicate devices and raises of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 8 2002
Datascope Corporation c/o Mr. Russell Olsen Director, Quality and Regulatory Affairs Patient Monitoring Division 800 MacArthur Blvd. Mahwah, NJ 07430-0619
Re: K020550
Trade Name: Passport 2® with View 12™ ECG Analysis Module Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: DSI Dated: June 6, 2002 Received: June 7, 2002
Dear Mr. Olsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Russell Olsen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. 2 uckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATION FOR USE STATEMENT
The indications for use for the Passport 2® with View 12™ ECG Analysis Module include the monitoring of the following human physiological parameters:
- ECG waveform derived from 3, 5 or 12 lead measurements ● Heart Rate derived from selected sources (SpO2, ECG, IBP, NIBP)
- Blood Oxygenation (SpO2 )* measurement/waveform ●
- ST Segment Analysis derived from 5 or 12 lead measurements
- Lethal Arrhythmia Detection derived from 5 or 12 lead measurements ● Non Invasive Blood Pressure (NIBP) measurement Invasive Blood Pressure (IBP) measurement/waveform measurable at two sites Respiration Rate/ waveform derived from ECG or CO2
- CO2, , Inspired and end tidal microstream/waveform . Temperature measurement via YSI 400/700 series probes
- Interpretation of Resting 12 lead ECG
The target populations are adult, pediatric and neonate with the exception of the:
- Lethal Arrhythmia Detection and ST Segment Analysis, for which the target populations . are adult and pediatric only, and
- Interpretation of Resting 12 Lead ECG, for which the target population is adult only. .
The monitor is intended for use in the health care facility setting.
The device has the capability of interfacing with Datascope's Gas Module II, displaying the measurements of Anesthetic Gases, O3, N2O, and CO2.
- The device monitors the SpO2 parameter via the Masimo SET® 2000 Pulse Oximeter Technology and Accessories (K990966). The Masimo SET® 2000 Pulse Oximeter Technology and Accessories are indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO3) and pulse rate (measured by an SpO3 sensor) and are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Prescription Use $\checkmark$
510(k) Number K020530
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.